IMMULITE TOXOPLASMA IGM, MODEL LKTM1 AND LKTM2, IMMULITE 2000 TOXOPLASMA IGM, MODEL L2KTM2 AND L2KTM6
Applicant
Diagnostic Products Corp.
Product Code
LGD · Microbiology
Decision Date
Jan 10, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3780
Device Class
Class 2
Indications for Use
IMMULITE® Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening. IMMULITE® 2000 Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® 2000 Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® 2000 Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening.
Device Story
IMMULITE and IMMULITE 2000 Toxoplasma IgM are automated, solid-phase, two-step chemiluminescent enzyme immunoassays. The device uses a polystyrene bead coated with partially purified Toxoplasma gondii antigen. Patient serum or plasma is incubated with the bead; Toxoplasma-specific IgM binds to the antigen. After a centrifugal wash, an alkaline phosphatase-labeled anti-human IgM antibody is added. A second incubation and wash follow. A phosphate ester of adamantyl dioxetane substrate is added; hydrolysis by alkaline phosphatase produces light. A luminometer measures photon output, which is proportional to the amount of bound IgM. Results are determined by comparing patient signal to a cutoff (adjustor) material. Used in clinical laboratories by technicians to aid physicians in diagnosing acute or recent Toxoplasma infection, particularly in pregnant women. The output provides a qualitative result that informs clinical decision-making regarding infection status.
Clinical Evidence
Clinical studies conducted at multiple sites in the U.S. compared IMMULITE/IMMULITE 2000 Toxoplasma IgM against predicate ELISAs. Studies included pregnant women and symptomatic/asymptomatic subjects. For IMMULITE, agreement with Kit A was 97.4% (all subjects) and 97.6% (pregnant); agreement with Kit B was 95.5% (all) and 95.2% (pregnant). For IMMULITE 2000, agreement with Kit A was 98.7% (all) and 98.8% (pregnant). CDC panel testing showed 97.9% and 99.0% total agreement for IMMULITE and IMMULITE 2000, respectively. Precision studies (within-run and total CVs) and interference testing (bilirubin, lipemia, hemoglobin) were performed.
Technological Characteristics
Solid-phase, two-step chemiluminescent enzyme immunoassay. Solid phase: polystyrene bead coated with partially purified Toxoplasma gondii antigen. Energy source: automated luminometer for photon detection. Substrate: phosphate ester of adamantyl dioxetane. Analyzers: IMMULITE and IMMULITE 2000. Connectivity: automated system. Sterilization: not specified. Software: embedded firmware for instrument control and signal processing.
Indications for Use
Indicated for presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. Used as an aid in diagnosis of acute, recent, or reactivated Toxoplasma infection when performed with Toxoplasma IgG assay. Not for blood/plasma donor screening.
Regulatory Classification
Identification
Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.
Zeus Scientific Toxo IgM ELISA Test System (K913787)
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 1 0 2002
# 012075 510(k) Summary of Safety and Effectiveness IMMULITE and IMMULITE 2000 Toxoplasma IgM
This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| Name:<br>Address: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, California 90045-5597 |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone Number:<br>Facsimile Number: | (310) 645-8200<br>(310) 645-9999 |
| Contact Person: | Edward M. Levine, Ph.D.<br>Director of Clinical Affairs |
| Date of Preparation: | January 8, 2002 |
| Device Name:<br>Trade: | IMMULITE® Toxoplasma IgM<br>IMMULITE® 2000 Toxoplasma IgM |
| Catalog Number: | LKTM1 (100 tests), LKTM2 (200 tests)<br>L2KTM2 (200 tests), L2KTM6 (600 tests) |
| Common: | Reagent system for the detection of IgM antibodies to<br>Toxoplasma gondii in human serum |
| Classification: | Class II device, 83-LGD (21CFR 866.3780) |
| Manufacturer: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, CA 90045<br>(The Quality System of Diagnostic Products Corporation is<br>registered to ISO 9001:1994) |
| Establishment Registration<br>Number | DPC's Registration Number 2017183 |
| Substantially Equivalent<br>Predicate Device: | Vidas (ELFA) Toxo IgM (K923166)<br>Zeus Scientific Toxo IgM ELISA Test System (K913787) |
| Description of Devices: | IMMULITE and IMMULITE 2000 Toxoplasma IgM are<br>clinical devices for use with their respective IMMULITE<br>and IMMULITE 2000 Automated Immunoassay Analyzers. |
{1}------------------------------------------------
### Intended Use of the Device:
IMMULITE® Toxoplasma IgM is a solid-phase chemiluminescent enzyme immunoassay for in vitro diagnostic use with the IMMULITE® Analyzer - for the presumptive qualitative detection of IgM antibodies to Toxoplasma in human serum or plasma (EDTA or heparinized), particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactivated Toxoplasma infection.
