← Product Code [KGO](/productcode/KGO) · K012104

# SNOW LOTUS POWDER-FREE LATEX SURGEON'S GLOVES (K012104)

_Beijing Reagent Latex Products Co., Ltd. · KGO · Aug 21, 2001 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012104

## Device Facts

- **Applicant:** Beijing Reagent Latex Products Co., Ltd.
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** Aug 21, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

A Powder Free Latex Surgical Glove is a disposable device intended in surgical settings for medical purpose that is worn on the hand to provide a barrier against potentially infectious materials and other contaminants.

## Device Story

Disposable powder-free latex surgical glove; worn on hands by surgeons/healthcare personnel in surgical settings. Acts as physical barrier against infectious materials and contaminants. Manufactured from natural rubber latex. Provides protection during surgical procedures. Device is non-sterile (sterility testing referenced per USP).

## Clinical Evidence

Bench testing only. Performance data confirms compliance with ASTM D3577-00 requirements for dimensions, physical properties (tensile strength, elongation at break, modulus), freedom from holes, and powder content. Biocompatibility confirmed.

## Technological Characteristics

Natural rubber latex surgical glove. Powder-free. Dimensions and physical properties meet ASTM D3577-00 standards. Tensile strength: 28MPa (before aging), 21MPa (after aging). Elongation at break: 830% (before aging), 800% (after aging). Modulus at 500% elongation: 3MPa. Powder content: 0.38mg/glove. Protein content: 31.4±0.05 µg/g. Biocompatible.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Class I Surgeon's Glove (Powder Free) 79KGO

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

BEIJING REAGENT LATEX PRODUCTS CO., LTD.

ADD: NO.30 SOUTH SANLITUN ROAD, DONGDAQIAO, CHAOWAL, BEUING, P.R.CHINA. POST CODE: 100020 FAX: 86-10-81501067 TEL: 86-10-81502573

KOI2104

### AUG 2 1 2001

510K SUMMARY as required by: 807.92 (c)

1.0APPLICANT

BEIJING REAGENT LATEX PRODUCTS CO., LTD. Name: NO. 30 SOUTH SANLITUN ROAD, DONGDAQIAO, CHAOWAI Add .: BEIJING, 100020 P.R.CHINA Tel NO. 86-10-81502573 Fax NO. 86-10-81501067

### Contact person: MRS. WANG YANNAN GENERAL MANAGER

- 2.0 Device trade name: SNOW LOTUS Common Name: Surgeon's Glove Classification Name: Powder Free Surgeon's Glove
3.0 Legally marketed device to which the company claiming equivalence:

Class I Surgeon's Glove (Powder Free) 79KGO that meets all the requirement of ASTM D3577.

4.0 Description of the device:

Class I Surgeon's Glove (Powder Free) 79KGO that meets all the requirement of ASTM D3577.

5.0 Intended use of the Device:

A Powder Free Latex Surgical Glove is a disposable device intended in surgical settings for medical purpose that is worn on the hand to provide a barrier against potentially infectious materials and other contaminants.

{1}------------------------------------------------

BEIJING REAGENT LATEX PRODUCTS CO., LTD. ADD: NO.30 SOUTH SANLITUN ROAD, DONGDAQIAO, CHAOWAI, BEIJING, P.R.CHINA. FAX: 86-10-81501067 POST CODE: 100020 TEL: 86-10-81502573

