← Product Code [JOW](/productcode/JOW) · K012606

# CLEARMEDICAL/KENDALL KNEE-HIGH SCD SLEEVES, CLEARMEDICAL/KENDALL THIGH-HIGH SCD SLEEVES, MEDIUM,SMALL, LARGE (K012606)

_Clearmedical, Inc. · JOW · Feb 6, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012606

## Device Facts

- **Applicant:** Clearmedical, Inc.
- **Product Code:** [JOW](/productcode/JOW.md)
- **Decision Date:** Feb 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ClearMedical/Kendall SCDs are intended as a single patient use accessory to a Kendall Controller System. The role of the SCD is to reduce the incidence of deep vein thrombosis, reduce pain and swelling after injury and surgery, and increase arterial blood flow. The SCD Sleeves are used in conjunction with the Controller System for non-ambulatory patients in a hospital environment.

## Device Story

ClearMedical/Kendall SCD Sleeves are single-patient-use accessories for Kendall Controller Systems; used for non-ambulatory hospital patients. Sleeves feature inflatable bladders made of polyester-cotton fabric, velcro straps, and polyethylene tubing with polypropylene snap-lock connectors. Device connects to a controller that inflates bladders to provide sequential limb compression. Mechanism promotes blood flow, reduces swelling, and prevents deep vein thrombosis/pulmonary embolism. Reprocessed devices are high-level disinfected to meet CDC/APIC standards.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance evaluation focused on infection control methods, fit/attachment, bladder function, velcro adhesion, and connector function to demonstrate substantial equivalence to predicate.

## Technological Characteristics

Materials: polyester-cotton fabric, polyethylene tubing, polypropylene snap-lock connector. Sensing/Actuation: pneumatic bladder inflation via external controller. Form factor: limb sleeve with velcro straps. Connectivity: mechanical connection to controller system. Sterilization: high-level disinfection (reprocessed).

## Regulatory Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

## Predicate Devices

- Kendall SCD Sleeves (Parts 5329, 5329W, 5330, 5330W, 5345, 5345W, 5480, and 5480W)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K02606

Image /page/0/Picture/1 description: The image shows a logo with a water droplet inside a diamond shape. To the right of the logo, the text "FEB 6 2002" is printed in a bold, sans-serif font. The text likely represents a date, with "FEB" indicating February, "6" indicating the day, and "2002" indicating the year.

510(k) Summary

Image /page/0/Picture/3 description: The image shows the word "ClearMed" in a bold, sans-serif font. The letters are black against a white background. A small trademark symbol is located to the right of the word.

ClearMedical, Inc. Submitter's name and Address: 1776 - 13610 Place NE Bellevue, WA 98005 Ph (425) 401-1414 Fax (425) 401-1515 3017110 FDA Registration Number: Richard Radford Contact Person: Director of Research and Product Development August 8, 2001 Date Summary Prepared:

Trade or Proprietary Name:

Common Name:

Classification Name:

Sleeve, Limb, Compressible (per 21 CFR section 870.5800)/JOW

Sequential Compression Sleeves

ClearMedical/Kendall Sequential Compression

## Equivalent Device

The ClearMedical/Kendall reprocessed Sequential Compression Device (SCD) Sleeves are substantially equivalent to the Kendall SCD Sleeves, Parts 5329, 5329W. 5330, 5330W. 5345, 5345W. 5480, and 5480W. This determination has been reached based on an evaluation and analysis of the predicate device's technical and promotional labeling and specific bench testing. For all established indicators of substantial equivalence, the ClearMedical devices demonstrated equality in safety and performance.

Device (SCD) Sleeves

#### Device Description:

The ClearMedical/Kendall SCDs are an accessory device to a SCD System which is used for the prevention of deep vein thrombosis and pulmonary embolism. The SCD Sleeve is designed with inflatable bladders that fill with air to

3

{1}------------------------------------------------

### 510(k) Summary (Cont'd)

provide compression to the legs. A connector tubing system attaches the SCD to the Controller svstem.

The SCD Sleeves are latex-free, made of a polyester-cotton fabric. They are designed with velcro hook straps, three bladders and molded polyethylene tubing with a polypropylene snap-lock connector. Kendall has recently changed the device from the color blue to white. ClearMedical will designate the white SCDs with a "w" after the part number (i.e. 5330W).

#### Intended Use:

The ClearMedical/Kendall SCDs are intended as a single patient use accessory to a Kendall Controller System. The role of the SCD is to reduce the incidence of deep vein thrombosis, reduce pain and swelling after injurv and surgery, and increase arterial blood flow. The SCD Sleeves are used in conjunction with the Controller System for non-ambulatory patients in a hospital environment.

#### Technological Characteristics of ClearMedical and Predicate Devices:

The predicate device and the ClearMedical/Kendall SCDs contain bladders that inflate with air to control circulation of a limb. Attached to the Sleeve is a connector tubing system that connects to the Controller System. In form, the predicate device and the ClearMedical reprocessed SCD are the same.

Technological indicators of substantial equivalence were identified and included methods of infection control, fit/attachment, bladder function, velcro adhesion and connector function.

The predicate device is delivered to the customer labeled 'non-sterile' whereas the ClearMedical/Kendall SCDs are delivered to the customer labeled "High Level-Disinfected". ClearMedical's infection control methods meet or exceed the CDC standards ("Guideline for Handwashing and Hospital Environmental Control", 1985, "APIC Guideline for Selection and Use of Disinfectants") for this class of device.

#### Summary of ClearMedical/Kendall SCD Performance:

Based on an assessment of bench tests and non-clinical performance data, we believe that in all relevant safety and performance indicators the ClearMedical/ Kendall SCD Sleeves demonstrates substantial equivalence to the predicate devices, the Kendall SCD Sleeves.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# FEB 0 6 2002

Mr. Richard Radford Director of Research and Product Development ClearMedical, Inc. 1776 136th Place NE Belluvue, WA 98005-2328

Re: K012606

Trade Name: ClearMedical/Kendall Sequential Compression Device (SCD) Sleeves Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: January 16, 2002 Received: January 17, 2002

Dear Mr. Radford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your deyice can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Richard Radford

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reaso be actived that I Dr usean that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter will and will you to FDA finding of substantial equivalence of your device to a legally premation noticate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you docure bpoints art 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Oiner general mironmation the mational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

cj

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Statement of Indications for Use

| 510(K) NUMBER (IF KNOWN): | K012606                                                   |
|---------------------------|-----------------------------------------------------------|
| DEVICE NAME:              | ClearMedical/Kendall Sequential Compression Devices (SCD) |

Part Numbers: 5329, 5330, 5345, 5480, 5329W, 5330W, 5345W, 5480W

# INDICATIONS FOR USE:

To reduce the incidence of deep vein thrombosis, reduce pain and swelling 10 rought or surgery and increase arterial blood flow.

(Please do not write below this line-continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number K012606

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter-Use _

---

**Source:** [https://fda-staging.innolitics.com/device/K012606](https://fda-staging.innolitics.com/device/K012606)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).