The Uniplant Dental Implant System is intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.
Device Story
Uniplant Dental Implant System consists of solid titanium alloy (Ti-6Al-4V) root-form implants and conical abutments. Implants feature screw-type design with either plasma-sprayed hydroxyapatite (HA) coating or resorbable blast media (RBM) surface treatment to facilitate bone attachment. System includes surgical instruments (drills, insertion modules, try-ins). Used by dentists/oral surgeons in clinical settings for surgical placement into maxillary/mandibular bone. Abutments attach to implants via threaded internal bore; prostheses fabricated via standard crown/bridge techniques. Device provides structural support for dental restorations, benefiting patients by replacing missing teeth and restoring oral function.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, material biocompatibility, and comparison of design characteristics to legally marketed predicate devices.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F1472). Surface: Plasma-sprayed HA (ASTM F1609) or RBM. Design: Two-stage, screw-type, root-form implant. Diameters: 3.4, 4.1, 5.1 mm. Lengths: 10, 12, 14 mm. Sterilization: Co60 gamma irradiation (25 kGy min).
Indications for Use
Indicated for edentulous or partially edentulous patients requiring surgical placement of dental implants in the maxillary or mandibular arch to support crowns, bridges, or overdentures.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
#### Uniplant Dental Implant System
Page 1
# ADMINISTRATIVE INFORMATION SEP 2 8 2001 Manufacturer Name: Universal Implant Systems, Inc. 1301 Ynez Place Coronado, CA 92118 Official Contact: Robert F. Mansueto, D.D.S. Telephone (619) 435-6227 FAX (619) 435-4717 Representative/Consultant: Floyd G. Larson PaxMed International 4329 Graydon Road San Diego, CA 92130 Telephone: (858) 792-1235 FAX: (858) 792-1236 E-mail: flarson@paxmed.com DEVICE NAME Classification Name: Endosseous dental implant
Common Name:
Trade/Proprietary Name:
Dental Implant
Uniplant Dental Implant System
### ESTABLISHMENT REGISTRATION NUMBER
Universal Implant Systems, Inc. is not yet registered with FDA.
## DEVICE CLASSIFICATION
Endosseous dental implants have been classified by FDA as Class III devices under a final order published in the Federal Register of August 12, 1987, as shown in 21 CFR 872.3640. Abutments to such implants are considered by FDA to be Class III devices inasmuch as they are used as accessories to or are used with endosseous dental implants. The device is reviewed by the Dental Products Panel and the Product Code for the device is DZE.
### CONFORMANCE WITH PERFORMANCE STANDARDS
No performance standards have been established under Section 514. Voluntary standards with which the Uniplant Dental Implant System complies include American Society for Testing and Materials (ASTM) designation F1472 (Standard Specification for Wrought Titanium 6Al-4V Alloy for Surgical Implant Applications (UNS R56400)) and ASTM F1609 (Standard Specification for Calcium Phosphate Coatings for Implantable Materials).
{1}------------------------------------------------
#### 510(k) Summary
## PACKAGING/LABELING/PRODUCT INFORMATION
Advertising material to be used for promotion of the Uniplant Dental Implant System will be consistent with the indications for use and other material shown herein.
Uniplant dental implants are packaged in a radiation sterilizable package consisting of an outer tamper evident container and an inner vial of glass or plastic. Sterilization is accomplished by means of Co6 gamma irradiation at a dose of 25 kGy (2.5 Mrad) minimum. Sterilization will be validated by the bioburden method. The sterility assurance level (SAL) that Universal Implant Systems intends to meet for the Uniplant dental implant is 106. The device is not represented to be "pyrogen free." Abutments and instruments will be packaged either sterile in a system similar to the implant packaging or non-sterile in plastic bags.
#### INTENDED USE
The Uniplant Dental Implant System is intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.
#### DEVICE DESCRIPTION
#### Design Characteristics
The Uniplant Dental Implant System is composed of a solid titanium alloy root-form implant and a solid conical abutment, with associated instruments. The implant is a screw-type, two-stage design, incorporating a titanium cover screw that seals the internal bore of the implant from the physiologic environment during healing. The implant is available in three diameters (3.4, 4.1 and 5.1 mm) and three implanted lengths (10 mm, 12 mm and 14 mm). The system includes surgical instruments such as drills, insertion modules and try-ins.
The system includes a solid conical abutment that is intended to the implant by engaging the threaded shaft of the abutment in the internal threaded bore of the implant. The abutment includes a cylindrical transmucosal portion and a tapered, grooved coronal portion, with an internal hex socket at the coronal end to facilitate attachment to the implant. A prosthesis may be fabricated by means of standard crown and bridge techniques for attachment to the abutment.
#### Material Composition
Implants and abutments for the Uniplant Dental Implant System are made from titaniumaluminum-vanadium alloy that meets ASTM designation F1472 (Standard Specification for Wrought Titanium -6Aluminum-4Vanadium Alloy for Surgical Implant Applications (UNS R 56400)). Uniplant implants have a smooth machined collar of 1.0 mm height and from that point to the apex are coated with a layer of plasma-sprayed hydroxyapatite (HA) to facilitate attachment of bone. An alternative surface treatment for the Uniplant is a resorbable blast media (RBM) treatment, in which acid soluble grit (hydroxyapatite) is used to roughen the surface in order to facilitate bone attachment. The surface is then treated with acid to remove any residual
{2}------------------------------------------------
any residual grit particles. The use of titanium, titanium alloy and HA coatings is widespread in commercially distributed, permanently implanted medical devices and the materials are widely considered to be biocompatible. Titanium is often used as a negative control in biocompatibility testing.
