CELL-DYN 3200 SYSTEM WITH ABSOLUTE AND PERCENT RETICULOCYTE
Applicant
Abbott Laboratories
Product Code
GKZ · Hematology
Decision Date
Sep 28, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The CELL-DYN 3200 System with Absolute and Percent Reticulocyte count is a multiparameter, automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories and physician office laboratories.
Device Story
The CELL-DYN 3200 is an automated hematology analyzer for clinical and physician office labs. It processes EDTA-anticoagulated whole blood specimens. The system consists of an analyzer module, a data module, and a display station. For reticulocyte analysis, blood is manually diluted and stained with New Methylene Blue N dye off-line for 15 minutes before aspiration. The analyzer uses a Helium-Neon laser as an optical light source; it counts, sizes, and classifies cells via optical laser light scatter (0, 10, and 90 degrees). The data module processes, stores, and reports results. Healthcare providers use the output to assess hematological parameters, aiding in clinical decision-making regarding patient blood health. The device provides automated quantification of cell populations, reducing manual labor and improving consistency in hematology testing.
Clinical Evidence
Bench testing only. Performance was evaluated through internal studies at Abbott Diagnostics Division. Metrics included background, carryover, linearity, precision, and accuracy. Results demonstrated that the device performs according to manufacturer specifications and is equivalent to the predicate device.
Technological Characteristics
Automated hematology analyzer using Helium-Neon laser optical light scatter (0, 10, 90 degrees). Analyzes EDTA-anticoagulated blood. Reticulocyte analysis requires off-line staining with New Methylene Blue N. System comprises an integrated analyzer/data module and a separate display station with a pressure-sensitive keypad. Class II device (21 CFR 864.5220).
Indications for Use
Indicated for the automated classification of formed elements in EDTA-anticoagulated whole blood, including white blood cells, red blood cells, hemoglobin, platelets, and reticulocytes (absolute and percent). Intended for use in clinical and physician office laboratories.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}------------------------------------------------
KD12934
| 510(k)<br>Summary | CELL-DYN 3200 System with Absolute and Percent Reticulocyte | |
|-------------------|-------------------------------------------------------------|--------------------------------------------------------------------------|
| | Submitted by: | Abbott Laboratories<br>5440 Patrick Henry Drive<br>Santa Clara, CA 95054 |
| | Contact Person: | John Dean<br>Phone: (408) 567-3430<br>Fax: (408) 567-3523 |
| | Date Prepared: | August 30, 2001 |
| | Proprietary Name | CELL-DYN 3200 System with Absolute and<br>Percent Reticulocyte |
| | Common Name: | Automated Hematology Analyzer |
| | Classification Name: | Automated Differential Cell Counter<br>(21 CRF 864.5220) |
| | Predicate Device: | CELL-DYN 4000 System, K961439/S1 |
#### Intended Use:
The CELL-DYN 3200 System with Absolute and Percent Reticulocyte count is a multiparameter, automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories and physician office laboratories.
### Device Description:
The CELL-DYN 3200 System has three main modules: 1) the Analyzer, which aspirates, dilutes and analyzes each whole blood specimen; 2) the Data Module, which automatically analyzes, stores, and reports specimen results; 3) the Display Station, which consists of a color monitor and pressure-sensitive keypad for selecting the displayed commands that operate the system. The Analyzer and Data Module are housed in a single chassis. The Display Station is a stand-alone module.
The analyzer counts, sizes and classifies blood cells by Optical Laser Light Scatter. The CELL-DYN 3200 System uses a Helium-Neon laser as the optical light source. The Optical Bench detects light in the form of scatter from blood cell surfaces and internal structures.
10
{1}------------------------------------------------
For the reticulocyte parameters, an off-line dilution of blood and Reticulocyte Reagent is prepared and stained for 15 minutes. The dilution is aspirated and the reticulocytes are counted in the WOC channel. Data are collected for scatter (0, 10, and 90 degree) as each cell passes through the laser beam.
The CELL-DYN 3200 System is designed to analyze EDTAanticoagulated whole blood specimens and report the additional Absolute and Percent Reticulocyte Parameters.
The CELL-DYN 3200 System is designed to classify the following formed elements of EDTA-anticoagulated blood:
| White blood Cell Parameters: | Red Blood Cell Parameters: |
|-----------------------------------------------|------------------------------------------------|
| WBC -- White Blood Cell or<br>Leukocyte Count | RBC -- Red Blood Cell or<br>Erythrocyte Count |
| NEU -- Neutrophil Absolute<br>Count | HCT -- Hematocrit |
| %N -- Neutrophil Percent | MCV -- Mean Cell Volume |
| LYM -- Lymphocyte Absolute<br>Count | RDW -- Red Cell Distribution<br>Width |
| %L -- Lymphocyte Percent | Hemoglobin Parameters: |
| MONO -- Monocyte Absolute<br>Count | HGB -- Hemoglobin<br>Concentration |
| %M -- Monocyte Percent | MCH -- Mean Cell Hemoglobin |
| EOS -- Eosinophil Absolute Count | MCHC -- Mean Cell Hemoglobin<br>Concentration |
| %E -- Eosinophil Percent | |
| BASO -- Basophil Absolute Count | |
| %B -- Basophil Percent | |
| Platelet Parameters:<br>PLT -- Platelet Count | Reticulocyte Parameters: |
| MPV -- Mean Platelet Volume | RETIC ABS-- Reticulocyte<br>Absolute |
| *PDW -- Platelet Distribution<br>Width | RETIC% -- Reticulocyte<br>Percent of RBC Count |
| *PCT -- Plateletcrit | |
* Clinical significance has not been established for these parameters. Therefore, they are not reportable in the U.S.
