The 3i Locator Abutment System for endosseous dental implants is appropriate for use with overdentures retained in whole or in part by endosseous implants in the mandible or maxilla.
Device Story
The 3i Locator Abutment System is a dental attachment system providing snap-fit retention for overdentures. It consists of a titanium alloy socket attached to an endosseous dental implant and a nylon component attached to the denture. The system functions by mating the nylon component into the titanium socket. It is used by dental professionals in a clinical setting to stabilize overdentures for patients with missing teeth. The device is a modification of the predicate Zest Locator Anchor System, adapted for use with 3i's Osseotite implants. It benefits patients by providing secure retention for removable dental prostheses.
Clinical Evidence
Bench testing only. The device underwent fatigue testing to 5 million cycles, demonstrating a fatigue strength of at least 175 pounds for the attachment components. No clinical data was required or provided.
Technological Characteristics
Materials: Titanium alloy (socket), Nylon 101 (component). Principle: Mechanical snap-fit retention. Form factor: Abutment for endosseous dental implants. Connectivity: None. Sterilization: Not specified (nylon components sterilized).
Indications for Use
Indicated for tissue-supported overdentures on 2-4 implants, partially edentulous overdentures with one or more implants, overdentures with minimum 3.17mm interarch distance, and overdentures with up to 40° divergence between implants. Contraindicated for multiple implants with >40° divergence and overdentures with >4 implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Zest Locator Anchor System
Submission Summary (Full Text)
{0}------------------------------------------------
Implant Innovations, Inc. Implant Innovations, Inc.
510(k) Premarket Notification for Modification to Locator® Anchor System
Image /page/0/Picture/2 description: The image is a logo for Implant Innovations, Inc., a Biomet company. The logo features a stylized "3i" in bold, black font. Below the "3i" are the words "IMPLANT INNOVATIONS, INC." in a smaller, sans-serif font, and below that is the text "A BIOMET COMPANY".
KO/2911
# Summary of Safety & Effectiveness
COMPANY: Implant Innovations, Inc. 4555 Riverside Drive Palm Beach Gardens, FL 33410
### CONTACT:
Jacquelyn A. Hughes, RAC Tacqueryn A. Regulatory Affairs & Quality Assurance Telephone: 561-776-6819 FAX: 561-776-6852 E-mail: jhughes@3implant.com
DATE PREPARED: August 29, 2001
NAME OF THE DEVICE: 3i Locator@Abutment System Classification: DZE Classinonene: Abutment for Endosseous Implant
PREDICATE DEVICES:
Zest Locator® Anchor System
### DEVICE DESCRIPTION:
DEVICE DESCRIPTION:
The 3i Locator Abutment System consists of a titanium alloy socket that is attached to a The 31 Locator Abutilent System consisted on umplants having an internal threaded socket. Both devices have a nylon component hat has a shape on one end that mates into socket. Both devices nave a hylon componities mass artached to the denture. This arrangement provides snap-fit retention of the system is identical to the arrangement provides shap-in relemon or the allow use with 37's Ossectite implant,
predicate devices, except for a modification to allow use with 31's from by 31 predicate devices, except for a modification of the system prior to distribution by 3i.
### INTENDED USE:
INTENDED USE:
The 3i Locator Implant Anchor Abutment for Endosseous Dental Implants is appropriate The 31 Locator Imprain Anchor Abdulentures retained in whole or in part by endosseous implants in the mandible or maxilla.
79
{1}------------------------------------------------
## INDICATIONS FOR USE:
- Tissue supported overdentures on 2 to 4 implants ●
- Partially edentulous overdentures with one or more implants .
- Farthany edentures with a minimum of 3.17mm of interarch distance .
- Overdentures with a milling greater than 40° of divergence between them .
### CONTRAINDICATIONS:
- INTRAINDICATIOND.
Multiple implants with greater than 40° of divergence between them ●
- Overdentures with more than 4 implants .
### PERFORMANCE DATA:
PERFORMANCE DATA.
