Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal, Abdominal, Intra-Operative, Pediatric, Small Organ (breast, thyroid, testicle), Neonatal Cephalic, Adult Cephalic, Cardiac, Trans-Rectal, Trans-Vaginal, Peripheral-Vascular, Muscular-Skeletal (Conventional, Superficial). Typical examinations performed using the system are: General abdominal and pelvic studies including organ surveys, assessment, and retro-peritoneal cavity studies. Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. Pediatric scans of organs and bony structures. Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery. Monitoring procedures for infertility studies (other than in vitro fertilization). First, second and third trimester pregnancy studies. Prostate, prostate biopsy guidance, and rectal wall studies. Neonatal head studies. Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. Cardiac studies in adults and children. Biopsy guidance for tissue or fluid sampling. Conventional podiatry scans.
Device Story
SA6000II is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via linear and curved linear array transducers (2.0-10 MHz) and displays B-mode, M-mode, or combined modes. Digital beamforming and digital dynamic receive focusing provide high-resolution imaging. Operators use a control panel to adjust transmit focusing, filtering, image enhancement, and dynamic window curves. Features include cine loop (64 frames), real-time zoom, and image filing. Optional biopsy guide adapters assist tissue sampling. Measurements include distance, area, circumference, volume, and obstetric calculations (e.g., BPD, HC, AC, FL). Used by healthcare professionals in clinical settings for diagnostic imaging and biopsy guidance. Output affects clinical decision-making by providing anatomical visualization and quantitative data for diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance verified through electromechanical safety testing (IEC 60601-1, UL 2601-1, CSA C22.2 No. 601.1) and acoustic output declaration per IEC 1157:1992.
Technological Characteristics
Mobile ultrasound system using piezoelectric transducers (2.0-10 MHz). Features digital beamforming and digital dynamic receive focusing. Supports B-mode, M-mode, and 3D imaging. Connectivity includes cine loop and image filing. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 1157. Biocompatible patient contact materials.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, and muscular-skeletal applications. Includes biopsy guidance.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
SEP 1 2 2001
K0/2887
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
Submitter's name, address, telephone number, contact person: 1)
> Gary J. Allsebrook, Consultant Regulatory Management Services 16303 Panoramic Way San Leandro CA 94578-1116 Telephone: 510-276-2648 Facsimile: 510-276-3559 Email: regman l @home.com
Prepared July 6, 2001
Name of the device, including the trade or proprietary name if applicable, the common or 2) usual name, and the classification name, if known:
> Common/Usual Name: Diagnostic Ultrasound System and Accessories
Proprietary Name:
SA600011 Diagnostic Ultrasound System and Transducers.
| Classification Names: | FR Number | Product Code |
|---------------------------------------|-----------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### Identification of the predicate or legally marketed device: 3)
Medison America, Inc. believes that SA6000II Ultrasound System is substantially equivalent to the currently marketed SA5500 System (K992761) and SA9900 System (K002185)
#### Device Description: 4)
The SA600011 scanner is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode, M-mode, or as a combination of these modes. SA600011 also gives the operator the ability to measure
{1}------------------------------------------------
anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals.
Five different models of transducers are available and any two may be connected at the same time. In addition to the initial operational settings for each transducer preprogrammed in the system, user-customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function. More detailed explanations of these functions and controls are included in the Operator's manual, and in the software/firmware documentation included in this 510(k) Notification. Patient contact materials are the same as those previously cleared in the predicate device, SA5500 Ultrasound System and Transducers (K992761) and SA9900 Ultrasound System and Transducers (K002185).
The SA6000II uses digital beamforming technology. The SA6000II supports a variety of Linear and Curved Linear Array probes for wide variety of applications. It is an ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 2.0 MHz to 10 MHz. These probes can be applied to a variety of fields such as fetal, abdominal, intra-operative, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, muscular-skeletal. The same clinical uses were cleared for the predicate device, SA5500 Ultrasound System (K992761) and SA9900 Ultrasound System (K002185).
SA6000II provides high quality images and various measuring functions. It can measure distances and calculate areas, circumferences and volumes, as well as calculate the date of delivery by using BPD (biparietal diameter), OFD (occipito-frontal diameter), HC (head circumference), AC (abdominal circumference), AD (abdominal diameter), FL (femur length), CRL (crown rump length), APD (anteroposterior abdominal diameter), TTD (transverse trunk diameter), GS (gestational sac), etc.
Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. The operating Modes of SA6000II are B, M, or as a combination of these modes. M-mode uses the sweep display method that has its images flow from the left to the right on the monitor. The SA6000II supports the Cine function (capable of storing up to 64 frames) and real-time zoom function to the region-of-interest. Management of patient history is possible by an image-filing function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing.
