K012131 · Pan Probe Biotech, Inc. · JXM · Jul 20, 2001 · Clinical Toxicology
Device Facts
Record ID
K012131
Device Name
LIVESURE BENZODIAZEPINES SCREEN TESTS
Applicant
Pan Probe Biotech, Inc.
Product Code
JXM · Clinical Toxicology
Decision Date
Jul 20, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3170
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Pan Probe Biotech LiveSure™ Benzodiazepines Screen Test Card and Test Strip devices are rapid, qualitative, competitive lateral flow immuno-chromatographic competitive urinary rapid in vitro diagnostic (IVD) assays for detection of Benzodiazepines, analogs, and metabolites (BZD) in human urine at a cut-off of 300 ng BZD/ml. The cut-off for both LiveSure™ Test Card and Test Strip device methods has been set at 300 ng BZD drug/ml based upon calibration using Oxazepam standards with a GC/MS method for the quantitation of the Oxazepam standard and urine test solutions. These IVD tests are intended for visual, qualitative screening, for professional use only, and are not intended for quantitative results, nor for over the counter sales. Pan Probe Biotech LiveSure™ BZD Screen Tests for BZD provide only preliminary qualitative analytical data. A more specific quantitative alternative method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
Device Story
Lateral flow competitive immunoassay; detects Benzodiazepines in human urine. Device consists of membrane pre-coated with Oxazepam drug-conjugate and absorbent pad with anti-BZD monoclonal antibody-colloidal gold conjugate. In absence of BZD, gold conjugate binds to test region forming visible pink band. In presence of BZD, drug competes for antibody binding sites, preventing test band formation. Control region confirms procedural validity. Used by professionals in clinical/forensic settings for preliminary screening. Results interpreted visually; positive results require GC/MS confirmation. Benefits include rapid, preliminary identification of drug presence to guide clinical decision-making.
Clinical Evidence
Clinical study of 387 urine samples compared LiveSure™ devices against GC/MS and EMIT® II. Results: 100% agreement with GC/MS positive results; 99.3% agreement with GC/MS negative results. Overall accuracy 99.5% (385/387) against GC/MS. Specificity of negatives compared to EMIT® II was 99.5%.
Indicated for professional, qualitative screening of human urine for Benzodiazepines, analogs, and metabolites at a 300 ng/ml cut-off. Not for OTC use or quantitative results.
Regulatory Classification
Identification
A benzodiazepine test system is a device intended to measure any of the benzodiazepine compounds, sedative and hypnotic drugs, in blood, plasma, and urine. The benzodiazepine compounds include chlordiazepoxide, diazepam, oxazepam, chlorzepate, flurazepam, and nitrazepam. Measurements obtained by this device are used in the diagnosis and treatment of benzodiazepine use or overdose and in monitoring levels of benzodiazepines to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). A benzodiazepine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
EMIT® II Assay
Submission Summary (Full Text)
{0}------------------------------------------------
510k Submission for
## LiveSure™ BENZODIAZEPINES SCREEN TESTS
Pan Probe Biotech, Inc.
Revision E, July 16, 2001
#### SUMMARY STATEMENT OF SAFETY AND EFFECTIVENESS
The sponsor, Pan Probe Biotech, Inc., has developed, manufactured, and tested under Good Laboratory The openod, I an Frob diagnostic (IVD) devices for qualitative testing of urine samples for the presence of Benzodiazepines, analogs, and metabolites in an IVD screening format.
phesinde name of the devices are the Pan Probe Biotech LiveSure™ Benzodiazepines Screen Test Card The Trade Thame of the de noor and of Benzodiazepines Test Systems, and a classification as and rost billp; framig a PPR 862.3170, with product code: JXM. These IVD devices are intended for medical/forensic screening of urines for Benzodiazepines, analogs, and metabolites.
The Pan Probe Biotech LiveSure™ Benzodiazepines Screen Test Card and Test Strip (i.e., LiveSure™ The Fair - robes) devices are rapid qualitative competitive chromatographic IVD immunoassays, in which Donzoally labeled drug conjugate competes with any Benzodiazepine (BZD) drugs, analogs or metabolites that may be present in test urine samples for limited specific antibody binding sites. I liveSure™ Benzodiazepines devices have a unique membrane pre-coated with a gold conjugate immunoassay indicator that is used is pre-labeled with specific monoclonal antibody from mouse directed imministed on the strip and Test Card consists of a membrane absorbent pad having a goldarobe-conjugate pre-labeled with specific monoclonal antibody from mouse that is directed against BZD drugs, and a chromatographic membrane pre-coated with a chemically modified Benzodiazepine (Oxazepam) drug-conjugate as a capture reagent. The Test region of each device has been layered with a Benzodiazepine [Oxazepam] drug-conjugate as a 1* capture reagent, while the Process Control region has been pre-coated with a 200 anti-mouse antibody reagent derived from goat. A pink colored anti-BZD monoconal antibody-colloidal gold conjugate pad is placed to the right of a test strip. In the absence of BZD drugs, analogs or metabolites in urine, pink colored antibody-colloidal gold conjugates move chromatographically along with the urinary samples on the membrane by capillary action. Antibodycolloidal gold conjugate binds to BZD-drug conjugate, forming an antibody-antigen complex. This antibody-BZD-drug conjugate appears as second visible pink colored band and captured reagent at the test region. Any BZD drugs, analogs or metabolites that are present in a sample urine act as antigens, competing with BZD-drug conjugate at the test band region for limited BZD-antibody binding sites on antibody-colloidal gold conjugate. When a sufficient concentration of urinary BZD drugs, analogs or metabolites are present, these analytes block the limited antibody binding sites. This blockage-binding prevents attachment of pink colored antibody-colloidal gold conjugate at the BZD-drug conjugate zone located at the test band region. To serve as a procedural control, a pink colored band in a control region will always appear, regardless of the presence of BZD in urine samples. Thus, negative urine samples produces two pink colored bands, while positive urine samples produce only one pink colored band.
