VALIDATION OF THE BREATHID SYSTEM, FOR THE DETECTION OF HELICBACTOR PYLORI
Applicant
Oridion Medical 1987 , Ltd.
Product Code
MSQ · Microbiology
Decision Date
Jul 9, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3110
Device Class
Class 1
Indications for Use
The BreathID™ System is used to diagnose and monitor Helicobacter pylori infection by measuring changes in the 13CO2/12CO2 ratio in a patient's breath following the ingestion of 13C urea. The Oridion BreathID™ system continually and non-invasively measures changes in the 13CO2/12CO2 ratio in exhaled breath, which may be indicative of increased urease production associated with active Helicobacter pylori infection in the stomach. The Oridion BreathID™ System is to be used as an aid for initial diagnosis and post treatment monitoring of Helicobacter pylori infection.
Device Story
BreathID™ system is a non-invasive diagnostic tool for H. pylori detection. Input: patient exhaled breath collected continuously via nasal cannula (FilterLine™) after ingestion of 13C-enriched urea tablet and Citrica test meal. Operation: device measures and computes changes in 13CO2/12CO2 ratio using proprietary Molecular Correlation Spectroscopy (MCS) technology. Output: real-time ratio measurements displayed to clinician. Usage: clinic setting; operated by trained personnel under physician/nurse supervision. Clinical decision-making: results aid in initial diagnosis and post-treatment monitoring of H. pylori infection. Benefit: non-invasive, continuous monitoring, avoids biopsy.
Clinical Evidence
Pivotal study (Sept 1999-June 2000) evaluated 315 pre-therapy and 77 post-therapy subjects. Pre-therapy: sensitivity 100%, specificity 99.2% vs. endoscopic/CLOtest® results. Post-therapy: sensitivity 100%, specificity 100% vs. endoscopic/CLOtest® results. Comparison to histology showed 95.8% sensitivity/97.7% specificity (pre-therapy) and 100% sensitivity/95.2% specificity (post-therapy). Study included patients on PPI/H2 therapy with no reported false negatives. Two mild adverse events reported.
Technological Characteristics
System utilizes Molecular Correlation Spectroscopy (MCS) for gas analysis. Continuous sampling via Oridion Nasal Filterline. Energy source: electrical. Form factor: benchtop measurement device. Reagents: 13C-urea tablet (75mg) and Citrica powder. Connectivity: not specified. Sterilization: not specified.
Indications for Use
Indicated for adult patients for the qualitative detection of urease associated with Helicobacter pylori in the stomach to aid in initial diagnosis and post-treatment monitoring of H. pylori infection.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
{0}------------------------------------------------
K011668
JUL - 9 2001
Image /page/0/Picture/2 description: The image shows the logo for Oridion. The logo consists of a stylized globe made of swirling lines above the word "Oridion" in a serif font. The globe is black and white, and the word "Oridion" is in black.
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR807.92
The assigned 510(k) number is
### PRODUCT NAME
#### TRADE/PROPRIETARY NAME
BreathID™ System
#### COMMON NAME
13C-Urea Breath Test for the Presence of Helicobacter pylori
### CLASSIFICATION NAME
MSQ Campylobacter pylori. The urea breath test was recently reclassified from LYR to MSQ
## ESTABLISHMENT ADDRESS:
Oridion Medical 1987 Ltd. 7 HaMarpe St. Har Hotzvim Science Based Industrial Park POB 45025 91450 Jerusalem, Israel
#### ESTABLISHMENT REGISTRATION NUMBER
NUMBER: 8044004
#### DEVICE LISTING FDA FORM 2892:
B073773
{1}------------------------------------------------
Oridion
.
