COMBITROL TS MULTI ANALYTE CONTROL, LEVELS 1,2,3, AUTOTROL TS MULTI ANALYTE CONTROL, LEVELS 1,2,3
K012431 · Bionostics, Inc. · JJY · Aug 28, 2001 · Clinical Chemistry
Device Facts
Record ID
K012431
Device Name
COMBITROL TS MULTI ANALYTE CONTROL, LEVELS 1,2,3, AUTOTROL TS MULTI ANALYTE CONTROL, LEVELS 1,2,3
Applicant
Bionostics, Inc.
Product Code
JJY · Clinical Chemistry
Decision Date
Aug 28, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
COMBITROL TS / AUTOTROL TS assayed controls are intended to be used to monitor and evaluate the analytical performance of the Roche OMNI, alpha series instruments for the analytes listed in the package insert.
Device Story
COMBITROL TS and AUTOTROL TS are aqueous liquid multi-analyte controls used for quality assurance in clinical laboratories. The device contains a stable suspension of polystyrene microbeads that reflect and absorb red and infrared light, simulating the optical properties of erythrocytes to allow for the measurement of total hemoglobin (tHb), hematocrit, and oxygen saturation (SO2) in the same manner as whole blood. The controls are provided in three levels with varying microbead concentrations to simulate low, medium, and high hematocrit samples. The device also contains clinically relevant quantities of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, lithium, ionized magnesium, urea, glucose, lactate, and creatinine. Laboratory personnel use these controls to perform periodic quality control checks on Roche OMNI, alpha series analyzers to ensure analytical performance and detect systematic deviations.
Clinical Evidence
No clinical data. Bench testing only. Stability claims are supported by accelerated aging studies on labile analytes and historical experience with similar formulations.
Technological Characteristics
HEPES-based aqueous liquid control; contains polystyrene microbeads for optical simulation of erythrocytes (tHb, hematocrit, SO2). Three-level concentration format. Packaged in 1.7 mL glass ampoules. Designed for use with Roche OMNI, alpha series analyzers.
Indications for Use
Indicated for use in clinical laboratories as a quality control product for Roche OMNI, alpha series instruments to evaluate test precision and detect systematic analytical deviations for analytes listed in the package insert.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
AVL OPTI-check PLUS Multi-Analyte Control (K001632)
Submission Summary (Full Text)
{0}------------------------------------------------
AUG 2 8 2001
Ko 12431
# 510(k) Summary*
BIONOSTICS
- Submitter's name, address (a) (1) Bionostics, Inc. 2 Craig Road Acton, MA 10720
Contact Person Kathleen Storro Director, QA & Regulatory Affairs (978) 263-3856 x 220
Date of preparation of this summary: 27 July 2001
- (2) Device trade or proprietary name:
COMBITROL TS and AUTOTROL TS Multi Analyte Controls
## Device common or usual name or classification name:
Multi Analyte Control Solution, All Types (Assayed and Unassayed)
| PRODUCT NOMENCLATURE | CLASSIFICATION<br>NUMBER | CLASS | PANEL |
|--------------------------------|--------------------------|-------|-----------|
| MULTI ANALYTE CONTROL SOLUTION | 862.1660 75 JJY | I | CHEMISTRY |
#### Substantial Equivalence (3)
COMBITROL TS and AUTOTROL TS Multi Analyte Controls are substantially equivalent in function, safety and efficacy to a number of currently marketed devices known as 'Combi' or 'Multi-Analyte' control solutions. In example:
Comparison of COMBITROL TS / AUTOTROL TS to predicate devices for substantial equivalency
| Characteristic | Predicate Devices | Modified Device | |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Name: | COMBITROL PLUS<br>Multi-Analyte Control | AVL OPTI-check PLUS<br>Multi-Analyte Control | COMBITROL TS<br>AUTOTROL TS<br>Multi Analyte Controls |
| 510(k), Date: | K972868, 08/28/97<br>K913133, 09/27/91 | K001632, 06/19/00 | |
| Number of levels: | 3 | 3 | 3 |
| Analytes: | pH, PCO2, PO2, Na+,<br>K+, Cl-, iCa++<br>Li+, iMg++, tHb, Hb<br>derivatives, Urea,<br>Glucose, Lactate,<br>Creatinine | pH, PCO2, PO2, Na+, K+,<br>Cl-, iCa++, Urea, Glucose<br>tHb, SO2 | pH, PCO2, PO2, Na+, K+,<br>Cl-, iCa++, Li+, iMg++, tHb,<br>Hct, SO2, Urea, Glucose,<br>Lactate, Creatinine, |
| Container: | glass ampoule | glass ampoule | glass ampoule |
| Fill volume: | 1.7 mL | 1.7 mL | 1.7 mL |
| Color: | red | milky | milky |
| Matrix: | HEPES based aqueous<br>with dyes to simulate<br>Hb and derivatives | HEPES based aqueous<br>with polystyrene beads to<br>simulate Hb and SO2 | HEPES based aqueous<br>with polystyrene beads to<br>simulate Hb and SO2 |