IMMULITE® 2000 Toxoplasma IgM is a solid-phase chemiluminescent enzyme immunoassay for in vitro diagnostic use with the IMMULITE® 2000 Analyzer – for the presumptive qualitative detection of IgM antibodies to Toxoplasma in human serum or plasma (EDTA or heparinized), particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® 2000 Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactivated Toxoplasma infection.
### Performance Equivalence:
Diagnostic Products Corporation (DPC) asserts that IMMULITE Toxoplasma IgM and IMMULITE 2000 Toxoplasma IgM produce substantially equivalent results to other commercially marketed assays, such Vidas (ELFA) Toxo IgM or Zeus Scientific Toxo IgM ELISA Test System.
### Technology Comparison:
Provided below is a comparison of DPC's IMMULITE and IMMULITE 2000 Toxoplasma IgM technology vs. the Zeus Scientific Toxo IgM ELISA Test System and Vidas Toxo IgM technology.
IMMULITE Toxoplasma IgM is a solid-phase, two-step chemiluminescent enzyme immunoassay. The solid phase, a polystyrene bead enclosed within an IMMULITE Test Unit, is coated with partially purified Toxoplasma gondii antigen.
Prediluted patient sample (1-in-21 dilution) and a protein-based buffer are simultaneously introduced into the Test Unit, and incubated for approximately 30 minutes at 37℃ with intermittent agitation. During this time. Toxoplasma -specific IgM in the sample binds to the Toxoplasma antigen-coated bead. Unbound serum is then removed by a centrifugal wash.
An alkaline phosphatase-labeled anti-human IgM antibody is introduced, and the Test Unit is incubated for approximately another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Test Unit is incubated for a further 10 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving
{2}------------------------------------------------
precision by providing a window for multiple readings. The bound complex -- and thus also the photon output, as measured by the luminometer -- is related to the presence of Toxoplasma InM in the sample. A result is then obtained by comparing the patient result to the response of the cutoff material (i.e., the adjustor).
IMMULITE 2000 Toxoplasma IgM is a solid-phase, two-step chemiluminescent enzyme immunoassay. The solid phase, a polystyrene bead added to an IMMULITE 2000 Reaction Tube, is coated with partially purified Toxoplasma gondii antigen.
Prediluted patient sample (1-in-20 dilution) and a protein-based buffer are simultaneously introduced into the Reaction Tube, and incubated for approximately 30 minutes at 37℃ with intermittent agitation. During this time. Toxoplasma -specific IgM in the sample binds to the Toxoplasma antigen-coated bead. Unbound serum is then removed by a centrifyigal wash.
