For in vitro diagnostic use only. The VITROS Magnetic HDL-Cholesterol Reagent and VITROS CHOL Slides quantitatively measure HDL cholesterol (HDLC) concentration in serum and plasma. VITROS Calibrator Kit 2 For in vitro diagnostic use only. VITROS Calibrator Kit 2 is intended for use in calibration of the VITROS Chemistry Systems for the quantitative measurement of CHOL, Cl-, ECO2, HDLC, K+, Na+, and TRIG.
Device Story
Device consists of Magnetic HDL-Cholesterol Reagent, VITROS CHOL slides, and Calibrator Kit 2. Used on VITROS 250, 550, 750, and 950 Series Analyzers. Principle: HDL separated by precipitation of LDL/VLDL using dextran sulfate and magnesium chloride; polymer-coated iron particles enhance precipitation; magnetic field removes non-HDL lipoproteins. Remaining HDL cholesterol measured via dry, multilayered slide using reflectance spectrophotometry. Operated by laboratory personnel in clinical settings. Output provides quantitative HDL-cholesterol concentration, aiding clinicians in assessing coronary heart disease risk.
Clinical Evidence
Bench testing only. Performance was validated using patient and quality control samples spanning the assay range (3.0-100.0 mg/dL) to demonstrate substantial equivalence to the predicate device.
Technological Characteristics
Dry, multilayered slide technology; reflectance spectrophotometry. Reagents include dextran sulfate (50,000 MW), magnesium chloride, and polymer-coated magnetic particles. Instrumentation: VITROS 250, 550, 750, 950 Series Analyzers. Incubation: 5 minutes at 37°C.
Indications for Use
Indicated for the quantitative measurement of HDL-cholesterol in human serum and plasma to evaluate the risk of coronary heart disease (CHD).
Regulatory Classification
Identification
A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
Predicate Devices
VITROS Magnetic HDL-Cholesterol Reagent and VITROS CHOL Slides (K984303)
Submission Summary (Full Text)
{0}------------------------------------------------
## Summary Information
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: 012593
- Ortho-Clinical Diagnostics, Inc. 1. Submitter name, 100 Indigo Creek Drive address. Rochester, New York 14626-5101 contact (716) 453-4041
Marlene A. Shulman Contact Person:
- Date Special 510(k) prepared: August 9, 2001 2. Preparation date
- 3. Device Trade or Proprietary Name: VITROS Chemistry Products Magnetic HDL-Cholesterol Reagent name VITROS CHOL Slide VITROS Chemistry Products Calibrator Kit 2 Common Name: HDLC Precipitating Reagent Classification Name: LDL & VLDL Precipitation, HDL
- The VITROS Chemistry Products Magnetic HDL-Cholesterol Reagent and 4. Predicate VITROS CHOL Slides (modified) and VITROS Chemistry Products device Calibrator Kit 2 are substantially equivalent to the VITROS Chemistry Products Magnetic HDL-Cholesterol Reagent and VITROS CHOL Slides (current slide) and VITROS Chemistry Products Calibrator Kit 2.
Continued on next page
3
{1}------------------------------------------------
#### 510(k) Summary, Continued
- The VITROS Chemistry Products Magnetic HDL-Cholesterol Reagent and 7. Comparison VITROS CHOL Slides (modified) and VITROS Chemistry Products to Predicate VITROS CHOE Shaos (instally equivalent to VITROS Chemistry Products Device Canorator Rit 2 are Sabbaniano, expent and VITROS CHOL Slides for use with Magnetic HDL-Cholesteror Reagues cleared by the FDA (K984303 January 28, 1999) for in vitro diagnostic use.
Table 1 lists the characteristics of the tests performed using the VITROS Fable I hists the entarasterol Reagent and VITROS CHOL Slide (current) and Magnetic HDL Cholesterol Reagent and VITROS CHOL Slides (modified)
| Table I List of Device Characteristics. Comparison to | | | |
|--------------------------------------------------------|-----------------------------------------------------|-------------------------------------------------|--|
| Device | Predicate Device | New Device | |
| Characteristic | VITROS Magnetic HDL-Cholesterol | VITROS Magnetic HDL-Cholesterol | |
| | Reagent and VITROS CHOL Slide | Reagent and VITROS CHOL Slide | |
| | (Current) | (Modified) | |
| Sample volume | 10 uL | 5.5 µL | |
| Quantity of | Triton X-100 0.91 mg; cholesterol oxidase | Triton X-100 0.73 mg; cholesterol oxidase | |
| Reactive | (Nocardia or Cellulomonas, E.C.1.1.3.6) 0.45 U; | (Nocardia or Cellulomonas, E.C.1.1.3.6) 0.36 U; | |
| Ingredients per | cholesterol ester hydrolase (Candida rugosa or | cholesterol ester hydrolase (Candida rugosa or | |
| CHOL slide (test) | Pseudomonas, E.C.3.1.1.13) 2.25 U; peroxidase | Pseudomonas, E.C.3.1.1.13) 1.80 U; peroxidase | |
| | (horseradish root, E.C.1.11.1.7) 5.85 U; and 2- | (horseradish root, E.C.1.11.1.7) 4.68 U; and 2- | |
| | (3,5-dimethoxy-4-hydroxyphenyl)-4,5-bis(4- | (3,5-dimethoxy-4-hydroxyphenyl)-4,5-bis(4- | |
| | dimethylaminophenyl) imidazole (leuco dye) 0.25 | dimethylaminophenyl) imidazole (leuco dye) 0.13 | |
| | mg. | mg. | |
| Intended Use | For in vitro diagnostic use only. | No Change. | |
| | The VITROS Magnetic HDL reagent and | | |
| | VITROS Cholesterol slides quantitatively measure | | |
| | HDL-cholesterol (HDLC) in serum and plasma. | | |
| Concentrations of | Triton X-100 0.81 mg; cholesterol oxidase | No Change. | |
| CHOL Reactive | (Nocardia or Cellulomonas, E.C.1.1.3.6) 0.4 U; | | |
| Ingredients per | cholesterol ester hydrolase (Candida rugosa or | | |
| cm-squared | Pseudomonas, E.C.3.1.1.13) 2.0 U; peroxidase | | |
| | (horseradish root, E.C.1.11.1.7) 5.2 U; and 2-(3,5- | | |
| | dimethoxy-4-hydroxyphenyl)-4,5-bis(4- | | |
