The GUARDIAN™ Limb Salvage Hip Components: Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and, 5) treatment of fractures that are unmanageable using other techniques. Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal, distal and/or total femur is required with the following conditions: 1) patients suffering from severe arthropathy of the hip that does not respond to any conservative therapy or better alternative surgical treatment; 2) surgical intervention for severe trauma, revision hip arthroplasties, and/or Oncology indications; and, 3) matastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors). The GUARDIAN™ Limb Salvage Knee Components: Indicated for cemented use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and, 5) treatment of fractures that are unmanageable using other techniques. Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal tibia is required with the following conditions: 1) patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment; 2) surgical intervention for severe trauma, revision knee arthroplasties, and/or Oncology indications; and, 3) matastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors).
Device Story
Modular orthopedic reconstruction system for lower limb salvage; components include femoral neck, mid-sections, stems, distal hinge femur, tibial hinge assembly, axial pin, and tibial sleeves. Used in proximal femur, distal femur, total femur, proximal tibia, and hinged knee reconstruction. System provides structural replacement following radical resection or severe joint arthropathy. Operated by orthopedic surgeons in clinical/surgical settings. Output is a reconstructed skeletal structure restoring limb function and reducing pain. Benefits include salvage of limbs otherwise requiring amputation due to trauma or oncology.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing and design comparison.
Technological Characteristics
Modular orthopedic prosthesis system; components include femoral neck, mid-section, stem, distal hinge femur, tibial hinge assembly, axial pin, and tibial sleeve. Materials are metal/polymer. Designed for cemented use. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring total hip or knee arthroplasty or radical resection/replacement of proximal/distal/total femur or proximal tibia due to degenerative joint disease (osteoarthritis, rheumatoid arthritis, avascular necrosis), functional deformity, failed prior revision, fracture, severe trauma, or oncology indications (metastatic disease, bone tumors).
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Lacey Rotating Hinge Knee
DCW Modular Distal Femoral System
S.O.S.™ Proximal Femur
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text includes 'K013035' on the top line and '10F3' on the bottom line. The handwriting is somewhat stylized, with some characters more clearly formed than others.
Image /page/0/Picture/1 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" on the left, followed by the word "WRIGHT" in large, bold letters. Below "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." with a line underneath. The address "5677 AIRLINE ROAD ARLINGTON, TN 38002" and phone number "901-867-9971" are printed below the line.
DEC 0 7 2001
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the GUARDIAN™ Limb Salvage System.
| Submitted By: | Wright Medical Technology, Inc. |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | September 7, 2001 |
| Contact Person: | Ehab M. Esmail<br>Manager, Regulatory Affairs |
| Proprietary Name: | GUARDIAN™ Limb Salvage System |
| Common Name: | Limb Salvage System |
| Classification Name and Reference: | 21 CFR 888.3350 Prosthesis, Hip, Semi-<br>Constrained, Metal/Polymer Cemented- Class II<br>21 CFR 888.3510 Prosthesis, Knee, Femorotibial,<br>Constrained, Cemented, Metal/Polymer – Class II |
Device Product Code and Panel Code:
Orthopedics/87/ JDI, KRO
#### DEVICE INFORMATION
#### A. INTENDED USE
The indications for use for the GUARDIAN™ Limb Salvage System will be substantially equivalent to the indication for use listed under competitive devices previously cleared for market and identical to the indication for use previously submitted under Lacey Rotating Hinge Knee, DCW Modular Distal Femoral System, and. S.O.S.™ Proximal Femur.
Image /page/0/Picture/11 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font, with a line underneath.
Image /page/0/Picture/12 description: The image shows a black and white logo of a globe. The globe is tilted to the right and shows the continents of Asia and Australia. The logo is simple and stylized, with a thick black line representing the outline of the globe. The continents are filled in with black, and the oceans are white.
{1}------------------------------------------------
K013035
2 OF 3
## GUARDIAN™ Limb Salvage Hip Components:
Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2) inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and,
- 5) treatment of fractures that are unmanageable using other techniques.
Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal, distal and/or total femur is required with the following conditions:
- 1) patients suffering from severe arthropathy of the hip that does not respond to any conservative therapy or better alternative surgical treatment;
- 2) surgical intervention for severe trauma, revision hip arthroplasties, and/or Oncology indications; and,
- 3) matastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors).
### The GUARDIAN™ Limb Salvage Knee Components:
Indicated for cemented use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2) inflammatory degenerative joint disease including rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and,
- 5) treatment of fractures that are unmanageable using other techniques.
Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal tibia is required with the following conditions:
- 1) patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment;
- 2) surgical intervention for severe trauma, revision knee arthroplasties, and/or Oncology indications; and,
- 3) matastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors).
Image /page/1/Picture/24 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font.
Image /page/1/Picture/25 description: The image shows a black and white drawing of a globe. The globe is tilted on its axis and shows the continents of Asia and Australia. The globe is surrounded by a circular frame, and there is some text at the bottom of the image, but it is unreadable.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text includes 'K013035' and '30F-3', written in a casual, slightly slanted style. The handwriting is clear, but the overall presentation suggests it may be a label or identifier noted quickly.