An alkaline phosphatase-labeled anti-human IgM antibody is introduced, and the Reaction Tube is incubated for approximately another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Reaction Tube is incubated for a further 5 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex -- and thus also the photon output, as measured by the luminometer -- is related to the presence of Toxoplasma IgM in the sample. A result is then obtained by comparing the patient result to the response of the cutoff material (i.e., the adjustor)
Zeus Scientific Toxo IgM ELISA Test System is an enzyme-linked immunosorbent assay (ELISA) designed to detect IgM class antibodies to T. gondii antigen. Patient serum samples to be assayed for antibody are first diluted and incubated in microwells coated with T. gondii antigen. Any antigen specific IgM antibody in the sample will bind to the immobilized antigen. The plate is washed to remove unbound antibody and other serum components.
Peroxidase conjugated goat anti-human IgM is added to the wells and the plate is incubated. The conjugate will react with IgM antibody immobilized on the solid phase in step 2. The wells are washed to remove unreacted conjugate.
The microtiter wells containing immobilized peroxidase conjugate are incubated with peroxidase substrate solution. Hydrolysis of the substrate by peroxidase produces a color change. After a period of time, the reaction is stopped and the color intensity of the solution is measured photometrically. The color intensity of the solution is directly related to the antibody concentration in the test sample.
Vidas Toxo IgM is an enzyme-linked fluorescent immunoassay (ELFA) performed in an automated Vidas instrument. All assay steps and assay temperature are controlled by the instrument. A pipette tip-like disposable device, the Solid Phase Receptacle (SPR) serves as the solid phase as well as the pipettor for the assay. The SPR is coated with goat anti-u chain
{3}------------------------------------------------
antibodies. The Vidas TXM assay configuration prevents nonspecific reactions with the SPR. Reagents for the assay are in the sealed TXM Reagent Strips.
After a sample dilution step, the sample is cycled in and out of the SPR for a specific length of time. IgM antibodies present in the specimen will bind to the anti-u chain antibodies coating the interior of the SPR. Unbound sample components are washed away. An Immunocomplex of T. gondii antigen and mouse monoclonal anti-P30 antibodies conjugated with alkaline phosphatase is cycled in and out of the SPR and will attach to the human anti-T. gondii IgM bound to the SPR wall. A final wash step removes unbound conjugate.
A fluorescent substrate is introduced into the SPR. Enzyme remaining on the wall of the SPR will catalyze the conversion of the substrate to the fluorescent product. The intensity of the fluorescence is measured by the optical scanner in the instrument.
#### Expected Values
Individuals infected with the Toxoplasma organism will typically exhibit detectable levels of IgM antibody immediately before or soon after the onset of symptoms.2 IgM titers normally decline within four to six months, but may persist at low levels up to a year. 4 Patients with active toxoplasma chorioretinitis usually have undetectable levels of IgM.
The prevalence of Toxoplasma infection can vary depending on a number of factors such as age, gender, geographical location, socio-economic status, race, type of test used, specimen collection and handling procedures, and clinical and epidemiological history of individual patients. There are approximately 3000 cases of congenital toxoplasmosis reported per year, with an average of 0.6 cases per 1000 pregnancies in the United States.
#### IMMULITE Toxoplasma IgM
Studies with presumed healthy, asymptomatic subjects and individuals suspected of acute toxoplasma viral infection were conducted at two clinical sites in the southern and northeastern United States. The study in the southern United States consisted of 172 specimens from 93 pregnant women and 79 individuals with various conditions. IMMULITE Toxoplasma IgM tests on these samples vielded the following results:
| Subjects | Total<br>n | Positive | | Negative | | Indeterminate | |
|----------|------------|----------|-----|----------|-----|---------------|----|
| | | n | % | n | % | n | % |
| Pregnant | 93 | 22 | 24% | 64 | 69% | 7 | 8% |
| Various | 79 | 14 | 18% | 62 | 78% | 3 | 4% |
| All | 172 | 36 | 21% | 126 | 73% | 10 | 6% |
{4}------------------------------------------------
Image /page/4/Figure/0 description: The image is a bar graph showing frequency on the y-axis and ratio on the x-axis. The frequency ranges from 0 to 70, while the ratio ranges from 0.0 to 5.0. The bar graph shows that the frequency is highest at a ratio of 0.0, with a frequency of around 60, and then decreases as the ratio increases.