In-house testing of LiveSure™ Benzodiazepines Screen Test Card and Test Strip devices against EMIT® II Assay as a predicate device provided data essentially showing equivalency between these devices and the predicate EMIT® II Assay. Additionally, independent clinical testing of 387 urine samples against LiveSure™ Benzodiazepines Screen Test Card and Test Strip devices, as well as EMIT® II Assay at an external reference laboratory resulted in a 100% percent with all GC/MS quantitative positive results. Moreover, LiveSure™ Benzodiazepines Test Card or Strip gave both 99.3% agreement with GC/MS negative results. In comparing the Test Card and Test Strip positives with EMIT® II positives, both 98.3% respective agreement with EMIT® II was found. Specificity of Test Card and Test Strip negatives with EMIT® II negatives was shown to be 99.5%, respectively. In terms of overall accuracy of values at and below the ±25% range of the NIDA/SAMHSA cut-off of 300 ng/ml of Benzodiazepines, however, the LiveSure™ Benzodiazepines Screen Test Card and Strip yielded no false positives or no FP, but EMIT®Il resulted in 1 FP values for urine samples with GC/MS results below ng/ml. Finally, the LiveSure™ Benzodiazepines Test Card and the Test Strip gave overall accuracy results of 385/387(99.5%) for both, versus GC/MS data, whereas 385/387(99.5%) accuracy was obtained with EMIT®II. Thus, as judged against GC/MS results from an independent laboratory, the LiveSure™ Benzodiazepines Test Card and Test Strip were determined to be equivalent in performance to each other and somewhat superior in capability versus assays with EMIT®II.
Additional information on this submission may be obtained by contacting Alice Yu, Vice President, Pan Probe Biotech, Inc. at: 1- 858-689-9936 or by fax at 1-858-689-6896.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered in the image and is the only element present.
Image /page/1/Picture/9 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized depiction of a human figure embracing three people, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Pan Probe Biotech, Inc. c/o James M. Barquest, Ph.D. California Department of Health Services Food and Drug Branch PO Box 942732 601 North Seventh Street (MS 357) Sacramento, CA 94234-7320
510(k) Number: K012131 Re: Trade/Device Name: Pan Probe Biotech LiveSure™ Benzodiazepines Screen Tests Regulation Number: 862.3170 Regulatory Class: II Product Code: JXM Dated: July 3, 2001 Received: July 9, 2001
JUL 2 0 2001
Dear Dr.Barquest:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nothication. The I'DA intellig of succion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and If you desire specific advice for your astic devices), please contact the Office of Compliance at additionally 807.10 for in the claims on the promotion and advertising of your device, (301) 594-4588. Additionally, for quest at (301) 594-4639. Also, please note the regulation
complease contact the Office of Compliance at (301) 594-4639. Also, please of Co please contact the Office of Compilance (21CFR 807.97). Other general entitled, "Misoranding by reference to president to president from the Division of Small mionitation on your responsionnes and in the more (800) 638-2041 or (301) 443-6597 or at its Miandracturers "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510k Submission for
### LiveSure™ BENZODIAZEPINE SCREEN TESTS
Pan Probe Biotech, Inc.
Proprietary Information
Revision E, July 16, 2001
## 510(k) Number (if known): Not yet assigned
# Device Name: Pan Probe Biotech LiveSure™ Benzodiazepines Screen Tests
## INDICATIONS FOR USE STATEMENT:
The Pan Probe Biotech LiveSure™ Benzodiazepines Screen Test Card and Test Strip devices are The Fan Froot Diotoon ENGGIF ENGGIF - Bonabative lateral flow immuno-chromatographic competitive urinary rapid in Thro Glaghoolio (112) qualifanes drugs, and metabolites (BZD) in human urine at assays for detoon of Donestials problems (Substance Abuse and Mental Health the Nox (Hattonal mother of Brag + 300 ng BZD/ml. The cut-off for both LiveSure™ Test Card oon Test Strip device methods has been set at 300 ng BZD drug/ml based upon calibration using and root benzodiazepine/BZD drug, and using Oxazepam standards with a GC/MS method for the quantitation of all the Oxazepam standard and urine test solutions. These IVD tests are intended for visual, qualitative screening, for professional use only, and are not intended for quantitative results, nor for over the counter sales. Pan Probe Biotech LiveSure™ BZD Screen Tests for BZD provide only preliminary qualitative analytical data. A more specific quantitative alternative method must be used in order to obtain a confirmed analytical result. NIDA quartitudio anomano meeting mass spectrometry (GC/MS) as the preferred confirmatory method. Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Jean Cooger
n Sign-Off)
Clinical Laboratory Devices
K012131
Prescription Use: or (Per 21 CFR 801.109)
Over-the-Counter Use: (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.