K013137 · Natus Medical, Inc. · GWJ · Oct 16, 2001 · Neurology
Device Facts
Record ID
K013137
Device Name
ALGO 3 NEWBORN HEARING SCREENER
Applicant
Natus Medical, Inc.
Product Code
GWJ · Neurology
Decision Date
Oct 16, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ALGO® 3 Newborn Hearing Screener is a mobile, noninvasive instrument used to screen infants for hearing loss. The screener uses AABR® technology. The screener is intended for babies between the ages of 34 weeks (gestational age) and 6 months. Babies should be well enough to be ready for discharge from the hospital, and should be asleep or in a quiet state at the time of screening. The screener is simple to operate. It does not require special technical skills or interpretation of results. Basic training with the equipment is sufficient to learn how to screen infants who are in good health and about to be discharged from the hospital. A typical screening process can be completed in 15 minutes or less. Sites appropriate for screening include the well-baby nursery, NICU, mother's bedside, audiology suite, outpatient clinic, or doctor's office.
Device Story
ALGO 3 screens for hearing loss by detecting auditory brainstem response (ABR). Device generates 35 or 40 dB nHL stimulus clicks delivered via transducer to infant ears; sensors on skin capture brain wave responses. System uses automated signal processing to isolate ABR from background noise; compares response to template derived from normal-hearing infants. Operates in nurseries, NICUs, or clinics; used by trained non-medical personnel. Output is binary 'PASS' or 'REFER' result printed for medical chart. No user-adjustable parameters; click rate, intensity, and thresholds are pre-programmed. Integrated software tracks data and generates reports. Benefits include objective, non-invasive, rapid screening without requiring specialized interpretation.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance comparison to predicate device.
Technological Characteristics
Mobile system comprising laptop, printer, and cable assemblies (ATA, PCA, Preamp). Uses disposable FlexiCouplers earphones and Jelly Tab sensors. Employs automated auditory brainstem response (AABR) technology. Pre-programmed screening parameters (click rate, intensity, impedance/noise thresholds). Software-based signal processing for ABR detection. Standalone operation with integrated data tracking software.
Indications for Use
Indicated for hearing loss screening in infants aged 34 weeks gestational age to 6 months. Patients must be medically stable for hospital discharge and in a quiet or asleep state during screening.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Predicate Devices
ALGO-2 Newborn Hearing Screener (KA36039)
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 1 6 2001
Image /page/0/Picture/3 description: The image shows the word "natus" in a bold, sans-serif font. The letters are black against a white background. A small registered trademark symbol is located to the right of the letter "s".
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ K013137
#### Submitter: A.
Natus Medical, Inc. 1501 Industrial Road San Carlos, CA 94070
phone: (650) 802-0400 (650) 802-6621 fax:
Contact: Sheila Ramerman
Date Prepared: October 12, 2001
## B. Device Names:
| Classification name | Stimulator, Auditory Evoked Response |
|---------------------|--------------------------------------|
| Common/usual name | Hearing Screener |
| Proprietary name | ALGO® 3 Newborn Hearing Screener |
- C. Predicate Device:
ALGO-2 Newborn Hearing Screener, KA36039
#### D. Device Description:
Like the predicate device, the ALGO 3 detects the auditory brainstem response (ABR) to a series of stimulus clicks ('clicks'), which screens the entire hearing pathway from the outer ear to the brainstem. The ABR is not affected by the status of the middle ear, and evaluation of middle ear status is not required prior to ABR detection. The ALGO 3 generates soft clicks at 35 dB nHL ('normal hearing level' scale) which are delivered to the infant's ears through an Acoustic Transducer Assembly cable (ATA) to disposable earphones (Flexicouplers™ Disposable Earphones); a 35 dB nHL stimulus will detect a 30 dB nHL hearing loss. Each click evokes a series of identifiable brain waves from the auditory brainstem area of the infant's brain. Sensors (Jelly Tab™ Sensors) applied to the infant's skin pick up the brain wave response and transmit the signals to the screener via the Patient Cable Attachment (PCA) and Preamplifier (Preamp) assemblies. The screener uses advanced signal processing technology to separate the ABR from background noise and from other
Natus Medical Inc. 1501 Industrial Road, San Carlos, CA94070-4111 650-802-0400 FAX 650-802-0401
{1}------------------------------------------------
brain activity. The ALGO 3 automatically compares the baby's ABR to a template of ABRs derived from normal hearing infants (automated auditory brainstem response technology), denved from normal learing analis (actornation a different is printed if the ALGO 3
AABR), and prints a "PASS" or "REFER" result. A "PASS" result is printed in the AADN), and prints a - 1785 - or - 100. mfidence, that an ABR is present and consistent with its template. A "REFER" result is printed if the ALGO 3 cannot detect an ABR response with sufficient statistical confidence or one that is consistent with the template. A builten with sufficient statistical volinesive or self-adhesive label, which can be affixed directly to the infant's medical chart.