## Observed signal/cutoff ratios for all samples
The study in the northeastern United States consisted of 168 specimens from 91 pregnant women and 77 individuals with various conditions. IMMULITE Toxoplasma IgM tests on these samples yielded the following results:
| Subjects | Total<br>n | Positive | | Negative | | Indeterminate | |
|----------|------------|----------|-----|----------|-----|---------------|----|
| | | n | % | n | % | n | % |
| Pregnant | 91 | 20 | 22% | 65 | 71% | 6 | 7% |
| Various | 77 | 14 | 18% | 60 | 78% | 3 | 4% |
| All | 168 | 34 | 20% | 125 | 74% | 9 | 5% |
Observed signal/cutoff ratios for all samples
Image /page/4/Figure/5 description: This image is a bar graph that shows frequency on the y-axis and ratio on the x-axis. The x-axis ranges from 0.0 to 5.0, and the y-axis ranges from 0 to 90. The bar at ratio 0.0 has a frequency of approximately 84, while the bar at ratio 1.0 has a frequency of approximately 21.
{5}------------------------------------------------
### IMMULITE 2000 Toxoplasma IgM
ININGEFFE 2000 renthy, asymptomatic subjects and individuals suspected of acute toxoplasma viral infection were conducted at one clinical sites in the southern United States. The study consisted of 172 specimens from 93 pregnant women and 79 individuals with various study consisted or 172 proplasma IgM tests on these samples yielded the following results:
| Subjects | Total<br>n | Positive | | Negative | | Indeterminate | |
|----------|------------|----------|-----|----------|-----|---------------|-----|
| | | n | % | n | % | n | % |
| Pregnant | 93 | 23 | 25% | 61 | 66% | 9 | 10% |
| Various | 79 | 16 | 20% | 59 | 75% | 4 | 5% |
| All | 172 | 39 | 23% | 120 | 70% | 13 | 8% |
80 70 60 ૨૦ requenc 40 30 20 10 0 5.0 0.0 1.0
#### Observed signal/cutoff ratios for all samples
#### Performance Characteristics
#### Clinical Performance
In a clinical study in the southern United States, a total of 172 frozen samples from apparently healthy male and female subjects, pregnant women and patients suspected of being toxoplasma IgM positive were tested by IMMULITE Toxoplasma IgM and by a commercially available enzyme immunoassay (Kit A- Zeus Scientific Toxo IgM ELISA Test System). These samples were also tested for Toxoplasma IgG. The IMMULITE Toxoplasma IgM results were compared to the results of Kit A.
Comparison for all subjects
| Kit A | Positive | Indeterm | Negative |
|----------------------------------------------------------|----------|----------|----------|
| Positive | 36 | 9 | 4 |
| Indeterm | 0 | 1 | 6 |
| Negative | 0 | 0 | 116 |
| Positive Agreement: 90.0% (36/40, 95% CI: 76.3% - 97.2%) | | | |
| Negative Agreement: 100% (116/116, 95% CI: 96.9% - 100%) | | | |
| Agreement: 97.4% (152/156, 95% CI: 93.6% - 99.3%) | | | |
{6}------------------------------------------------
Comparison for pregnant subjects
| Kit A | Pos | Ind | Neg |
|----------------------------------------------------------|-----|-----|-----|
| Pos | 22 | 7 | 2 |
| Ind | 0 | 0 | 4 |
| Neg | 0 | 0 | 58 |
| Positive Agreement: 91.7% (22/24, 95% CI: 73.0% - 99.0%) | | | |
| Negative Agreement: 100% (58/58, 95% CI: 93.8% - 100%) | | | |
| Agreement: 97.6% (80/82, 95% CI: 91.5% - 99.7%) | | | |
In another clinical study in the northeastern United States, a total of 168 frozen samples from apparently healthy male and female subjects, pregnant women and patients suspected of being toxoplasma IgM positive were tested by IMMULITE Toxoplasma IgM and another commercially available enzyme immunoassay (Kit B - Vidas Toxo IgM). These samples were also tested for Toxoplasma IgG. The IMMULITE Toxoplasma IgM results were compared to the results of Kit B.
