K011569 · Bionx Implants, Inc. · JEY · Feb 22, 2002 · Dental
Device Facts
Record ID
K011569
Device Name
BIOSORBFX O/M 2.0/2/4 SYSTEM
Applicant
Bionx Implants, Inc.
Product Code
JEY · Dental
Decision Date
Feb 22, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BioSorbFX O/M 2.0/2.4 System is intended for use in trauma and reconstructive procedures in the midface, maxilla and mandible. Specifically, the device is indicated for use in surgical repair procedures in the treatment of trauma to the midface, maxilla and mandible, and in orthognathic and reconstructive procedures of the midface, maxilla or mandible. The BioSorbFX O/M 2.0/2.4 System stabilizes bone during healing, in conjunction with appropriate postoperative immobilization. When used in the mandible, the BioSorbFX O/M System must be used in conjunction with appropriate maxillomandibular fixation (MMF). The BioSorbFX O/M 2.0/2.4 System implants are not intended for use in and are contraindicated for: 1. Mandibular tumor resection. 2. Situations where internal fixation is otherwise contraindicated e.g., active or potential infection; patient conditions, including blood supply limitations, insufficient quantity or quality of bone; and where patient co-operation cannot be guaranteed (e.g., alcoholism). 3. Significant comminuted fractures, including significant bone loss, of the mandible. 4. Intermaxillary fixation without an appropriate external fixation by other means.
Device Story
BioSorbFX O/M 2.0/2.4 is a bioabsorbable bone fixation system comprising plates and threaded fasteners (screws). Used in midface, maxilla, and mandible surgery; stabilizes bone during healing. Implants are composed of self-reinforced copolymer (70% poly-L-lactide, 30% poly-D,L-lactide). Strength retention lasts 18-36 weeks; full biodegradation occurs within 2-3 years. System includes instrument set (drills, taps, screwdrivers, plate benders, retractors). Operated by surgeons in clinical/OR settings. Mandibular applications require concurrent maxillomandibular fixation (MMF). Provides mechanical stabilization to facilitate bone healing; reduces need for secondary removal surgery due to bioabsorbable nature.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics, materials, and design principles to legally marketed predicate devices.
Technological Characteristics
Bioabsorbable bone plates and fasteners (2.0, 2.4, 2.8mm). Material: Self-reinforced copolymer (70% poly-L-lactide, 30% poly-D,L-lactide). Strength retention: 18-36 weeks. Biodegradation: 2-3 years. Includes manual surgical instruments (drills, taps, benders).
Indications for Use
Indicated for surgical repair of trauma to midface, maxilla, and mandible, and for orthognathic and reconstructive procedures in these areas. Contraindicated for mandibular tumor resection, cases with active/potential infection, poor bone quality/quantity, compromised blood supply, non-compliant patients (e.g., alcoholism), significant comminuted mandibular fractures with bone loss, or cases lacking appropriate external fixation.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
Bionx Implants, Inc. BioSorbFX™ 1.5/2.0 Bioabsorbable Fixation System (K982139)
Bionx Implants, Inc. BioSorbFX™ O/M 2.0/2.4 Bioabsorbable Fixation System (K982721)
{0}------------------------------------------------
K011569
FEB 2 2 2002
# 510(k) Summary Bionx Implants Inc.'s BioSorbFX™ O/M 2.0/2.4 Bioabsorbable Fixation System
Submitter's Name, Address, Telephone Number, and Contact Person
Bionx Implants, Inc. 1777 Sentry Parkway West Gwynedd Hall, Suite 400 Bluc Bell, PA 19422
Contacts:
Gerard S. Carlozzi President and Chief Executive Officer (215) 643-5000 Phone: Facsimile: (215) 653-0984
Bionx Implants Ltd. Tuija Annala Quality Manager P.O.Box 3 FIN-33721 Tampere Finland Phone: 358-3-316 5679 358-3-316 2629 Facsimile:
February 22, 2002 Date prepared:
Name of the Device: BioSorbFX ()/M 2.0/2.4 Bioabsorbable Fixation System
#### Common or Usual Name:
Bioabsorbable Craniofacial Bone Plates and Plate Fasteners
Classification Name: Bone Plate and fasteners (Product Code 76 JEY)
#### Predicate Devices:
- 1. Bionx Inplants, Inc. BioSorbFX™ 1.5/2.0 Bioabsorbable Fixation! System ("BioSorbFX"M 1.5/2.0 System") (K982139)
- 12. Bionx Implants, Inc. BioSorbFX™ O/M 2.0/2.4 Bioabsorbable Fixation System ("BioSorbFXTM O/M 2.0/2.4 System") (K982721)
- 3. Biomet Inc. 2.5mm LactoSorb® Screws (K981666)
- 4. Biomet Inc. LactoSorb@ Sheets (K992158)
- 5.1 Synthes SMF Resorbable Meshes and Sheets (K003786)
- 6. Lorenz Resorbable Distractor System (K002083)
- 7. Macropore MacroporeMX Mandibular Fixation System (K000696)
{1}------------------------------------------------
## 8. Macropore Protective Sheet (K983360)
#### Intended Use
. .
