The Duracon® Posteriorly Stabilizer (PS) Lipped Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be used for revision of a failed prosthesis.
Device Story
Duracon® PS Lipped Tibial Insert is a component of a total knee prosthesis system. It functions as a bearing surface for cemented total knee arthroplasty. Device features include an anterior/posterior (A/P) lipped design for enhanced stability in full extension and a patellar cut-out for improved tracking. It utilizes a snap-fit locking mechanism with an auxiliary locking screw for attachment to Duracon® tibial baseplates. Used by orthopedic surgeons in clinical/hospital settings during primary or revision knee replacement procedures. The device replaces damaged joint surfaces, restores joint anatomy, and provides stability in cases of cruciate ligament deficiency or patellar absence. It is intended to improve patient mobility and reduce pain associated with degenerative or traumatic joint conditions.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Polymer tibial insert component for a semi-constrained, cemented knee prosthesis. Features snap-fit locking mechanism with locking screw. Includes A/P lipped geometry and patellar cut-out. Materials and manufacturing methods are identical to predicate Duracon® Stabilized inserts. Designed for use with existing Duracon® femoral and patellar components.
Indications for Use
Indicated for patients with noninflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), rheumatoid arthritis, functional deformity, post-traumatic loss of joint anatomy (patello-femoral erosion, dysfunction, or prior patellectomy), or irreparable knee fracture requiring primary or revision cemented total knee arthroplasty. Contraindicated in patients with overt infection, distant foci of infection, rapid disease progression, skeletal immaturity, or poor bone stock. Increased risk in uncooperative patients, those with neurological disorders, osteoporosis, metabolic bone disorders, osteomalacia, or previous arthrodesis.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
OCT 1 0 2001
510(k) Summarv Duracon® PS Lipped Tibial Inserts
Image /page/0/Picture/2 description: The image shows handwritten text that appears to be a combination of letters and numbers. The first line reads 'K012172', and the second line reads '1 OF 4'. The handwriting is somewhat stylized, with distinct shapes for each character, and the text is presented in a clear, legible manner against a white background.
#### 510(k) Summary Duracon® PS Lipped Tibial Insert
#### Submission Information
| Name and Address of the Sponsor<br>of the 510(k) Submission: | Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401-1677 |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mary-Catherine Dillon<br>Regulatory Affairs Specialist<br>and<br>Susan Krasny, PhD<br>Director, Regulatory Affairs |
| Date of Summary Preparation: | July 10, 2001 |
| Device Identification | |
| Proprietary Name: | Duracon® PS Lipped Tibial Insert |
| Common Name: | Knee Prosthesis |
| Classification Name and Reference: | Knee Joint, Patellofemorotibial,<br>Polymer/Metal/Polymer, Semi-<br>Constrained, Cemented Prosthesis<br>21 CFR '888.3560 |
#### Predicate Device Identification
The Duracon® PS Lipped Tibial Bearing Inserts are substantially equivalent to the Duracon® Stabilized Tibial Inserts cleared via K932070 and K936292.
#### Device Description
The Duracon® PS Lipped tibial inserts share the same critical design features as the predicate Duracon® Stabilized tibial inserts. They both have the same snap-fit locking
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows handwritten text on a white background. The text on the top line reads 'K012172'. The text on the bottom line reads '20F4'. The handwriting appears to be somewhat stylized and the numbers are clearly written.
mechanism, augmented by a locking screw, for mating with the current Duracon® tibial baseplates (K915512). The subject Duracon inserts are intended to be used with current Duracon® femoral and patellar components. The subject inserts will be available in the same range of sizes and thicknesses and are manufactured from the same materials as the current Duracon® Stabilized inserts.
The subject inserts have an anterior/posterior (A/P) lipped feature to enhance femoral/tibial implant anterior stability in full extension. In addition, the subject inserts incorporate a patellar cut-out feature to improve patellar tracking.
#### Intended Use
The intended use of the Duracon® PS Lipped tibial inserts is identical to that of the predicate Duracon® Stabilized tibial bearing inserts. As with the predicate inserts, the modified inserts are single use devices for use as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be used for revision of a failed prosthesis.
#### Indications
- Noninflammatory degenerative joint disease including osteoarthritis, traumatic . arthritis or avascular necrosis,
- . Rheumatoid arthritis,
- . Correction of functional deformity,
- Revision procedures where other treatments or devices have failed, u
{2}------------------------------------------------
K012172
3 OF 4
510(k) Summary Duracon® PS Lipped Tibial Inserts
- adon't To Elped notal more
Post-traumatic loss of joint anatomy, particularly when there is patello-femoral
erosion, dysfunction or prior patellectomy; and,
- Irreparable fracture of the knee .
# Contraindications
#### Absolute contraindications include:
- 6 Overt infection,
- Distant foci of infections (which may cause hematogenous spread to the implant I site),
- Rapid disease progression as manifested by joint destruction or bone absorption . apparent on roentgenogram,
- l Skeletally immature patients
- Cases where there is poor bone stock which would make the procedure # unjustifiable,
### Conditions presenting an increased risk of failure include:
- Uncooperative patient or patient with neurological disorders who is incapable of l following instructions,
- I Osteoporosis,
- Metabolic disorders which may impair bone formation, l
- . Osteomalacia, and
- 트 Previous arthrodesis
# Performance Data
Mechanical testing has been performed to demonstrate the substantial equivalence of the subject inserts to the predicate inserts.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows handwritten text on a white background. The text "K012172" is written in the top half of the image. Below that, the text "4 OF 4" is written in a smaller font size.
# Statement of Technological Comparison
The fundamental scientific technology of the current Duracon® PS Lipped tibial inserts has not changed with regard to the modified inserts. The modified inserts employ the same basic design concepts, the same materials, and the same manufacturing methods.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge of the circle. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 0 2001
Ms. Mary-Catherine Dillon Regulatory Affairs Howmedica Osteonics Corp. 59 Route 17 Allendale, New Jersey 07401-1677
Re: K012172 Trade Name: Duracon® Posteriorly Stabilized Lipped Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial, polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: June 22, 2001 Received: July 12, 2001
Dear Ms. Dillon:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
#### Page 2 - Ms. Mary-Catherine Dillon
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sincerely yours,
Mark N Millerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K O la | 7 à
Device Name: Duracon® Posteriorly Stabilized Lipped Tibial Insert
The Duracon® Posteriorly Stabilizer (PS) Lipped Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be used for revision of a failed prosthesis.
#### Indications
- Noninflammatory degenerative joint disease including osteoarthritis, traumatic 미 arthritis or avascular necrosis,
- Rheumatoid arthritis, 트
- Correction of functional deformity, l
- Revision procedures where other treatments or devices have failed, 1
- Post-traumatic loss of joint anatomy, particularly when there is patello-femoral . erosion, dysfunction or prior patellectomy; and,
- . Irreparable fracture of the knee
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(CFR 801.109) | X | OR |
|-----------------------------------|---|----|
|-----------------------------------|---|----|
Over-the Counter Use (Per 21
Musth N millersen
Division Sign-Off)
eral. Restorative
**510(k) Number** K012172
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.