DC - CRISTAL

K011674 · Austenal, Inc. · EIH · Sep 4, 2001 · Dental

Device Facts

Record IDK011674
Device NameDC - CRISTAL
ApplicantAustenal, Inc.
Product CodeEIH · Dental
Decision DateSep 4, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

The indications for use for DC-Cristall is as a substructure for single unit porcelain fused ceramic fixed dental restorations; namely crowns. Substructures of DC-Cristall are machined using the DCS Precident CAD/CAM System. The porcelain veneer is applied using dental porcelains currently on the market.

Device Story

DC-Cristall is a glass ceramic material supplied in block form; serves as feedstock for the DCS Precident CAD/CAM System. The material is machined into substructures for single unit porcelain fused ceramic crowns. Dental professionals apply porcelain veneers to these substructures using commercially available dental porcelains. The device functions as a restorative material component within a CAD/CAM workflow to provide structural support for dental crowns.

Clinical Evidence

Bench testing only; compliance with ISO 6872:1999 Dental Ceramics.

Technological Characteristics

Glass ceramic material in block form. Designed for use with DCS Precident CAD/CAM System. Complies with ISO 6872:1999 Dental Ceramics.

Indications for Use

Indicated for use as a substructure for single unit porcelain fused ceramic fixed dental restorations (crowns) in patients requiring dental restoration.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Submission Summary (Full Text)

{0}------------------------------------------------ ## KO11674 ## 510(k) Premarket Notification Summary Austenal Inc. Name & Address: AUSTENAL, INC. 5101 S. Keeler Ave Chicago, IL 60632-4287 (773) 735-0600 FAX (773) 735-3940 Contact: Ronald Dudek Date Prepared: August 23, 2001 Trade or Proprietary Name: DC-Cristall for the DCS Precident CAD/CAM System 76EIH -Powder, Porcelain, Classification Name: Note: At this time a classification does not exist for a solid ceramic material. 76EIH is the closest based on intended use. Device Description: The DC-Cristall material is a glass ceramic material, furnished in a block form, designed for use as feed stock in the DCS Precident CAD/CAM System. Intended Use: The indications for use for DC-Cristall is as a substructure for single unit porcelain fused ceramic fixed dental restorations; namely crowns. Substructures of DC-Cristall are machined using the DCS Precident CAD/CAM System. The porcelain veneer is applied using dental porcelains currently on the market. Technological Characteristics: DC-Cristall is a ceramic material that is similar in composition to other ceramic materials that are currently offered in the dental marketplace for the fabrication of single unit crowns. DC-Cristall is substantially equivalent in concept to predicate devices Substantial Equivalence: (i.e. ceramic materials that are used in the CAD/CAM system) intended for use as the substructure of porcelain fused to ceramic crowns. The safety and efficacy of DC-Cristall is assured by compliance with ISO 6872:1999 Dental Ceramics. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP = 4 2001 Mr. Ronald Dudek Director of Technical Resources Austenal, Incorporated 4101 West 51st Street Chicago, Illinois 60632-4287 K011674 Re : DC-Cristall Trade/Device Name: 872.6660 Requlation Number: Requlatory Class: II Product Code: EIH Dated: August 23, 2001 Received: August 27, 2001 Dear Mr. Dudek: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the marker is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements {2}------------------------------------------------ Page 2 - Mr. Dudek concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Timothy L. Wheatridge Timo hy A. Ulatowski Directbr Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K011674 Page 1 of 1 ## INDICATIONS FOR USE STATEMENT As Required by CFR 807.87 (e) K011674 510(k) Number (if known): Device Name: DC-Cristall Indications For Use: The indications for use for DC-Cristall is as a substructure for single unit porcelain fused ceramic fixed dental restorations; namely crowns. Substructures of DC-Cristall are machined using the DCS Precident CAD/CAM System. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The Counter Use (Optional format 1-2-96) Susan Hauser (Division Sign-Off) Division of Dental, Infection Control, and General Hospital I 510(k) Number _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...