The SA6000II has been designed to meet the following electromechanical safety standards:
- EN 60601-1 (IEC 601-1.) European Norm, Medical Electrical Equipment (
- UL 2601-1, Underwriters Laboratories Standards, Medical Electrical Equipment (
- C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment く
- CEI/IEC 1157:1992, International Electrotechnical Commission, Requirements for the ( declaration of the acoustic output of medical diagnostic ultrasonic equipment
{2}------------------------------------------------
- EN 60601-1-2 (IEC 60601-1-2,) European Norm, Collateral Standard: Electromagnetic ( Compatibility
- Compliant with the European Medical Device Directive Certificate issued by TUV. (
#### 5) Intended Use:
SA6000II intended uses as defined FDA guidance documents are:
- Fetal OB/GYN (
- ( Abdominal
- ( Pediatric
- Small Organ (breast, thyroid, testicle) (
- Neonatal Cephalic {
- ( Adult Cephalic
- Cardiac (
- { Trans-Rectal
- { Trans-Vaginal
- ( Peripheral-Vascular
- Muscular-Skeletal (Conventional, Superficial) {
Typical examinations performed using the system are:
- General abdominal and pelvic studies including organ surveys, assessment, and retro-peritoneal ( cavity studies.
- Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. (
- ( Pediatric scans of organs and bony structures.
- Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery. (
- Monitoring procedures for infertility studies (other than in vitro fertilization). (
- ( First, second and third trimester pregnancy studies.
- Prostate, prostate biopsy guidance, and rectal wall studies. (
- ( Neonatal head studies.
- Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. (
- Cardiac studies in adults and children. (
- ( Biopsy guidance for tissue or fluid sampling.
- ( Conventional podiatry scans.
{3}------------------------------------------------
.
#### Technological Characteristics: 6)
This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and disnayed as 2D and M-mode, Spectral Doppler, Color Doppler, Power Doppler, or 3D images. Transducer patient contact materials are biocompatible.
The device's acoustic output limits are:
All Applications:
| | (Maximum Range) |
|-------|-----------------|
| ISPTA | 720 mW/cm2 |
| MI | 1.9 |
The limits are the same as predicate Track I devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 2 2001
Medison America, Inc. % Mr. Mark Job Program Manager TUV Product Service, Inc. 1775 Old Highway 8 NW. Suite 104 NEW BRIGHTON MN 55112-1891
Re: K012887
Trade Name: SonoAce SA-6000II Diagnostic Ultrasound System Regulatory Class: II/21 CFR 892.1560 Product Code: 90 IYO Dated: August 24, 2001 Received: August 28, 2001
Dear Mr. Job:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SonoAce SA-6000II Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
L5-9EC/7.5MHz/Linear Array L5-9ER/7.5MHz/Linear Array C2-4ES/3.0MHz/Curved Linear Array C3-7ED/4.5MHz/Curved Linear Array EC4-9ES/6.5MHz/Curved Linear Array
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations
{5}------------------------------------------------
affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{6}------------------------------------------------
Page 3 – Mr. Job
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
01288 510(k) Number: / Device Name: SA600011 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B-M) | Other<br>(Specify) |
|-----------------------------------|---|----|----|-----|-----|-------------------|---------------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | | | | | | P | Notes 3, 5, 6 |
| Abdominal (See Note 1) | | P | P | | | | | | P | Notes 3, 5, 6 |
| Intra-Operative<br>(See Note 4) | | P | P | | | | | | P | Note 5, 6 |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | | | | | | P | Notes 3, 5, 6 |
| Small Organ<br>(See Note 2) | | P | P | | | | | | P | Notes 3, 5, 6 |
| Neonatal Cephalic | | P | P | | | | | | P | Note 5 |
| Adult Cephalic | | P1 | P1 | | | | | | P1 | Note 5 |
| Cardiac | | P | P | | | | | | P | Note 5 |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | P | P | | | | | | P | Notes 3, 5 |
