← Product Code [LYY](/productcode/LYY) · K012876

# DERMAGRIP POWDER FREE LATEX EXAMINATION GLOVES WITH ALOE VERA PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS) (K012876)

_Wrp Specialty Products Sdn. Bhd. · LYY · Oct 1, 2001 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012876

## Device Facts

- **Applicant:** Wrp Specialty Products Sdn. Bhd.
- **Product Code:** [LYY](/productcode/LYY.md)
- **Decision Date:** Oct 1, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Powder Free Latex Examination Glove with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less), Non Sterile are disposable devices intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable, non-sterile, powder-free latex examination glove; coated with Aloe Vera; features low protein content (≤ 50 µg/g). Used by healthcare professionals in clinical settings to provide a protective barrier between examiner and patient. Prevents cross-contamination during examinations. Manufactured to meet ASTM D 3578-00 standards for physical properties, dimensions, and freedom from pinholes. Biocompatibility confirmed via primary skin irritation and dermal sensitization testing.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including dimensions, physical properties, freedom from pinholes (ASTM D 3578-00), powder-free residue (ASTM D 6124-00), protein content (ASTM D 5712-99), and biocompatibility (primary skin irritation and dermal sensitization).

## Technological Characteristics

Material: Latex. Powder-free (ASTM D 6124-00). Protein content: < 50 µg/g (ASTM D 5712-99). Meets ASTM D 3578-00 for dimensions, physical properties, and pinhole AQL. Biocompatible (non-irritant, non-sensitizer).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KO12876

# 510(k) SUMMARY

# 1.0 Submitter:

| Name:      | WRP Specialty Products Sdn. Bhd.                                                                              |  |  |
|------------|---------------------------------------------------------------------------------------------------------------|--|--|
| Address:   | Lot 11, Jalan 2, Kawasan Perusahaan Bandar Baru Salak<br>Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA |  |  |
| Phone No.: | +60 3 8706 9788                                                                                               |  |  |
| Fax No.:   | +60 3 8706 5020                                                                                               |  |  |

Date of Summary Prepared: 14th August, 2001

# 2.0 Contact Person:

| Name:      | Mr. Mohd Haizan Hussein |
|------------|-------------------------|
| Phone No.: | +60 3 8706 9788         |
| Fax No.:   | +60 3 8706 5020         |

#### Name of the device: 3.0

| Trade Name:          | Powder Free Latex Examination Glove with Aloe Vera and<br>Protein Content Labeling Claim (50 micrograms or less),<br>Non Sterile |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Examination Gloves                                                                                                               |
| Classification Name: | Patient Examination Gloves (per 21 CFR 880.6250)                                                                                 |

# Identification of The Legally Marketed Device: 4.0

Class I patient examination gloves, 80LYY, powder free, that meets all the requirements of ASTM standard D 3578 - 00 and FDA 21 CFR 800.20.

#### 5.0 Description of The Device:

.

The Powder Free Latex Examination Glove with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less), Non Sterile meets all the requirements of ASTM Standard D 3578 - 00 and FDA 21 CFR 800.20.

Page 1 of 3

{1}------------------------------------------------

K012876

### Intended Use of the Device: 6.0

The Powder Free Latex Examination Glove with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less), Non Sterile are disposable devices intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

#### Summary of The Technological Characteristics of The Device: 7.0

The Powder Free Latex Examination Glove with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less), Non Sterile are summarized with the following technological characteristics compared to ASTM or equivalent standards.

| CHARACTERISTICS       | STANDARDS                                                     | DEVICE<br>PERFORMANCE                                                           |
|-----------------------|---------------------------------------------------------------|---------------------------------------------------------------------------------|
| Dimensions            | ASTM D 3578 - 00                                              | Meets                                                                           |
| Physical Properties   | ASTM D 3578 - 00                                              | Meets                                                                           |
| Freedom from pinholes | ASTM D 3578 - 00<br>FDA 21 CFR 800.20                         | Meets                                                                           |
| Powder-Free           | ASTM D 6124 - 00                                              | < 2 mg/glove                                                                    |
| Protein Level         | ASTM D 5712 - 99                                              | < 50  µg/g                                                                      |
| Biocompatability      | Primary Skin Irritation in<br>Rabbits<br>Dermal Sensitization | Passes<br>(Not a primary skin irritant)<br>Passes<br>(Not a contact sensitizer) |

### Substantial Equivalent Based on Assessment of Non-Clinical Performance Data 8.0

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

#### 9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

{2}------------------------------------------------

K012876

# 10.0 Conclusion

It can be concluded that the Powder Free Latex Examination Glove with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less), Non Sterile will perform according to the glove performance standards referenced in section 7 above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.

Page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with flowing lines, representing the department's mission related to health and human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT - 1 2001

Mr. Mohd Haizan Hussein Manager, QA/RA WRP Specialty Products Sdn. Bhd. Lot 11, Jalan 2, Kawasan Perusahaan Bander Baru, Salak Tinggi, Sepang, Selangor MALAYSIA

Re: K012876

Trade/Device Name: Dermagrip Powder Free Latex Examination Gloves with Aloe Vera and Protein Content Labeling Claim (50 Micrograms or Less) Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: August 14, 2001 Received: August 27, 2001

Dear Mr. Hussein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timo Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

# INDICATIONS FOR USE

Applicant:

WRP Specialty Products Sdn Bhd

K012876

510(k) Number (if known):

(

Device Name:

# POWDER FREE LATEX EXAMINATION GLOVES WITH ALOE VERA AND PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS), NON STERILE

Indications For Use:

The Powder Free Latex Examination Gloves with Aloe Vera and Protein Content The Fowder I roo Daton Endimans or less), Non Sterile is a disposable device Labeling Claim (20 micrograms of 1000), on the examiner's hand or finger to prevent contamination between patient and examiner.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter _____________________________________________________________________________________________________________________________________________________________ OR

Olin S. him

(Division Sign-Off) Division of Dental, Infection Contr and General Hospital 510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K012876](https://fda-staging.innolitics.com/device/K012876)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).