DOXADENT DENTAL COMPOSITE, MODELS A, B, C, 12A, 12B, 12C, 6A, 6B, AND 6C
K011839 · Doxa Certex AB · EBF · Jan 17, 2002 · Dental
Device Facts
Record ID
K011839
Device Name
DOXADENT DENTAL COMPOSITE, MODELS A, B, C, 12A, 12B, 12C, 6A, 6B, AND 6C
Applicant
Doxa Certex AB
Product Code
EBF · Dental
Decision Date
Jan 17, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DoxaDent is a dental ceramic composed of calcium aluminate cement and oxides that is intended to restore carious lesions or structural defects in teeth. Specifically, DoxaDent is indicated for use as a restorative dental material for permanent application in the treatment of class I, II, and V cavities.
Device Story
DoxaDent is a calcium aluminate-based dental ceramic restorative material. It is applied by dental professionals to treat class I, II, and V cavities. The material consists of a blend of calcium aluminate cement, silica, zirconium oxide, and iron oxide colorants, with alkali salt accelerators to control hydration. It functions as a permanent restorative material for structural tooth defects. Clinical benefits include restoration of tooth integrity and function. The device is used in a clinical setting by dentists.
Clinical Evidence
Bench testing included flexural properties, shade, color stability, radio-opacity, working/setting times, marginal adaptation in extracted teeth, microstructure, compressive strength, acid erosion (EN 29917:1994/ISO 9917:1991), dimensional stability, wear, leakage/pH, electrical pulp sensitivity, and hardness. Biocompatibility and chemical stability were evaluated. Clinical data consisted of X-ray analysis of implants and evaluation of gingival reactions and plaque formation.
Technological Characteristics
Dental ceramic composed of calcium aluminate cement, silica, zirconium oxide, and iron oxide colorants. Particle sizes: 0.5-5 um and 0.02-0.2 um. Additives: LiCl (<0.1% w/w) as accelerator. Acid erosion testing per EN 29917:1994/ISO 9917:1991. Material is a permanent restorative cement.
Indications for Use
Indicated for permanent restoration of class I, II, and V dental cavities in patients requiring treatment for carious lesions or structural tooth defects.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
K011839
# 510(k) SUMMARY
# Doxa Certex AB's DoxaDent™
JAN 1 7 2002
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Doxa Certex AB Axel Johanssons gata 4 - 6, SE-75451 Uppsala, Sweden +46-18-4782000 Phone: Facsimile: +46-18-4782071
Contact Person: Nils-Otto Ahnfelt
Date Prepared: June 12, 2001
# Name of Device and Name/Address of Sponsor
DoxaDent™
Doxa Certex AB Axel Johanssons gata 4 - 6 SE-75451 Uppsala, Sweden
### Common or Usual Name
DoxaDent Dental Ceramic
## Classification Name
Tooth Shade Resin Materials
#### Predicate Devices
- > Dentsply AP Dyract® Compomer dental composites
- > Bisco, Inc.'s PYRAMID™, ADVENT™, and RENEW™ dental composites
- > Coltene/Whaledent's Coltosol Temporary Endodontic Filling
{1}------------------------------------------------
# Intended Use/Indications for Use
DoxaDent is a dental ceramic composed of calcium aluminate cement and oxides that is intended to restore carious lesions or structural defects in teeth. Specifically, DoxaDent is indicated for use as a restorative dental material for permanent application in the treatment of class I, II, and V cavities.
# Technological Characteristics and Substantial Equivalence
DoxaDent™ is a dental ceramic composed of a calcium aluminate cement and oxides, i.e., silica and zirconium oxide, and iron oxide colorants, as a filler material, in a blend of fine irregularly shaped particles ranging from 0.5 to 5 um in diameter and microfine particles having a diameter of 0.02 to 0.2 um. Trace concentrations of iron oxide additives (< 0.05% w/w) and alkali salt accelerators (LiCl, < 0.1% w/w) are included to provide for coloring and to control the rate of hydration, respectively.
DoxaDent™ is substantially equivalent to a combination of Dentsply AP Dyract® Compomer, Bisco, Inc.'s PYRAMID™, ADVENT™, and RENEW™ dental composites, as well as Coltène/Whaledent's Coltosol Temporary Endodontic Filling. Any minor difference do not raises any new issues of safety or effectiveness.
#### Performance Data
The following performance test were conducted by Doxa Certex AB on the DoxaDent™:
- · Flexural properties
- ◆ Shade
- � Color stability
- Radio-opacity ◆
- Working and setting times �
- Study on the marginal adaptation of � DoxaDent™ in extracted teeth
- Microstructure of DoxaDent™ ◆ ◆
- Determination of compressive strength
- Some aspects of biocompatibility and chemical stability of a calcium-aluminate-hydrate-based dental restorative material, submitted for publication to the 2001 IADR meeting in Chiba, -Japan
{2}------------------------------------------------
- ◆ Acid erosion by impinging jet technique according to EN 29917:1994/ISO 9917:1991 and Technical Corrigendum 1:1993 Dental waterbased cements
- Dimensional stability �
- ♦ Wear testing
- ◆ Leakage (Leaching) and pH
- Electrical pulp sensitivity test ◆ ◆
- ♦ Hardness
In all instances, the DoxaDent™ functioned as intended and performance observed was as expected.
### Clinical Data
DoxaCertex AB performed X-ray analysis of DoxaDent implants and evaluation of gingival reactions and plaque formation
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 7 2002
Doxa Certex AB C/O Mr. Jonathan S. Kahan Hogan and Harton 555 13th Street Northwest Washington, District of Columbia 20004-1109
Re: K011839
Trade/Device Name: DoxaDent™ Regulation Number: 872.3690 Regulation Name: Tooth Shade Resin Materials Regulatory Class: II Product Code: EBF Dated: October 31, 2001 Received: October 31, 2001
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
#### Page 2 - Mr. Kahan
You must comply with ail the Act's requirements, including, but not limited to: registration r od intist comply with and 10 ; labeling (21 CFR Part 801); good manufacturing practice alle libang (2) 02 10 11 the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 11ms letter will and will be rogen of the FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the and additionally in the at (301) 594-4613 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
K011839 510(k) Number (if known):
DoxaDent™ Device Name:
Indications for Use:
DoxaDent is a dental ceramic composed of calcium aluminate cement and oxides that is intended to restore carious lesions or structural defects in teeth. Specifically, DoxaDent is indicated for use as a restorative dental material for permanent application in the treatment of class I, II, and V cavities.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use**
(Per 21 CFR 801.109)
Over-The-Counter Use _ (Optional Format 1-2-96)
wnges 'Division Sign-Off) Division of Dental, Infection Control, and General Hospita 510(k) Number _
OR
\\\DC - 69389/1 - #1335262 v1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.