ThoraScan provides the radiologists or pneumologists the ability to acquire chest x-ray images by filmless radiography (digital radiography), based on a Csl/CCD slot scan detector produced by Thomson TE. The x-ray transmission profile of the chest is converted into an electronic, digital image in real-time. Chest images become available for preview by the x-ray technician on the operator's workstation only seconds after the x-ray exposure. After operator verification by the tech, digital (DICOM) images can be stored on electronic media, as e.g. CD-ROM, magnetic disk, or be exported to a (DICOM/PACS) network, c.q. clinical review station or to a film printer. The Nucletron Oldelft ThoraScan is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.
Device Story
ThoraScan is a digital chest X-ray system utilizing slot-scan technology. Input: X-ray transmission profile of the chest. Transformation: Cesium iodide (CsI) scintillator coupled to CCDs converts X-rays into real-time digital images. Output: Digital images transmitted via DICOM network. Usage: Clinical or mass-chest radiography settings; operated by X-ray technicians. Workflow: Images previewed on operator workstation (OWS) seconds post-exposure; verified by technician; stored on electronic media or exported to PACS/clinical review stations/film printers. Benefit: Provides filmless digital radiography with improved image quality compared to predicate.
Clinical Evidence
No formal clinical trial; qualitative assessment performed. PA and lateral radiograms were evaluated by radiologists, who reported superior image quality compared to the predicate Digidelca-C. Bench testing included resolution and contrast measurements demonstrating improved performance.
Technological Characteristics
Digital chest X-ray system; slot-scan technology. Detector: CCDs coupled to cesium iodide (CsI) scintillator. Components: Camera stand, tube stand, control unit, X-ray tube, collimator, generator, operator workstation (OWS). Connectivity: DICOM network interface. Standards: 21 CFR 1020.30/31, UL-187, IEC601-1, 93/42/EEC, 89/336/EEC.
Indications for Use
Indicated for clinical or mass-chest radiography in patients requiring chest x-ray examinations.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
510(K) Nucletron [Oldelft ThoraScan] [June 27th, 2001]
SEP 1 9 2001
Image /page/0/Picture/2 description: The image is a black and white graphic. It features a square border, inside of which is a black oval shape. Two white circles are positioned on opposite sides of the oval, partially overlapping it. The overall design is simple and geometric.
K012165/S1
NUCLETRON B.V.
Waardgelder 1 3905 TH Veenendaal 3900 AX Veenendaal P.O.Box 930 The Netherlands Phone +31 318 557133 Fax +31 318 550485
Department of Health and Human Services Center of Device and Radiological Health Office of Device Evaluation Special 510(k) section
ucletron
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by section 807.92(c)
## Submitter of 510(k): a.
| Company name: | Nucletron Corporation |
|----------------------|-----------------------------|
| Registration number: | 1121753 |
| Address: | 7080 Columbia Gateway Drive |
| | Columbia, MD 21046-2133 |
| Phone: | 410-312-4100 |
| Fax: | 410-312-4197 |
| Correspondent: | Lisa Dimmick |
## Device Name: b.
| Trade/Proprietary Name: | Oldelft ThoraScan (internally also known as |
|-------------------------|---------------------------------------------|
| Name: | ThoraScan and Digidelca-T) |
| Common/Usual Name: | Digital Chest X-ray System |
| Classification Name: | Stationary X-ray System (90-KPR) |
| Classification | Class II |
| CFR Citation | 21 CRF 892.1680 Tier 1 submission |
| Panel | Radiology |
## Legally Marketed Device(s) C.
Our device is substantially equivalent to the legally marketed predicate device cited in the table below:
| Manufacturer | Device | 510(k) # |
|--------------|-----------|----------|
| Nucletron BV | Digidelca | K980296 |
## Description: ರ.
ThoraScan is a digital X-ray chest camera based on slot-scan technology and is designed to perform radiographic chest examinations. The detector is comprised of CCD's directly coupled to a cesium iodide scintillator. The digital image is sent through a DICOM network. The ThoraScan consists of a camera stand with detector, tube stand, control unit, x-ray tube, collimator, generator and the OWS (Operators workstation).
{1}------------------------------------------------
# 510(K) Nucletron [Oldelft ThoraScan] [June 27th, 2001]
#### Intended use e.
Clinical or mass-chest radiography
# Summary of technological considerations f.
The ThoraScan is substantially equivalent to the predicate device (Digidelca). Compared to the Digidelca the slot-scan detector and the control unit have been improved.
The ThoraScan presents no new safety concerns. It will comply with the x-ray requirements of 21CFR 1020.30/31 as well as safety requirements of UL -187, IEC601-1 rood collateral standards, as well as the requirements of the Medical Device Directive (93/42/EEC) and the EMC Directive (89/336/EEC).
## Non-clinical tests ்.
Prototype tests of the detector include a.o. resolution and contrast measurements which are reported in Tab 10 of this application and which show superior results compared to the Digidelca-C.
## Clinical tests h.
PA and lateral radiograms were judged by several radiologist and found the digital images had better image quality than the Digidelca-C
#### Conclusions i.
Nucletron considers the ThoraScan to be equivalent with the predicate device. The ThoraScan provides radiograms that result in better imaging performance than Digidelca.
Jtc
Name: Rudolf Scholte Title: Business Manager Nucletron B.V. Veenendaal, The Netherlands
28/06/01
Date
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines above them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Public Health Service
SEP 1 9 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lisa Cole Dimmick Director of Regulatory Affairs Nucletron Corporation 7080 Columbia Gateway Drive COLUMBIA MD 21046-2133
Re: K012165
Trade/Device Name: Oldelft Thorascan Number 180T Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: II Product Code: 90 KPR Dated: August 24, 2001 Received: August 27, 2001
Dear Ms. Dimmick
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
proceed to the market.
{4}------------------------------------------------
510(K) Nucletron [Oldelft ThoraScan] [June 27, 2001]
Image /page/4/Picture/1 description: The image shows the word "Nucletron" in bold, black font. To the left of the word is a square logo. The logo contains a black circle with two white circles inside of it.
K012165
NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal 3900 AX Veenendaal P.O.Box 930 The Netherlands Phone +31 318 557133 Fax +31 318 550485
Department of Health and Human Services Center of Device and Radiological Health Office of Device Evaluation Special 510(k) section
# Statement of intended use
Device Name:
Oldelft ThoraScan
# Intended Use:
ThoraScan provides the radiologists or pneumologists the ability to acquire chest x-ray images by filmless radiography (digital radiography), based on a Csl/CCD slot scan detector produced by Thomson TE.
The x-ray transmission profile of the chest is converted into an electronic, digital image in real-time. Chest images become available for preview by the x-ray technician on the operator's workstation only seconds after the x-ray exposure.
operator o wonecation one by the tech, digital (DICOM) images can be stored on electronic media, as e.g. CD-ROM, magnetic disk, or be exported to a (DICOM/PACS) network, c.q. clinical review station or to a film printer.
Note: PACS, networks, clinical review stations and (laser-) film printers are not considered part of the Digidelca system for which this application is filed, only external interfaces tot this type of equipment are defined.
# Prescription use:
The Nucletron Oldelft ThoraScan is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.
ui
Name: Title: Business Manager Nucletron B.V. Veenendaal, The Netherlands
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K012165
28/6/01.
Date
*Prescription Use* ✓
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.