| | dimethylaminophenyl) imidazole (leuco dye) 0.17 | | |
| | mg. | | |
| Magnetic HDL- | Dextran sulfate (50,000 MW); Magnesium | No Change. | |
| Cholesterol | Chloride, Magnetically responsive particles. | | |
| Reagent Reactive | | | |
| Ingredients. | | | |
#### vice Characteristics: Comparison to Predicate Device
{2}------------------------------------------------
## 510(k) Summary, Continued
| Device<br>Characteristic | Predicate Device<br>VITROS Magnetic HDL-Cholesterol<br>Reagent and VITROS CHOL Slide<br>(Current) | New Device<br>VITROS Magnetic HDL-Cholesterol<br>Reagent and VITROS CHOL Slide<br>(Modified) |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Basic principle | HDL is separated by precipitation of LDL and<br>VLDL by using dextran sulfate (MW 50,000) and<br>magnesium chloride. The reagent also contains<br>iron particles that are coated with a polymer. The<br>polymer enhances the precipitation of the non-<br>HDL lipoproteins onto the iron particles. The<br>precipitated non-HDL lipoproteins are removed<br>from the supernate by applying a magnetic field.<br>Following pretreatment of the sample, HDL<br>cholesterol is measured by a dry, multilayered<br>slide utilizing reflectance spectrophotometry | No Change. |
| Sample type | Serum , plasma | No Change. |
| Assay Range<br>Serum, Plasma | 3.0- 100.0 mg/dL | No Change. |
| Instrumentation | VITROS 250, 550, 750 and 950 Series Analyzers | No Change. |
| Incubation time<br>and temperature | 5 minutes at 37°C | No Change. |
#### 8. Conclusions
The information presented in this pre-market notification demonstrates that The information probential with on of the VITROS HDL-Cholesterol Reagent and CHOL the performance or ar use with human serum and plasma is substantially equivalent to the cleared predicate device.
Equivalence was demonstrated using magnetic HDL-Cholesterol Reagent and manufactured slides along with patient and quality control samples with measured HDL-cholesterol values spanning the assay range.
The information presented in this premarket notification provide a reasonable The miorination prosition Magnetic HDL-Cholesterol Reagent and CHOL assuration that the for use with human serum and plasma is safe and effective for the stated intended use.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the head, body, and tail feathers of the bird.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# AUG 3 0 2001
Ms. Marlene A. Shulman Regulatory Affairs Associate Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, NY 14626-5101
K012593 Re:
K012393
Trade/Device Name: VITROS Chemistry Products Magnetic HDL-Cholesterol Reagent, VITROS CHOL Slide, and VITROS Chemistry Products Calibrator Kit 2
Regulation Number: 21 CFR 862.1150
Regulatory Class: II Product Code: JIS Regulation Number: 21 CFR 862.1475 Regulatory Class: I, reserved Product Code: LBR Dated: August 9, 2001 Received: August 10, 2001
Dear Mr. Shulman:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have reviewed your boourer b r (=) ce (= ) = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Americans, or to devices and into t (Act). You may, therefore, market the device, subject to I coord i ood, Drug, and Goodions of the Act. The general controls provisions of the Act include the general controls pro About of usting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 Children Approval), it the Code of Federal Regulations, Title 21, Parts 800 to 895. allecting your determination assumes compliance with the Good Manufacturing A Substantally equivalent avoilable (GMP) regulation (21 CFR Part 820) and that, through I ractive for Mourcal toos, the Food and Drug Administration (FDA) will verify such perfour offirms, Failure to comply with the GMP regulation may result in regulatory action. In assumptions: That are to centry ther announcements concerning your device in the Federal adultion, I Dri may paches response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the ally Controll Journageation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket This lotter will and in your distantial equivalence of your device to a legally marketed nouthoute device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 11 you desire books an in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (201) 594-1500. I radiation f Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small monfacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Statement of Intended Use
Page 1 of 1
| 510(k) Number (if known): | K012593 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | VITROS Chemistry Products Magnetic HDL-Cholesterol<br>Reagent<br>VITROS CHOL Slide<br>VITROS Chemistry Products Calibrator Kit 2 |
| Intended Use: | For in vitro diagnostic use only.<br>The VITROS Magnetic HDL-Cholesterol Reagent and VITROS<br>CHOL Slides quantitatively measure HDL cholesterol (HDLC)<br>concentration in serum and plasma. |
| | VITROS Calibrator Kit 2 |
| | For in vitro diagnostic use only.<br>VITROS Calibrator Kit 2 is intended for use in calibration of the<br>VITROS Chemistry Systems for the quantitative measurement<br>of CHOL, Cl-, ECO2, HDLC, K+, Na+, and TRIG. |
| Summary and Explanation of<br>Test: | HDL cholesterol is used to evaluate the risk of developing<br>coronary heart disease (CHD). The risk of CHD increases with<br>lower HDL cholesterol concentrations. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use __
Keria Alexander for Joan Cooper
(Division Sign-Off)
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K012593
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.