| Trans-Vaginal | | P | P | | | | | | P | Notes 3, 5 |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | P | P | | | | | | P | Note 5 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | P | P | | | | | | P | Notes 3, 5, 6 |
| Muscular-Skeletal<br>Superficial | | P1 | P1 | | | | | | P1 | Notes 3, 5, 6 |
| Others (Specify) | | | | | | | | | | |
# Mode of Operation
N = new indication; P = previously cleared in K992761, SA5500 Uttrasound System, P1 = previously cleared in K002185, SA9900 Ultrasound System ; E = added under Appendix E
Other Indications or Modes
Note 1: Abdominal, Solid organs, aneurysms(P)
Note 2: For example: thyroid, breast, testes (P)
Note 3: Includes imaging for guidance of biopsy (P, P1)
Note 4: Intra-Abdominal Organs(P)
Note 5: 3D Imaging (P, P1)
Note 6: Harmonic Imaging (P1)
# Concurrence of CDRH, Office of Device Evaluation(ODE)
Nancy brogdon
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K01288
ection 4.3, Page 1
1107
Indications for Use
{8}------------------------------------------------
### 510(k) Number:
#### Device Name: SA6000II Ultrasound System L5-9EC (Linear Array 7.5MHz/40mm/128elements) Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color Doppler* | Power (Amp) Doppler | Color Velocity Imaging | Combined (B-M) | Other (Specify) |
|--------------------------------|---|---|---|-----|-----|----------------|---------------------|------------------------|----------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal (See Note 1) | | N | N | | | | | | N | Notes 3, 5 |
| Intra-Operative (See Note 4) | | N | N | | | | | | N | Notes 5 |
| Intra-Operative Neurological | | | | | | | | | | |
| Pediatric | | N | N | | | | | | N | Notes 3, 5 |
| Small Organ (See Note 2) | | N | N | | | | | | N | Notes 3, 5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | N | N | | | | | | N | Note 5 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal Conventional | | N | N | | | | | | N | Notes 3, 5 |
| Muscular-Skeletal Superficial | | N | N | | | | | | N | Notes 3, 5 |
| Others(Specify) | | | | | | | | | | |
Mode of Operation
N = new indication; P = previously cleared in K992761, SonoAce 5500 Ultrasound System ; E = added under Appendix E
Other Indications or Modes
Note 1: Abdominal, Solid organs, aneurysms
Note 2: For example: thyroid, breast, testes
Note 3: Includes imaging for guidance of biopsy
Note 4: Intra-Abdominal Organs
Note 5: 3D Imaging
Note 6: Harmonic Imaging
Concurrence of CDRH, Office of Device Evaluation(ODE)
Nancy Langdon
(Divisi Divis and Ra 510(k) Nu
Prescription Use (Per 21 CFR 801.109)
Indications for Use
ection 4.3, Page 2
{9}------------------------------------------------
### 510(k) Number:
# Device Name: SAS000Il Ultrasound System
L5-9ER (Linear Array 7.5MHz/50mm/128elements) Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B-M) | Other<br>(Specify) |
|-----------------------------------|---|---|---|-----|-----|-------------------|---------------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal (See Note 1) | | N | N | | | | | | N | Notes 3, 5 |
| Intra-Operative<br>(See Note 4) | | N | N | | | | | | N | Notes 5 |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | | | | | | N | Notes 3, 5 |
| Small Organ<br>(See Note 2) | | N | N | | | | | | N | Notes 3, 5 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | N | N | | | | | | N | Note 5 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | N | N | | | | | | N | Notes 3, 5 |
| Muscular-Skeletal<br>Superficial | | N | N | | | | | | N | Notes 3, 5 |
| Others(Specify) | | | | | | | | | | |
# Mode of Operation
N = new indication; P = previously cleared in K992761, SonoAce 5500 Ultrasound System ; E = added under Appendix E
Other Indications or Modes
Note 1: Abdominal, Solid organs, aneurysms
Note 2: For example: thyroid, breast, testes
Note 3: Includes imaging for guidance of biopsy
Note 4: Intra-Abdominal Organs
Note 5: 3D Imaging
Note 6: Harmonic Imaging
### Concurrence of CDRH, Office of Device Evaluation(ODE)
Nancy Hodgson
Prescription Use (Per 21 CFR 801.109)
Division of Reproductive, Abdominal, minal.
