The Reprocessed Sealed Electrophysiology Diagnostic Catheters, are intended for temporary use during electrophysiology studies for intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias; and for electrophysiologic mapping of cardiac structures during these evaluations.
Device Story
Reprocessed single-use electrophysiology diagnostic catheters; high-torque shaft with proximal handle; steerable or non-steerable; 4F to 7F diameter; 65-120 cm length; 2-20 platinum radiopaque electrodes. Used in clinical electrophysiology labs by physicians. Catheter connects via cables to external stimulators and electrophysiologic recorders. Device facilitates intracardiac signal acquisition and electrical stimulation for arrhythmia evaluation and cardiac mapping. Reprocessing restores device to original manufacturer specifications; includes cleaning, sterilization, and functional testing. Benefits include clinical utility of diagnostic catheters for cardiac rhythm management.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including functional design testing, process validation for cleaning and sterilization, and packaging validation. Manufacturing process includes 100% visual and functional testing of all reprocessed units.
Technological Characteristics
High-torque shaft, handle, 2-20 platinum radiopaque electrodes. Diameter 4F-7F, length 65-120 cm. Deflectable tip. Connects to external stimulators/recorders via manufacturer-recommended cables. Sterilization and cleaning processes validated for reprocessing.
Indications for Use
Indicated for patients undergoing electrophysiology studies for the evaluation of cardiac arrhythmias, requiring intracardiac sensing, recording, stimulation, temporary pacing, or electrophysiologic mapping of cardiac structures.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, composed of three overlapping profiles.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 2004
SterilMed, Inc. c/o Dr. Bruce Lester Vice President Research and Development 11400 7310 Avenue North Minneapolis, MN 55369
Re: K012678 - Supplemental Validation Submission Trade Name: See Enclosed List Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Regulatory Class: Class II (two) Product Code: NLH Dated: August 13, 2001 Received: August 14, 2001
Dear Dr. Lester:
The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on August 14, 2002 . We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
{1}------------------------------------------------
#### Page 2 - K012678
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Bhimmer for
Roy D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{2}------------------------------------------------
# Page 4 – K012678
# Enclosure – List of Devices
| Orig models found to be | SE |
|-------------------------|----|
| Daig (132) | |
| 401206 | |
| 401207 | |
| 401209 | |
| 401210 | |
| 401211 | |
| 401212 | |
| 401221 | |
| 401222 | |
| 401223 | |
| 401224 | |
| 401225 | |
| 401226 | |
| 401227 | |
| 401228 | |
| 401259 | |
| 401260 | |
| 401261 | |
| 401262 | |
| 401271 | |
| 401272 | |
| 401273 | |
| 401274 | |
| 401275 | |
| 401276 | |
| 401277 | |
| 401278 | |
| 401279 | |
| 401281 | |
| 401282 | |
| 401283 | |
| 401284 | |
| 401285 | |
| 401287 | |
| 401288 | |
| 401290 | |
| 401291 | |
| 401293 | |
| 401294 | |
| Orig models found to be | SE |
| 401296 | |
| 401297 | |
| 401298 | |
| 401299 | |
| 401300 | |
| 401301 | |
| 401305 | |
| 401306 | |
| 401307 | |
| 401308 | |
| 401309 | |
| 401310 | |
| 401311 | |
| 401312 | |
| 401313 | |
| 401317 | |
| 401318 | |
| 401325 | |
| 401328 | |
| 401329 | |
| 401353 | |
| 401354 | |
| 401355 | |
| 401356 | |
| 401357 | |
| 401359 | |
| 401360 | |
| 401362 | |
| 401366 | |
| 401374 | |
| 401375 | |
| 401376 | |
| 401379 | |
| 401380 | |
| 401381 | |
| 401382 | |
| 401383 | |
| 401384 | |
| 401385 | |
| 401386 | |
| 401387 | |
| Orig models found to be | SE |
| 401388 | |
| 401389 | |
| 401390 | |
| 401392 | |
| 401393 | |
| 401397 | |
| 401399 | |
| 401400 | |
| 401401 | |
| 401430 | |
| 401433 | |
| 401434 | |
| 401435 | |
| 401436 | |
| 401438 | |
| 401440 | |
| 401441 | |
| 401442 | |
| 401443 | |
| 401444 | |
| 401445 | |
| 401448 | |
| 401449 | |
| 401450 | |
| 401451 | |
| 401453 | |
| 401466 | |
| 401468 | |
| 401474 | |
| 401475 | |
| 401527 | |
| 401528 | |
| 401530 | |
| 401531 | |
| 401532 | |
| 401534 | |
| 401688 | |
| 401859 | |
| 401860 | |
| 401863 | |
| 401864 | |
| Orig models found to be | SE |
| 401865 | |
| 401871 | |
| 401872 | |
| 401873 | |
| 401874 | |
| 401876 | |
| 401877 | |
| 401878 | |
| 401890 | |
| 401891 | |
| 401893 | |
| 401894 | |
{3}------------------------------------------------
## Page 5 - K012678
:
.
