K011781 · Surgi-Vision, Inc. · MOS · Aug 31, 2001 · Radiology
Device Facts
Record ID
K011781
Device Name
INTERCEPT-URETHRAL MICROCOIL
Applicant
Surgi-Vision, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 31, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Intercept-Urethral Microcoil is recommended for high-resolution Magnetic Resonance Imaging of the male and female urethra and surrounding tissue. The single use, disposable coil was designed to be inserted in the urethra of the patient during MRI scans in order to obtain improved image quality in the anatomical regions surrounding the urethra including the prostate in males. The flexible coil facilitates placement of the coil in the anatomy. This product is to be used with a 1.5T GE Signa MRI machine. The signals picked up by the coil are conducted through a small coaxial cable to a connection with the standard surface coil connector for MRI systems. The coil and cable are completely sealed inside the insulating layer.
Device Story
Intracavitary specialty coil; inserted into male/female urethra during MRI scans; captures anatomical signals from surrounding tissue; signals transmitted via coaxial cable to standard MRI surface coil connector; used with 1.5T GE Signa MRI machine; provides improved image quality for prostate and urethral regions; single-use disposable device; flexible design facilitates anatomical placement; sealed in insulating layer to prevent direct patient contact.
Clinical Evidence
Bench testing only. Evaluated RF heating (no excess heating vs. control), peripheral nerve stimulation (no increased susceptibility), imaging performance in male/female urethra, and mechanical properties (stiffness, tensile, tracking, torque). All performance specifications met.
Technological Characteristics
Intracavitary specialty MRI coil; flexible form factor; single-use disposable; coaxial cable interface; compatible with 1.5T GE Signa MRI systems; sealed in insulating layer; passive sensing principle.
Indications for Use
Indicated for high-resolution MRI of male and female urethra and surrounding tissue, including prostate in males, during MRI scans.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# Kol
## 510(k) Summary
## Intercept-Urethral Microcoil
Common/Classification Name: Accessory to Magnetic Resonance Diagnostic Device, 21 CFR 892.1000
> Surgi-Vision, Inc. 20 Firstfield Road, Suite 200 Gaithersburg, MD 20878
Prepared: June 7, 2001 Contact: Nancy E. Taylor,
#### LEGALLY MARKETED DEVICE A.
The Intercept-Urethral Microcoil was cleared for marketing as the The Intercept-Orchhar Microcon November 28, 2000 in prospect Surgi-Vision, Inc. submits the present notification K0022916. notification Koozzo For Suignifications for use for use for the Surgipremarket notheation of dadinoner of the Intercept-Urethral Microcoil.
#### DEVICE DESCRIPTION B.
The Intercept-Urethral Microcoil (IUMC) is an intracavitary specialty The Intercept of ordinal Micros the anatomical regions surrounding the coll for use in NH muging or he coil are conducted through a urethia. The orghale pronterfacing network to a connection with the standard surface coil connector for MRI systems.
The imaging coil and the coaxial cable are completely sealed inside the The imaging our and the coil from direct contact with any insulating slevvo. This lociated ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . is advanced into the subject's urethra and positioned in the region of ls auvanced into the babjoor atching and decoupling circuit is located interest: "The balan, values outside the body.
#### INTENDED USE C.
The Intercept-Urethral Microcoil is recommended for high-resolution Magnetic Resonance Imaging of the male and female urethra and
Surgi-Vision, Inc. Intercept-Urethral Microcoil June 7, 2001
Exhibit 1-Page 1
{1}------------------------------------------------
surrounding tissue. The single use, disposable coil was designed to be inserted in the urethra of the patient during MRI scans in order to obtain improved image quality in the anatomical regions surrounding the urethra including the prostate in males. The flexible coil facilitates placement of the coil in the anatomy. This product is to be used with a 1.5T GE Signa MRI machine. The signals picked up by the coil are conducted through a small coaxial cable to a connection with the standard surface coil connector for MRI systems. The coil and cable are completely sealed inside the insulating layer.
