The Lorenz Resorbable Laryngotracheal Reconstruction Implants are intended to be used for the temporary treatment and repair of laryngotracheal fractures and laryngotracheomalacia and lumen augmentation in the temporary treatment and repair of glottic stenosis.
Device Story
Lorenz LactoSorb® Laryngotracheal Reconstruction (LTR) Implants consist of resorbable mesh sheets and graft plates used for temporary airway support. Mesh sheets facilitate repair of laryngotracheal fractures and laryngotracheomalacia; graft plates provide lumen augmentation for glottic stenosis. Implants are secured to tracheal cartilage using sutures and/or Lactosorb® screws. Device is intended for surgical use by clinicians. Resorbable material provides structural support during healing, potentially avoiding interference with growth. Clinical benefit includes tissue tolerance and adequate support for bone/cartilage healing.
Clinical Evidence
Animal and clinical studies demonstrated Lactosorb® implants were well-tolerated by tissue and provided adequate support for healing of bone and cartilage. Results indicated potential advantage of lack of interference with growth compared to non-resorbable stents.
Technological Characteristics
Resorbable synthetic polymer (Lactosorb®). Form factors: 0.75mm thick mesh sheets (50x50mm) and graft plates (0.5mm thick, 10x15mm base). Fixation via sutures and 1.5mm diameter Lactosorb® screws (3-8mm lengths).
Indications for Use
Indicated for patients requiring temporary treatment and repair of laryngotracheal fractures, laryngotracheomalacia, and glottic stenosis.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
{0}------------------------------------------------
K012409
## OCT 1 8 2001
## Summary of Safety and Effectiveness
#### Specification Holder:
Walter Lorenz Surgical, Inc. 1520 Tradeport Drive Jacksonville, FL 32218-2480 Establishment Registration: 1032347 Walter Lorenz Surgical, Inc. is a wholly owned subsidiary of Biomet, Inc.
| Contact Person: | Sheryl Malmberg<br>Walter Lorenz Surgical, Inc.<br>Jacksonville, FL 32218<br>Phone: (904) 741-4400 ext. 255<br>Fax: (904) 741-4500 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------|
|-----------------|------------------------------------------------------------------------------------------------------------------------------------|
Lorenz LactoSorb® Laryngotracheal Reconstruction Implants Proprietary Name:
Ear Nose & Throat Synthetic Polymer Material Common Name:
Classification Name: 21 CFR 874.3620
Legally Marketed Devices to which substantial Equivalence is Claimed: Eliachar Laryngeal foam Stent (K000154), Lorenz Stent (K896667), and Montgomery Laryngeal Stent (K842287).
Device Description: The Lorenz Lactosorb® Laryngotracheal Reconstruction (LTR) Implants consist of mesh sheets for the temporary treatment and repair of laryngotracheal fractures and laryngotracheomalacia, and graft plates for lumen augmentation in the temporary treatment and repair of glottic stenosis. The mesh sheets are 0.75mm thick and 50mm wide by 50mm long. The mesh sheets are offered with two different hole configurations to facilitate attachment to the tracheal cartilage using suture and/or screws. The graft plate is a one-piece device consisting of a rectangular base with an elliptical top. The top is available in widths of 3mm, 4mm, 5mm, and 6mm to allow for various amounts of augmentation of the lumen. The base is 0.5mm thick and 10mm wide by 15mm long. The base has holes around the outer edges to allow for fixation to the tracheal cartilage using suture and/or screws. The Lactosorb® Screws and Lactosorb® Direct Drive Screws have a 1.5 diameter with a range of lengths available from 3 to 8mm.
Intended Use: The Lorenz Lactosorb® Resorbable Laryngotracheal Reconstruction Implants are intended to be used for the temporary treatment and repair of laryngotracheal fractures and laryngotracheomalacia and lumen augmentation in the temporary treatment and repair of glottic stenosis.
