K012469 · Biomet Manufacturing, Inc. · HWC · Dec 3, 2001 · Orthopedic
Device Facts
Record ID
K012469
Device Name
RESORBABLE NO PROFILE SCREW AND WASHER
Applicant
Biomet Manufacturing, Inc.
Product Code
HWC · Orthopedic
Decision Date
Dec 3, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Resorbable No Profile Screw and Washer is indicated for the following procedures: 1. ACL and PCL reconstruction 2. Medial collateral ligament repair 3. Lateral collateral ligament repair 4. Posterior oblique ligament repair 5. Iliotibial band tenodesis reconstruction 6. Patellar ligament and tendon repair This device is also intended to be used as back-up fixation in ACL reconstruction in conjunction with other marketed devices in order to provide additional fixation strength in instances of questionable bone quality.
Device Story
Resorbable No Profile Screw and Washer provides mechanical fixation for ligament and tendon repairs. Device consists of 6.5 mm diameter PLLA/PGA copolymer screw (25-55 mm lengths) and 18 mm diameter washer. Used by surgeons in orthopedic procedures; functions as primary or back-up fixation in ACL reconstruction to augment fixation strength in poor bone quality. Device is implanted during surgery; provides stability to soft tissue grafts during healing process. Resorbable material eliminates need for secondary removal surgery.
Clinical Evidence
No clinical data. Substantial equivalence established via bench mechanical testing.
Technological Characteristics
Material: PLLA/PGA copolymer. Dimensions: 6.5 mm diameter screw (25-55 mm length), 18 mm diameter washer. Principle: Mechanical bone fixation fastener. Resorbable material composition.
Indications for Use
Indicated for patients undergoing ACL/PCL reconstruction, medial/lateral collateral ligament repair, posterior oblique ligament repair, iliotibial band tenodesis reconstruction, or patellar ligament/tendon repair. Also indicated for back-up fixation in ACL reconstruction where bone quality is questionable.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K01246951
Page 1 of 2
DEC 0 3 2001
Image /page/0/Picture/2 description: The image shows the logo for Biomet INC. The logo is in black and white and features the word "BIOMET" in a bold, sans-serif font. The letters are connected to each other, and the "INC" is stacked vertically to the right of the "T". Below the logo, the words "CORPORATE HEADQUARTERS" are printed in a smaller, serif font.
## SUMMMARY OF SAFETY AND EFFECTIVENESS
| Applicant or Sponsor: | Arthrotek, Inc.<br>(A wholly owned subsidiary of Biomet, Inc.)<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587<br>Establishment Registration No.: 1825034 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Sara B. Shultz<br>Biomet Manufacturing, Corp.<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587<br>Phone: (219) 267-6639<br>FAX: (219) 372-1683 |
| Proprietary Name: | Resorbable No Profile Screw and Washer |
| Common or Usual Name: | resorbable screw and washer |
| Classification Name: | Screw, Fixation, Bone, Non-spinal, Non-metallic<br>(888.3040)<br>Washer, Bolt Nut, Non-spinal, Non-metallic<br>(888.3030) |
## Device Product Code: 87HWC and HTN
Legally Marketed Devices To Which Substantial Equivalence is Claimed: Arthrotek Interference Screw (Biomet, Inc., K982497), Harpoon Suture Anchor (Biomet, Inc., K943806/K973775), Sutureless Anchor (Innovasive Devices, K984490), EndoPearl™ with Threader (Linvatec Corp., K993339)
Indications for Use: The Resorbable No Profile Screw and Washer is indicated for the following procedures:
- 1. ACL and PCL reconstruction
- 2. Medial collateral ligament repair
- 3. Lateral collateral ligament repair
- 4. Posterior oblique ligament repair
- 5. Iliotibial band tenodesis reconstruction
- 6. Patellar ligament and tendon repair
MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587
SHIPPING ADDRESS Airport Industrial Park Warsaw, IN 46580
OFFICE 219.267.6639
FAX ■
Image /page/0/Picture/18 description: The image shows the number 000250, with the last digit being a 1. The numbers are in a bold, sans-serif font. The image is in black and white.
E-MAIL biomet@biomet.com
{1}------------------------------------------------
K012469581
page 2 of 2
Image /page/1/Picture/1 description: The image shows the logo for Biomet Inc. The logo consists of the word "BIOMET" in a bold, sans-serif font, with the letters connected to form a single shape. To the right of the word "BIOMET" is the word "INC" in a smaller font, stacked vertically. Below the logo is the phrase "CORPORATE HEADQUARTERS" in a smaller, serif font.
This device is also intended to be used as back-up fixation in ACL reconstruction in conjunction with other marketed devices in order to provide additional fixation strength in instances of questionable bone quality.
Device Description: The Resorbable No Profile Screw and Washer consists of a resorbable 6.5 mm screw that varies in length from 25 mm to 55 mm (5 mm increments) and a 18 mm diameter washer. This device is comprised of a PLLA/PGA copolymer.
The device was designed to be used in conjunction with marketed devices such as resorbable or allograft interference screws in ACL reconstruction. The purpose of this back-up fixation will be to provide additional fixation strength in instances of questionable bone quality.
Summary of Technologies: The Resorbable No Profile Screw and Washer technological characteristics (materials, design, sizes, and indications) are similar to or identical to the predicate devices.
Non-Clinical Testing: Mechanical testing was performed to establish substantial equivalence to the predicate devices.
Clinical Testing: Clinical testing was not used to establish substantial equivalence to predicate devices.
> MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587
> > 용
SHIPPING ADDRESS Airport Industrial Park Warsaw, IN 46580
()FFICE 219.267.6639 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
E-MAIL biomet@biomet.com
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines above two wavy lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 0 3 2001
Biomet Manufacturing, Corporation c/o Ms. Sara B. Shultz Regulatory Specialist P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K012469
Trade/Device Name: Resorbable No Profile Screw and Washer Regulation Number: 888.3040 and 888.3030 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener and Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: HWC and HTN Dated: October 24, 2001
Received: October 25, 2001
Dear Ms. Shultz:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your bection of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to conninerce provision to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, devices that have boon recuire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicrororo, mains of the Act include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your device is olassined (600 as 10,000) of on the major regulations affecting your device can may or sabject to back as a sateral Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r tease oe actives and i be reviewan that your device complies with other requirements of the Act
{3}------------------------------------------------
## Page 2 - Ms. Sarah B. Shultz
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listingPage (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
b. Mali N. Millison
Celia M. Witten, Ph.D., M.D. Director Divison of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K0124691
Page **_ of _**
## 510(k) NUMBER (IF KNOWN):__KOLA 469
DEVICE NAME: Resorbable No Profile Screw and Washer
INDICATIONS FOR USE:
The Resorbable No Profile Screw and Washer is indicated for the following procedures:
- 1. ACL and PCL reconstruction
- 2. Medial collateral ligament repair
- 3. Lateral collateral ligament repair
- 4. Posterior oblique ligament repair
- 5. Iliotibial band tenodesis reconstruction
- 6. Patellar ligament and tendon repair
This device is also intended to be used as back-up fixation in ACL reconstruction in conjunction with other marketed devices in order to provide additional fixation strength in instances of questionable bone quality.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-96)
(Division Sign-Off
Division of General, Restorative and Neurological Devices
510(k) Number
0000003
K01246
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.