K012129 · Ortho Development Corp. · GEI · Aug 22, 2001 · General, Plastic Surgery
Device Facts
Record ID
K012129
Device Name
ORTHO DEVELOPMENT ORION-I EMF SYSTEM
Applicant
Ortho Development Corp.
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 22, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ORION-I EMF System is intended to be used for general surgical purposes in coagulation, cutting and ablation of soft tissues, including but not limited to orthopedic, arthroscopic, spinal, and neurosurgical procedures. Arthroscopic procedures may include those on the shoulder, elbow, wrist, hip, knee, and ankle. The ORION-I EMF System is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
Device Story
ORION-I EMF System is a line-voltage powered, bipolar radiofrequency (RF) generator; delivers up to 16 watts of power to electrode tip. Used for coagulation, cutting, and vaporization (ablation) of soft tissue. Includes sterile, single-use hand-pieces and electrodes. Operated by qualified medical personnel in clinical settings (orthopedic, arthroscopic, spinal, neurosurgical). Device transforms electrical energy into RF output at the electrode tip to achieve tissue effects. Physician controls application directly via hand-piece; output facilitates surgical tissue management, potentially benefiting patient through precise tissue ablation/coagulation.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, function, and material comparisons to legally marketed predicate devices.
Technological Characteristics
Bipolar radiofrequency generator; line-voltage powered; max power output 16 watts. Includes sterile, single-use hand-pieces and electrodes. No software or complex algorithms described.
Indications for Use
Indicated for general surgical coagulation, cutting, and soft tissue ablation in orthopedic, arthroscopic, spinal, and neurosurgical procedures. Contraindicated when, in the physician's judgment, an electrosurgical procedure is contrary to the patient's best interest.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
## 510(k) Summary
## AUG 2 2 2001
## Ortho Development Corporation ORION-I EMF System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K012129 |
|--------------------------------|---------|
|--------------------------------|---------|
A. Submitter:
Ortho Development Corporation 106 West 12200 South Draper, UT 84020
Phone: (801) 553-9991 (801) 553-9993 Fax:
Carol Freasier Contact: Regulatory Affairs
Date Prepared: June 22, 2001
- B. Device Names:
ORION-I EMF System Trade Name: Electrosurgical generator and accessories Common/usual Name: Electrosurgical and Coagulation Unit and Accessories Classification Name:
#### C. Predicate Devices:
.
- K001986, K001407 Ellman Surgitron IEC II RF generator .
- ArthroCare System 2000 RF generator K001588, K992581
- K001986 Ellman Surgitron Hand-pieces ●
- K982176 Kirwan Surgical Hand-pieces .
- K982176 Kirwan Surgical Electrodes .
#### D. Device Description:
Dovios Doodlip.com and accessories, sterile, single-use hand-pieces and sterile, single-use electrodes. The and accessementor is a line-voltage powered, bipolar radiofrequency generator capable of delivering up to 16 watts of power at the electrode tip. The ORION-I utilizes controlled of dollivering up to 10 waste of perfor was and vaporization (ablation) of soft tissue during a variety of electrosurgical procedures in general surgical use, including orthopedic, arthroscopic, spinal, and neurosurgical procedures.
{1}------------------------------------------------
- E. Intended Use:
The ORION-I EMF System is intended to be used for general surgical purposes in The Ortion FEM Oyelowanonization (ablation) of soft tissues, including but not limited to orthopedic, arthroscopic, spinal, and neurosurgical procedures. Arthroscopic procedures may include those on the shoulder, elbow, wrist, hip, knee, and ankle. The ORION-I EMF System is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
#### Contraindications
The use of the ORION-I EMF System is contraindicated, when in the judgment of the r he use of the Ortrol rrincedure would be contrary to the best interest of the patient.
- F. Comparison with the Predicate Device:
Oompanson with the Prodited between the Ortho Development ORION-I EMF System There are no orghineant anversely affect the use of the use of the device. The EMF and the prodicate do food which treatle predicate devices in design, function, materials, and indications for use/intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the first line and "Service" on the second line. The text appears to be extracted from a document or sign.
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a flowing, calligraphic style.
# AUG 2 2 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ortho Development Corporation c/o Michael Kwan, Ph.D. Principal Reviewer and Program Coordinator Underwriters Laboratories, Inc. 1655 Scott Boulevard Santa Clara, California 95050
Re: K012129
Trade/Device Name: ORION-1 EMF System Regulation Number: 878.4400 Regulatory Class: II Product Code: GEI Dated: August 3, 2001 Received: August 10, 2001
Dear Dr. Kwan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Michael Kwan, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
L. Mark N. Millburn
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KO12129 510(k) Number (if known): __
Device Name: ORION-I EMF System
Indications for Use:
The ORION-I EMF System is intended to be used for general surgical purposes in coagulation, cutting and ablation of soft tissues, including but not limited to orthopedic, arthroscopic, spinal, curring and ablation of ook abouting rocedures may include those on the shoulder, elbow, wrist, hip, knee, and ankle.
The ORION-I EMF System is intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
### Contraindications
The use of the ORION-I EMF System is contraindicated, when in the judgment of the physician, an electrosurgical procedure would be contrary to the best interest of the patient.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
for Mark N Wilkerson
(Division Sign-Off)
ivision of General, Restorative d Neurological Devices
Number ***_****_\$\qquad\$ \$KO*{12/3}9\$
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.