The Duracon® Constrained Posterior Stabilizer (PS) Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, Duracon® PS and TS femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be present.
Device Story
Duracon® Constrained PS Tibial Insert is a component of a total knee system; used in primary or revision cemented total knee arthroplasty. Device features an intercondylar post that fits into the intercondylar cavity of Duracon® TS or PS femoral components; includes a metallic support bracket imbedded in the tibial insert with two metallic feet engaging the baseplate. Locking mechanism is secured by a hex head locking screw. Designed for cases with joint surface destruction, bone deformity, or inadequate/absent cruciate ligaments; provides stability when anterior-posterior stability is impaired, particularly following patellectomy. Used by orthopedic surgeons in clinical settings to restore joint function and stability.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by mechanical testing comparing the subject device to the predicate.
Technological Characteristics
Polymer/metal/polymer semi-constrained cemented knee prosthesis. Features include an intercondylar post, embedded metallic support bracket, and hex head locking screw for baseplate fixation. Designed for use with Duracon® tibial baseplates, wedges, and femoral components.
Indications for Use
Indicated for patients with noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), rheumatoid arthritis, functional deformity, post-traumatic loss of joint anatomy (patello-femoral erosion/dysfunction/patellectomy), or irreparable knee fracture requiring primary or revision cemented total knee arthroplasty. Contraindicated in patients with overt infection, distant foci of infection, rapid disease progression, skeletal immaturity, or poor bone stock.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
KC012776 ા of 3
# NOV 1 6 2001
510(k) Summary Duracon® Constrained PS Tibial Inserts
### 510(k) Summary Duracon® Constrained PS Tibial Insert
#### Submission Information
| Name and Address of the Sponsor of the 510(k) Submission: | Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401-1677 |
|-----------------------------------------------------------|------------------------------------------------------------------|
| Contact Person: | Margaret F. Crowe Regulatory Affairs Consultant (201) 934-4359 |
| Date of Summary Preparation: | August 10, 2001 |
**Date of Summary Preparation**
August 10, 2001
### Device Identification
| Proprietary Name: | Duracon® Constrained PS Tibial Insert |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Knee Prosthesis |
| Classification Name and Reference: | Knee Joint, Patellofemorotibial,<br>Polymer/Metal/Polymer, Semi-<br>Constrained, Cemented Prosthesis<br>21 CFR '888.3560 |
### Predicate Device Identification
The Duracon® Constrained PS Tibial Inserts are substantially equivalent to the Duracon® TS tibial inserts found substantially equivalent in premarket notification K973164.
## Device Description
The Duracon® Constrained PS tibial inserts share critical design features as the predicate Duracon® TS tibial inserts. Both designs incorporate an intercondylar post
1a
{1}------------------------------------------------
# KQ12776 2 of 3
#### 510(k) Summary Duracon® Constrained PS Tibial Inserts
which fits into the intercondylar cavity of the Duracon® TS or PS femoral component. Both designs have a metallic support bracket imbedded into the tibial insert at the time of manufacture. This bracket has two metallic feet which engage the posterior tabs of the baseplate during assembly. The insert-baseplate locking mechanism is augmented by a hex head locking screw in both designs.
## Intended Use
The intended use of the Duracon® Constrained PS tibial inserts is identical to that of the predicate Duracon® TS tibial bearing inserts. As with the predicate inserts, the new inserts are single use devices for use as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be present.
## Indications
- Noninflammatory degenerative joint disease including osteoarthritis, traumatic t arthritis or avascular necrosis,
- . Rheumatoid arthritis,
- Correction of functional deformity, 을
- Revision procedures where other treatments or devices have failed, .
- Post-traumatic loss of joint anatomy, particularly when there is patello-femoral t erosion, dysfunction or prior patellectomy; and,
- . Irreparable fracture of the knee
{2}------------------------------------------------
KO12776 3073
510(k) Summary Duracon® Constrained PS Tibial Inserts
# Contraindications
# Absolute contraindications include:
- 프 Overt infection,
- Distant foci of infections (which may cause hematogenous spread to the implant . site).
- Rapid disease progression as manifested by joint destruction or bone absorption apparent on roentgenogram,
- . Skeletally immature patients
- Cases where there is poor bone stock which would make the procedure . unjustifiable.
# Conditions presenting an increased risk of failure include:
- Uncooperative patient or patient with neurological disorders who is incapable of 트 following instructions,
- . Osteoporosis,
- 트 Metabolic disorders which may impair bone formation,
- 트 Osteomalacia, and
- 트 Previous arthrodesis
# Performance Data
Mechanical testing has been performed to demonstrate the substantial equivalence of the subject inserts to the predicate inserts.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 2001
Ms. Margaret Crowe Regulatory Affairs Consultant Howmedica Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677
Re: K012776
Trade Name: Duracon® Constrained Posterior Stabilized (PS) Tibial Inserts Regulation Number: 888.3560 Regulation Name: Prothesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Regulatory Class: II Product Code: JWH Dated: August 17, 2001 Received: August 20, 2001
Dear Ms.Crowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Ms. Margaret Crowe
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
fo Mark n Millburn
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 1012776
### 510(k) Number (if known): K
NOV 1 6 2001
Duracon® Constrained PS Tibial Insert Device Name:
The Duracon® Constrained Posterior Stabilizer (PS) Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, Duracon® PS and TS femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be present. More specific indications/contraindications are listed below:
#### Indications
- Noninflammatory degenerative joint disease including osteoarthritis, traumatic 트 arthritis or avascular necrosis,
- Rheumatoid arthritis, ■
- Correction of functional deformity, .
- Revision procedures where other treatments or devices have failed, 사
- Post-traumatic loss of joint anatomy, particularly when there is patello-femoral 트 erosion, dysfunction or prior patellectomy; and,
- в Irreparable fracture of the knee
to Mark N. Mulkerson
" ><<<<<<rive
K012776
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.