# 6.0 Technological characteristics of the device compared to predicate device.

| Measured Parameters of Latex Surgical Gloves Powder Free<br>manufactured by Beijing Reagent Products Co., Ltd. |        |            | ASTM D3577-00<br>Requirement for Latex<br>Surgical Gloves<br>Powder Free |
|----------------------------------------------------------------------------------------------------------------|--------|------------|--------------------------------------------------------------------------|
| Characteristics                                                                                                | Size   | Value      |                                                                          |
| Length                                                                                                         | 6.5    | Min.265 mm | Min. 265mm                                                               |
|                                                                                                                | 7      | Min.270 mm | Min. 265mm                                                               |
|                                                                                                                | 7.5    | Min.270 mm | Min. 265mm                                                               |
|                                                                                                                | 8      | Min.270 mm | Min. 265mm                                                               |
|                                                                                                                | 8.5    | Min.280 mm | Min. 265mm                                                               |
|                                                                                                                | 9      | Min.280 mm | Min. 265mm                                                               |
| Width                                                                                                          | 6.5    | 77-88 mm   | 83±6 mm                                                                  |
|                                                                                                                | 7      | 84-94 mm   | 89±6 mm                                                                  |
|                                                                                                                | 7.5    | 90-100 mm  | 95±6 mm                                                                  |
|                                                                                                                | 8      | 96-108 mm  | 102±6 mm                                                                 |
|                                                                                                                | 8.5    | 102-114 mm | 108±6 mm                                                                 |
|                                                                                                                | 9      | 108-120 mm | 114±6 mm                                                                 |
| Thickness                                                                                                      | Cuff   | 0.14mm     | Min.0.10 mm                                                              |
|                                                                                                                | Palm   | 0.18mm     | Min.0.10 mm                                                              |
|                                                                                                                | Finger | 0.22mm     | Min.0.10 mm                                                              |

### 7.0 PHYSICAL PROPERTIES

|                               | BEFORE AGING |             | AFTER AGING |            |
|-------------------------------|--------------|-------------|-------------|------------|
| Characteristics               | Value        | ASTM D3577  | Value       | ASTM D3577 |
| Tensile Strength              | 28MPa        | 24MPa       | 21MPa       | 18MPa      |
| Elongation at break %         | 830%         | 750%        | 800%        | 560%       |
| Modulus at 500%<br>elongation | 3MPa         | 5.5Mpa(max) |             |            |

{2}------------------------------------------------

BEIJING REAGENT LATEX PRODUCTS CO., LTD.

K012104

ADD: NO.30 SOUTH SANLITUN ROAD, DONGDAQIAO, CHAOWAL, BEIJING, P.R.CHINA. FAX: 86-10-81501067 TEL: 86-10-81502573

POST CODE: 100020

# 8.0 PERFORMANCE REQUIREMENT

| Characteristics       | Related<br>defects                                                         | Inspection<br>Level | Level as per<br>ASTM | AQL | AQL as per<br>ASTM D3577 |
|-----------------------|----------------------------------------------------------------------------|---------------------|----------------------|-----|--------------------------|
| Sterility             | Fails sterility                                                            | As per ISO          | As per USP           | N/A | N/A                      |
| Freedom from<br>Holes | Holes                                                                      | G-I                 | G-I                  | 1.5 | 1.5                      |
| Dimension             | Width<br>Length<br>Thickness                                               | S-2                 | S-2                  | 4   | 4                        |
| Physical<br>Property  | Tensile<br>strength<br>Elongation<br>At break<br>before and<br>after aging | S-2                 | S-2                  | 4   | 4                        |

## POWDER CONTENT

| VALUE        | ASTM REQUIREMENT |
|--------------|------------------|
| 0.38mg/glove | 2mg/glove max    |

### PROTEIN CONTENT

| And Address and OTY                                                                                                                                                            | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| would of control contrôleand concession of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of t | CARD A - CANNON - CA & CANADA                                                                                                                                                  |
|                                                                                                                                                                                | And Career of Career of Career of Children                                                                                                                                     |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | WALLE CLAST COLLECT COLLECTION                                                                                                                                                 |
| -                                                                                                                                                                              | ויחוץ זו זר ארויזית זו  זוחייא                                                                                                                                                 |
| --------                                                                                                                                                                       | THE STEWAN                                                                                                                                                                     |
| 1                                                                                                                                                                              | STATE OF COLORIES OF COLLECTION OF                                                                                                                                             |
| FOR THE FOR COLLECT CALLERS COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION CONTRACTION CONTRACTION CONTRACTION  | .                                                                                                                                                                              |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| LAND AND COLLECT CHANNER CHANNER CHARRET                                                                                                                                       | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| (                                                                                                                                                                              |                                                                                                                                                                                |
|                                                                                                                                                                                | LEBELLER MAN                                                                                                                                                                   |
|                                                                                                                                                                                |                                                                                                                                                                                |
|                                                                                                                                                                                | ------------------------------------                                                                                                                                           |

