← Product Code [NLM](/productcode/NLM) · K012644

# REPROCESSED TROCARS AND CANNULAS (K012644)

_Surgical Instruments Service and Savings, Inc. · NLM · Jul 29, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012644

## Device Facts

- **Applicant:** Surgical Instruments Service and Savings, Inc.
- **Product Code:** [NLM](/productcode/NLM.md)
- **Decision Date:** Jul 29, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Reprocessed Trocars and Cannulas are intended to open and maintain a port of entry during endoscopic procedures.

## Device Story

Reprocessed endoscopic/laparoscopic trocars and cannulas; used to create and maintain access ports during minimally invasive surgery. Devices consist of obturators and sleeves/cannulas with various tip configurations (bladed, blunt, dilating). Reprocessing restores single-use devices to original functional specifications for reuse in clinical settings. Operated by surgeons during endoscopic procedures to facilitate instrument passage into body cavities. Benefit includes cost-effective access to surgical instrumentation while maintaining clinical performance standards equivalent to original devices.

## Clinical Evidence

No clinical data. Substantial equivalence established via supplemental validation data demonstrating that reprocessed devices meet original manufacturer specifications.

## Technological Characteristics

Mechanical surgical access devices. Components include trocars (obturators) and cannulas (sleeves). Tip configurations: bladed with shield, blunt, or dilating. Diameters range from 5mm to 12mm. Materials and specifications are consistent with original predicate devices. Non-powered, manual instrumentation. Sterilization is performed as part of the reprocessing cycle.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is simple and recognizable, conveying a sense of authority and purpose.

Public Health Service

MAY - 2 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MediSISS % Ms. Brandi J. James Director of Technical Services 2747 SW 6th Street Redmond, Oregon 97756

Re: K012644 - Supplemental Validation Submission Trade/Device Name: See Enclosed List Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NLM Dated: August 10, 2006 Received: August 14, 2006

Dear Ms. James:

The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on July 29, 2002. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements 

{1}------------------------------------------------

Page 2 - Ms. Brandi J. James

as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Nef
Osseser

Enclosure

{2}------------------------------------------------

Page 3 - Ms. Brandi J. James

## Indications for Use

510 (k) Number (if known) : K012644

Device Name:

Reprocessed Endoscopic/Laparoscopic Trocars and Cannulas

Indications for Use:

Reprocessed Trocars and Cannulas are intended to open and maintain a port of entry during endoscopic procedures.

Prescription Use AND/OR AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED )

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number: `ko 12644`
Page 1 of *_*_

{3}------------------------------------------------

## Page 4 - Ms. Brandi J. James

| Manufacturer | Model   | Description                                                       | Diameter<br>Size | Tip Configuration |
|--------------|---------|-------------------------------------------------------------------|------------------|-------------------|
| AutoSuture   | 179094  | VERSAPORT™ RPF* V2 RPF* V2 Trocar<br>System, w/Radiolucent Sleeve | 5mm              | Bladed w/Shield   |
| AutoSuture   | 179095P | VERSAPORT™ Plus V2 Trocar,<br>w/Radiolucent Sleeve                | 11mm             | Bladed w/Shield   |
| Ethicon      | 355L    | ENDOPATH TRISTAR Trocar w/Stability<br>Sleeve                     | 5mm              | Bladed w/Shield   |
| Ethicon      | 512B    | ENDOPATH TRISTAR Blunt Tip<br>Bladeless Trocar                    | 12mm             | Blunt             |
| Ethicon      | 355LD   | ENDOPATH Dilating Tip Trocar,<br>w/Stability Sleeve               | 5mm              | Bladed w/Shield   |
| Ethicon      | 355SD   | ENDOPATH Dilating Tip Trocar<br>w/Stability Sleeve                | 5mm              | Bladed w/Shield   |
| Ethicon      | 512SD   | ENDOPATH Dilating Tip Trocar with<br>Stability Sleeve             | 12mm             | Bladed w/Shield   |
| Ethicon      | 578SD   | ENDOPATH Dilating Tip Trocar<br>w/Stability Sleeve                | 7-8mm            | Bladed w/Shield   |
| Ethicon      | 511SD   | ENDOPATH Dilating Tip Trocar<br>w/Stability Sleeve                | 11mm             | Bladed w/Shield   |
| Ethicon      | 35HL    | ENDOPATH Bladeless Trocar w/Smooth<br>Sleeve, Handled             | 5mm              | Blunt             |
| Ethicon      | 35LNA   | ENDOPATH Resposable Bladeless Trocar<br>with Housing              | 5mm              | Blunt             |
| Ethicon      | 35NLT   | ENDOPATH Bladeless Trocar with<br>Stability Sleeve                | 5mm              | Blunt             |
| Ethicon      | 511H    | ENDOPATH Bladeless Trocar w/Smooth<br>Sleeve, Handled             | 11mm             | Blunt             |
| Ethicon      | 511NT   | ENDOPATH, Bladeless Trocar w/Stability<br>Sleeve                  | 11mm             | Blunt             |
| Ethicon      | 511O    | ENDOPATH, Bladeless Trocar, w/Smooth<br>Sleeve                    | 11mm             | Blunt             |
| Ethicon      | 512HA   | ENDOPATH Resposable Bladeless<br>Obturator and Housing, Handled   | 12mm             | Blunt             |
| Ethicon      | 35LST   | ENDOPATH, Integrated Stability Trocar<br>Sleeve                   | 5mm              | Cannula Only      |

.

Reprocessed Endoscopic/Laparoscopic Trocars and Cannulas found to be substantially equivalent:

---

**Source:** [https://fda-staging.innolitics.com/device/K012644](https://fda-staging.innolitics.com/device/K012644)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).