The ALGO 3 is also capable of screening using clicks at 40 dB nHL. Some newborn hearing The 2000 5 is also capable of the (international programs, for example), and some programs profer to belocit ar to we the screening threshold at 40 dB nHL rather than at 35. The incorporation of a 40 dB screening application in the ALGO 3 device makes the device more useful to more users.
The ALGO 3 device is easy enough for trained volunteers and other trained non-medical personnel to use. Interpretation of screening results is not required, as the screener automatically determines if the response is consist with a template and automatically generates a "PASS" or "REFER" result, as previously discussed. In addition, training is minimal because of the easy-to-navigate graphical user interface, HELP topics in software, and a user tutorial in the ALGO 3 software.
Screening results can be tracked using the integrated ALGO DataBook® NHS Data Tracking System software. The DataBook utility retrieves and displays patient and screening data stored in ALGO 3 software. Data can be viewed on the ALGO 3's display; can be saved as an ASCII file on diskette for data management and reporting purposes; and can be exported to other databases for data management and reporting purposes.
There are no user-adjustable parameters, so that each screen is performed under exactly the same conditions. Click rate, click intensity, impedance threshold, and noise rejection thresholds are all pre-programmed into the device software, and users cannot change these screening parameters.
The ALGO 3 is composed of the following components, as is the Predicate Device:
- Screening module, with laptop computer �
- . Printer module, with label printer
- Cable assemblies (ATA, PCA, Preamp) .
- Cart for transport and component storage .
- Screening and accessory supplies 방
Accessories available for use with the ALGO 3 include single-use, disposable FlexiCouplers Disposable Earphones, and single-use, disposable Jelly Tab Sensors.
#### E. Intended Use:
The ALGO® 3 Newborn Hearing Screener is a mobile, noninvasive instrument used to screen infants for hearing loss. The screener uses AABR® technology. The screener is intended for babies between the ages of 34 weeks (gestational age) and 6 months. Babies
{2}------------------------------------------------
should be well enough to be ready for discharge from the hospital, and should be asleep or in a quiet state at the time of screening.
The screener is simple to operate. It does not require special technical skills or interpretation The solconer is smiple to openate. Annoment is sufficient to learn how to screen infants who of results. Dasie anning out to be discharged from the hospital. A typical screening process are in good neard and about to oc also. Sites appropriate for screening include the wellbaby nursery, NICU, mother's bedside, audiology suite, outpatient clinic, or doctor's office.
### Comparison with the Predicate Device: F.
The ALGO 3 Newborn Hearing Screener is a hardware, software, supplies, and materials The ALGO 3 NOW Dom Hearing Screener (K936039). The ALGO 3 and the modition of and Hearing Screeners have the same intended use and use the same operating principle.
Based on the data and information presented here, the modified ALGO 3 Newborn Hearing Screener and Accessories are substantially equivalent to the ALGO-2 Newborn Hearing Screener and Accessories manufactured and distributed by Natus Medical, Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol of three wavy lines, resembling a person with outstretched arms, which is the department's official symbol.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 6 2001
Ms. Sheila Ramerman Director, Regulatory Affairs Natus Medical, Inc. 1501 Industrial Road San Carlos, California 94070-4111
Re: K013137
Trade Name: ALGO® 3 Newborn Hearing Screener Regulation Number: 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ Dated: September 18, 2001 Received: September 19, 2001
Dear Ms. Ramerman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Sheila Ramerman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Walker, MP
Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name: _ ALGO® 3 Newborn Hearing Screener and Accessories
Indications for Use:
The ALGO® 3 Newborn Hearing Screener is a mobile, noninvasive instrument used to screen infants for hearing loss. The screener uses AABR® technology. The screener is intended for babies between the ages of 34 weeks (gestational age) and 6 months. Babies should be well enough to be ready for discharge from the hospital, and should be asleep or in a quiet state at the time of screening.
The screener is simple to operate. It does not require special technical skills or interpretation of results. Basic training with the equipment is sufficient to learn how to screen infants who are in good health and about to be discharged from the hospital. A typical screening process can be completed in 15 minutes or less. Sites appropriate for screening include the well-baby nursery, NICU, mother's bedside, audiology suite, outpatient clinic, or doctor's office.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | OR | Over-The-Counter |
|----------------------|-----------------------------------------------------------|--------------------------|
| (Per 21 CFR 801.109) | Use | (Optional Format 1-2-96) |
| | (Division Sign-Off) | |
| | Division of General, Restorative and Neurological Devices | |
| 510(k) Number | K013137 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.