Comparison for all subjects
| Kit B | Positive | Indeterm | Negative |
|------------------------------------------------------------|----------|----------|----------|
| Positive | 28 | 4 | 1 |
| Indeterm | 0 | 0 | 2 |
| Negative | 6 | 5 | 122 |
| Positive Agreement: 96.6% (28/29, 95% CI: 82.2% - 99.9%) | | | |
| Negative Agreement: 95.3% (122/128, 95% CI: 90.1% - 98.3%) | | | |
| Agreement: 95.5% (150/157, 95% CI: 91.0% - 98.2%) | | | |
Comparison for pregnant subjects
| | Pos | Ind | Neg |
|----------------------------------------------------------|-----|-----|-----|
| Pos | 16 | 4 | 0 |
| Ind | 0 | 0 | 1 |
| Neg | 4 | 2 | 64 |
| Positive Agreement: 100% (16/16, 95% CI: 79.4% - 100%) | | | |
| Negative Agreement: 94.1% (64/68, 95% CI: 85.6% - 98.4%) | | | |
| Agreement: 95.2% (80/84, 95% CI: 88.3% - 98.7%) | | | |
In a clinical study in the southern United States, a total of 172 frozen samples from normal male and female subjects, pregnant women and patients suspected of being toxoplasma IgM positive were tested by IMMULITE 2000 toxoplasma IgM, and by a commercially available enzyme immunoassays (Kit A- Zeus Scientific Toxo IgM ELISA Test System). These samples were also tested for Toxoplasma IgG. The IMMULITE 2000 Toxoplasma IgM results were compared to the results of Kit A.
{7}------------------------------------------------
#### Comparison for all subjects
| Kit A | Positive | Indeterm | Negative |
|----------------------------------------------------------|----------|----------|----------|
| Positive | 38 | 9 | 2 |
| Indeterm | 1 | 3 | 3 |
| Negative | 0 | 1 | 115 |
| Positive Agreement: 95.0% (38/40, 95% CI: 83.1% - 99.4%) | | | |
| Negative Agreement: 100% (115/115, 95% CI: 96.8% - 100%) | | | |
| Agreement: 98.7% (153/155, 95% CI: 95.4% - 99.8%) | | | |
Comparison for pregnant subjects:
| | | IMMULITE 2000 Toxoplasma IgM | | | |
|----------------------------------------------------------|-----|-------------------------------------|-----|--|--|
| Kit A | Pos | Ind | Neg | | |
| Pos | 23 | 7 | 1 | | |
| Ind | 0 | 2 | 2 | | |
| Neg | 0 | 0 | 58 | | |
| Positive Agreement: 95.8% (23/24, 95% CI: 78.9% - 99.9%) | | | | | |
| Negative Agreement: 100% (58/58, 95% CI: 93.8% - 100%) | | | | | |
| Agreement: | | 98.8% (81/82, 95% CI: 93.4% - 100%) | | | |
In a study at DPC, IMMULITE 2000 Toxoplasma IgM was compared to IMMULITE Toxoplasma IgM on 291 samples:
| | Positive | Indeterm | Negative |
|----------------------------------------------------------|----------|----------|----------|
| Positive | 15 | 0 | 0 |
| Indeterm | 3 | 0 | 0 |
| Negative | 0 | 5 | 268 |
| Positive Agreement: 100% (15/15, 95% CI: 78.2% - 100%) | | | |
| Negative Agreement: 100% (268/268, 95% CI: 98.6% - 100%) | | | |
| Agreement: 100% (283/283, 95% CI: 98.7% - 100%) | | | |
Indeterminate results were excluded from calculations.
