← Product Code [JWH](/productcode/JWH) · K012095

# THE LPS - CONTINUUM TM HYBRID TIBIA, MODELS 32-5886-2Y-XX (ZIMMER) AND 05-119-XXYY1-0 (IMPLEX) (K012095)

_Implex Corp. · JWH · Jul 26, 2001 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K012095

## Device Facts

- **Applicant:** Implex Corp.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Jul 26, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The LPS - Continuum® TM Hybrid Tibia is intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates cemented total knee arthroplasty. In addition, the LPS - Continuum TM Hybrid Tibia is indicated for use in the presence of knee instability caused by a compromised or non-functional posterior cruciate ligament.

## Device Story

LPS-Continuum Hybrid Tibia is a prosthetic tibial component for total knee arthroplasty; manufactured from Trabecular Metal (Hedrocel Porous Tantalum) with direct compression-molded ultra-high molecular weight polyethylene (UHMWPE). Device functions as an articular surface and tibial component; intended for use with Zimmer NexGen LPS Femoral components. Used by orthopedic surgeons in clinical settings to replace damaged knee joint surfaces; provides structural support and articulation. Benefits patient by restoring joint function and stability in cases of severe degeneration or ligamentous compromise.

## Technological Characteristics

Materials: Trabecular Metal (Hedrocel Porous Tantalum) and ultra-high molecular weight polyethylene (UHMWPE). Form factor: Tibial component for total knee arthroplasty. Designed for cemented fixation.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Reference Devices

- Zimmer NexGen LPS Femoral components

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Implex Corp.

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

| The LPS - Continuum TM Hybrid Tibia |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name:                     | Implex Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Submitter Address:                  | 80 Commerce Drive<br>Allendale, New Jersey 07401-1600                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Contact Person:                     | Robert Poggie                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Phone Number:                       | (201) 818-1800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Fax Number:                         | (201) 995-9722                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date Prepared:                      | June 29, 2001                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Trade Name:                  | The LPS - Continuum TM Hybrid Tibia                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Common<br>Name:              | Articular Surface and Tibial Components                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Classification Number<br>and Name:  | 21 CFR § 888.3560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Substantial<br>Equivalence:         | The term "substantial equivalence" as used in this<br>510(k) notification is limited to the definition of<br>substantial equivalence found in the Federal Food,<br>Drug and Cosmetic Act, as amended and as applied<br>under 21 CFR 807, Subpart E under which a device<br>can be marketed without premarket approval or<br>reclassification. A determination of substantial<br>equivalency under this notification is not intended to<br>have any bearing whatsoever on the resolution of<br>patent infringement suits or any other patent matters.<br>No statements related to, or in support of substantial<br>equivalence herein shall be construed as an<br>admission against interest under the US Patent Laws<br>or their application by the courts. |
| Device Description:                 | The LPS - Continuum TM Hybrid Tibia is manufactured<br>from Trabecular Metal (Hedrocel Porous Tantalum) with<br>direct compression molded ultra-high molecular weight<br>polyethylene (UHMWPE).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

These tibial components are intended for use with Zimmer NexGen LPS Femoral components.

{1}------------------------------------------------

#### 510(k) Summary (Continued)

#### Indications for Use:

The LPS - Continuum® TM Hybrid Tibia is intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates cemented total knee arthroplasty. In addition, the LPS - Continuum TM Hybrid Tibia is indicated for use in the presence of knee instability caused by a compromised or non-functional posterior cruciate ligament.

#### Conclusion:

The LPS - Continuum TM Hybrid Tibia is substantially equivalent to the identified predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three horizontal lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 6 2001

Dr. Robert A. Poggie Director of Applied Research Implex Corp. 80 Commerce Drive. Allendale, New Jersey 07401

Re: K012095

Trade Name: The LPS-Continuum ™ Hybrid Tibia Regulatory Class: II Regulation Number: 888.3560 Product Code: JWH Dated: July 3, 2001 Received: July 5, 2001

Dear Dr. Poggie:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

### Page 2 - Dr. Robert A. Poggie

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific a in vitro diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in This anagliestions on the promotion and advertising of your device, (2017) 594-4639. Rudinonary, 10. queens at (301) 594-4639. Also, please note the regulation prease connact the office or Scrence to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small monifacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Machluurks

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known) :

l2.PT

Device Name:

# The LPS- Continuum® TM Hybrid Tibia

Indications For Use:

The LPS - Continuum® TM Hybrid Tibia is intended for use where severe degeneration, trauma, or other pathology of the knee joint indicates cemented total knee arthroplasty. In addition, the LPS -Continuum® TM Hybrid Tibia is indicated for use in the presence of knee instability caused by a compromised or non-functional posterior cruciate ligament.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

Bhatthee to tai iun

(Division Sign-O Division of General, Restorative and Neurological Devices Counter Use 510(k) Number KO12099

Prescription Use (Per 21 CFR 801.109)

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K012095](https://fda-staging.innolitics.com/device/K012095)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