Table of comparison to legally marketed predicate devices
| Characteristics | BreathID™ System | CLOtest®<br>K882199 | MERETEK UBT®<br>Breath Test K952220 | MERETEK UBT® Breath<br>Test K K972352 |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Test measurement<br>device | Oridion BreathID™ Test<br>Device | Visual observation<br>of color change | Gas Isotope Ratio Mass<br>Spectrometer | Gas Isotope Ratio Mass<br>Spectrometer |
| Test Sample | Gas Sample<br>continuously<br>transported to test<br>measurement device<br>by Oridion nasal<br>cannula Filterline<br>K980324 | Sample is biopsy<br>specimen | Gas sample stored in<br>specially designed<br>breath collection bag | Gas sample stored in<br>specially designed breath<br>collection bag |
| 13C-Urea | Raw material supplier<br>CIL, 75mg Tablet<br>dissolved in water.<br>Manufacturer and<br>Packager CIL (NDA<br>#21-314 submitted) | NA | Raw material supplier<br>ISOTEC, 125mg powder<br>(Pranactin®) (in a glass<br>vial) dissolved in water.<br>Manufacturer and<br>Packager unknown | Raw material supplier<br>ISOTEC, 125mg powder<br>(Pranactin®) (in a glass<br>vial) dissolved in water.<br>Manufacturer and<br>Packager unknown |
| Applicable pre and<br>post-treatment | Yes | Yes | No | Yes |
| Characteristics | BreathID™ System | CLOtest® K882199 | MERETEK UBT® Breath Test K952220 | MERETEK UBT® Breath Test K K972352 |
| Test Meal | Citrica 4.5 gr dissolved in water | NA | Ensure pudding | Ensure pudding |
| Test results time | 10-30 minutes | NA | Sample must be sent to lab for measurement, could be hours to days | Sample must be sent to lab for measurement, could be hours to days |
| Breath collection | Continuous over test time of 10-30 minutes | NA | One sample before ingestion of 13C-Urea and one sample after 30 minutes. | One sample before ingestion of 13C-Urea and one sample after 30 minutes. |
| Cut off point | 5.0 delta per mil above baseline (post dose minus pre dose) | NA | 2.4 delta per mil above baseline (post dose minus pre dose) | 2.4 delta per mil above baseline (post dose minus pre dose) |
| Intended use | See Below (page 16) | See Below (page 16) | See Below (page 16) | See Below (page 16) |
| Organism | Hp in vivo | Hp in tissue biopsy | Hp in vivo | Hp in vivo |
| Reagent | 13C-Urea | Urea | 13C-Urea | 13C-Urea |
| Result | 13CO2/12CO2 ratio – Molecular Correlation Spectroscopy (MCS) | CO2+NH3 Color change | 13CO2+NH3 Gas Isotope Ratio Mass Spectrometer | 13CO2+NH3 Gas Isotope Ratio Mass Spectrometer |
Page No. 4 of 11
Oridion Medical 1987 Ltd. • POB 45025 • Jerusalem 91450 Israel • Voice:+972 2 589-9115 • Fax:+972 2 582-8873
.
,"
{2}------------------------------------------------
の
Oridion
Oridion Medical 1987 Ltd. • POB 45025 • Jerusalem 91450 Israel • Voice:+972 2 589-9115 • Fax:+972 2 582-8873
Page No. 5 of 11
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Oridion. The logo consists of a stylized globe made of horizontal lines, positioned above the word "Oridion" in a serif typeface. The globe is black and white, and the text is black.
## Intended Use
- Oridion BreathID™ system .
The intended use of the Oridion BreathID™ system is to non invasively measure, in a The intended ase of the Ondion Broad.nl " of catio of exhaled breath after drinking a test drink which includes 13C enriched urea. The system measures urease associated with Helicobacter pylori infection in the stomach to aid in the initial diagnosis and post treatment monitoring of Helicobacter pylori.infection. The detection is accomplished by measuring monitoring of Hollobactor ps 1302 and 1202 using Oridion proprietary (MCS) gas onanges in the ratio botwoon - ove of change in the 1302/202 ratio may be indicative of a physiological or metabolic change in the patient's condition. The system is for use by trained operators under the supervision of physicians, nurses or other healthcare professionals.
