The Ex-Press™ Miniature Glaucoma Implant is intended to reduce intraocular pressure in patients with glaucoma where medical and conventional surgical treatments have failed.
Device Story
Ex-PRESS™ Miniature Glaucoma Implant is a stainless steel drainage tube (2.96mm length, 0.4mm diameter) designed to relieve intraocular pressure (IOP) in glaucoma patients. The device features a blunt needle-shaped penetrating tip for anterior chamber insertion, a distal flange acting as a reservoir, and a spur for anchoring. An internal stainless steel wire restricts aqueous humor flow; two versions (R-30, R-50) offer different flow resistances based on wire diameter. Three transverse openings provide secondary drainage if the primary port occludes. Used by ophthalmologists in surgical settings, the device shunts aqueous humor from the anterior chamber to a subconjunctival bleb. By lowering IOP, the device aims to prevent further glaucomatous damage. It is provided sterile with a disposable introducer.
Clinical Evidence
Prospective, open-label, multi-center study of 113 patients with open-angle glaucoma across 14 international sites. Per-protocol cohort (n=58) followed for 1 year. Primary endpoints: IOP reduction >20% from baseline (77% success) and IOP <21 mmHg (83% success). Mean glaucoma medication usage decreased from 1.55 to 0.52. Adverse events included filtering surgery (12.4%), device removal (12.4%), and bleb revision (10.6%).
Technological Characteristics
Medical grade stainless steel construction. Dimensions: 2.96mm length, 0.4mm diameter. Sensing/actuation: Passive mechanical drainage via internal wire-restricted flow path. Models R-30 and R-50 differ by internal wire diameter. Sterilized. No software or electronic components.
Indications for Use
Indicated for patients with glaucoma who have failed medical therapy and conventional surgical intervention to reduce intraocular pressure.
Regulatory Classification
Identification
An aqueous shunt is an implantable device intended to reduce intraocular pressure in the anterior chamber of the eye in patients with neovascular glaucoma or with glaucoma when medical and conventional surgical treatments have failed.
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) “Aqueous Shunts—510(k) Submissions.”
{0}------------------------------------------------
# K012852
#### 510(k) Summary
#### SUBMITTER:
#### Submitted on behalf of:
Company Name: Address:
Phone: Fax:
CONTACT PERSON:
Optonol, Ltd. Communication Center Neve Ilan 90850 Israel (972) 2 5349666 (972) 2 5349660
Richard E. Lippman, O.D., F.A.A.O. Official Representative and Correspondent C.L. McIntosh, Inc. 12300 Twinbrook Parkway Suite 230 Rockville, MD 20852
#### DATE SUMMARY PREPARED:
February, 2002
TRADE NAME: COMMON NAME: Ex-PRESS™ Miniature Glaucoma Implant Eye valve implant
## SUBSTANTIALLY EQUIVALENT TO:
The Ex-PRESS™ Miniature Glaucoma Implant is a drainage device for the anterior chamber of the eye to relieve intraocular pressure in patients with glaucoma from failed medical therapy and failed prior surgical intervention. The device is substantially equivalent to the Ahmed Glaucoma Valve (K980657), the OptiMed Glaucoma Shunt (K903462) and the Baerveldt Glaucoma Shunt (K905129 and K955455).
### DESCRIPTION of the DEVICE:
The Ex-PRESS™ Miniature Glaucoma Implant device is a stainless steel tube with a blunt needle shaped penetrating tip at one end and a flat, angled flange at the opposite (distal) end. The device is 2.96mm in length and 0.4mm in diameter. The device consists of a tube whose purpose is to capture aqueous fluid from the anterior chamber of the eve and transport the fluid to the distal end and out of the device from which the fluid then moves into a conjunctival bleb. There are three transverse reserve openings near the proximal end of the device to serve as ports for aqueous fluid in case of occlusion of the primary opening. The flange at the distal end of the device serves as
{1}------------------------------------------------
a reservoir and prevents intrusion into the eye. The spur along the bottom of the device prevents extrusion of the device and serves as an anchor. The material of the device is a medical grade stainless steel composition similar in characteristics to other medical implant devices for different indications for use. The device has a stainless steel wire embedded in the tube transverse to the shaft. The purpose of the wire is to restrict the flow passing through the tube.
The device drains aqueous humor from the anterior chamber to relieve excessive intraocular pressure associated with glaucoma. The devices available are the R-30 and R-50 versions that represent different flow characteristics created by different wire diameters welded inside the device. The higher the R- number, the lower the resistance of the device to flow.
#### INDICATIONS FOR USE:
The Ex-PRESS™ Miniature Glaucoma Implant is intended to reduce intraocular pressure in patients with glaucoma where medical and conventional surgical treatments have failed.
#### PARAMETERS AVAILABLE:
The Ex-PRESS™ Miniature Glaucoma Implant is available in the following models: R-30 and R-50.