### BIOCOMPATIBILITY:

| VALUE                   | FDA REQUIREMENT         |
|-------------------------|-------------------------|
| Biologically Compatible | Biologically Compatible |

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a sequence of alphanumeric characters, specifically "K012104". The characters are handwritten and appear to be on a white background. The numbers are clearly written and easy to read.

BEIJING REAGENT LATEX PRODUCTS CO., LTD. ADD: NO.30 SOUTH SANLITUN ROAD, DONGDAQIAO, CHAOWAI, BEIJING, P.R.CHINA. POST CODE: 100020 TEL: 86-10-81502573 FAX: 86-10-81501067

### 9.0 Performance Data:

The performance test data of the powder free surgical gloves manufactured by Beijing Reagent Latex Products Co., Ltd. is given below:

| Measured Parameters of Latex Surgical Gloves Powder Free |        |            |
|----------------------------------------------------------|--------|------------|
| manufactured by Beijing Reagent Latex Products Co., Ltd. |        |            |
| Characteristics                                          | Size   | value      |
| Length                                                   | 6.5    | Min.265 mm |
|                                                          | 7      | Min.270 mm |
|                                                          | 7.5    | Min.270 mm |
|                                                          | 8      | Min.270 mm |
|                                                          | 8.5    | Min.280 mm |
|                                                          | 9      | Min.280 mm |
| Width                                                    | 6.5    | 77-88 mm   |
|                                                          | 7      | 84-94 mm   |
|                                                          | 7.5    | 90-100 mm  |
|                                                          | 8      | 96-108 mm  |
|                                                          | 8.5    | 102-114 mm |
|                                                          | 9      | 108-120 mm |
| Thickness                                                | Cuff   | 0.14mm     |
|                                                          | Palm   | 0.18mm     |
|                                                          | Finger | 0.22mm     |

#### PHYSICAL PROPERTIES

| Characteristic             | Before Aging | After Aging |
|----------------------------|--------------|-------------|
| Tensile Strength           | 28MPa        | 21MPa       |
| Elongation at break %      | 830%         | 800%        |
| Modulus at 500% elongation | 3MPa         | 3MPa        |

{4}------------------------------------------------

BEIJING REAGENT LATEX PRODUCTS CO., LTD.

Image /page/4/Picture/1 description: The image shows a sequence of handwritten numbers. The numbers appear to be "1012104". The numbers are written in black ink on a white background. The handwriting is somewhat stylized, with some numbers slightly overlapping.

ADD: NO.30 SOUTH SANLITUN ROAD, DONGDAQIAO, CHAOWAI, BEIJING, P.R.CHINA. POST CODE: 100020 TEL: 86-10-81502573 FAX: 86-10-81501067

### PERFORMANCE REQUIREMENT

| Characteristics       | Related defects                                                            | LEVEL | AQL |
|-----------------------|----------------------------------------------------------------------------|-------|-----|
| Sterility             | Fails sterility                                                            | N/A   | N/A |
| Dimension             | Width<br>Length<br>Thickness                                               | S-2   | 4.0 |
| Freedom<br>from Holes | Holes                                                                      | S-4   | 1.5 |
| Physical<br>Property  | Tensile<br>strength<br>Elongation<br>At break<br>before and<br>after aging | S-2   | 4.0 |

### POWDER CONTENT: PROTEIN CONTENT: 31.4±0.05 Hg/g BIOCOMPATIBILITY: Biologically Compatible

10. Clinical Data: N/A

### 11. CONCLUSION OF PERFORMANCE TEST DATA:

The Powder Free Surgical Glove manufactured by BEIJING REAGENT LATEX PRODUCTS CO., LTD. Meet or exceed the ASTM D 3577 Meet FDA Pin holes Requirement Meet labeling claim as shown by the Requirment

