← Product Code [FCG](/productcode/FCG) · K013069

# BIOSPEED (K013069)

_H.S. Hospital Services S.P.A. · FCG · Dec 12, 2001 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K013069

## Device Facts

- **Applicant:** H.S. Hospital Services S.P.A.
- **Product Code:** [FCG](/productcode/FCG.md)
- **Decision Date:** Dec 12, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

This biopsy device can be used in Fluoroscopic, CT, Mammographic and Laparoscopic procedures to obtain biopsies of various tissues including those from Breast, Kidney, Liver, Prostate.

## Device Story

BIOSPEED™ is an automated, disposable spring-loaded biopsy needle. Used by clinicians in fluoroscopic, CT, mammographic, and laparoscopic procedures to obtain tissue samples from breast, kidney, liver, and prostate. Device operates via spring-loaded mechanism to capture tissue core. Output is a physical tissue specimen for diagnostic evaluation. Benefits include standardized, minimally invasive tissue acquisition.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Automated, disposable spring-loaded biopsy needle. Mechanical actuation. Form factor designed for use in fluoroscopic, CT, mammographic, and laparoscopic procedures.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- PRO-B Biopsy Needle ([K951598](/device/K951598.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with the letters "HG" in a bold, stylized font. The letters are black and have a textured appearance, possibly due to the image being a scan or photocopy. There are horizontal lines below the letters, adding to the overall design.

## PREMARKET NOTIFICATION SUBMISSION - 510 (k)

BIOSPEED ™

Owner Operator # 8010312

## KO13069 DEC 1 2 2001 510 (k) SUMMARY

| Applicant           | : H.S. Hospital Service S.p.A.<br>Via Naro, 81 – 00040 Pomezia (Roma) Italy                                                                                              |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person      | : MMC International, LLC<br>Mr. Lucio Improta<br>10147 Umberland Place – Boca Raton, FL 33428<br>Tel. (561) 477-1671 - Fax. (561) 477-0863<br>e-mail : mmcintern@aol.com |
| Submission Date     | : September 03, 2001                                                                                                                                                     |
| Trade Name          | : BIOSPEED™ Spring Loaded Biopsy Needle                                                                                                                                  |
| Common Name         | : Spring Loaded Biopsy Needle                                                                                                                                            |
| Classification Name | : 876.1075 - Biopsy instrument                                                                                                                                           |

Substantial Equivalence :

| Company Name   | Product Name        | 510(k) # |
|----------------|---------------------|----------|
| Promedical Ltd | PRO-B Biopsy Needle | K951598  |

Indication for use :

This biopsy device can be used in Fluoroscopic, CT, Mammographic and Laparoscopic procedures to obtain biopsies of various tissues including those from Breast, Kidney, Liver, Prostate.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 2 2001

H.S. Hospital Service S.p.A. c/o Mr. Lucio Improta MMC International, LLC 10147 Umberland Place Boca Raton, Florida 33428

Re: K013069

K015009
Trade/Device Name: BIOSPEED™ Spring Loaded Biopsy Needle Regulation Number: 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: FCG Dated: September 3, 2001 Received: September 13, 2001

Dear Mr. Improta:

We have reviewed your Section 510(k) premarket notification of intent to market the indication we nave reviewed your becaon. 910(s) pecifically equivalent (for the indications felerenced above and nave determined in asketed predicate devices marketed in interstate for use stated in the encrosuly to regars neatment date of the Medical Device Ameral Food France commerce prior to May 26, 1976, the encordance with the provisions of the Federal Food, Drug, devices that have been recalised in acceracy was a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval applicati and Cosmelle Act (Act) that do not require approval controls provisions of the Act. The 1 ou may, dierelore, market the act include requirements for annual registration, listing of general controls provisions of the 110 labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classifica (sec above) into existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulations ED may be subject to such additional controller Entisting and to 898. In addition, FDA may be found in the Ood of Pouchas concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of a basedianted on the requirements of the Act that TDA has made a dolorimiation administered by other Federal agencies. You must in and ligit or any Federal statutes and regulations annuting, but not limited to: registration and listing (21 comply with an the Act 3 requirements, merceing, and manufacturing practice requirements as set CFR Part 607), adomig (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QD) regulations (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Lucio Improta

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and w you to ough and and equivalence of your device to a legally premailed.com "The results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you decline specific and 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 Of 11 and 1659. Additionally, for questions on the promotion and advertising of Compliance at (301) 594-1697 office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). I cigulation childred, "Thiorians" responsibilities under the Act may be obtained from the Other general mistmances, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Walk, is

Image /page/2/Picture/5 description: The image shows a handwritten symbol or character. It appears to be a cursive letter or a stylized design. The symbol has a loop at the top and a curved line extending downwards, with a small stroke or mark at the bottom left.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo with a bold, stylized design. Below the logo, the text "Owner Operator # 8010312" is printed in a simple, sans-serif font. The logo itself appears to be a stylized representation of letters or symbols, with a thick, horizontal line pattern filling the shapes. The overall impression is a combination of a graphic element and identifying text, likely for a business or individual operating under that number.

## PREMARKET NOTIFICATION SUBMISSION - 510 (k)

Data: 09-03-2001

BIOSPEED ™

510 (k) # KO13069

DEVICE NAME

BIOSPEED™ Biopsy Needle

## INDICATION FOR USE

This device is an automated, disposable biopsy needle and can be used in Fluoroscopic, CT and Mammographic procedures to obtain biopsies of various tissues including those from Prostate Kidney, Breast and Liver.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED LLEASE DO NOT MIGTE DEDO 4 THIS EXTE SOULD -----------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Wale

(Division Sign-Of! (Division Sign-Of.)
Division of Geners | Restorative
Division of General | Aces Division of Gener
and Neurological Levices

510(k) Number K013067

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K013069](https://fda-staging.innolitics.com/device/K013069)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).