#### PRECLINICAL INFORMATION
A battery of preclinical tests including biocompatibility evaluations of intramuscular implantation in rabbits, cytotoxicity tests, animal ocular implantations for safety and tolerance, and an LAL Endotoxicity Test were conducted. In each of these tests the results indicate that the device is nontoxic and the tests indicate that no additional questions of safety are raised.
Physical evaluations including dimensional and surface quality tests, pressure/flow characteristics and structural integrity laboratory bench tests were conducted and determined as acceptable for this device as required in the FDA Guidance: Aqueous Shunts- 510(k) Submissions, November 16. 1998.
#### CLINICAL TESTING
The clinical study was a prospective, open-label, multi-center study conducted at 14 international sites in 8 countries outside of the United States. There were 113 open-angle glaucoma patients enrolled into 3 protocols: [1] Failed medical therapy or laser trabeculoplasty, {2} Failed medical therapy or laser trabeculoplasty patients who underwent a combined Ex-PRESS™ implantation and cataract surgery and, [3] Failed filtering surgery (trabeculectomy).
The studies were conducted in accordance with the Helsinki Declarations for protection of human subjects and the applicable laws and regulations in each country. Study sites were monitored and managed by Optonol Ltd.
{2}------------------------------------------------
There was a 1-year follow-up period with intermediate scheduled examinations that included tonometry, gonioscopy, and slit-lamp examination.
The safety and effectiveness evaluation was done on a total of 113 patients implanted with the R-30 and R-50, 58 consecutive patients completed a one-year follow-up (defined as Per-Protocol cohort).
The cumulative probability of success for the Per Protocol cohort (R-30 and R-50, n=58) at 1 year (Kaplan-Meier survival curve) was 77%, where success was defined as an IOP reduction greater than 20% from baseline with or without medications (overall success). The cumulative probability of success for the Per Protocol cohort (R-30 and R-50, n=58) at 1 year (Kaplan-Meier survival curve) was 83%, where success was defined as an IOP less than 21 mmHg with or without medications (overall success).
The overall success for the Per Protocol cohort (R-30 and R-50, n=58) at 1 year was 80.4% (weighted mean), where overall success was defined as an IOP reduction greater than 20% from baseline with or without medications. The overall success for the Per Protocol cohort (R-30 and R-50, n=58) at 1 year was 75.9%, where overall success was defined as an IOP < 21 mmHg with or without medications.
The overall average number of glaucoma medications dropped significantly from 1.55 preoperation to 0.52 one-year post-operation.
Comparison of the Ex-PRESS™ efficacy results to other glaucoma drainage devices based on the literature indicates a comparable safety and efficacy profile.
For safety, the most common adverse events for the intent to treat cohort (n=113) were: additional glaucoma filtering surgery (12.4%), device removal (12.4%), revision of bleb without antimetabolites (10.6%), device-iris touch (9.7%), clinically insignificant immediate transient postoperative hyphema [<2mm blood in the anterior chamber] (8.9%), post-operative transient shallow anterior chamber (8.0%), post operative hypotony (7.1%), revision of bleb with antimetabolites (7.1%), tenon cyst (6.2%), corneal complications (6.2%), anterior chamber reformation (5.3%) and device exposure (5.3%). The safety profile of the two versions was similar.
{3}------------------------------------------------
## LABELING
The Ex-PRESS™ Miniature Glaucoma Implant is provided sterile to the practitioner in a sealed pouch contained in a sized container surrounded by an outer carton. Included in the package is the disposable device introducer used to assist in implanting the Ex-PRESS™ Device in the eye.
The package comes complete with Instructions for Use for the Practitioner.
The company address is as follows:
Optonol, Ltd. Communication Center Neve Ilan 90850 Israel (972) 2 5349666
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract image of an eagle.
## MAR 2 6 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Optonol, Ltd. c/o CL McIntosh Richard E. Lippman, O.D., F.A.A.O. 12300 Twinbrook Parkway, Suite 625 Rockville, MD 20852
Re: K012852
Trade/Device Name: Ex-Press Miniature Glaucoma Implant, Models R30 and R50 Regulation Number: 21 CFR 886.3920 Regulation Name: Aqueous shunt Regulatory Class: Class II Product Code: KYF Dated: March 11, 2002 Received: March 12, 2002
Dear Dr. Lippman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
### Indications Statement
| 510/1 N<br>Number<br>1 | | | |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--|
| Device | ਾ MM<br>were of the will and the direct and<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>There to Million in a Million Children and the<br>I an analam i a sumura a comment | Children's a lot all call call class | |
#### Indications for Use:
The Ex-Press™ Miniature Glaucoma Implant is intended to reduce intraocular pressure in patients with glaucoma where medical and conventional surgical treatments have failed.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over -The-Counter Use ________________________________________________________________________________________________________________________________________________________
(Oprtional Format 1-2-96)
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number Jo 12 85
UUUU. ..
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.