← Product Code [HRX](/productcode/HRX) · K012611

# REPROCESSED DYONICS ARTHROSCOPIC SHAVERS (K012611)

_Alliance Medical, Inc. · HRX · Nov 7, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012611

## Device Facts

- **Applicant:** Alliance Medical, Inc.
- **Product Code:** [HRX](/productcode/HRX.md)
- **Decision Date:** Nov 7, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Reprocessed Arthroscopic Shavers are intended for resecting tissue and bone found in articular body cavities during orthopedic, maxillofacial, hand, foot and plastic surgery in patients requiring arthroscopic or orthopedic surgery

## Device Story

Reprocessed arthroscopic shavers consist of a bur or blade at the end of a long rod rotating within a hollow stainless steel housing. The housing features a distal window allowing the bur to cut target structures while protecting adjacent tissue. The device attaches to a motorized handpiece that drives the internal blade and provides suction to remove cut tissue from the surgical site. Used in OR settings by surgeons for arthroscopic, orthopedic, maxillofacial, hand, foot, and plastic surgeries. The device facilitates tissue/bone resection, loose fragment removal, and debris shaving. Reprocessing restores the device to original performance specifications.

## Clinical Evidence

Bench testing only. Performance was demonstrated through biocompatibility validation and functional testing to ensure the reprocessed devices meet original specifications.

## Technological Characteristics

Stainless steel housing with internal rotating bur/blade. Mechanical actuation via motorized handpiece. Suction-assisted tissue removal. Design, materials, and dimensions are identical to original predicate devices.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Smith & Nephew Dyonics® Disposable Arthorscopy Blade ([K833587](/device/K833587.md))
- Smith & Nephew Dyonics® Arthroscopic Surgical Blade ([K904284](/device/K904284.md))
- Smith & Nephew Dyonics® Disposable Arthroscopic Surgery Blades ([K934229](/device/K934229.md))
- Smith & Nephew Dyonics® EP-1 Shaver ([K953096](/device/K953096.md))
- Smith & Nephew Dyonics® Disposable Arthroscopic Blades ([K953695](/device/K953695.md))
- Smith & Nephew Dyonics® Intra-Articular Shaver ([K771218](/device/K771218.md))
- Smith & Nephew Dyonics® Intra-Articular Surgical System ([K820367](/device/K820367.md))
- Smith & Nephew Dyonics® Endoscopic Surgery Blades ([K971253](/device/K971253.md))
- Smith & Nephew Dyonics® Modified Uses of the Arthroscopic Surgical System, Trimmer Blade and Full Radius Blade ([K900070](/device/K900070.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 0 7 2001

Submitter:

Image /page/0/Picture/1 description: The image shows the logo for Alliance Medical Corporation. The logo consists of three curved lines above the company name. The lines are thick and black, and they are arranged in a way that suggests movement or progress. The text "ALLIANCE MEDICAL CORPORATION" is written in a simple, sans-serif font.

K012611

## 510(k) Summary of Safety and Effectiveness

10232 South 51st Street Phoenix, Arizona 85044

TEL 480.763.5300
FAX 480.763.5310
Toll Free 888.888.3433
Toll Free 888.888.3433 -medical.com

| Submitter:           | 10232 South 51st Street<br>Phoenix, Arizona 85044                                        |  |
|----------------------|------------------------------------------------------------------------------------------|--|
| Contact:             | Don Selvey<br>Vice President, Regulatory Affairs and Quality Assurance<br>(480) 763-5300 |  |
| Date of preparation: | 10 August 2001                                                                           |  |
| Name of device:      | Reprocessed Arthroscopic Shavers                                                         |  |
| Common Name:         | Arthroscopic Shaver                                                                      |  |
| Classification Name: | Arthroscope                                                                              |  |