This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
{1}------------------------------------------------
#### Description of the new device (4)
Description of the new device
COMBITROL TS / AUTOTROL TS is a specially formulated aqueous liquid material intended to for use to monitor all analytes measured by the Roche OMNI, alpha series analyzers. It contains a stable suspension of polystyrene microbeads which series and partially absorb red and infrared light similar to erythrocytes, allowing true simulation of the measurement of tHb and SO2 in exactly the same manner as these analytes are determined in whole blood by the Roche OMNI, alpha series analyzers. The three control levels contain three different concentrations of microbeads to simulate low, medium, and high hematocrit blood samples. metobeaus to simulate tons in the provide a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program.
COMBITROL TS / AUTOTROL TS contain clinically relevant quantities of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, lithium, ionized 1 002, 1 02, 80drains) lactate, urea, creatinine and suitable concentrations of microbeads to simulate clinically relevant values of tHb, hematocrit and oxygen saturation.
- Intended use of the device (5)
COMBITROL TS / AUTOTROL TS assayed controls are intended to be used to Combilition 10 / 110 / 110 / 110 / 110 Roche OMNI, alpha series instruments for the analytes listed in the package insert ..
- Technological characteristics of the device. (6)
Technological characteribaTS assayed control is intended to be used to monitor and evaluate the analytical performance of the Roche OMNI, alpha series instruments for the analytes listed in the package insert is technologically equivalent to currently marketed products to which substantial equivalence is claimed. It contains a low concentration, stable suspension of polystyrene microbeads which reflect and partially absorb red and infrared light similar to erythrocytes, allowing true simulation of the measurement of tHb, hematocrit and SO2 in exactly the same manner as these analytes are determined in whole blood by the Roche OMNI, alpha series instruments.
- Summary of non-clinical tests submitted with the premarket notification for the (b) (1) device.
Accelerated aging studies on most labile analytes, together with experience with other products with similar formulations support stability claim.
- Summary of clinical tests submitted with the premarket notification for the device. (b) (2) N/A
- Conclusions drawn from the clinical and non-clinical trials. (b) (3) Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AUG 2 8 2001
Ms. Kathleen Storro Director, QA and Regulatory Affairs Bionostics, Inc. 2 Craig Road Acton, MA 01720
Re: K012431
Trade/Device Name: COMBITROL TS / AUTOTROL TS Multi-Analyte Controls Regulation Number: 21 CFR 862.1660 Regulatory Class: I, reserved Product Code: JJS, JJY Dated: July 27, 2001 Received: July 31, 2001
Dear Ms. Storro:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include are gone an other annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 recultures) in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such perious on. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices ---Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number: K012431
### COMBITROL TS / AUTOTROL TS Device Name: Multi-Analyte Controls
COMBITROL TS / AUTOTROL TS assayed controls are intended to be used to monitor and evaluate the analytical performance of the Roche OMNI, alpha series instruments for the analytes listed in the package insert.
For In Vitro Diagnostic Use
## Indications for Use:
As a part of the quality control program in institutions reporting those analytes listed in the package insert, COMBITROL TS and AUTOTROL TS Multi-Analyte Controls should be used in Roche OMNI, alpha series instruments to evaluate test precision and to detect systematic analytical deviations in those laboratories choosing to use a traditional, liquid, quality control product.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | <div></div> |
|------------------------------------------|-------------|
|------------------------------------------|-------------|
OR
| (Division Sign-Off) | Kesia Alexander for Jan Cooper (Optional Format 1-2-96) |
|-----------------------------------------|---------------------------------------------------------|
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K012431 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.