### EQUIVALENCE TO MARKETED PRODUCT
Universal Implant Systems, Inc. submits the following information to demonstrate that, for the purposes of FDA's regulation of medical devices, the Uniplant Dental Implant System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices: "O" Company Threaded Implants and Fixed C&B Abutment (K923889, K950066, K961665, K973926), the Steri-Oss Replace HA-Coated Implant (K962845), and the Simpler Implant System (K974401, K974402, K974856, K974857).
#### Intended Uses
The indications for use for the Uniplant Dental Implant System and the predicate devices are substantially the same. All are intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.
#### Design and Materials
The design and functional characteristics of the Uniplant Dental Implant System and the "O" Company Threaded Implant are similar in their use of a root-form threaded design. The threads are intended to provide macroretention in bone and to increase the surface area for bone attachment relative to a smooth design. The threads of the Steri-Oss Replace implant and the Simpler threaded implant serve a similar function. The "O" Company Implant and Fixed C&B Abutment shares with the Uniplant Dental Implant System the method of attaching the abutment to the implant using a threaded abutment shaft. The Uniplant shares the use of Ti-6Al-4V with numerous marketed implants. It shares the use of an HA coating with the "O" Company Threaded Implant, the Steri-Oss Replace HA-Coated Implant, the Simpler HA-coated cylinder implant and numerous other marketed implants.
#### Hydroxyapatite Coating Characteristics
The characteristics of the HA coating used on the Uniplant Dental Implant System are shown in vendor master files. The Uniplant Dental Implant System HA coating has crystallinity, purity and mechanical properties at least as high as those of the coatings on currently marketed devices and at least as high as the suggestions of the FDA Guidance Document.
{3}------------------------------------------------
# SUMMARY: TABLE OF SUBSTANTIAL EQUIVALENCE
The Uniplant Dental Implant System is substantially equivalent to the "O" Company Threaded Implants and Fixed C&B Abutment, the Steri-Oss Replace HA-Coated Implant, and the Simpler Implant System in the following respects:
| | Subject<br>Device | Predicate Devices | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------------|
| | Uniplant<br>Dental<br>Implant<br>System | "O" Company<br>Threaded<br>Implants<br>(K923889,<br>K950066,<br>K961665,<br>K973926) | Steri-Oss<br>Replace<br>HA-Coated<br>Implant<br>(K962845) | Simpler<br>Implant<br>System<br>(K974401,<br>K974402,<br>K974856,<br>K974857) |
| INTENDED USE | | | | |
| Surgical placement in the maxillary<br>and/or mandibular arch to support crowns,<br>bridges, or overdentures in edentulous or<br>partially edentulous patients | YES | YES | YES | YES |
| DESIGN | | | | |
| Body shape | cylindrical | cylindrical | tapered | cylindrical |
| Surface design | threaded | threaded | threaded | threaded or<br>push-in |
| Single-stage or two-stage design | two-stage | two-stage | two-stage | single-stage or<br>two-stage |
| Method of abutment attachment to<br>implant | threaded<br>connection | threaded<br>connection | screw | threaded<br>connection,<br>screw, cement |
| Available diameters, mm | 3.4, 4.1, 5.1 | 3.25, 4.0 | 6.0, 5.0, 4.3 | 3.25, 4.0 |
| Available lengths, mm | 10, 12, 14 | 8, 10, 12, 14,<br>16 | 10, 13, 16 | 8, 10, 13, 15 |
| Solid abutment for cemented restorations | YES | YES | YES | YES |
| MATERIALS | | | | |
| Implant body | Ti-6Al-4V | CP Ti | CP Ti | Ti-6Al-4V |
| Abutment | Ti-6Al-4V | Ti-6Al-4V | Ti alloy | Ti-6Al-4V |
| Surface characteristics / coating | RBM or<br>HA | Machined or<br>HA | HA | Grit blasted or<br>HA |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol on the right and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle symbol is composed of three curved lines representing the wings and body of the eagle.
Public Health Service
SEP 2 8 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Universal Implant Systems, Incorporated C/O Mr. Floyd G. Larson Consultant Paxmed International 4329 Graydon Road San Diego, California 92130
K011574 Re : Uniplant Dental Implant System Trade/Device Name: Regulation Number: 872.3640 Regulatory Class: III Product Code: DZE Dated: May 18, 2001 Received: May 22, 2001
Dear Mr. Larson:
We have reviewed your Section 510(k) notification of incent to we have reviewed your referenced above and we have determined the market the device icrores and indications for use device in the enclosure) to legally marketed predicate devices stated in the enously commerce prior to May 28, 1976, the markeed in theorbeat of the Medical Device Amendments, or to devices ehatement ade or lassified in accordance with the provisions of that have been rooman, and Cosmetic Act (Act). You may, the reactar rood, bray, subject to the general controls encreasions of the Act. The general controls provisions of the provibions or chirements for annual registration, listing of Act include roquirefacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be (Spectal controls, or oral controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good equivacins accountice requirements, as set forth in the Quality Manufacturing Fraound for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the (zi Crk Fare Gro, aistration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further regaracery accerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Larson
Please note: this response to your premarket notification rieast note. Childress any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as first recei will as to (k) premarket notification. The FDA finding acperstantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in xitro regaroric devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of advertibility of Jour 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" Other general information on your (21CFR 807.97). responsibilities under the Act may be obtained from the Division responsiblances International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Clus-ton
Timothy A. Ulatowski
Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K011574
Device Name: Uniplant Dental Implant System
Indications for Use:
Intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
==============================================================================================================================================================================
Prescription Use \$\checkmark\$
---
MCAdyhm fes MSR
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital 510(k) Number
OR
Over-The-Counter Use
iv
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.