#### Performance Data
Precision (Serum): Precision studies for IMMULITE Toxoplasma IgM assay were conducted at three different sites: in-house at DPC (Site 1) and at two sites in the southern and northeasten United States (Sites 2 and 3). At Site 1, samples were assayed in duplicate over the course of 20 days, two runs per day, for a total of 40 runs and 80 replicates. (See "Site 1" table). At Sites 2 and 3, samples were assayed in triplicate over the course of 5 days, one run per day, for a total of 5 runs and 15 replicates. (See "Site 2" and "Site 3" tables). The means, within-run and total CVs were calculated by the Analysis of Variance. Results are expressed as a signal-to-cutoff ratio. Precision statistics are summarized below.
{8}------------------------------------------------
| | | Within-Run | | Total | |
|---|-------|------------|------|-------|-------|
| | Mean | SD | CV | SD | CV |
| 1 | 3.03 | 0.241 | 8.0% | 0.337 | 11.1% |
| 2 | 1.42 | 0.110 | 7.8% | 0.155 | 10.9% |
| 3 | 0.928 | 0.055 | 5.9% | 0.089 | 9.5% |
| 4 | 0.579 | 0.033 | 5.7% | 0.055 | 9.4% |
| 5 | 0.25* | | | | |
### IMMULITE Toxoplasma IgM Precision - Serum (ratio): Site 1
* Consistently at a very low ratio
#### IMMULITE Toxoplasma IgM Precision - Serum (ratio): Site 2
| | Mean | Within-Run | | Total | |
|---|--------|------------|------|-------|------|
| | | SD | CV | SD | CV |
| 1 | 3.30 | 0.184 | 5.6% | 0.203 | 6.2% |
| 2 | 1.56 | 0.113 | 7.2% | 0.110 | 7.1% |
| 3 | 0.980 | 0.068 | 6.9% | 0.063 | 6.4% |
| 4 | 0.650 | 0.032 | 4.9% | 0.030 | 4.6% |
| 5 | 0.284* | | | | |
* Consistently at a very low ratio
#### IMMULITE Toxoplasma IgM Precision - Serum (ratio): Site 3
| | Within-Run | | | Total | |
|---|------------|-------|------|-------|------|
| | Mean | SD | CV | SD | CV |
| 1 | 3.10 | 0.110 | 3.5% | 0.152 | 4.9% |
| 2 | 1.38 | 0.099 | 7.2% | 0.097 | 7.0% |
| 3 | 0.910 | 0.029 | 3.2% | 0.056 | 6.2% |
| 4 | 0.600 | 0.022 | 3.7% | 0.030 | 5.0% |
| 5 | 0.266* | - | - | - | - |
* Consistently at a very low ratio
Precision (Serum): Precision studies for IMMULITE 2000 Toxoplasma IgM assay were conducted at two different sites: in-house at DPC (Site 1) and in the southern United States (Site 2). At both sites, samples were assayed in triplicate over the course of 5 days, one run per day, for a total of 5 runs and 15 replicates. (See "Site 1" and "Site 2" tables). The means, within-run and total CVs were calculated by the Analysis of Variance. Results are expressed as a signal-tocutoff ratio.