The Meretek UBT® Breath test (K972352) e
The intended use of the Meretek test is for use in the qualitative detection of urease associated with Helicobacter pylori in the human stomach and as an aid in the initial diagnosis and post-treatment monitoring of Helicobacter pylori infection in adult patients. This test is essentially equivalent to the test described in Meretek K952220 except that the intended use labeling has been expanded to include post-treatment monitoring of Helicobacter pvlori.
Intended Use (Meretek)K952220 .
The intended use is to non invasively detect urease associated with Helicobacter pylori infection in the stomach, and to aid in the initial diagnosis and post treatment monitoring of Helicobacter pylori. infection.
Intended Use (CLOtest®) .
The intended use is to detect urease associated with Helicobacter pylori infection in the stomach, and to aid in the initial diagnosis and post treatment monitoring of Helicobacter pylori.infection.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Oridion" in a serif font, with a stylized globe-like symbol above it. The globe is made up of horizontal lines that curve around to create a spherical shape. The text is in a dark color, likely black, and the globe symbol has a similar dark tone.
## ORIDION BreathID™ SYSTEM DESCRIPTION
The BreathID™ system is a non-invasive breath test system for detecting urease associated with Helicobacter pylori. The system consists of:
- 1) A medical device (BreathID™ system) to measure and compute the ratio between 1202 and 13CO2 in the patient's exhalation.
- 2) A Test Kit.
#### Test Device
The BreathID™ system is based on Oridion's proprietary CO2 measurement technology. The device will be used to measure and compute changes in the ratio between 1202 and 1302 in the patient's exhalation. The Oridion BreathID test device measurement is made by continuous sampling of the breath.
The CO2 produced in normal breathing contains approximately 99% 12O2 and 1% 13CO2 (12C and 13C are stable isotopes of carbon). The Oridion BreathID™ system measures the changes in ratio between 1302 and 1202 using our proprietary technology. The system is for use by trained operators under the supervision of physicians, nurses or other healthcare professionals.
#### Test Kit
Part of the BreathID™ system is the Test Kit (IDkit™). The Test Kit is used to perform the test for the presence of Helicobacter pylori. The drug in the Test Kit is 13 - Urea. The nasal cannula device used in the kit is an Oridion Nasal Filterline.
#### The Test Kit consists of:
- 1) Oridion Nasal FilterLine™
- 2) A packaged tablet of 13C-urea
- 3) A package of powdered Citrica
- 4) A drinking straw
- 5) Package insert
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Oridion. The logo consists of a stylized globe made up of horizontal lines, positioned above the word "Oridion" in a serif typeface. The globe is textured with varying line thickness, giving it a three-dimensional appearance.
## CLINICAL STUDY SUMMARY
Dates: (Pivotal study)
September 1999-June 2000
#### Subjects:
315 subjects pre-therapy and 77 subjects post-therapy
Objective:
- To evaluate the sensitivity and specificity of the BreathID™ system to detect . the presence of Helicobacter pylori pre-treatment.
#### Results:
Comparison of BreathID™ system results to endoscopic results -- Pre-therapy 1Sensitivity 100%
Specificity 99.2%
Helicobacter pylori positive is defined as positive CLOtest® and positive histology; Helicobacter pylori negative is defined as negative CLOtest® and negative histology. 24 hr CLOtest® results were used to evaluate efficacy.
## Comparison of BreathID™ system results to CLOtest® results -- Pre-therapy
Relative Sensitivity 100%
Relative Specificity 99.2%
Comparison of BreathID™ system results to Histology results - Pre-therapy Sensitivity 95.8%
Specificity 97.7%
## Objective:
- To evaluate the sensitivity and specificity of the BreathID™ system to detect . the presence of Helicobacter pyfori post- treatment, and to evaluate the ability of Oridion's BreathID™ system to monitor the efficacy of treatment.
