RESPIEVENTS SOFTWARE IS EVENT DATA PROCESSING SOFWARE FOR A PC
K012020 · Non-Invasive Monitoring Systems, Inc. · DQK · Aug 21, 2001 · Cardiovascular
Device Facts
Record ID
K012020
Device Name
RESPIEVENTS SOFTWARE IS EVENT DATA PROCESSING SOFWARE FOR A PC
Applicant
Non-Invasive Monitoring Systems, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Aug 21, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
RespiEvents™ is a software package running on a personal computer that is intended to provide analysis of breathing patterns from RespiTrace technology, aid in identifying and classifying apneas, displaying heart rate changes from electrocardiographic waveforms, logging values from pulse oximetry, and displaying signals from physiologic recording devices such as a body position sensor and impedance pneumograph, in the wake and sleeping states as well as activities of daily living.
Device Story
RespiEvents™ software (v5.2) runs on PC; processes physiologic data from RespiTrace™ technology, ECG, pulse oximetry, body position sensors, and impedance pneumographs. Captures full-fidelity data; provides playback and analysis of breathing patterns and apneas; displays heart rate and physiologic signals. Used by qualified personnel in clinical or home settings to assist in identifying/classifying apneas. Output presented via Windows-based GUI; aids healthcare providers in clinical decision-making regarding respiratory status. No alarm functions provided.
Clinical Evidence
Bench testing only. Verification and validation performed using simulated input signals, manual operations, and recordings of actual physiological waveforms to confirm compliance with user requirements and performance specifications.
Technological Characteristics
PC-based software; Windows 95/98 compatible; Windows User Interface. Processes physiologic signals (respiratory, ECG, pulse oximetry, impedance, position). No changes to core algorithms or fundamental scientific technology from predicate.
Indications for Use
Indicated for patients requiring analysis of breathing patterns, apnea identification/classification, heart rate monitoring via ECG, pulse oximetry logging, and physiologic signal display (body position, impedance pneumography) during wake, sleep, and daily activities.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
K012020
Image /page/0/Picture/1 description: The image shows the logo for Non-Invasive Monitoring Systems Inc. The logo features the company name in two lines, with the acronym "Nims" in a bold, stylized font above the full name. The word "Nims" is set against a black background.
Special 510(k) Submission
#### 510(k) Summary APPENDIX G
AUG 2 1 2001
510(k) Summary For Non-Invasive Monitoring Systems, Inc. RespiEvents™ software, Version 5.2
June 24, 2001 DATE THIS SUMMARY WAS PREPARED:
SUBMITTER'S NAME AND ADDRESS:
Non-Invasive Monitoring Systems, Inc. 1840 West Avenue Miami Beach, Florida 33139
### CONTACT PERSON:
Allan F. Brack, Chief Executive Officer Telephone: (305) 534-3694 Facsimile: (305) 534-9368
#### DEVICE NAME:
Proprietary or Trade Name: RespiEvents™ software, Version 5.2 Event data processing software for a PC Common Name: Programmable Diagnostic Computer Classification Name:
### PREDICATE DEVICE:
The legally marketed device to which equivalence is claimed is:
RespiEvents™ software, Version 4.2, K001369, cleared on July 26, 2000.
- 74 DQK * Product Code:
- 870.1425 * CFR Section:
### DEVICE DESCRIPTION:
RespiEvents™ is a software analysis program running on a personal computer. RespiEvents™ presents waveforms collected from RespiTrace™ technology-based RespiEvelts "prosents wavelorm" and RespiTrace PT™ and is intended to provide analysis of:
> 1840 West Avenue, Miami Beach, FL 33139, USA Tel: 1 305 534 3694 - Fax: 1 305 534 9368 http://www.nims-inc.com - nims@respitrace.com
Non-Invasive Monitoring Systems, Inc. RespiEvents™ v5.2
June 2001 APPENDIX G, page G01
{1}------------------------------------------------
- Breathing patterns from Respitrace™ technology, ● ·
- Aid in identifying and classifying apneas, .
- Displaying heart rate changes from electrocardiographic waveforms, .
- Logging values from pulse oximetry, and .
- Displaying signals from physiologic recording devices .
The purpose of RespiEvents™ software is to capture full fidelity Respitrace™ data on a continuous basis, and to provide playback and analysis of the captured data.
The system presents the information from physiologic recording devices such as a body position sensor and impedance pneumograph, in the wake and sleeping states. The system does not present alarms, alarm functions, or alarm management.
# INTENDED USE:
RespiEvents™ is a software package running on a personal computer that is intended to provide analysis of breathing patterns from RespiTrace technology, aid in identifying and classifying apneas, displaying heart rate changes from electrocardiographic waveforms, logging values from pulse oximetry, and displaying signals from physiologic recording devices such as a body position sensor and impedance pneumograph, in the wake and sleeping states as well as activities of daily living.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
This submission is for modifications to the RespiEvents™ software, version 4.2 that was cleared for marketing under Premarket Notification K001369. The RespiEvents™ software, version 5.2 is a modification to the predicate device which does not change the intended use or the fundamental scientific technology.
The RespiEvents™ software has been modified to operate on PC running Microsoft Windows 95 or 98 and presents the user with an ergonomically acceptable typical Windows User Interface. The software also has several re-named traces for greater user clarity. No modifications were made to the core software, algorithms, or fundamental scientific technology.
### NONCLINICAL TESTS USED IN DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
The new software has undergone complete verification and validation test scenarios to show that all functions are in compliance with the user requirements, safety standards, and performance specifications. This involved the use of simulated input signals, manual operations and recordings of actual physiological waveforms.
# CONCLUSIONS FROM NONCLINICAL TESTING:
The testing of the RespiEvents™ software, version 5.2 demonstrates the performance is substantially equivalent to the predicate device. When used in accordance with the directions for use, by qualified personnel, RespiEvents™ software, version 5.2, is safe and effective, as indicated, for its intended use.
Non-Invasive Monitoring Systems, Inc. RespiEvents™ v5.2
June 2001 APPENDIX G, page G02
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its body and wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 1 2001
Mr. Allan F. Brack Chief Executive Officer Non-Invasive Monitoring Systems, Inc. 1840 West Avenue Miami Beach, FL 33139
Re: K012020
Trade Name: RespiEvents™ Software, Version 5.2 Regulation Number: 21 CFR 870.1425 Regulatory Class: II (two) Product Code: 74 DQK Dated: August 3, 2001 Received: August 6, 2001
Dear Mr. Brack:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Allan F. Brack
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4645. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Dank Mella
Tames E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT APPENDIX B
RespiEvents™ software, version 5.2 Indications for Use
510(k) Number:
K012020
Device Name:
RespiEvents™ software, version 5.2
Indications for Use:
RespiEvents™ is a software package running on a personal computer that is intended to provide analysis of breathing patterns from RespiTrace technology, aid in identifying and classifying apneas, displaying heart rate changes from electrocardiographic waveforms, logging values from pulse oximetry, and displaying signals from physiologic recording devices such as a body position sensor and impedance pneumograph, in the wake and sleeping states as well as activities of daily living.
This is the same intended use as most recently cleared for the NIMS RespiEvents DOS software, version 4.2, 510(k): K001369, dated July 26, 2000.
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
**Division of Cardiovascular & Respiratory Devices**
**510(k) Number** K012030
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.