{9}------------------------------------------------
| | Mean | Within-Run | | Total | |
|---|--------|------------|-------|-------|-------|
| | | SD | CV | SD | CV |
| 1 | 3.90 | 0.081 | 2.1% | 0.094 | 2.4% |
| 2 | 2.07 | 0.150 | 7.2% | 0.149 | 7.2% |
| 3 | 1.34 | 0.074 | 5.5% | 0.071 | 5.3% |
| 4 | 0.810 | 0.092 | 11.4% | 0.082 | 10.1% |
| 5 | 0.262* | – | – | – | – |
### IMMULITE 2000 Toxoplasma IgM Precision - Serum (ratio): Site 1
*Consistently at a very low ratio
#### IMMULITE 2000 Toxoplasma IgM Precision - Serum (ratio): Site 2
| | | Within-Run | | | Total | |
|---|--------|------------|------|--|-------|-------|
| | Mean | SD | CV | | SD | CV |
| 1 | 3.20 | 0.130 | 4.1% | | 0.138 | 4.3% |
| 2 | 1.72 | 0.109 | 6.3% | | 0.113 | 6.6% |
| 3 | 1.18 | 0.044 | 3.7% | | 0.078 | 6.6% |
| 4 | 0.720 | 0.031 | 4.3% | | 0.104 | 14.4% |
| 5 | 0.300* | | | | | |
*Consistently at a very low ratio
Precision (Plasma): Precision studies for IMMULITE Toxoplasma IgM and IMMULITE 2000 Toxoplasma IgM assays on plasma samples (EDTA and heparin) were conducted at DPC by testing samples in triplicate over the course of 3 days, two runs per day, for a total of 6 runs and 18 replicates. The means, within-run and total CVs were calculated by the Analysis of Variance. Results are expressed as a signal-to-cutoff ratio. Precision statistics are summarized below.
### IMMULITE Toxoplasma IgM Precision - EDTA (ratio):
| | | Within-Run | | Total | |
|---|-------|------------|------|-------|-------|
| | Mean | SD | CV | SD | CV |
| 1 | 0.262 | 0.023 | 8.8% | 0.033 | 12.6% |
| 2 | 1.11 | 0.052 | 4.7% | 0.069 | 6.2% |
| 3 | 1.33 | 0.075 | 5.6% | 0.077 | 5.8% |
| 4 | 1.66 | 0.087 | 5.2% | 0.083 | 5.0% |
{10}------------------------------------------------
| | Mean | Within-Run | | Total | |
|---|-------|------------|------|--------|------|
| | | SD | CV | SD | CV |
| 1 | 0.259 | 0.0191 | 7.4% | 0.0183 | 7.1% |
| 2 | 1.00 | 0.06 | 6.0% | 0.052 | 5.2% |
| 3 | 1.21 | 0.097 | 8.0% | 0.084 | 6.9% |
| 4 | 1.51 | 0.084 | 5.6% | 0.073 | 4.8% |
### IMMULITE 2000 Toxoplasma IgM Precision - EDTA (ratio):
#### IMMULITE Toxoplasma IgM Precision - Heparin (ratio):
| | Within-Run | | | Total | |
|---|------------|-------|-------|-------|-------|
| | Mean | SD | CV | SD | CV |
| 1 | 0.67 | 0.077 | 11.5% | 0.081 | 12.1% |
| 2 | 1.18 | 0.089 | 7.5% | 0.123 | 10.4% |
| 3 | 1.42 | 0.112 | 7.9% | 0.136 | 9.6% |
| 4 | 1.71 | 0.137 | 8.0 | 0.161 | 9.4% |
#### IMMULITE 2000 Toxoplasma IgM Precision - Heparin (ratio):
| | Mean | Within-Run | | | Total | |
|---|------|------------|-------|-------|-------|----|
| | | SD | CV | | SD | CV |
| 1 | 0.65 | 0.068 | 10.5% | 0.07 | 10.8% | |
| 2 | 1.2 | 0.061 | 5.1% | 0.057 | 4.8% | |
| 3 | 1.32 | 0.073 | 5.5% | 0.071 | 5.4% | |
| 4 | 1.54 | 0.084 | 5.5% | 0.078 | 5.1% | |
Crossreactivity: Crossreactivity: A study was conducted to evaluate whether the measurement of Toxoplasma IgM antibody is affected by closely related microorganisms. Ninety-three seronegative sera containing antibodies to Varicella Zoster Virus (n=3), Measles (n=10), Cytomegalovirus (CMV) (n=10), Herpes Simplex Virus (n=10), Toxoplasma (n=10), mvcloplasma pneumoniae (n=10), Epstein-Barr Virus (n=10), Syphilis (n=10) and Parvovirus (n=10) and rheumatoid factor (n=10) were tested by IMMULITE and IMMULITE 2000 Toxoplasma IgM and all yielded negative results.
