← Product Code [EMA](/productcode/EMA) · K012316

# PERMACEM / PERMACEM DUAL (K012316)

_Dmg USA, Inc. · EMA · Aug 23, 2001 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012316

## Device Facts

- **Applicant:** Dmg USA, Inc.
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Aug 23, 2001
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

PermaCem is a chemical cure compomer cement for permanent luting of metal, resin or ceramic crowns, bridges, inlays and onlays. PermaCem Dual is a dual cure compomer cement for permanent luting of metal, resin or ceramic crowns, bridges, inlays and onlays. PermaCem and PermaCem Dual can also be used for: Luting of orthodontic anchors and brackets. Luting of abutments for dentures. Core build-up. Splinting of teeth in combination with wires, Kevlar or Ribbond-type materials. Repair material for resin-based provisionals. Bite registration material. Occlusal individualisation of plastic bite rails.

## Device Story

PermaCem and PermaCem Dual are two-component, radiopaque, compomer-based dental luting materials. PermaCem is chemical-curing; PermaCem Dual is dual-curing (chemical and light-cure). Both are provided in automix or handmix delivery systems. Used by dental professionals for permanent cementation of dental restorations (crowns, bridges, inlays, onlays), orthodontic appliances, and denture abutments, as well as for core build-ups, splinting, and bite registration. The material is applied to the tooth structure or restoration interface to provide mechanical retention and sealing. Benefits include versatility in restorative procedures and radiopacity for radiographic assessment.

## Clinical Evidence

Bench testing only. Conformity demonstrated via compliance with ISO 4049 (Dentistry - Polymer-based filling, restorative and luting materials) and FDA guidance for dental cements.

## Technological Characteristics

Two-component, radiopaque, compomer-based luting material. Available in automix or handmix delivery. Chemical-cure (PermaCem) or dual-cure (PermaCem Dual) polymerization. Complies with ISO 4049 standards.

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- Advance ([K940914](/device/K940914.md))
- Optibond ([K940513](/device/K940513.md))
- C&B Luting Composite ([K940030](/device/K940030.md))
- Flexi-Flow ([K922249](/device/K922249.md))
- C&B Metabond ([K960464](/device/K960464.md))
- Duolink ([K943596](/device/K943596.md))
- Variolink ([K931309](/device/K931309.md))
- Dyract Cem (Unknown)
- Fuji Duet (Fuji Plus) ([K946100](/device/K946100.md))
- Resinomer ([K924151](/device/K924151.md))
- Cement-It (Unknown)
- ABC Dual Adhesive Bridge Cement (Unknown)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows a sequence of handwritten alphanumeric characters. The characters are 'K012316' and are written in a bold, sans-serif style. The characters are all connected and are written in black ink.

# AUG 2 3 2001

| Trade Name:          | PermaCem / PermaCem Dual                                                                                        |
|----------------------|-----------------------------------------------------------------------------------------------------------------|
| Sponsor:             | DMG USA, Inc.<br>414 South State Street<br>Dover, DE 19901                                                      |
|                      | Registration # not yet assigned                                                                                 |
| Device Generic Name: | Dental luting material                                                                                          |
| Classification:      | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II. |

#### Predicate Devices:

| Product Name                    | 510(k) # | Manufacturer         |
|---------------------------------|----------|----------------------|
| Advance                         | K 940914 | Dentsply Int'l.      |
| Optibond                        | K940513  | Kerr Mfg. Co.        |
| C&B Luting Composite            | K940030  | Bisco                |
| Flexi-Flow                      | K922249  | EDS                  |
| C&B Metabond                    | K960464  | Parkell              |
| Duolink                         | K943596  | Bisco                |
| Variolink                       | K931309  | Vivadent             |
| Dyract Cem                      | Unknown  | Dentsply             |
| Fuji Duet (Fuji Plus)           | K946100  | GC                   |
| Resinomer                       | K924151  | Bisco                |
| Cement-It                       | Unknown  | Jeneric/Pentron Inc. |
| ABC Dual Adhesive Bridge Cement | Unknown  | Vivadent             |

