Heparin Lock Flush Solution, USP syringes are intended for use in maintaining patency of vascular access devices (VAD'S).
Device Story
Pre-filled syringe containing Heparin Sodium (10u/mL or 100u/mL) in 0.9% Sodium Chloride solution; used to flush and maintain patency of vascular access devices (VADs). Device consists of three-piece piston design with polypropylene barrel; aseptically filled. Available in various volumes (2mL to 6mL) and configurations, including kits with saline syringes and blunt plastic cannulas. Operated by clinicians in healthcare settings to prevent occlusion of vascular access lines.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design similarities to legally marketed predicate devices.
Technological Characteristics
Three-piece piston syringe; polypropylene barrel; sterile Heparin Sodium (10 or 100U/mL) and 0.9% Sodium Chloride solution; aseptically filled; manual operation; various fill volumes (2-6mL) in 3, 5, or 10mL syringe sizes.
Indications for Use
Indicated for maintaining patency of vascular access devices (VADs) in patients requiring such access.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
1 Becton Drive Franklin Lakes, New Jersey 07417 tel: 201.847.6800 www.bd.com
K04967
JAN 2 3 2002
Image /page/0/Picture/3 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a stylized sun-like symbol on the left and the letters "BD" on the right. The sun-like symbol has a circle with rays emanating from the top and a stylized figure at the bottom. The letters "BD" are in a bold, sans-serif font.
Indispensable to human health
## Summary of Safety and Effectiveness for the Heparin Lock Flush Solution, USP BD Pre-Filled Heparin Lock Flush Syringe
1 BD Contact person:
Gregory W. Morgan Head of Regulatory BD Medical Surgical - Mail Code 226 1 Becton Drive Franklin Lakes, NJ 07417-1880 Phone (201) 847-4344 (201) 847-4855 Fax
- 2 Device Name: Heparin Lock Flush Solution, USP BD Pre-Filled Heparin Lock Flush Syringe
- 3 Predicate Device(s):
- 3.1 Vital Signs Heparin Lock Flush Device K953120
- 3.2 BD Preefil™ Normal Saline Flush Syringe K982558
- 3.3 Baxter Heparin Lock Flush Syringe K003245
- 4 Product Description / Function:
Product sizes/reorder numbers: The following sizes/reorder numbers of the Heparin Lock Flush Solution, USP syringes to be offered are:
| BD | |
|--------|--------------------------------------------|
| Cat. # | Description |
| 306509 | Heparin, 10u/mL, 6mL Fill In 10mL Syringe |
| 306510 | Heparin, 10u/mL, 5mL Fill In 10mL Syringe |
| 306511 | Heparin, 10u/mL, 5mL Fill In 5mL Syringe |
| 306521 | Heparin, 10u/mL, 3mL Fill In 10mL Syringe |
| 306512 | Heparin, 10u/mL, 3mL Fill In 3mL Syringe |
| 306527 | Heparin, 10u/mL, 2mL Fill In 3mL Syringe |
| 306513 | Heparin, 100u/mL, 5mL Fill In 10mL Syringe |
| 306514 | Heparin, 100u/mL, 3mL Fill In 10mL Syringe |
{1}------------------------------------------------
| BD<br>Cat. # | Description |
|--------------|----------------------------------------------------------------------------------------------------------------------------------------|
| 306515 | Heparin, 100u/mL, 5mL Fill In 5mL Syringe |
| 306516 | Heparin, 100u/mL, 3mL Fill In 5mL Syringe |
| 306517 | Heparin, 100u/mL, 3mL Fill In 3mL Syringe |
| 306528 | Heparin, 100u/mL, 2mL Fill In 3mL Syringe |
| 306525 | Heparin, 10u/mL, 5mL Fill In 10mL Syringe With Blunt Plastic Cannula |
| 306531 | Heparin, 100u/mL, 5mL Fill In 10mL Syringe With Blunt Plastic Cannula |
| 306529 | Heparin, 100u/mL, 5mL Fill In 5mL Syringe With Blunt Plastic Cannula |
| 306537 | SASH Kit (2 - 0.9% Sodium Chloride Syringes and 1 - 100u/mL Heparin Lock<br>Flush Syringe), 5mL Fill In 10mL Syringe - 100u/mL Heparin |
| 306538 | SASH Kit (2 - 0.9% Sodium Chloride Syringes and 1 - 10u/mL Heparin Lock<br>Flush Syringe), 3mL Fill In 3mL Syringe - 10u/mL Heparin |
| 306539 | SASH Kit (2 - 0.9% Sodium Chloride Syringes and 1 - 100u/mL Heparin Lock<br>Flush Syringe), 3mL Fill In 5mL Syringe - 100u/mL Heparin |
Intended Uses: Heparin Lock Flush Solution, USP syringes are intended for use in maintaining patency of vascular access devices (VAD'S).
- 5 Equivalence determination:
The elements of comparison between the Heparin Lock Flush Solution. USP syringes with the BD Preefil™ Normal Saline Flush syringe, the Baxter Heparin Lock Flush syringe and Vital Signs Heparin Lock Flush Device predicate devices are as follows:
- The Heparin Lock Flush Solution, USP syringe contains sterile flush (0.9% ● Sodium Chloride Injection, USP and Heparin Sodium 10 and 100U/mL solution as does the predicate devices).
- The Heparin Lock Flush Solution, USP syringe has a sterile solution contents . and fluid path as do the predicate devices.
- The Heparin Lock Flush Solution, USP syringe is aseptically filled as the . predicate device, BD Preefil TM.
- The Heparin Lock Flush Solution, USP syringe is a three-piece piston design . as are the predicate devices, BD Preefil™and the Baxter Heparin Lock Flush syringe.
- . The Heparin Lock Flush Solution, USP syringe barrel is molded from polypropylene as are the predicate devices, BD Preefil™and the Baxter Heparin Lock Flush syringe.
- . The Heparin Lock Flush Solution, USP syringe has a similar package unit design as the predicate device, BD Preefil ™.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three stylized human figures.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 3 2002
Mr. Gregory W. Morgan BD Medical Surgical BD Medical Injections Systems 1 Becton Drive, MC 226 Franklin Lakes, New Jersey 07417
Re: K011967
Trade/Device Name: BD Pre-Filled Heparin Lock Flush Syringe Regulation Number: 880.5200 Regulation Name: Heparin Lock Flush Syringe Regulatory Class: II Product Code: NGT Dated: November 7, 2001 Received: November 8, 2001
Dear Mr. Morgan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Mr. Morgan
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/adrh/dsma/dsmamain.html
Sincerely yours,
Timo Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
The Heparin Lock Flush Solution, USP syringes are intended for use in maintaining
patency of vascular access devices (VAD'S).
Patricie Cucenate
Division Sign. Oto
sion Sign-Off) sion of Dental, Infection Corn Fral Hospital T
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.