** For Laboratory Use Only, therefore it is not reportable.
{2}------------------------------------------------
### Similarities and Differences:
The CELL-DYN 3200 and the CELL-DYN 4000 System are similar in that both systems enumerate reticulocytes in EDTA-anticoagulated whole blood using New Methylene Blue N dye and optical laser scatter. The two Reticulocyte Reagents enable both the CELL-DYN 3200 and the CELL-DYN 4000 Systems to count and classify reticulocytes. Blood specimens for the CELL-DYN 3200 are externally stained in an off-line preparation step with New Methylene Blue N that requires incubation prior to measurement, while the CELL-DYN 4000 system automatically aliquots Reticulocyte Reagent containing the dye when CBC + RETC or CBC + RETC, Resistant RBC test is selected. They are different in that the Analyzer and Data Module of the CELL-DYN 3200 are housed in a single chassis while the CELL-DYN 4000 has a stand-alone data station consisting of a monitor, CPU and keyboard.
## Equivalency Data:
The data compiled to support the claim that the CELL-DYN 3200 System with Reticulocyte Percent (RETIC %) and Reticulocyte Absolute (RETIC ABS) parameters is substantially equivalent to the CELL-DYN 4000 System includes: background, carryover, linearity, precision, and accuracy. The data supports the claim that the CELL-DYN 3200 System with RETIC % and RETIC ABS parameters is substantially equivalent to the CELL-DYN 4000 System.
The background, carryover, linearity, precision, and accuracy data shows performance to manufacturer's specifications. Data was collected at an internal Abbott Diagnostics Division site.
### Conclusion:
The CELL-DYN 3200 System with RETIC % and RETIC ABS parameters demonstrates substantial equivalence to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three lines forming its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. John Dean Manager, Regulatory Affairs Abbott Laboratories 5440 Patrick Henry Drive Santa Clara, California 95054
Re: K012934
Trade/Device Name: CELL-DYN® 3200 System with Absolute and Percent Reticulocyte Regulation Number: 21 CFR § 864.5220 Regulation Name: Automated Differential Cell Counter Regulatory Class: II Product Code: GKZ Dated: August 30, 2001 Received: August 31, 2001
SEP 2 8 2001
Dear Mr. Dean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
K012934 5 10(k) Number (if known):
Device Name: CELL-DYN® 3200 System with Absolute and Percent Reticulocyte
Indications for Use:
The CELL-DYN 3200® System with Absolute and Percent Reticulocyte is an automated, multiparameter hematology analyzer designed to classify the following formed elements of EDTA-anticoagulated blood:
| White blood Cell Parameters: | Red Blood Cell Parameters: |
|--------------------------------------------|---------------------------------------------|
| WBC -- White Blood Cell or Leukocyte Count | RBC -- Red Blood Cell or Erythrocyte Count |
| NEU -- Neutrophil Absolute Count | HCT -- Hematocrit |
| %N -- Neutrophil Percent | MCV -- Mean Cell Volume |
| LYM -- Lymphocyte Absolute Count | RDW -- Red Cell Distribution Width |
| %L -- Lymphocyte Percent | Hemoglobin Parameters: |
| MONO -- Monocyte Absolute Count | HGB -- Hemoglobin Concentration |
| %M -- Monocyte Percent | MCH -- Mean Cell Hemoglobin |
| EOS -- Eosinophil Absolute Count | MCHC -- Mean Cell Hemoglobin Concentration |
| %E -- Eosinophil Percent | Reticulocyte Parameters: |
| BASO -- Basophil Absolute Count | RETIC ABS -- Reticulocyte Absolute |
| %B -- Basophil Percent | RETIC% -- Reticulocyte Percent of RBC Count |
| Platelet Parameters: | |
| PLT -- Platelet Count | |
| MPV -- Mean Platelet Volume | |
| *PDW -- Platelet Distribution Width | |
| *PCT -- Plateletcrit | |
* Clinical significance has not been established for these parameters. Therefore, they are not reportable in the U.S.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
| OR | Over-The Counter Use |
|---------------------|-----------------------------------------|
| (Division/Sign-Off) | |
| | Division of Clinical Laboratory Devices |
| 510(k) Number | K 010934 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.