#### 12. ANY OTHER INFORMATION

Any other information required by FDA regarding product safety and effectiveness will be provided on request.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is enclosed within a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the top and sides of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 1 2001

Mr. Wang Yannan General Manager Beijing Reagent Latex Products No. 30 South Sanlitun Road, Dongdaqiao Chaowai, Beijing, P.R. CHINA

K012104 Re : "Snow Lotus" Powder-Free Latex Trade/Device Name: Surgeon's Gloves 878.4460 Regulation Number: Regulatory Class: I Product Code: KGO Dated: November 30, 2001 Received: July 5, 2001

Dear Mr. Yannan:

We have reviewed your Section 510(k) notification of intent to we have reviewed your beccion 310mm, and we have determined the market the device fererenced above the indications for device is substancially equivalently marketed predicate use stated in the encreate, commerce prior to May 28, 1976, devices marketed in incerbead as Device Amendments, or to the enactment date of the neasified in accordance with the devices that have been recrabbilding, and Cosmetic Act (Act).
provisions of the Federal Food, and inchingt to the general provisions of the reactar toon andy, subject to the general controls provisions of the Act. The general controls controls provisions of choude requirements for annual provisions of the Act therauc requeed manufacturing practice, registration, firsting or as resinst misbranding and adulteration.

If your device is classified (see above) into either class II II your device is crabblied (boo wearket Approval), it may (Special Controls) or Stimbling Controls. Existing major be subject to such addrerends ice can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A rederal Regulations, freic ar, farob assumes compliance with Subscancially equivalenceuring Practice requirements, as set the Current Good Manaracter Regulation (QS) for Medical Forch In the Quarty of CFR Part 820) and that, Devices. Ocheral regarations, the Food and Drug Chrough periodic QB Inbecount) such assumptions. Failure to

{6}------------------------------------------------

#### Page 2 - Yannan

comply with the GMP regulation may result in regulatory Ith the GMP regulation may publish further announcements action. action. In addition, FDA may purchasis manufactor of the mote:
concerning your device in the tendersion whicssion does concerning your device in the readlication submission does this response to your premaince neer have under sections 531 not affect any obligation you might under the Electronic through 542 of the Act 101 devices and 1 and 1 mm . . . Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as This lecter will afrow your smarket notification.
described in your 510(k) premarket notification. The FDA described in your 510\x) prematice notribution to a legally
finding of substantial equivalence of your leggification for your finding of substancial equivalines of plassification for your marketed predicate device resures in a olaboration.
device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling If you desire specific advice rol additionally 809.10 for in regulation (21 cm Fare over are over act the Office of
vitro diagnostic devices), please contact the for quogti Compliance at (301) 594-4622. Additionally, for questions on Compliance at (301) 354 ising of your device, please contact the promotion and advertising or your actise.
the Office of Compliance at (301) 594-4639. Also, please note ene regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general premarket notification (zick over the Act may be Information on your respon of Small Manufacturers obtained from the Division or among comment its toll-free number International and Comsumer Abbrando at its internet address (800) 650 2041 01 - (202) /dsma/dsmamain.html".

Sincerely yours,

Patricia Cuenite/for

Timothy A. Ulatowski Director Director and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# Indication for Use Statement

Applicant: Beijing Reagent Latex Products

510(k) Number (if known): _KOLZI_O 4

Device Name: Powder Free Latex Surgeon's Gloves

#### Indications For Use:

A surgeon's glove is a device made of natural or synthetic rubber intended to be worn by
the are and A surgeon's grove 15 a device intraction wound from contamination. (21CFR 878.4460)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Qiu S. Lim

(Division Sign-Off) (Division Sign-On)
Division of Dental, Infection Control,
Division of Dental, Infection Devices Division oral Hospital Device 510(k) Number --

---

**Source:** [https://fda-staging.innolitics.com/device/K012104](https://fda-staging.innolitics.com/device/K012104)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).