#### Results:
Comparison of BreathID™ system results to Endoscopic results - Post-therapy Sensitivity 100%
Specificity 100%
Comparison of BreathID™ results to CLOtest® results - Post-therapy
Relative Sensitivity 100%
Relative Specificity 100%
## Comparison of BreathID™ results to Histology results - Post-therapy Sensitivity 100%
Specificity 95.2%
<sup>1</sup> The limits of the 95% 2 sided confidence interval are calculated using exact method Oridion Medical 1987 Ltd. • POB 45025 • Jerusalem 91450 Israel • Voice:+972 2 589-9115 • Fax:+972 2 582-8873
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Oridion. The logo consists of a stylized globe made up of horizontal lines, positioned above the word "Oridion" in a serif typeface. The globe is shaded to give it a three-dimensional appearance, and the word "Oridion" is in a relatively large font.
## Comparison of BreathID™ results to Meretek results – Post-therapy
Relative sensitivity 93.3% Relative specificity 100%
## Objective:
- To evaluate the sensitivity and specificity of the BreathID™ system to detect . the presence of Helicobacter pylori with a "variable time" breath test procedure.
## Results :
Clinical data demonstrated the equivalence of the standard test to the varying time tests regarding the diagnostic results. These results did not depend on clinical stage or on the specific medical center.
## Analysis of the influence of PPI and H2
This multi-site study had no exclusionary criteria for and was indifferent to PPI/H² therapy. No difference was noted between the predicate devices and the experimental device (BreathID™ system), regarding testing accuracy and concurrent therapy. Out of 317 diagnostic patients, 233 (73.5%) were using PPI/H2. Amongst all Helicobacter pylori-positive test results (32), 61.5% were being therapeutically treated by PPI/H². Of these Helicobacter pylori-positive patients using PPI/H², 48.4% had taken the PPI/H² 24 hours prior to testing, while 71.0% had taken their medication within 48 hours of testing.
Percentage of Helicobacter pylori Positive Pre-Therapy Patients According to Days without Medication
| 1711 | Call of the cause of the | And In the later of the later of | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A CONSULTERS OF CONSULTION OF<br>AND A BE THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE FORM THE<br>A LINE OF LE LE CONN CHIEF CHARRES CHARRES | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | and and and and the property of the first of the first of the first of | |
| A AM A C A C C A C C C C C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C<br>- | Section of the country of the bring to the commission in the commission in comments of the comments of the comments of the comments of the comments of the comments of the com<br>Callery of Children of Children of | A LA MANAGER LA METH THE LEAST PRODUCTION CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CONTRACT CARATION CONTRACT CARATION CONTRACT CARATI | Company Call LE BELL LE BELL TOR MERICA CA<br>Annual program free of the art the contract and the contraction of the first and the contribution of | 1 | No and Expection and a mark partie |
| None | And IT's " STYPE PITC<br>and consisted by the state of the comments of | A H A C B & A L A S A C<br>and the American Calif. A A A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A B A<br>AND THE LABOR LE &<br>Call College Company | and studio de collection and consisted and consisted and any and the production of the program and the company of the company of the company of the comments of the comments o | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>Charles Concession College<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
## Conclusions:
There were no false negatives reported for any of the subjects taking PPI or H². The results for subjects taking either PPI or H2 therapy (shown according to therapy) are shown in the tables above.
## ADVERSE EVENTS:
There were only two adverse events in two subjects in this study, both of which were judged as mild. One adverse event was judged not to be related to the device, and possibly related to the procedure; and the other was judged as possibly related to the device and the procedure. Both subjects recovered without treatment.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Oridion" in a bold, serif font. Above the word is a logo that resembles a sphere made of horizontal lines. The lines are not perfectly straight, giving the sphere a slightly organic or textured appearance. The logo is positioned directly above the center of the word "Oridion".