510(k) Summary

# Product Description:

#### PermaCem:

PermaCem is a chemical-curing, radiopaque two-component luting material available in automix delivery systems or as a handmix material, designed for the permanent luting of metals, ceramics, and resin-based materials, which may be luted to tooth structures or in various combinations to each other. Possible uses include:

- . Luting of orthodontic anchors and brackets
- . Luting of abutments to dentures
- . Core build-up material
- Splinting of teeth in combination with wires, Kevlar or Ribbond-type materials .
- . Repair material for provisionals
- . Bite registration material
- . Build up material for plastic bite rails (occlusal individualisation).

#### PermaCem Dual:

PermaCem Dual is a dual cure (chemical and/or light cure), radiopaque two-component luting material available in automix delivery systems or as a handmix material, designed for the permanent luting of

{1}------------------------------------------------

metals, ceramics, and resin-based materials, which may be luted to tooth structures or in various combinations to each other. Possible uses include:

- Core build-up material .
- Splinting of teeth in combination with wires, Kevlar or Ribbond-type materials .
- Repair material for provisionals .
- Bite registration material ●
- Build up material for plastic bite rails (occlusal individualisation). .

#### Indications for Use:

PermaCem is a chemical cure compomer cement for permanent luting of metal, resin or ceramic crowns, bridges, inlays and onlays.

PermaCem Dual is a dual cure compomer cement for permanent luting of metal, resin or ceramic crowns, bridges, inlays and onlays.

PermaCem and PermaCem Dual can also be used for:

- . Luting of orthodontic anchors and brackets
- Luting of abutments for dentures .
- Core build-up
- Splinting of teeth in combination with wires, Kevlar or Ribbond-type materials
- Repair material for resin-based provisionals
- . Bite registration material
- . Occlusal individualisation of plastic bite rails

## Safety and Performance:

This submission is an Abbreviated 510(k) as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), DMG USA, Inc. has provided information to demonstrate conformity with FDA's guidance document entitled Dental Cements - Premarket Notification, August 1998 and ISO 4049 - Dentistry - Polymer-based filling, restorative and luting materials.

## Conclusion:

Based on their indications for use, technological characteristics, and comparison to predicate devices, the PermaCem / PermaCem Dual materials have been shown to be safe and effective for their intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three wavy lines or stripes, which are meant to represent a stylized caduceus, a symbol often associated with healthcare.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 3 2001

Ms. Pamela Papineau Consultant DMG USA, Incorporated 414 South State Street Dover, Delaware 19901

Re : K012316 Trade/Device Name: Permacem/Permacem Dual Requlation Number: 872.3275 Regulatory Class: II Product Code: EMA Dated: June 22, 2001 Received: July 23, 2001

Dear Ms. Papineau:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action.

{3}------------------------------------------------

Page 2 - Ms. Papineau

concerning your device in the Federal Register. Please note: this response to your premarket notification submission does ente feed any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

A. Ulatowski Timo Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DMG USA Inc. Abbreviated 510(k) Premarket Notification

K012316

0000

PermaCem / PermaCem |

Page 1

510(k) Number (if known): _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Device Name: PermaCem / PermaCem Dual

Indications for Use:

PermaCem is a chemical cure compomer cement for permanent luting of metal, resin, or ceramic crowns, bridges, inlays and onlays.

PermaCem Dual is a dual cure compomer cement for permanent luting of metal, resin, or ceramic crowns, bridges, inlays and onlays.

PermaCem Dual can also be used for:

- Luting of orthodontic anchors and brackets .
- Luting of abutments for dentures ●
- Core build-up .
- Splinting of teeth in combination with wires, Kevlar or Ribbond-type . materials
- Repair material for resin-based provisionals .
- . Bite registration material
- Occlusal individualisation of plastic bite rails .

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use **__**
(Per 21 CFR 801.109)

OR Over-the Counter Use **__**

(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number **__** *K012316*

---

**Source:** [https://fda-staging.innolitics.com/device/K012316](https://fda-staging.innolitics.com/device/K012316)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).