and Radiological Devices
510(k) Number
KO 1288
Section 4.3, Page 3
{10}------------------------------------------------
### 510{k) Number:
## Device Name: SA6000Il Ultrasound System
C2-4ES (Curved Linear Array 3.0MHz/20R/128elements) Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B-M) | Other<br>(Specify) |
|-----------------------------------|---|---|---|-----|-----|-------------------|---------------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | | | | | | N | Note 5 |
| Abdominal (See Note 1) | | N | N | | | | | | N | Note 5 |
| Intra-Operative<br>(See Note 4) | | N | N | | | | | | N | Note 5 |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | | | | | | N | Note 5 |
| Small Organ<br>(See Note 2) | | N | N | | | | | | N | Note 5 |
| Neonatal Cephalic | | N | N | | | | | | N | Note 5 |
| Adult Cephalic | | N | N | | | | | | N | Note 5 |
| Cardiac | | N | N | | | | | | N | Note 5 |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | N | N | | | | | | N | Note 5 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | |
## Mode of Operation
N = new indication; P = previously cleared in K992761, SonoAce 5500 Ultrasound System ; E = added under Appendix E
Other Indications or Modes
Note 1: Abdominal, Solid organs, aneurysms
Note 2: For example: thyroid, breast, testes
Note 3: Includes imaging for guidance of biopsy
Note 4: Intra-Abdominal Organs
Note 5: 3D Imaging
Note 6: Harmonic Imaging
## Concurrence of CDRH, Office of Device Evaluation(ODE)
Nancy bradsha
(Division Sign-Off)
Prescription Use (Per 21 CFR 801.109)
Division of Reproduct and Radiological Device 510(k) Number
Indications for Use
Section 4.3, Page 4
{11}------------------------------------------------
### 510(k) Number:
#### Device Name: SA600011 Ultrasound System C3-7ED (Curved Linear Array 4.5MHz/50R/128elements) Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B-M) | Other<br>(Specify) |
|-----------------------------------|---|---|---|-----|-----|-------------------|---------------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | | | | | | N | Notes 3, 5, 6 |
| Abdominal (See Note 1) | | N | N | | | | | | N | Notes 3, 5, 6 |
| Intra-Operative<br>(See Note 4) | | N | N | | | | | | N | Notes 5, 6 |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | | | | | | N | Notes 3, 5, 6 |
| Small Organ<br>(See Note 2) | | N | N | | | | | | N | Notes 3, 5, 6 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | N | N | | | | | | N | Notes 3, 5, 6 |
| Muscular-Skeletal<br>Superficial | | N | N | | | | | | N | Notes 3, 5, 6 |
| Others(Specify) | | | | | | | | | | |
Mode of Operation
N = new indication; P = previously cleared in K992761, SonoAce 5500 Ultrasound System ; E = added under Appendix E
Other Indications or Modes
Note 1: Abdominal, Solid organs, aneurysms
Note 2: For example: thyroid, breast, testes
Note 3: Includes imaging for guidance of biopsy
Note 4: Intra-Abdominal Organs
Note 5: 3D Imaging
Note 6: Harmonic Imaging
# Concurrence of CDRH, Office of Device Evaluation(ODE)
Naveen C. Beoydore
(Division Sign-Off)
Division of Reproductive, and Radiological Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
Section 4.3, Page 5
Indications for Use
{12}------------------------------------------------
### 510(k) Number:
#### Device Name: SA60001! Ultrasound System EC4-9ES (Curved Linear Array 6.5MlHz/10R/128elements) Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B-M) | Other<br>(Specify) |
|-----------------------------------|---|---|---|-----|-----|-------------------|---------------------------|------------------------------|-------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | | | | | | N | Notes 3, 5 |
| Abdominal (See Note 1) | | N | N | | | | | | N | Notes 3, 5 |
| Intra-Operative<br>(See Note 4) | | N | N | | | | | | N | Notes 5 |
| Intra-Operative<br>Neurological | | | | | | | | | | |
| Pediatric | | N | N | | | | | | N | Notes 3, 5 |
| Small Organ<br>(See Note 2) | | N | N | | | | | | N | Notes 3, 5 |
| Neonatal Cephalic | | N | N | | | | | | N | Notes 5 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | N | N | | | | | | N | Notes 3, 5 |
| Trans-Vaginal | | N | N | | | | | | N | Notes 3, 5 |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral -Vascular | | N | N | | | | | | N | Notes 3, 5 |
| Laparoscopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | N | N | | | | | | N | Notes 3, 5 |
| Muscular-Skeletal<br>Superficial | | N | N | | | | | | N | Notes 3, 5 |
| Others(Specify) | | | | | | | | | | |
Mode of Operation
N = new indication; P = previously cleared in K992761, SonoAce 5500 Ultrasound System; E = added under Appendix E
Other Indications or Modes
Note 1: Abdominal, Solid organs, aneurysms
Note 2: For example: thyroid, breast, testes
Note 3: Includes imaging for guidance of biopsy
Note 4: Intra-Abdominal Organs
Note 5: 3D Imaging
Note 6: Harmonic Imaging
### Concurrence of CDRH, Office of Device Evaluation(ODE)
Yarriyl Shroader
(Division Sign-Off)
Division of Reproductive and Radiological Device 510(k) Number
Prescription Use (Per 21 CFR 801.109)
Indications for Use
Section 4.3, Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.