Alliance Medical Corporation

### Reprocessed device(s):

| Manufacturer | Description   | Model   |
|--------------|---------------|---------|
| Dyonics®     | Cutter        | 3439    |
| Dyonics®     | Cutter        | 3440    |
| Dyonics®     | Trimmer       | 3441    |
| Dyonics®     | Full Radius   | 3442    |
| Dyonics®     | Full Radius   | 3443    |
| Dyonics®     | Full Radius   | 3444    |
| Dyonics®     | Turbo Whisker | 3446    |
| Dyonics®     | Turbo Trimmer | 3529    |
| Dyonics®     | Incisor       | 3810    |
| Dyonics®     | Synovator     | 3826    |
| Dyonics®     | Synovator     | 4190    |
| Dyonics®     | Incisor       | 4191    |
| Dyonics®     | Razor Cut     | 4222    |
| Dyonics®     | Incisor       | 4223    |
| Dyonics®     | Razor Cut     | 4280    |
| Dyonics®     | Razor Cut     | 4284    |
| Dyonics®     | Full Radius   | 7205305 |
| Dyonics®     | Full Radius   | 7205306 |
| Dyonics®     | Full Radius   | 7205307 |
| Dyonics®     | Cutter        | 7205308 |
| Dyonics®     | Cutter        | 7205309 |
| Dyonics®     | Synovator     | 7205310 |
| Dyonics®     | Synovator     | 7205311 |
| Dyonics®     | Incisor       | 7205312 |
| Dyonics®     | Incisor       | 7205313 |
| Dyonics®     | Incisor       | 7205314 |

The starter of the

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows three curved lines that are stacked on top of each other. The lines are thick and black, and they appear to be slightly blurred. The lines are arranged in a way that creates a sense of movement or flow. The background of the image is white.

# A L L I A N C E
MEDICAL CORPORATION

| Dyonics® | Turbo Trimmer | 7205315 |
|----------|---------------|---------|
| Dyonics® | Turbo Whisker | 7205316 |
| Dyonics® | Razorcut      | 7205317 |
| Dyonics® | Razorcut      | 7205318 |
| Dyonics® | Razorcut      | 7205319 |
| Dyonics® | Incisor       | 7205345 |
| Dyonics® | Incisor       | 7205459 |

| Predicate device(s): | K833587                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Disposable Arthorscopy Blade                                                           |  |  |  |  |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--|--|--|--|
|                      | K904284                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Arthroscopic Surgical Blade                                                            |  |  |  |  |
|                      | K934229                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Disposable Arthroscopic Surgery Blades                                                 |  |  |  |  |
|                      | K953096                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® EP-1 Shaver                                                                            |  |  |  |  |
|                      | K953695                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Disposable Arthroscopic Blades                                                         |  |  |  |  |
|                      | K771218                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Intra-Articular Shaver                                                                 |  |  |  |  |
|                      | K820367                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Intra-Articular Surgical System                                                        |  |  |  |  |
|                      | K971253                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Endoscopic Surgery Blades                                                              |  |  |  |  |
|                      | K900070                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Smith & Nephew Dyonics® Modified Uses of the Arthroscopic Surgical System, Trimmer Blade and Full Radius Blade |  |  |  |  |
| Device description:  | Arthroscopic shavers can be used to abrade, cut and excise tissue and bone; remove loose fragments; and shave away debris in arthroscopic surgeries, as well as surgeries of the jaw and sinuses.<br><br>The arthroscopic shaver components reprocessed by Alliance Medical Corporation include a bur or blade at the end of a long rod that rotates within a long hollow stainless steel housing. The housing has a window cut out on one side of the distal end, allowing the bur to cut one structure while the adjacent one is still protected by the housing on the opposite side of the bur or blade. This system attaches to a motorized handpiece that drives the internal bur or blade inside the outer housing and provides suction to pull the cut tissue away from the surgical site. |                                                                                                                |  |  |  |  |
| Intended use:        | Reprocessed Arthroscopic Shavers are intended for resecting tissue and bone found in articular body cavities during orthopedic, maxillofacial, hand, foot and plastic surgery in patients requiring arthroscopic or orthopedic surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                |  |  |  |  |

.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with three curved lines above the word "ALLIANCE". The three lines are arranged in a parallel fashion, with each line curving upwards and to the right, resembling a stylized representation of motion or energy. The word "ALLIANCE" is written in a bold, sans-serif font, positioned directly below the curved lines, suggesting that the lines are a symbol associated with the organization or entity named "ALLIANCE".