: 上一篇:
,
.
and the comments of the comments of the comments of
:
.
{4}------------------------------------------------
### Page 6 – K012678
:
.
.
. . . . . . .
:
:
.
{5}------------------------------------------------
# Page 7 – K012678
{6}------------------------------------------------
# Indications for Use Page
Device Name: Reprocessed Sealed Electrophysiology Diagnostic Catheters
Indications for Use:
The Reprocessed Sealed Electrophysiology Diagnostic Catheters, are intended for temporary use during electrophysiology studies for intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias; and for electrophysiologic mapping of cardiac structures during these evaluations.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thos. A. for Bran Zucher
(Division Sign-Off)
Division of Cardiovascular
and Respiratory Devices
510(k) Number K012628
Prescription Use
(Per 21 CFR 801.109)
SterilMed, Inc. Reprocessed Sealed EP Diagnostic Catheters Confidential
Premarket Notification
Page I
{7}------------------------------------------------
# AUG 1 4 2002
# SECTION 2. SUMMARY AND CERTIFICATION
#### 510(K) SUMMARY A.
・
| Submitter: | SterilMed, Inc. |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Patrick Fleischhacker<br>11400 73rd Avenue North<br>Minneapolis MN, 55369<br>Ph: 888-856-4870<br>Fax: 763-488-3350 |
| Date Prepared: | August 13, 2001 |
| Trade Name: | SterilMed Reprocessed Sealed Electrophysiology<br>Diagnostic Catheters |
| Classification Name:<br>and Number: | Class II, 21 CFR 870.1220 |
| Product Code: | DRF |
| Predicate Device(s): | The SterilMed reprocessed sealed electrophysiology<br>diagnostic catheter is substantially equivalent to: the A20<br>(K953768), manufactured by Biosense Webster (formerly<br>known as Cordis Webster), the Bard Dynamic Tip Catheter<br>(K912213), and the Bard Intracardiac Electrode Catheter<br>manufactured by C.R. Bard Inc., the Stablemapr EP<br>Diagnostic Catheter (K981642), manufactured by<br>Medtronic Inc., and the original manufacturer's devices. |
| Device Description: | The device consists of a reprocessed catheter that has a<br>high-torque shaft with a handle at the proximal end, and<br>may or may not be steerable. These catheters have an outer<br>diameter of 4F to 7F, a length ranging from 65 to 120 cm,<br>with 2-20 platinum, radiopaque electrodes along the tip<br>shaft and a variety of inter-electrode spacings and curve<br>styles at the tip. The tip generally is deflectable. Specific<br>cables, as recommended by the original manufacturer,<br>connect to the handle and interface between the catheter<br>and an external stimulator and/or an electrophysiologic<br>recorder. It should be noted that this submission pertains to<br>the catheter only. It does not include any other components<br>in a system such as, connector cables, external stimulators,<br>or electrophysiologic recorders. |
{8}------------------------------------------------
The reprocessed sealed electrophysiology diagnostic Intended Use: catheters, are intended for temporary use during electrophysiology studies for intracardiac sensing, recording, stimulation and temporary pacing during the evaluation of cardiac arrhythmias; and for electrophysiologic mapping of cardiac structures during these evaluations. Functional and Representative samples of reprocessed sealed Safety Testing: electrophysiology diagnostic catheters underwent design testing to demonstrate appropriate functional characteristics. Process validation testing was done to validate the cleaning and sterilization procedures as well as the device's packaging. In addition, the manufacturing process includes visual and functional testing of all products produced. The electrophysiology diagnostic catheters sealed and Conclusion: reprocessed by SterilMed are substantially equivalent to the A20 (K953768), manufactured by Biosense Webster (formerly known as Cordis Webster), the Bard Dynamic Tip Catheter (K912213), and the Bard Intracardiac Electrode Catheter manufactured by C.R. Bard Inc., the Stablemapr EP Diagnostic Catheter (K981642), manufactured by Medtronic Inc., and the reprocessed devices' counterparts from the original manufacturer. This conclusion is based upon the fact that these devices' are essentially identical to the predicate devices in terms of functional design, materials, indications for use, and construction.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.