The Apex Modular™ Alumina Femoral Head is intended for use in combination with the Apex Modular Hip Stem as the femoral component in total hip replacement procedures. This ceramic head is intended to articulate with a polyethylene or metal-backed polyethylene acetabular cup component. This prosthesis is intended for single use implantation, and may be used for the following conditions, as appropriate: - Non-inflammatory degenerative joint disease, including osteoarthritis and . avascular necrosis; - Rheumatoid arthritis; . - Correction of functional deformity; . - Congenital dislocation; . - Revision procedures where other treatments or devices have failed; . - Femoral neck and trochanteric fractures of the proximal femur. ●
Device Story
Modular alumina ceramic femoral head; component of total hip replacement system. Input: mechanical interface with Apex Modular Hip Stem via 12/14 taper. Operation: provides articulating surface for polyethylene or metal-backed polyethylene acetabular cup. Used in surgical setting by orthopedic surgeons. Output: physical articulation surface; restores hip joint function. Benefits: high-purity alumina ceramic material provides wear resistance; modularity allows for offset adjustments (-3.5 to +4 mm) to accommodate patient anatomy.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence established via mechanical burst and fatigue testing of alumina heads on the Apex Modular 12/14 taper.
Technological Characteristics
Material: Biolox® forte (high purity aluminum oxide ceramic). Design: Modular femoral head, 28 mm and 32 mm diameters, 12/14 taper interface. Offsets: -3.5, 0, +3.5 mm (28 mm) and -4, 0, +4 mm (32 mm). Single-use, non-porous prosthesis.
Indications for Use
Indicated for patients requiring total hip replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, congenital dislocation, revision of failed prior treatments, or femoral neck/trochanteric fractures of the proximal femur.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
Smith & Nephew Biolox Alumina Ceramic Heads (K981847 and K991162)
Plus Orthopedics Biolox® forte Alumina Heads (K990261)
Submission Summary (Full Text)
{0}------------------------------------------------
# Apex Modular™ Alumina Femoral Head
| 1. Submitter: | Apex Surgical, LLC<br>12 Harding Street<br>Suite 202<br>Lakeville, MA 02347 | Contact |
|---------------|-----------------------------------------------------------------------------|---------|
|---------------|-----------------------------------------------------------------------------|---------|
#### 2. Device Name
| Proprietary Name: | Apex Modular™ Alumina Femoral Head |
|----------------------|----------------------------------------------------------------------------------------------|
| Common Name: | Ceramic femoral head |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis |
| Regulatory Class: | Class II per 21 CFR §888.3353 |
#### 3. Intended Use
The Apex Modular™ Alumina Femoral Head is intended for use in combination with the Apex Modular Hip Stem as the femoral component in total hip replacement procedures. This ceramic head is intended to articulate with a polyethylene or metal-backed polyethylene acetabular cup component. This prosthesis is intended for single use implantation, and may be used for the following conditions, as appropriate:
- Non-inflammatory degenerative joint disease, including osteoarthritis and . avascular necrosis;
- Rheumatoid arthritis; .
- Correction of functional deformity; .
- Congenital dislocation; .
- Revision procedures where other treatments or devices have failed; .
- Femoral neck and trochanteric fractures of the proximal femur. ●
#### 4. Device Description
The Apex Modular™ Alumina Heads are manufactured of Biolox® forte alumina (high purity aluminum oxide ceramic) by CeramTec AG (Biolox® is a registered trademark of CeramTec AG). The bore on this ball was designed and tested for compatibility with the neck taper on the Apex Modular Hip Stem (K000788). These modular heads are available in 28 mm and 32 mm diameters, with three offset options for each head size: -3.5, 0, and +3.5 mm (28 mm diameter heads) and -4, 0, and +4 mm (32 mm diameter heads).
### August 27, 2000
Edward J. Cheal, Ph.D. Managing Director (508) 947-6500 (voice) (208) 248-8227 (fax)
510(K) Summary KO12918
NOV 2 7 2001
Image /page/0/Picture/18 description: The image shows the logo for Apex Surgical. The logo consists of a large, bold letter "A" followed by the words "Apex Surgical" in a slightly smaller font. The letters are all black, and the word "Surgical" has a trademark symbol next to it.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Apex Surgical. The logo consists of a large, stylized letter "A" on the left, followed by the words "Apex Surgical" in a bold, sans-serif font. The letters are all black, and there is a small trademark symbol to the upper right of the word "Surgical".
### 5. Predicate Device Comparison
Substantial equivalence is claimed to the Smith & Nephew Biolox Alumina Ceramic Heads and the Plus Orthopedics Biolox® forte Alumina Ceramic Heads. The table below r rodus and the features and characteristics of the Apex Modular Alumina Femoral Head to these predicate devices.
| | Apex Modular<br>Alumina<br>Femoral Head | Smith &<br>Nephew Biolox<br>Alumina<br>Ceramic Heads<br>(K981847 and<br>K991162) | Plus<br>Orthopedics<br>Biolox® forte<br>Alumina Heads<br>(K990261) |
|-----------------------------------------------|-----------------------------------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------|
| INTENDED USE | | | |
| Primary and revision<br>total hip replacement | Yes | Yes | Yes |
| DESIGN | | | |
| Taper design | 12/14 | 12/14 | 12/14 |
| Head diameters | 28 and 32 mm | 28 and 32 mm | 28 and 32 mm |
| Offsets, 28 mm heads | -3.5, 0, +3.5 mm | 0, +3.5, +7 mm | -3.5, 0, +3.5 mm |
| Offsets, 32 mm heads | -4, 0, +4 mm | 0, +4, +8 mm | -4, 0, +4 mm |
| MATERIALS | | | |
| Ceramic head | Alumina | Alumina | Alumina |
| Stem trunion | Titanium alloy | Ti alloy or CoCr | Ti alloy or CoCr |
These predicate heads are produced by the same vendor (CeramTec AG), using the same material (Biolox® forte), as the subject device. Burst and fatigue testing of these alumina heads on the Apex Modular 12/14 taper have been completed as per the relevant FDA guidance document.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/13 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, with flowing lines representing movement or connection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 7 2001
Edward J. Cheal, Ph.D. Managing Director Apex Surgical, LLC 12 Harding Street, Suite 202 Lakeville, Massachusetts 02347
Re: K012918
Trade/Device Name: Apex Modular Alumina Femoral Head Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis
Regulatory Class: II Product Code: LZO Dated: August 27, 2001 Received: August 30, 2001
Dear Dr. Cheal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Edward J. Cheal, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Milken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use Device Name: Apex Modular™ Alumina Femoral Head KO12918
The Apex Modular™ Alumina Femoral Head is intended for use in combination with the The Apex Modular Hip Stem as the femoral component in total hip replacement procedures. Apex Modular Hip Otom as the rended to articulate with a polyethylene or metal-backed This ceramic nead in intended to annonent. This prosthesis is intended for single use implantation, and may be used for the following conditions, as appropriate:
- ation, and may be decenerative joint disease, including osteoarthritis and . avascular necrosis;
- Rheumatoid arthritis; .
- Correction of functional deformity; ●
- Congenital dislocation; .
- Oongonitar arocedures where other treatments or devices have failed; .
- Femoral neck and trochanteric fractures of the proximal femur. .
(Please do not write below this line - Continue on another page if necessary) Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N Millman
sion of General, Restorative
51 ( ) Number.
OR
Prescription Use X (Per 21 CFR §801.109) Over-the-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.