TOWEL DRAPE, MODELS 4409, 4410

K012957 · Dynarex Corp. · KKX · Mar 6, 2002 · General, Plastic Surgery

Device Facts

Record IDK012957
Device NameTOWEL DRAPE, MODELS 4409, 4410
ApplicantDynarex Corp.
Product CodeKKX · General, Plastic Surgery
Decision DateMar 6, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4370
Device ClassClass 2

Indications for Use

The Dynarex Towel Drape is a device consisting of natural or synthetic materials intended to be used as a patient covering during General Surgical Procedures.

Device Story

The Dynarex Towel Drape is a single-use, disposable sterile barrier device used during general surgical procedures. It is composed of nonwoven fabric or polyethylene film, which are impervious to fluids, serving to isolate the operative site from the surrounding area. Some models feature adhesive tape to secure the drape to the periphery of the operative site and may include attached pouches for the collection of liquid effluents. The device is provided sterile, having been subjected to gamma radiation to achieve a Sterility Assurance Level (SAL) of 10^-6. It is intended for use by healthcare professionals in clinical or surgical settings to maintain a sterile field and manage surgical fluids.

Clinical Evidence

Bench testing only. The device is validated for sterility using gamma radiation to achieve a Sterility Assurance Level (SAL) of 10^-6, in compliance with ANSI/AAMI/ISO 11137-1994 requirements for validation and routine control of radiation sterilization.

Technological Characteristics

Materials: Nonwoven fabric or polyethylene film. Form factor: Towel drape, potentially with adhesive tape and fluid collection pouches. Sterilization: Gamma radiation, SAL 10^-6, validated per ANSI/AAMI/ISO 11137-1994. Single-use, disposable.

Indications for Use

Indicated for use as a protective patient covering during general surgical procedures to isolate the operative site from the surrounding area.

Regulatory Classification

Identification

A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.

Special Controls

*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Dynarex Corporation (Pre-market Notification) Page 14 KO12957 Appendix A (Summary of Safety And Effectiveness) 6 2002 MAR ### Submitter: John Gagliardi, President (contact person) MidWest Process Innovation, LLC 7736 Woodside Court Maineville, OH 45039 513-573-0085 (Telephone and fax) or 513-573-0519 (Telephone and fax) JGAGL777@One.Net Trade Name: Towel Drape Common Name: Towel Drape Classification Name: Surgical Towel Drape ## Summary of Safety and Effectiveness: The Dynarex Towel Drape is substantially equivalent in function The Dynarex Towel Drape Is substantially equivals in Invelec International Drape K911039 and intended use to the DEKA Medical Surgical Drape -- K90210, Invotes International and intended use to the DERA Medical Surgical Drape - RoobErry In Solor and the 18" x 26" Towel Drape, manufactured for Dynarex Corporation (see example of labeling in Appendix D). These products are presently on the market. #### Specifically: Specifically. the Dynarex Towel Drape is exactly similar in functional design, performs the the Dynarex Tower Drape is exactly of hiller in rainer presently distributed devices. same functions and has the same intended use as these presently distributively same functions and has the same intended abo as those problems. The packaging methods and packaging materials are exactly the same, respectively. The Dynarex Towel Drape is composed of nonwoven fabric or polyethylene film and are of The Dynarex Towel Drape is composed of nomfordinations or procedures. These devices use nonvarious collingurations approphate for General Suring surgical procedures. woven and film materials as a protective patient covering during surgest them, thus f woven and film materials as a procective pacific surgitive argust them, thus function These barrier materials are essentially impervious to anarex Towel Drapes may feature to isolate the operative site from the surrounding area. Dynarefiya site Fluid to isolate the operative site from the surrounding around of the operative site. Fluid tape adhesive to temporarily bind the drape to the peripher wound solid, and tape admesive to temporarily bind the drape for collection of operative wound solid and collection pouches are attached to the drupe for collected to a sterilizing dose of gamma liquid effluents. The Dynarex Towel Drape will be subjected to a sterilizing dose of ilquid ematines. The Dynarex a Sterility Assurance Level (SAL) of 10-6. This towel drape is a single use, disposable provided in a sterile prockage. The This tower drape is a shigic use, disposuble pro validate the sterilization process are in method of sterilization and the method used to validate the study of the clubs method of Stemization and the method used to validation of Health Care Products-----------------------------------------------------------------------------------------------Compliance with ANSI/Arily 1991 1139 17 000. Requirements for Validation and Routho Corner Towel Drape is 10th This is assured through assurance level (SAL) intenthot is achieved with each lot of sterilized product. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three diagonal lines representing the wings of a bird or the staff of Hermes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular fashion around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 6 2002 Dynarex Corporation C/O John Gagliardi Midwest Process Innovation 7736 Woodside Court Mainsville, Ohio 45039 Re: K012957 Trade/Device Name: Dynarex Towel Drape Regulation Number: 878.4370 Regulation Name: Towel Drape Regulatory Class: II Product Code: KKX Dated: January 28, 2002 Received: January 31, 2002 Dear Mr. Gagliardi: We have reviewed your Section 510(k) premarket notification of intent to market the device we have reviewed your secures some the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device interstate conmisered prior to may 20, 20, 20, 20, 20, 2011, 2017 10, 2019 10, 2019 10, 2019 Amendinone, or to avriles and Cosmetic Act (Act) that do not require approval of a premarket the Four I vou, Drag, and Sounday, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include controls provise for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (I MIT), it may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device can be round the further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {2}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours Tim Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Dynarex Corporation_Pre-market Notification Image /page/3/Picture/2 description: The image is a black and white photograph with a grainy texture. The image is mostly white with some black spots scattered throughout. The black spots are concentrated in the lower left and upper right corners of the image. The image is abstract and does not depict any recognizable objects or scenes. ### Device Name: Dynarex Towel Drape KO12957 Indications for Use: The Dynarex Towel Drape is a device consisting of natural or synthetic materials intended to be used as a patient covering during General Surgical Procedures. Clins S. Lin ("Tivision Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 8 100k) Number _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...