{1}------------------------------------------------
#### Summary of Technologies:
| Comparison<br>Information | Lorenz Lactosorb®<br>Laryngotracheal<br>Reconstruction<br>Implants | Eliachar Laryngeal<br>Foam Stent | Lorenz Stent | Montgomery<br>Lryngeal Stent |
|---------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Walter Lorenz<br>Surgical, Inc. | Hood Laboratories | Walter Lorenz<br>Surgical, Inc. | Boston Medical<br>Products, Inc. |
| 510(k) | New | K000154 | K896667 | K842287 |
| Indications for Use | Reconstruction for<br>stenosis,<br>tracheomalacia, and<br>laryngeal fracture. | Reconstruction for<br>stenosis and<br>laryngeal fracture. | Glottic and<br>subglottic stenosis<br>aud upper tracheal<br>stenois with or<br>without laryngeal<br>stenosis | For the prevention<br>and treatment of<br>laryngeal stenosis<br>when the glottic<br>stenosis involves<br>the midglottis,<br>posterior glottis,<br>supraglottis, and<br>subglottis,<br>singularly or in<br>combination. |
| Design | Prefabricated | Prefabricated | Prefabricated | Prefabricated |
| Material | Lactosorb® | Silicone Foam | PTFE | Silicone |
| Sizes | 50 x 50mm mesh,<br>50 x 50mm sheet<br>15 x 10mm plate<br>with 4 available top<br>width sizes (3,4,5,<br>and 6mm) | 2 sizes: 25mm,<br>40mm, 15mm and<br>25mm, 52mm,<br>17mm | 16 sizes, ranging<br>from 6 mm to<br>18mm | Child, Small Adult,<br>Medium Adult, and<br>Large Adult |
| Method of Placement | Screws and/or<br>Sutures | Adherence to<br>laryngeal contours<br>with strap for<br>additional control<br>and anchoring | Sutures | Sutures with<br>silicone buttons |
| Method of<br>Deployment | Not applicable | Not applicable | Not applicable | Not applicable |
Non - Clinical Testing: Animal studies have been undertaken demonstrating the acceptability of Lactosorb® implants for laryngotracheal reconstruction. From these studies it has been concluded that Lactosorb® implants represent an advancement in the management of tracheomalacia and other conditions requiring temporary airway stenting, including graft procedures.
Studies have been undertaken demonstrating the acceptability of Lactosorb® Clinical Testing: implants for laryngotracheal reconstruction. From these studies it has been concluded that Lactosorb® implants were well tolerated by tissue and provided adequate support for healing of bone and cartilage, along with the potential advantage of a lack of interference with growth.
Lactosorb is a trademark of Biomet, Inc.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with a human profile visible in the negative space between the wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 8 2001
Walter Lorenz Surgical, Inc. c/o Ms. Sheryl Malmerg Regulatory Manager 1520 Tradeport Drive Jacksonville, FL 32218-2480
Re: K012409
Trade/Device Name: Lorenz Resorbable Laryngotracheal Reconstruction Implants Regulation Number: 21 CFR 874.3620 Regulation Name: Ear Nose and Throat Synthetic Polymer Material Regulatory Class: II Product Code: NHB Dated: July 19, 2001 Received: July 30, 2001
Dear Ms. Malmerg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally presidents in a classification for your device and thus, permits your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: Lorenz Resorbable Laryngotracheal Reconstruction Implants
## Indications For Use:
The Lorenz Resorbable Laryngotracheal Reconstruction Implants are intended to be used for the The Lorenz Resorbable Laryligotracheal Recomed Seconders, Maryneofracheomalacia, and lumen temporaly treatment and repair of raily agently of glottic stenosis.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <div> <span style="text-decoration: overline;">✓</span> </div> | OR | Over-The-Counter-Use (Optional Format 1-2-96) |
|---------------------------------------|----------------------------------------------------------------|----|-----------------------------------------------|
|---------------------------------------|----------------------------------------------------------------|----|-----------------------------------------------|
| (Division Sign-Off) | <div> <img alt="Signature" src="signature.png" width="100"/> </div> |
|--------------------------------|---------------------------------------------------------------------|
| Division of Ophthalmic Devices | |
| 510(k) Number | 2012409 |
001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.