← Product Code [FZP](/productcode/FZP) · K012317

# COALESCENT SURGICAL U-CLIP (K012317)

_Coalescent Surgical · FZP · Aug 31, 2001 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012317

## Device Facts

- **Applicant:** Coalescent Surgical
- **Product Code:** [FZP](/productcode/FZP.md)
- **Decision Date:** Aug 31, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Coalescent Surgical U-Clip™ is intended for endoscopic and non-endoscopic use in tissue and prosthetic material approximation and the creation of anastomoses in blood vessels, grafts and other tubular structures.

## Device Story

The Coalescent Surgical U-Clip™ is a single self-closing vascular clip designed for tissue approximation and vascular anastomosis. The device features a vascular clip attached to a needle via a flexible member, allowing for precise placement prior to closure. It is intended for use in both endoscopic and non-endoscopic surgical procedures. The device is fabricated from medical and implantable grade materials. It is operated by a surgeon to approximate tissue or create anastomoses in blood vessels, grafts, and other tubular structures, providing a mechanical alternative to traditional suturing.

## Clinical Evidence

Bench testing only. In vitro studies were conducted to evaluate functional characteristics. Results confirmed that the device met all internal specification requirements, external standard requirements, and predicate performance expectations.

## Technological Characteristics

Single self-closing vascular clip; includes needle and flexible member; fabricated from medical and implantable grade materials. Mechanical device; no energy source or software.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Coalescent Surgical Sutured-Clip™ ([K994160](/device/K994160.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

> KO12317 510(k) Number:

#### Applicant Information:

- July 16, 2001 Date Prepared:

| Name:    | Coalescent Surgical, Inc.                                   |
|----------|-------------------------------------------------------------|
| Address: | 559 E. Weddell Drive<br>Sunnyvale, CA 94089<br>408-743-9794 |

| Contact Person:   | Michael A. Daniel |
|-------------------|-------------------|
| Phone Number:     | (415) 407-0223    |
| Facsimile Number: | (925) 932-5706    |

#### Device Information:

| Classification:      | Class II Implantable Clips                                 |
|----------------------|------------------------------------------------------------|
| Trade Name:          | Coalescent Surgical U-Clip™                                |
| Common Name:         | Implantable Clip, Vascular Clip                            |
| Classification Name: | Surgical Devices: Implantable Clip, 79FZP, 21 CFR 878.4300 |

### Predicate Devices:

The Coalescent Surgical U-Clip™ is substantially equivalent in intended use and method of operation to the following predicate devices:

- Coalescent Surgical Sutured-Clip™ 510(k) K994160 .
### Device Description:

The Coalescent Surgical U-Clip™ is a single self-closing clip for vascular anastomosis and tissue approximation applications. The U-Clip consists of a specially designed vascular clip with a needle connected to one end via a flexible member. This design allows precise placement of clips prior to closure. The device is fabricated from standard medical and implantable grade materials.

{1}------------------------------------------------

### 510(k) SUMMARY

# (Continued)

### Intended Use:

The Coalescent Surgical U-Clip™ is intended for endoscopic and non-endoscopic use in tissue and prosthetic material approximation and the creation of anastomoses in blood vessels, grafts and other tubular structures.

## Comparison to Predicate Device(s):

The Coalescent Surgical U-Clip™ is substantially equivalent to the Coalescent Surgical Sutured-Clip (K994160). These devices are intended for application of vascular clips to tissue for purposes of performing vascular anastomosis and tissue and graft approximation.

### In Vitro Test Data:

Design analysis and in vitro data confirm that functional characteristics are substantially equivalent to the predicate device cited. Testing included in vitro studies. All data fell well within, both, internal specification requirements, as well as external standard requirements and predicate performance expectations.

#### Summary:

Based upon the product technical information, intended use, performance and biocompatibility information provided in this pre-market notification, the Coalescent Surgical U-Clip™ has been shown to be substantially equivalent to currently marketed predicate device.

8/3

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around a central emblem. The emblem consists of a stylized representation of three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 3 1 2001

Mr. Michael A. Daniel Regulatory and Clinical Affairs Coalescent Surgical 559 East Weddell Drive Sunnyvale, California 94089

Re: K012317

Trade/Device Name: Coalescent Surgical U-Clip™ Regulation Number: 878.4300 Regulatory Class: II Product Code: FZP Dated: July 16, 2001 Received: July 23, 2001

Dear Mr. Daniel:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Michael A. Daniel

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Lise W. L.

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): __ TBD

Coalescent Surgical U-Clip™ Device Name:

# Indications For Use:

:

The Coalescent Surgical U-Clip™ is intended for endoscopic and non-endoscopic use in The Coalescont Burghout o Onpvessels, grafts and other tubular structures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
  

| Prescription Use<br>(Per 21 CFR 801.109) | <div style="height:20px;"></div> <div style="border: 1px solid black; height: 20px;"></div> |
|------------------------------------------|---------------------------------------------------------------------------------------------|
|------------------------------------------|---------------------------------------------------------------------------------------------|

OR

| Over-The-Counter Use | <div style="height:20px;"></div> <div style="border: 1px solid black; height: 20px;"></div> |
|----------------------|---------------------------------------------------------------------------------------------|
|----------------------|---------------------------------------------------------------------------------------------|

(Optional Format 1-2-96)

| (Division Sign-Off)                                          | <div style="height:20px;"></div> <div style="border: 1px solid black; height: 20px;"></div> |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Division of General, Restorative<br>and Neurological Devices |                                                                                             |
| 510(k) Number                                                | K012517                                                                                     |

510(k) Number -

0 0 0 0 0 0 9

---

**Source:** [https://fda-staging.innolitics.com/device/K012317](https://fda-staging.innolitics.com/device/K012317)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com