← Product Code [MRW](/productcode/MRW) · K012773

# DISCOVERY FACET SCREW FIXATION SYSTEM (K012773)

_Depuyacromed · MRW · Nov 16, 2001 · OR · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012773

## Device Facts

- **Applicant:** Depuyacromed
- **Product Code:** [MRW](/productcode/MRW.md)
- **Decision Date:** Nov 16, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** OR
- **Attributes:** Therapeutic

## Indications for Use

The Discovery Facet Screw Fixation System is intended to stabilize the spine as an aid to fusion by two different fixation methods: Transfacet fixation - The screws are inserted bilaterally through the superior side of the facet, across the facet joint and into the inferior pedicle. Translaminar-facet fixation - The screws are inserted bilaterally through the lateral aspect of the spinous process, through the lamina, through the superior side of the facet, across the facet joint and into the inferior pedicle. For both methods, this system is indicated for the posterior surgical treatment of any or all of the following at the L1 to S1 (inclusive) spinal levels: 1) Trauma, including spinal fractures and/or dislocations; 2) Spondylolisthesis; 3) Spondylolysis; 4) Pseudoarthrosis or failed previous fusions which are symptomatic or which may cause secondary instability or deformity; 5) Degenerative diseases which include: (a) degenerative disc disease (ddd) as defined by neck and/or back pain of discogenic origin as confirmed by patient history with degeneration of the disc as confirmed by radiographic studies and/or (b) degenerative disease of the facets with instability

## Device Story

Discovery Facet Screw Fixation System; orthopedic implant for spinal stabilization. System comprises broad-headed screws and washers; used to compress bone grafts and stabilize facet joints. Surgical implantation via posterior approach; two methods: transfacet or translaminar-facet fixation. Screws inserted bilaterally across facet joint into inferior pedicle. Used by surgeons in clinical/OR settings to enhance spinal fusion. Provides mechanical immobilization of facet joints; aids in treating spinal instability, trauma, and degenerative conditions. Benefits patient by restoring spinal stability and facilitating fusion.

## Clinical Evidence

No clinical data. Evidence consists of bench testing only, specifically static and fatigue 3-Point Bend Testing and Cantilever Testing to evaluate mechanical performance.

## Technological Characteristics

Manufactured from ASTM F-136 implant grade titanium alloy. System includes fully threaded and lag screw styles with washers. Mechanical fixation device; no energy source or software components.

## Predicate Devices

- Preamendment Townley Bone Graft Screw
- Townley Transfacetpedicular Screw System ([K003928](/device/K003928.md))
- Townley Transfacet/Intrapedicular Screw ([K994308](/device/K994308.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

|              | VIII.                | 510(k) Summary                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|--------------|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|              | SUBMITTER:           | DePuy AcroMed™, Inc.<br>325 Paramount Drive<br>Raynham, MA 02780                                                                                 | NOV 1 6 2001<br>K012773                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| NOV 1 6 2001 | CONTACT PERSON:      | Karen F. Jurczak                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|              | DATE PREPARED:       | August 17, 2001                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|              | CLASSIFICATION NAME: |                                                                                                                                                  | System, Facet Screw Spinal Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|              | PROPRIETARY NAME:    | DISCOVERY Facet Screw                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|              | PREDICATE DEVICES:   | Preamendment Townley Bone Graft Screw<br>Townley Transfacetpedicular Screw System (K003928)<br>Townley Transfacet/Intrapedicular Screw (K994308) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|              | INTENDED USE:        |                                                                                                                                                  | The Discovery Facet Screw Fixation System is intended to<br>stabilize the spine as an aid to fusion by two different<br>fixation methods:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|              |                      |                                                                                                                                                  | Transfacet fixation - The screws are inserted bilaterally<br>through the superior side of the facet, across the facet joint<br>and into the inferior pedicle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|              |                      |                                                                                                                                                  | Translaminar-facet fixation - The screws are inserted<br>bilaterally through the lateral aspect of the spinous process,<br>through the lamina, through the superior side of the facet,<br>across the facet joint and into the inferior pedicle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|              |                      |                                                                                                                                                  | For both methods, this system is indicated for the posterior<br>surgical treatment of any or all of the following at the L1 to<br>S1 (inclusive) spinal levels: 1) Trauma, including spinal<br>fractures and/or dislocations; 2) Spondylolisthesis; 3)<br>Spondylolysis; 4) Pseudoarthrosis or failed previous fusions<br>which are symptomatic or which may cause secondary<br>instability or deformity; 5) Degenerative diseases which<br>include: (a) degenerative disc disease (ddd) as defined by<br>neck and/or back pain of discogenic origin as confirmed by<br>patient history with degeneration of the disc as confirmed<br>by radiographic studies and/or (b) degenerative disease of<br>the facets with instability |
|              | MATERIALS:           | Manufactured from ASTM F-136 implant grade titanium<br>alloy.                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

.

\$P^{1/2}\$

:

{1}------------------------------------------------

The Discovery Facet Screw Fixation System consists of DESCRIPTION: screws and washers designed to compress bone grafts and/or fractures. The screws are intended only for use in combination with the washer. The system includes two screw styles: fully threaded and lag.

> The DISCOVERY Facet Screw is a broad-headed screw that is designed to compact or stabilize adjacent facet articular processes to enhance spinal fusion and stability. The system is intended to stabilize the spine as an aid to fusion through bilateral immobilization of the facet joints. The screws are inserted posteriorly through the superior side of the facet, across the facet joint, and into the pedicle.

Biomechanical testing, including static and fatigue 3-Point PERFORMANCE DATA: Bend Testing and Cantilever Testing, were conducted.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stripes representing the department's mission to protect the health of all Americans and provide essential human services.

## Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Frank Maas Manager, Regulatory Affairs DePuy Acromed, Inc. 325 Paramount Drive Raynham, Massachusetts 02767

NOV 1 6 2001

Re: K012773 Trade/Device Name: Discovery Facet Screw Fixation System Regulatory Number: N/A Regulation Name: N/A Regulatory Class: unclassified Product Code: MR W Dated: August 17, 2001 Received: August 20, 2001

Dear Mr. Maas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a 

{3}------------------------------------------------

legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Q. Mark N. Millman

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## IV. Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

NOV 1 6 2001

DISCOVERY Facet Screw Fixation System Device Name:

Indications For Use:

The Discovery Facet Screw Fixation System is intended to stabilize the spine as an aid to fusion by two different fixation methods:

Transfacet fixation - The screws are inserted bilaterally through the superior side of the facet, across the facet joint and into the inferior pedicle.

Translaminar-facet fixation - The screws are inserted bilaterally through the lateral aspect of the spinous process, through the lamina, through the superior side of the facet across the facet joint and into the inferior pedicle.

For both methods, this system is indicated for the posterior surgical treatment of any or all of the following at the L1 to S1 (inclusive) spinal levels: 1) Trauma, including spinal fractures and/or dislocations; 2) Spondylolisthesis; 3) Spondylolysis; 4) Pseudoarthrosis or failed previous fusions which are symptomatic or which may cause secondary instability or deformity; 5) Degenerative diseases which include: (a) degenerative disc disease (ddd) as defined by neck and/or back pain of discogenic origin as confirmed by patient history with degeneration of the disc as confirmed by radiographic studies and/or (b) degenerative disease of the facets with instability

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use: X OR Over-The-Counter Use:
(Per 21 CFR 801.109)

for
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
510(k) Number. K012773

---

**Source:** [https://fda-staging.innolitics.com/device/K012773](https://fda-staging.innolitics.com/device/K012773)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