All previously submitted bench testing of the Zest Locator Anchor System has All previously submitted belief tosting of an essess in retention and withstanding occlusal forces. The system was also fatigue tested to 5 million cycles, and demonstrated that the fatigue strength of the attachment components is still at least 175 pounds. There are no modifications to the system in the 31 Locator Abutment System that would demonstrate modifications to the system in the St Locations (replacement males) were fatigue different results. Stermized Nyfon 101 components (2001) of non-sterilized components.
In conclusion, this 510(k) supports the substantial equivalence of the 3i Locator In Concrusion, this 510(x) Seppedicate, the Zest Locator Anchor System.
80
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and a pair of wings at the top.
Public Health Service
SEP 2 8 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Jacqueline A. Hughes Director of R/A & Q/A Implant Innovations, Incorporated 4555 Riverside Drive Palm Beach Gardens, Florida 33410
Re: K012911
K012911
Trade/Device Name: 31 Locater Abutment System Regulation Number: 872.3640 Regulation Names Endosseous Dental Implant Regulatory Class: III Product Code: DZE Dated: August 29, 2001 Received: August 30, 2001
Dear Ms. Hughes:
We have reviewed your Section 510(k) premarket notification of intent to market the device the We have reviewed your Section 310(x) pendice is substantially equivalent (for the referenced above and have determined the device to legally marketed predicate devices marketed in
indications for use stated in the enclosure was an anteger dies of the Meal indications for use stated in the enclosules to the enactment date of the Medical Device interstate commerce prior to May 20, 1970, the condance with the provisions of
Amendments, or to devices that have been recorders approval of a premarket Amendments, or to devices that have occi receive approval of a prematice approval of a premarket
the Federal Food, Drug, and Cosmetic Act (Act) that douice, subject to the ge the Federal Food, Drug, and Cosment Act (100) market the device, subject to the general
approval application (PMA). You may, therefore, market the device, subject to the ge approval application (FMA). Tou may, and controls provisions of the Act include controls provisions of the Act. The general of devices, good manufacturing practice,
requirements for annual registration, listing of devices, good manufacturing practice, requirements for annual regional misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III
ns affect If your device is classined (see above) into entrols. Existing major regulations affecting (PMA), it may be subject to such additional controllations, Title 21, Parts 800 to 898. In
your device can be found in the Code of Federal Regulations, Title 21, Federal your device can be found in the Code of Peacharies, would been and one in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that FDA S Issualice of a savice complies with other requirements
mean that FDA has made a determination that your device complies with other Federal agenc mean that FDA has made a delemination that your as a secures and regulations and registes.
{3}------------------------------------------------
#### Page 2 - Ms. Hughes
You must comply with all the Act's requirements, including, but not limited to: registration and I ou must comply while an labeling (21 CFR Part 801); good manufacturing practice nstilly (21 CFR Part 807), accems (QS) regulation (21 CFR Part 820); and if requirements as set form in the quality by examp (section 531-542 of the Act, 21); CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your 510(k) premated notification. The FDA finding of substantial equivalence of your device to a legally marketed nounceaton. The PDA income of succeantary vour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific advice for your do to contact the Office of Compliance at additionally 809.10 for m vito diagnestions on the promotion and advertising of your device, (301) 594-4613. entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your responsional Assistance at its toll-free number (800) 638-2041 or Manufacturers, International ant ediress http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Implant Innovations, Inc. Implant Innovations, Inc.
510(k) Premarket Notification for Modification to Locator® Anchor System
510(k) Number (if known): K012291
Device Name: _________________________________________________________________________________________________________________________________________________________________
### Intended Use
The 3i Locator Abutment System for endosseous dental implants is appropriate for use The 3i Locator Abutment System for endosseous denar mipans in 1-11-11-11 in the mandible or maxilla.
#### Indications for Use
- Tissue supported overdentures on 2 to 4 implants .
- I issue supported ous overdentures with one or more implants �
- Partially edentureds overseinm of 3.17mm of interarch distance .
- Overdentures with a minimum of STFAMIL to advergence between them ●
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--|--------------------------------------------------------|
|--|--------------------------------------------------------|
*Susan Runner*
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
| 510(k) Number | K012911 |
|---------------|---------|
|---------------|---------|
| Prescription Use: | |
|----------------------|-----------------------|
| (Per 21 CFR 801.109) | OR |
| | Over the Counter Use: |
76
of 1
Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.