#### SUBSTANTIAL EQUIVALENCE SUMMARY D.
This premarket notification concerns only additional intended uses for the IUMC. Concurrently, the name will be changed from the Surgi-Vision Prostate Coil to the Intercept Urethral Microcoil.
In all other aspects, the IUMC is the device that was cleared for marketing by FDA on November 28, 2000, in premarket notification K002916.
#### TECHNOLOGICAL CHARACTERISTICS E.
See Section D, above.
#### E. TESTING
Surgi-Vision carried out testing and/or analysis of the Intercept-Urethral Microcoil that addressed the following issues:
- Possibility of excess RF heating; (1)
- Possibility of increased susceptibility of patients to peripheral (2) nerve stimulation;
- (3) Imaging performance; and
- (4) Mechanical testing.
The results of the heating experiments demonstrate that there is no Surgi-Vision, Inc. Exhibit 1-Page 2 Intercept-Urethral Microcoil June 7, 2001
{2}------------------------------------------------
excess heating when IUMC is positioned in a phantom that is representative of worst case clinical conditions. The change in temperature observed during use of the IUMC is not significantly different then that observed without the coil.
The calculations done to determine current leakage by the MRI pulsed gradient field demonstrate that there is no possibility of increased susceptibility of patients to nerve stimulation.
The imaging performance was evaluated in the male and female urethra. These images are shown in Exhibit 7.
Mechanical test results demonstrated that there are no safety issues and that all of the performance specifications were met. Stiffness, tensile, tracking, torque, and electrical testing were all completed to the satisfaction of the specifications. Mechanical flexing resulted in insignificant changes in the electrical properties of the coils. These results are described in Exhibit 8.
#### G. CONCLUSIONS
This pre-market submission has demonstrated the IUMC is safe and effective for the additional indications as required and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
Surgi-Vision, Inc. Intercept-Urethral Microcoil June 7, 2001
Exhibit 1-Page 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 3 1 2001
Ms. Nancy E. Taylor CEO/President Surgi-Vision Inc. 20 Firstfield Road, Suite 200 GAITHERSBURG MD 20878 Re: K011781
Intercept Urethral Microcoil (MRI specialty coil) Dated: June 7, 2001 Received: June 7, 2001 Regulatory Class: II 21 CFR 892.1000/Procode: 90 MOS
### Dear Ms. Taylor:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have is w marked for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate substance prior to May 28, 1976, the enactment date of the Medical Device that have been reclassified in econdance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to ate general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent decemination assumes compliance with the Current Good Manufacturing Practice requirements, 000 to 09:51 Pater Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic as so rolling the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please rotation response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
> (301) 594-4591 8xx.1xxx (301) 594-4616 876.2xxx, 3xxx, 4xxx, 5xxx (301) 594-4616 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx (301) 594-4654 892.2xxx, 3xxx, 4,xxx, 5xxx (301) 594-4692 Other
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(K) Number (if known): _ Ko | 1781
Device name: Intercept-Urethral Microcoil
Indications for Use:
The Intercept-Urethral Microcoil is recommended for high-resolution Magnetic Resonance Imaging of the male and female urethra and magnetis is tissue. The single use, disposable coil was designed to be inserted in the urethra of the patient during MRI scans in order to obtain improved image quality in the anatomical regions surrounding the urethra including the prostate in males. The flexible coil facilitates placement of the coil in the anatomy. This product is to be used with a 1.5T GE Signa MRI machine. The signals picked up by the coil are conducted through a small coaxial cable to a connection with the standard surface coil connector for MRI systems. The coil and cable are completely sealed inside the insulating layer.
(Please do not WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use
Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
Surgi-Vision, Inc.
Intercept-Urethral Microcoil
June 7, 2001
and Radiological Devices
510(k) Number
K011781
Page III-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.