#### ALLIANCE MEDICAL CORPORATION

.

| Indications<br>statement:         | Reprocessed arthroscopic shavers are indicated for use in<br>orthopedic surgical procedures of the joints, jaw or sinuses where<br>the cutting and removal of soft and hard tissue or bone is needed in<br>patients requiring orthopedic surgery.                                                                                                                                                                                                                                         |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics: | The design, materials, and intended use of the Reprocessed<br>Arthroscopic Shavers are identical to the predicate devices. The<br>mechanism of action of the Reprocessed Arthroscopic Shaver is<br>identical to the predicate devices in that the same standard<br>mechanical design, materials, shapes and sizes are utilized. There<br>are no changes to the claims, intended use, clinical applications,<br>patient population, performance specifications, or method of<br>operation. |
| Performance<br>data:              | Bench and laboratory testing was conducted to demonstrate<br>performance (safety and effectiveness) of the Reprocessed<br>Arthroscopic Shavers.<br><br>Biocompatibility Validation of reprocessing Function test(s)<br>Performance testing demonstrates that Reprocessed Arthroscopic<br>Shavers perform as originally intended.                                                                                                                                                          |
| Conclusion:                       | In accordance with the Federal Food, Drug and Cosmetic Act 21<br>CFR Part 807 and based on the information provided in this<br>premarket notification, Alliance Medical Corporation concludes<br>that the modified device (the Reprocessed Arthroscopic Shaver) is<br>safe, effective and substantially equivalent to the predicate devices<br>as described herein.                                                                                                                       |

:

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing segments.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 0 7 2001

Mr. Don Selvey Vice President, Regulatory Affairs and Quality Assurance Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044

Re: K012611

Trade/Device Name: Alliance Medical Reprocessed Dyonics® Arthroscopic Shavers Regulation Number: 888.1100 Regulation Name: Arthroscope and accessories Regulatory Class: II Product Code: HRX Dated: August 10, 2001 Received: August 13, 2001

Dear Mr. Selvey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Don Selvey

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Welker, us

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## NOV 0 7 2001

## INDICATIONS FOR USE

K DI2611 510(k) Number:

Device Name: Alliance Medical Corporation Reprocessed Arthroscopic Shavers

Indications for Use: The reprocessed arthroscopic shavers are indicated for use in orthopedic surgical procedures of the joints, jaw, or sinuses where the cutting and removal of soft and hard tissue or bone is needed in patients requiring orthopedic surgery.

| Manufacturer | Description   | Model   | Manufacturer | Description     |
|--------------|---------------|---------|--------------|-----------------|
| Dyonics      | Cutter        | 3439    | Dyonics      | Full Radius     |
| Dyonics      | Cutter        | 3440    | Dyonics      | Full Radius     |
| Dyonics      | Trimmer       | 3441    | Dyonics      | Cutter          |
| Dyonics      | Full Radius   | 3442    | Dyonics      | Cutter          |
| Dyonics      | Full Radius   | 3443    | Dyonics      | Synovator       |
| Dyonics      | Full Radius   | 3444    | Dyonics      | Synovator       |
| Dyonics      | Turbo Whisker | 3446    | Dyonics      | Incisor         |
| Dyonics      | Turbo Trimmer | 3529    | Dyonics      | Incisor         |
| Dyonics      | Incisor       | 3810    | Dyonics      | Incisor         |
| Dyonics      | Synovator     | 3826    | Dyonics      | Turbo Trimmer   |
| Dyonics      | Synovator     | 4190    | Dyonics      | Turbo Whisker   |
| Dyonics      | Incisor       | 4191    | Dyonics      | Razorcut        |
| Dyonics      | Razor Cut     | 4222    | Dyonics      | Razorcut        |
| Dyonics      | Incisor       | 4223    | Dyonics      | Razorcut        |
| Dyonics      | Razor Cut     | 4280    | Dyonics      | Orbit Synovator |
| Dyonics      | Razor Cut     | 4284    | Dyonics      | Incisor         |
| Dyonics      | Full Radius   | 7205305 | Dyonics      | Incisor         |

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use
Lisa Wal

(Division Sign-Off) Division of General, Restorative and Neurological Jevices

510(k) Number K612614

Model

7205306

7205311

7205316

7205317 7205318 7205319

7205321 7205345 7205459

---

**Source:** [https://fda-staging.innolitics.com/device/K012611](https://fda-staging.innolitics.com/device/K012611)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com