← Product Code [JWY](/productcode/JWY) · K012921

# SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE (K012921)

_Trek Diagnostic Systems, Inc. · JWY · Nov 2, 2001 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012921

## Device Facts

- **Applicant:** Trek Diagnostic Systems, Inc.
- **Product Code:** [JWY](/productcode/JWY.md)
- **Decision Date:** Nov 2, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Susceptibility plate is an in vitro diagnostic product for the susceptibility testing of Streptococcus pneumoniae and Haemophilus influenzae. This 510(k) is for the addition of Vancomycin in the dilution range of 0.06 - 4 ug/ml to the Sensititre Haemophilus/Streptococcus pneumoniae MIC panel for testing Streptococcus pneumoniae and Haemophilus influenzae isolates. The indications for use for Vancomycin is for: Streptococcus pneumoniae (including penicillin-resistant strains).

## Device Story

Sensititre HP MIC Susceptibility Plates are in vitro diagnostic devices used in clinical laboratories to determine the minimum inhibitory concentration (MIC) of antimicrobial agents against specific bacteria. This submission adds Vancomycin to existing MIC panels. The device consists of microtiter plates containing dehydrated antimicrobial agents. Clinical samples (bacterial isolates) are inoculated into the wells; the device measures bacterial growth inhibition to determine susceptibility. Results assist clinicians in selecting appropriate antibiotic therapy for patients infected with Streptococcus pneumoniae or Haemophilus influenzae. The device is intended for professional use in clinical laboratory settings.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing of the antimicrobial susceptibility test performance.

## Technological Characteristics

Microtiter plate format containing dehydrated antimicrobial agents (Vancomycin). Principle of operation is the microbroth dilution method for determining MIC. Designed for in vitro diagnostic use.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is facing to the right. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Cynthia C. Knapp NOV 0 2 2001 Director of Lab Services Trek Diagnostic Systems, Inc. 29299 Clemens Road, Suite 1-K Westlake, OH 44145

K012921 Re:

Trade/Device Name: Sensititre™ Haemophilus/Streptococcus pneumoniae (HP) MIC Susceptibility Plates, Vancomycin Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: II Product Code: JWY Dated: August 29, 2001 Received: August 31, 2001

Dear Ms. Knapp:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{1}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

NOV 0 2 2001

K012921

510 (k) Number (If known):

Device Name: Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Susceptibility Plates.

Indications For Use:

The Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Susceptibility plate is an in rific Schollae Fracmopillian Susceptibility testing of Streptococcus pneumoniae and Haemophilus influenzae.

This 510(k) is for the addition of Vancomycin in the dilution range of 0.06 - 4 ug/ml to the I mis Stortifits for the addition of cass pneumoniae MIC panel for testing Streptococcus Delistine Fracthophilus influenzae isolates. The "Indications for Use" and clinical pheumomac and Traomycin is for: Streptococcus pneumoniae (including penicillin-resistant strains )

## (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                      |
|--------------------------------------------------------|----------------------|
| <div><img alt="signature" src="signature.png"/></div>  |                      |
| (Division Sign-Off)                                    |                      |
| Division of Clinical Laboratory                        |                      |
| 510(k) Number                                          | K012921              |
| Prescription Use<br>(Per 21 CFR 801.109                |                      |
|                                                        | OR                   |
|                                                        | Over-The-Counter Use |

---

**Source:** [https://fda-staging.innolitics.com/device/K012921](https://fda-staging.innolitics.com/device/K012921)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).