K012598 · Sterilmed, Inc. · NUJ · Nov 7, 2001 · General, Plastic Surgery
Device Facts
Record ID
K012598
Device Name
REPROCESSED LAPAROSCOPIC ELECTRIC INSTUMENTS
Applicant
Sterilmed, Inc.
Product Code
NUJ · General, Plastic Surgery
Decision Date
Nov 7, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Laparoscopic Electric Instruments are designed for use in minimally invasive procedures and open surgical procedures to facilitate coagulation, transection, mobilization, and dissection of tissue.
Device Story
Reprocessed laparoscopic electric instruments consist of handle, rotating insulated shaft (5mm diameter, 19-33cm length), and distal tip (dissectors, scissors, or cutting forceps). Devices are monopolar or bipolar, featuring cautery connectors or cables. Operated by surgeons in OR settings during minimally invasive or open procedures. Device connects to RF generator; electrical current delivered to distal tip for tissue coagulation, transection, mobilization, and dissection. Reprocessing restores single-use devices to original functional state via validated cleaning and sterilization processes. Benefits include cost-effective access to surgical instrumentation.
Clinical Evidence
Bench testing only. Representative samples underwent functional testing to demonstrate appropriate characteristics. Process validation performed for cleaning, sterilization, and packaging. Manufacturing process includes 100% visual and functional testing.
Technological Characteristics
Monopolar or bipolar electrosurgical instruments. Components: handle, rotating insulated shaft, distal tip (dissectors, scissors, cutting forceps). Energy source: external RF generator. Connectivity: cautery connector or cable. Reprocessing involves validated cleaning and sterilization methods.
Indications for Use
Indicated for use in minimally invasive and open surgical procedures for tissue coagulation, transection, mobilization, and dissection.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The symbol is composed of three interconnected human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 3 1 2007
SterilMed, Inc. % Mr. Dennis Toussaint Director, Regulatory Affairs 11400 73rd Avenue North Maple Grove, Minnesota 55369
Re: K012598 - Supplemental Validation Submission Trade/Device Name: Reprocessed Electric Instruments (Enclosed List) Regulation Number: 21CFR878.4400 Regulation Name: Electrosurgical Device, Cutting & Coagulation & Accessories Regulatory Class: II Product Code: NUJ Dated: August 8, 2001 Received: August 10, 2001
Dear Mr. Toussaint:
The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on October 23, 2000. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if
{1}------------------------------------------------
Page 2 - Mr. Dennis Toussaint
applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Mark A. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure Intended Use form
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for SterilMed, INC. The logo is in a bold, sans-serif font. Below the company name, it says "Medical Device Reprocessing" and "Small Equipment & Instrument Repair". The text is black and the background is white.
SVS For 510(K): K012598 Reprocessed Laparoscopic Electric Instruments
## INDICATIONS FOR USE PAGE (K012598)
(Same as currently cleared 510(k))
Device Name: Reprocessed Laparoscopic Electric Instruments
Indications for Use: Laparoscopic Electric Instruments are designed for use in minimally invasive procedures and open surgical procedures to facilitate coagulation, transection, mobilization, and dissection of tissue.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Melkerson
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K012598
{3}------------------------------------------------
# List of thirteen (13) Reprocessed Laparoscopic Electric Instruments
of the SVS for K012598
| Manufacturer | Model # | Description |
|-----------------|---------|----------------------------------------|
| AutoSuture | 174301 | Endo Mini-Shears, 5mm, Monopolar, 31cm |
| | 174501 | EndoShears, 5mm, Monopolar, 19cm |
| | 174503 | Endo Mini-Shears, 5mm, Monopolar, 19cm |
| | 174505 | Endo Dissect, 5mm Monopolar, 19cm |
| | 176605 | EndoSciz, 5mm, Monopolar, 31cm |
| | 176643 | EndoShears, 5mm, Monopolar, 31cm |
| | 176645 | EndoDissect, 5mm, Monopolar, 31cm |
| Ethicon | 5DCS | Scissors, 5mm, Monopolar, 31cm |
| | 5DCD | Dissector, 5mm, Monopolar, 31 cm |
| Everest (Gyrus) | 3005 | Cutting Forceps, 5mm, Bipolar, 33cm |
| | 3025 | Cutting Forceps, 5mm, Bipolar, 24cm |
| Gyrus | 3005PK | Cutting Forceps, 5mm, Bipolar, 33cm |
| | 3025PK | Cutting Forceps, 5mm, Bipolar, 24cm |
:
:
{4}------------------------------------------------
### SUPPLEMENTAL VALIDATION SUBMISSION JUNE 27, 2006
For 510(K): K012598 Reprocessed Laparoscopic Electric Instruments
SVS for K012598, Reprocessed Lap. Ele. Instruments
II. SUMMARY AND CERTIFICATION
A. 510(k) Summary
| <strong>Labels</strong> | <strong>Values</strong> |
|-------------------------|-----------------------------------------------|
| Submitter: | SterilMed, Inc. |
| Contact Person: | Dennis Toussaint |
| | 11400 73rd Avenue North |
| | Maple Grove, MN 55369 |
| | Ph: 888-856-4870 |
| | Fax: 763-488-3350 |
| Date Prepared: | June 28, 2006 |
| Trade Name: | Reprocessed Laparoscopic Electric Instruments |
| Classification Name: | Endoscopic Electrosurgical Accessory |
| Classification Number: | Class II, 21 CFR 878.4400 |
| Product Code: | NUJ |
Predicate
Devices:
Device
Description:
Intended Use:
Functional and
Safety Testing:
Page 1 of 2
The reprocessed laparoscopic electric instruments are substantially
equivalent to BiCoag Forceps (K945975), manufactured by Everest;
Endopath Endoscopic Instruments (K984240) manufactured by
Ethicon; and counterpart devices from other original equipment
manufacturers.
Laparoscopic electric instruments are devices that are designed for use
via laparoscopes or open surgical procedures. The devices have a
handle, a rotating insulated shaft with a diameter of 5mm, a length of
19-33 cm and a distal tip. The distal tip of the devices consists of a
variety of configurations including: dissectors, scissors and cutting
forceps. The devices may be monopolar or bipolar and have a cautery
connector on the handle or a connector cable. The device is connected
via a cautery cable to a RF generator which provides electrical current
to the device.
Laparoscopic electric instruments are designed for use in minimally
invasive procedures and/or open surgical procedures to facilitate
coagulation, transaction, resection, mobilization and dissection of
tissue.
Representative samples of reprocessed laparoscopic electric
instruments underwent functional testing to demonstrate appropriate
functional characteristics. Process validation testing was done to
validate the cleaning and sterilization procedures as well as device
packaging. In addition, the manufacturing process includes visual and
functional testing of all products produced.
{5}------------------------------------------------
**STERILMED, INC.**
Medical Device Reprocessing
Small Equipment & Instrument Repair
## SUPPLEMENTAL VALIDATIC SUBMIS JUNE 27, 2006
For 510(K): K012598 Reprocessed Laparoscopic Electric Instruments
Conclusion:
The reprocessed laparoscopic electric instruments are substantially equivalent to BiCoag Forceps (K945975), and Endopath Endoscopic Instruments (K984240), manufactured by Ethicon; and counterpart devices from other original equipment manufacturers.
This conclusion is based upon the devices' similarities in functional design, materials, indications for use and methods of construction.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.