← Product Code [LMC](/productcode/LMC) · K012941

# DRYSTAR, MODEL 4500M (K012941)

_Agfa Corp. · LMC · Nov 29, 2001 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012941

## Device Facts

- **Applicant:** Agfa Corp.
- **Product Code:** [LMC](/productcode/LMC.md)
- **Decision Date:** Nov 29, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2040
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Drystar 4500M is a free-standing printer used to print diagnostic images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable, including digital mammography.

## Device Story

Drystar 4500M is a network-only, dry, black-and-white printer; utilizes direct thermal printing technology to produce continuous-tone diagnostic images. Input consists of digital medical images received via network; output is hard-copy transparent film. Device features dual film input trays (8x10 and 10x12 inch capacity); supports daylight loading. Specifically adapted for mammography through use of thicker film for wider optical density range and specialized border handling. Used in clinical settings; operated by healthcare professionals. Provides physical hard copies for diagnostic review on standard view boxes; facilitates clinical decision-making by enabling traditional film-based interpretation of digital imaging data.

## Clinical Evidence

Bench testing only. Electrical safety tested per EN 60601-1-1 and UL-2601; electromagnetic compatibility tested per EN 60601-1-2. No clinical data provided.

## Technological Characteristics

Direct thermal printing; dry process. Network-only connectivity. Dual film input trays (8x10, 10x12). Uses thicker film for mammography to achieve wider optical density range. Electrical safety: EN 60601-1-1, UL-2601. EMC: EN 60601-1-2. Identical hardware/software architecture to Drystar 4500.

## Regulatory Identification

A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Drystar 4500 ([K010275](/device/K010275.md))
- Kodak Dryview 8610 Laser Imager for Mammography ([K002146](/device/K002146.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 2 9 2001

510(k) Summary

## Drystar 4500M

K012941

Common/Classification Name: Medical Image Hard Copy Device 21 CFR 892.2040

> Agfa Corporation 10 South Academy Street Greenville, SC 29602-9048

Contact: Jeff Jedlicka, Prepared: August 30, 2001

#### LEGALLY MARKETED PREDICATE DEVICES A.

On April 27, 2001, FDA cleared the Drystar 4500 printer under 510(k) number K010275. The Drystar 4500M, the subject of the present 510(k), is very similar to the currently marketed Drystar 4500 in both hardware and software. From the point of view of the hardware and software, the Drystar 4500M is substantially equivalent to the Drystar 4500. From the point of view of the intended use of the device, the printing of mammography images, the device is substantially equivalent to the Kodak Dryview 8610 Laser Imager for Mammography, which was cleared by FDA on 27 September 2000 as K002146. The Dryview 8610 also has a similar dry process thermal imaging technology to that of the Drystar 4500M.

#### B. DEVICE DESCRIPTION

The Drystar 4500M is a dry, B/W printer, using the direct thermal printing principle to establish continuous-tone images with medical diagnostic image quality. The printer can be sold with one or two film input trays; one can contain 8 x 10 film and the other 10 x 12 film. Film may be loaded in full daylight. The printer is a network-only printer.

The Drystar 4500M uses a thicker film for mammography than is used for regular medical images in order to provide a wider range of optical densities. The printer also handles borders for mammography images in a different manner than for regular medical images. Otherwise, the device is very similar to the Drystar 4500.

{1}------------------------------------------------

### INTENDED USE C.

The Drystar 4500M is a free-standing printer used to print diagnostic images on transparent film for viewing on a standard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable, including digital mammography.

#### SUBSTANTIAL EQUIVALENCE SUMMARY D.

The Drystar 4500M is a medical device and it has the same indications for use as the legally marketed Drystar 4500, except specific reference to mammography is added, as in the indications for use for the Kodak 8610 Dryview Printer. The Drystar 4500M has the identical technological characteristics as the Drystar 4500, and these are similar to those of the Kodak 8610. This premarket notification has described the characteristics of the Drystar 4500M in sufficient detail to assure substantial equivalence.

#### TECHNOLOGICAL CHARACTERISTICS ய்

The technological characteristics of the Drystar 4500M are identical to those of the predicate device, the Drystar 4500. The technological characteristics of the Drystar 4500M are also very similar to those of the Kodak 8610, another dry process printer.

#### TESTING F.

The device was tested for electrical safety according to EN 60601-1-1 and UL-2601, as described in the 510(k) for the Drystar 4500. The electrical systems are the same across the different models of the Drystar family of printers.

The device was tested for electromagnetic compatibility according to EN 60601-1-2, as described in the 510(k) for the Drystar 4500. The electrical systems and chassis are the same across the models of the Drystar family of printers.

#### CONCLUSIONS G.

This pre-market submission has demonstrated Substantial Equivalence as defined and understood in Sections 513(f)(1) and 513(l)(1) of the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

000020

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## NOV 2 9 2001

Agfa Corporation % T. Whit Athey, Ph.D. Senior Consultant C.L. McIntosh & Associates 12300 Twinbrook Pkwy., Suite 625 ROCKVILLE MD 20852

Re: K012941 Trade/Device Name: Drystar 4500M Printer Digital Image Hardcopy Device Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: 90 LMC Dated: August 30, 2001 Received: August 31, 2001

Dear Dr. Whit Athey:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave revioused your becated in a device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for aso stated in the encreated of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may , morelever manns of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours,

Nancy C. Snogdin

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Drystar 4500M ________________________________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The Drystar 4500M is a free-standing printer used to print diagnostic images on transparent The Drystar 4500N is a 150 bandard view box. It may be used in any situation in which a hard copy of an image generated by a medical imaging device is required or desirable, including digital mammography.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE)        |               |
|---------------------------------------------------------------|---------------|
|                                                               | Nancy Broglon |
| (Division Sign-Off)                                           |               |
| Division of Reproductive, Abdominal, and Radiological Devices |               |
| 510(k) Number                                                 | K012941       |

Prescription Use __ ما __ ما __ ما __ ما __ الما (Per 21 CFR 801.109) OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

00002:

---

**Source:** [https://fda-staging.innolitics.com/device/K012941](https://fda-staging.innolitics.com/device/K012941)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
