The Stryker Cross-Pinned Interference Screw System is intended for use in the surgical reconstruction of the anterior cruciate ligament (ACL) deficient knee to provide interference fixation of the various ACL allografts and autografts, including the patella bone-patellar tendontibial bone graft complex, the semi-tendinosus tendon graft, the semi-membranosus tendon graft, and the Achilles tendon graft.
Device Story
The Stryker Cross-Pinned Interference Screw is an orthopedic implant system used in ACL reconstruction surgery. It consists of an interference screw with a transverse hole and an optional cross pin for adjunctive fixation. The system includes specialized instruments, such as a drill guide, to ensure alignment for the cross pin. The device is used by surgeons in an operating room setting to secure ACL allografts or autografts within the knee. By providing both interference screw fixation and optional cross-pinning, the device enhances graft stability. Cadaver testing demonstrated increased yield and ultimate load strength compared to traditional interference screws, potentially improving clinical outcomes for patients undergoing ligament reconstruction.
Clinical Evidence
Bench testing only. Cadaver testing was performed to compare the strength of the cross-pinned interference screw against the predicate device, demonstrating a significant increase in both yield load and ultimate load.
Technological Characteristics
Material: Ti-6AI-4V ELI (ASTM F136). Components: Interference screw with transverse hole and optional cross pin. Sterilization: Gamma irradiation (EN 552) or Ethylene oxide (EN 550) to SAL 10^-6. Biocompatibility: ISO-10993 and G95-1 compliant. Packaging: ASTM 4169.
Indications for Use
Indicated for patients undergoing surgical reconstruction of an ACL-deficient knee requiring interference fixation of allografts or autografts (patellar tendon, semitendinosus, semimembranosus, or Achilles tendon).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The characters are 'K', 'O', '1', '2', '2', '7', and '0'. The characters are written in black ink on a white background. The handwriting is somewhat messy, but the characters are still legible.
Image /page/0/Picture/2 description: The image shows the logo for Stryker Endoscopy. The word "Stryker" is in a bold, sans-serif font, with the letters connected. Below the word "Stryker" is the word "ENDOSCOPY" in a smaller, sans-serif font. The logo is black and white.
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### Device Name
Classification Name:
Smooth or Threaded Metallic Bone Fixation Fasteners: 21 CFR §888.3040, Class II Threaded Fixation Pin (87 JDW) Stryker Cross-Pinned Interference Screw
# Common and Usual Name: Proprietary Name:
## Predicate Device
Stryker Wedge Interference Screw (#K972233) currently marketed by Stryker Endoscopy (Santa Clara, CA).
### Summary
This summary of 510(k) safety and effectiveness is being submitted in accordance with requirements of SMDA 1990.
The Stryker Cross-Pinned Interference Screw is intended for use in the surgical reconstruction of the anterior cruciate ligament (ACL) deficient knee to provide interference fixation of the various ACL allografts and autografts, including the patellar tendon-tibial bone graft complex, the semitendinosus tendon graft, the semi-membranosus tendon graft, and the Achilles tendon graft. The interference screw fixation technique is a common method in orthopedic surgery, and has been well published in professional journals such as Arthroscopy: The Journal of Arthroscopic and Related Surgery.
The Stryker Cross-Pinned Interference Screw consists of an implant and associated instruments for installation of the implant. The implant consists of two components: an interference screw and a cross pin. The interference screw has a transverse hole in it for receipt of the cross pin. The cross pin is available for optional use by the user, in the event that adjunctive fixation is deemed appropriate. A drill guide provides alignment for drilling a hole coincident with the hole in interference screw and for receiving the cross pin.
Both implants will be provided sterile for single-use applications (ASTM 4169). These devices will be sterilized by Gamma irradiation (EN 552) or Ethylene oxide (EN550), including limits for Ethylene Oxide residuals and validated to a sterility assurance level (SAL) of 10 *. The devices are biocompatible per ISO-10993 and G95-1. The Stryker Cross-Pinned Interference Screw is equivalent in intended use, safety, and efficacy to the predicate device. The cadaver testing showed a significant strength increase of both the yield load and the ultimate load. The material of construction is Ti-6AI-4 V ELI per ASTM F136.
The Stryker Cross-Pinned Interference Screw System does not raise new issues when compared to the currently marketed predicate device. Therefore, it is considered substantially equivalent to the Stryker Wedge Interference Screw System.
Contact:
Date: July 10, 2001
Alisa Miller Quality Engineer Stryker Endoscopy 2590 Walsh Avenue Santa Clara, CA 95051 (408) 567-9100 x.2259
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. The logo is simple, yet recognizable, and represents the department's mission to protect the health of all Americans.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 8 2001
Ms. Alisa Miller Quality Engineer Stryker Endoscopy 2590 Walsh Avenue Santa Clara, California 95051
Re: K012270
Trade/Device Name: Stryker Cross-Pinned Interference Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: JDW Dated: July 10, 2001 Received: July 19, 2001
Dear Ms. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Alisa Miller
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N Milkersen
Celia M. Witten. Ph.D., M.D. Director Divison of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
June 22, 2001
510(k) Number if known: K012270
# INDICATION FOR USE:
The Stryker Cross-Pinned Interference Screw System is intended for use in the surgical reconstruction of the anterior cruciate ligament (ACL) deficient knee to provide interference fixation of the various ACL allografts and autografts, including the patella bone-patellar tendontibial bone graft complex, the semi-tendinosus tendon graft, the semi-membranosus tendon graft, and the Achilles tendon graft.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) OR
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
Mark N. Mulhausen
Restorative and Neurological Devices 012270
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.