#### Conjugated or unconjugated bilirubin: no effect up to 20 mg/dL Interference: Lipemia: no effect of triglycerides up to 3000 mg/dL Hemoglobin: no effect up to 539 mg/dL
{11}------------------------------------------------
Anti-coagulants: Twenty-eight blood samples drawn into plain, heparinized and EDTA vacutainer tubes were assayed by the IMMULITE 2000 Toxoplasma IgM assay. Results (in S/CO ratio) from the anticoagulant tubes were compared with those from the serum tubes in regression analyses:
| Regressions: | Heparin = $0.96 \times (Serum) + 0.05$<br>EDTA = $0.93 \times (Serum) + 0.04$ | r = 0.991<br>r = 0.992 |
|---------------------|-------------------------------------------------------------------------------|------------------------|
| Means (S/CO ratio): | Serum = 0.80<br>Heparin = 0.82<br>EDTA= 0.79 | |
Gel Barrier: Twenty-eight blood samples drawn into plain and SST vacutainer tubes were assayed by the IMMULITE 2000 Toxoplasma IgM assay. Results (in S/CO ratio) from the SST tubes were compared with those from the serum tubes in a regression analysis:
| Regressions: $SST = 0.97 \times (Serum) + 0.08$ | $r = 0.994$ |
|-------------------------------------------------|-------------|
| Means (S/CO ratio): Serum = 0.80 | |
| SST = 0.85 | |
#### CDC Toxoplasma 1998 Human Serum Panel
The following information is from a serum panel obtained from the CDC and tested by DPC. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.
The panel consists of 32 positive and 65 negative samples. IMMULITE Toxoplasma IgM demonstrated 97.9% total agreement with the CDC results. There was 93.8% agreement with the positive specimens and 100% agreement with the negative specimens. IMMULITE 2000 Toxoplasma IgM demonstrated 99.0% total agreement with the CDC results. There was 96.9% agreement with the positive specimens and 100% agreement with the negative specimens.
#### Conclusion:
The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for the IMMULITE Toxoplasma IgM and IMMULITE 2000 Toxoplasma IgM assays.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in a sans-serif font and is evenly spaced around the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 1 0 2002
Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles, CA 90045-5597
Re: k012075
Trade/Device Name: IMMULITE® and IMMULITE® 2000 Toxoplasma IgM Regulation Number: 21 CFR 866.3780 Regulation Name: Toxoplasma gondii serological reagents Regulatory Class: Class II Product Code: LGD Dated: November 30, 2001 Received: December 3, 2001
#### Dear Dr. Levine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{13}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
#### 510(k) Number: Kolao 75 IMMULITE® Toxoplasma IgM and Device Name: IMMULITE® 2000 Toxoplasma IgM
### Indications For Use:
IMMULITE® Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening.
IMMULITE® 2000 Toxoplasma IgM - For in vitro diagnostic use with the IMMULITE® 2000 Analyzers - for the presumptive qualitative detection of IgM antibodies to Toxoplasma gondii in human serum, particularly for women of childbearing age. When performed in conjunction with a Toxoplasma IgG assay, the IMMULITE® 2000 Toxoplasma IgM can be used as an aid in the presumptive diagnosis of acute, recent or reactive Toxoplasma gondii infection. This product has not been cleared/approved by the FDA for blood/plasma donor screening.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubois
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K012075
\$\checkmark\$
**Prescription Use**
**(Per 21 CFR 801.109)**
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.