### B. DEVICE DESCRIPTION
The GUARDIAN™ Limb Salvage System consists of components that are used in the reconstruction of the lower limb. The reconstruction applications are proximal femur, distal femur, total femur, proximal tibia, and hinged knee. The components are femoral neck, mid-section, stem, distal hinge femur, tibial hinge assembly, axial pin, tibial sleeve spacer, tibial sleeve, male-male mid-section, resurfacing hinge femur, and proximal tibia.
| | RECONSTRUCTION APPLICATIONS | | | | |
|-------------------------|-----------------------------|-----------------|----------------|-------------------|----------------|
| COMPONENTS | PROXIMAL<br>FEMUR | DISTAL<br>FEMUR | TOTAL<br>FEMUR | PROXIMAL<br>TIBIA | HINGED<br>KNEE |
| FEMORAL NECK | ✓ | | ✓ | | |
| MID-SECTION | ✓ | ✓ | ✓ | ✓ | |
| STEM | ✓ | ✓ | ✓ | ✓ | |
| DISTAL HINGE FEMUR | | ✓ | ✓ | | |
| TIBIAL HINGE ASSEMBLY | | ✓ | ✓ | ✓ | ✓ |
| AXIAL PIN | | ✓ | ✓ | ✓ | ✓ |
| TIBIAL SLEEVE SPACER | | ✓ | ✓ | ✓ | ✓ |
| TIBIAL SLEEVE | | ✓ | ✓ | | |
| MALE-MALE MID-SECTION | | | ✓ | | |
| RESURFACING HINGE FEMUR | | | | ✓ | ✓ |
| PROXIMAL TIBIA | | | | ✓ | |
## C. SUBSTANTIAL EQUIVALENCE INFORMATION
The intended use, material, type of interface, and design features of GUARDIAN™ Limb Salvage System are substantially equivalent to the competitive devices previously cleared for market. The safety and effectiveness of the GUARDIAN™ Limb Salvage System are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification.
Image /page/2/Picture/7 description: The image shows a black and white logo of a globe. The globe is tilted and shows the continents of Asia and Australia. The words "World Trade Organization" are written below the globe. The logo is simple and recognizable.
and the comments of the comment of the starting the first and the first and
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three stripes representing the department's mission. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 0 7 2001
Mr. Ehab M. Esmail Manager, Regulatory Affairs Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002
Re: K013035
Trade/Device Name: GUARDIAN Limb Salvage System Regulation Number: 21 CFR 888.3510; 21 CFR 888.3350 Regulation Name: Er OFF oversion is eetal/polymer constrained cemented prosthesis; Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: KRO, JDI Dated: September 7, 2001 Received: September 10, 2001
Dear Mr. Esmail:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becared in added is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreaty to ttgansy the Medical Device Amendments, or to conninered pror to May 20, 1978, are oncernance with the provisions of the Federal Food, Drug, devices that have boon roomsomed in avere approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, therefore, market of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may or subject to back as back as Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loase of actived that I Dr over device complies with other requirements of the Act that I Dri has made a associations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OFF art 007), accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quint on control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Ehab M. Esmail
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N Millerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
page 1 of 2
for Mark n millumor
| (Divisior | |
|-------------------------|-------------|
| Divis. | Restorative |
| and Newblogica services | |
| 510(k) Number | K01305 |
| | 0 |
WRIGHT
MEDICAL TECHNOLOGY, INC.
5677 AIRLINE ROAD
ARLINGTON, TN 38002
901-867-9971
K013035
GUARDIAN™ Limb Salvage System
j Number
### INDICATIONS STATEMENT
### GUARDIAN™ Limb Salvage Hip Components:
OUARDIAN - Limis Burning arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2) inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and,
- 5) treatment of fractures that are unmanageable using other techniques.
Limb salvage system is also indicated for procedures where radical resection and replacement Entho sarvage system is asso total femur is required with the following conditions:
- 1) patients suffering from severe arthropathy of that does not respond to any conservative therapy or better alternative surgical treatment;
- ourgical intervention for severe trauma, revision hip arthroplasties, and/or Oncology indications.
- 3) matastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone tumors)
## The GUARDIAN™ Limb Salvage Knee Components:
Indicated for cemented use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions:
- 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2) inflammatory degenerative joint disease including rheumatoid arthritis;
- 3) correction of functional deformity;
- 4) revision procedures where other treatments or devices have failed; and
- 5) treatment of fractures that are unmanageable using other techniques.
Image /page/5/Picture/25 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a serif font. Below the word "WRIGHT" are the words "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font.
Image /page/5/Picture/26 description: The image shows a black and white drawing of a cell. The cell appears to be in the process of cell division, with the chromosomes visible inside the cell. The cell membrane is also visible, and it appears to be intact. There are some dark spots inside the cell.
INDICATIONS STATEMENT
{6}------------------------------------------------
page of 2
Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal tibia is required with the following conditions:
- 1) patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment;
- 2) surgical intervention for severe trauma, revision knee arthroplasties, and/or Oncology indications.
- matastatic diseases (e.g., osteosarcomas, chondrosarcomas, gaint cell tumors, bone 3) tumors)
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of General Restorative Devices 510(k) Number
Prescription Use
(Per21 CFR 801.109) X
OR
Over-The Counter Use (Optional Format 1-2-96)
Mark M. Millican
Restorative 510(K) V
Image /page/6/Picture/14 description: The image shows a black and white drawing of a globe. The globe appears to be a depiction of Earth, with visible landmasses and oceans. A curved line is above the globe, possibly representing an orbit or trajectory. The image has a grainy texture, suggesting it may be a scan or reproduction of an older illustration.
Image /page/6/Picture/15 description: The image shows the logo for Wright Medical Technology. The logo consists of a stylized "W" on the left, followed by the words "WRIGHT MEDICAL TECHNOLOGY" stacked on top of each other. Below the logo is the text "A Wright Medical Group Company."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.