## CUT-OFF POINT (THRESHOLD) DETERMINATION
The cut-off point (COP) is the level (threshold) used to discriminate between H. pylori infected and non-infected individuals. The threshold value (COP) for the BreathID™ System test is 5 DOB.
The threshold level of 5 DOB was confirmed to be the optimal Cut-off point (COP) for the BreathID System in a multi center study . The sensitivity and specificity achieved with this COP were found to be 100% and 99.2%, respectively, for pre-therapy patients and 95.5% and 100% for post-therapy patients.
Image /page/7/Figure/4 description: This image is a scatter plot titled "Values of 'Last-Delta' in Clinically Negative and Positive Subjects". The y-axis is labeled "Last Delta Units" and ranges from -1 to 59. The x-axis has two categories: "Negative" and "Positive", with data points scattered above each category. The data points for the "Positive" category are clustered much higher than those for the "Negative" category.
Figure 1 Cut-off point histogram
The histogram in Figure 1 shows graphically that the 5 DOB can distinguish very clearly between the infected (positive) and the uninfected (negative) populations.
Oridion Medical 1987 Ltd. • POB 45025 • Jerusalem 91450 Israel • Voice: +972 2 589-9115 • Fax: +972 2 582-8873
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Oridion. The logo consists of a stylized globe made up of horizontal lines, positioned above the word "Oridion" in a serif typeface. The globe is black and white, and the text is black.
## Reference Studies
In addition to the Pivotal study there were several reference studies (one still in addition to the 1 frotar other and 247 negative patients. The results reported were supportive of the conclusions reached in the pivotal study.
{9}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUL - 9 2001
Mr. Sandy Brown Regulatory Affairs Manager Oridion Medical 1987 Ltd. 7 HaMarpe Street Har Hotzvim Industrial Park P.O. Box 45025 Jerusalem, Israel
Re: K011668
Trade Name: Oridion BreathID™ System for Helicobacter pylori Regulation Number: 866.3110 Regulatory Class: I Product Code: MSQ Dated: May 14, 2001 Received: May 17, 2001
Dear Mr. Brown:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have and we have determined the device is substantially equivalent (for the indications for use above into we introllosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, provision in the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
In addition, we have determined that your product contains the following component subject to regulation as drugs: 13C-enriched urea tablet-75mg.
Our substantially equivalent determination does not apply to the drug component (NDA 21-314) Our product. For information on applicable Agency requirements for marketing this product, we suggest you contact:
> Mark Goldberger, M.D., M.P.H. Director Division of Special Pathogens and Immunologic Drug Products Center for Drug Evaluation and Research Food and Drug Administration 9201 Corporate Blvd. Rockville, Maryland 20850
{10}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification although we recommend that you first contact the Center for Drug Evaluation and Research before marketing your drug component[s]. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device or your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising, please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-302) at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Kano Aza
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and
Radiological Health
#### Enclosure
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "Oridion" in a bold, serif font. Above the word is a logo that appears to be a sphere made of horizontal lines. The lines are not perfectly straight, giving the sphere a slightly organic or textured look. The logo is positioned directly above the word, centered horizontally.
June 10, 2001
Indications For Use
510(k) Number (if known): K011668
BreathID™ System Device Name:
Indications For Use:
The BreathID™ System is used to diagnose and monitor Helicobacter pylori infection by measuring changes in the 1302/1202 ratio in a patient's breath following the ingestion of 13C urea.
The Oridion BreathID™ system continually and non-invasively measures changes in the 130% 200 exhaled breath, which may be indicative of increased urease production associated with active Helicobacter pylori infection in the stomach. The Oridion BreathID™ System is to be used as an aid for initial diagnosis and post treatment monitoring of Helicobacter pylori infection.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Woody Dubois
Division of Clinical Laboratory Devices 510(k) Number
Oridion Medical 1987 Ltd. • POB 45025 • Jerusalem 91450 Israel • Voice: +972 2 582-8873
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.