The BioSorbPX O/M 2.0/2.4 System is intended for use in traum: and reconstructive The Bloodbres in the midface, maxilla and mandible. Specifically, the device is indicated processives in and repair procedures in the treatment of trauma to the midface, maxilla and mandible, and in orthognathic and reconstructive procedures of the]midface, maxilla or mandible. The BioSorbFX O/M 2.0/2.4 System stallizes bone during healing, in conjunction with appropriate postoperative immobilization. Which used in the mandible, the BioSorbFX O/M System must be used in conjunction with appropriale maxillomandibular fixation (MMF).
The BioSorbFX O/M 2.0/2.4 System implants are not intended for use in and are contraindicated for:
- 1. Mandibular tumor resection.
- 2. Situations where internal fixation is otherwise contraindicated e.g., active or potential infection; patient conditions, including blood supply limitations, insufficient quantity or quality of bone; and where patient co-operation cannot be guaranteed (e.g., alcoholism).
- Significant comminuted fractures, including significant bone loss, of the 3. mandible.
- 4. Intermaxillary fixation without an appropriate external fixation by other uncans.
### Technological Characteristics and Substantial Equivalence
The BioSorbFX™ O/M 2.0/2.4 Bioabsorbable Fixation Systems consists of various place configurations, which are attached to the bone by 2.0, 2.4 and 2.8mm/threaded fasteners (i.c., screws) of various lengths. The specifications for the BioSprbFXTM O/M 2.0/2.4 System plates and fastencrs will require a self-reinforced copolymer derived from Llactide and D/L-lactide ("PLA copo."), which consists of 70 molar percent poly-L-lactide and 30 molar percent poly-D, L-lactide, with strength retention time 18-36 weeks.
Properly used, in the presence of adcquate immobilization, absortuble BioSprtFX'"> O/M 2.0/2.4 System implants stabilize bone during healing. The BioSorbFX™ O/M 2.0/2.4 System implants gradually lose their strength during 18-36, weeks. Biodegradation takes place within two to three years. The BioSorbFX™ Ont 2.0/2.4 System includes an instrument set containing bone drills, bone taps, screwdrivers, plate benders and cheek retractors. This instrument set is identical with the previously cleared BioSorbFX™ 2.0/2.4 System (K982721) and substantially equivalent with the previously cleared BioSorbFX™ 1.5/2.0 System (K982139) and BioSorb™ Endobrow ScrewTM (K972919).
The BioSorbFX™ O/M 2.0/2.4 System has the same intended use and printiples of operation and is substantially equivalent to the devices previously cleared by FDA operation the Bionx Implants Inc. BioSorbFX™ 2.0/2.4 System (K982721),
{2}------------------------------------------------
BioSorbFX™ 1.5/2.0 System (K982139), Biomet, Inc. 2.5mm LactoSorb® Screws (K981666) and LactoSorb® Sheets (K992158), Synthes SMF Resorbable Meshes and Sheets (K003786), Lorenz Resorbable Distractor System (K002083), Macropore Sheets (1000-78), Mandibular Fixation System (K000696) and Macropore Protective Sheet (K983360) -
The BioSorbFX O/M 2.0/2.4 System has following similarities with previously cleared BioSorbFX 2.0/2.4 System (K982139):
- Uses the same operating principles ー
- Incorporates the same basic design ।
- Is manufactured, packaged and sterilized using the same machinery and । processes.
The BioSorbFX O/M 2,0/2.4 System has following similarities with previously cleared Biomet Inc. 2.5mm LactoSorb® Screws (K981666) and LactoSorb@ Sheets (K992158), Synthes SMF Resorbable Meshes and Sheets (K003786), Lorenz Resorbable Distractor System (K002083), Macropore MacroporeMX Mandibular Fixation System (K000696) and Macropore Protective Sheet (K98.3360).
- Uses the same operating principles -
- Incorporates the same design principles -
- Is used for similar indications ।
In summary the BioSorbPX O/M 2.0/2.4 System described in this submission is, in our opinion, substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 2 2002
Bionx Implants, Incorporated Ms. Tuija Annala Director, Quality and Regulatory Affairs Bionx Implants, Limited Hermiankatu 6-8 L Tampere, FINLAND
Re: K011569
Trade/Device Name: BioSorbFX™ O/M 2.0/2/4 System Regulation Number: 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: November 22, 2001 Received: November 26, 2001
Dear Ms. Annala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Annala
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Swoon Quarr
Image /page/4/Picture/5 description: The image shows a black and white drawing of an abstract symbol. The symbol appears to be composed of curved and straight lines, forming a complex shape. The lines are thick and bold, creating a strong contrast against the white background. The symbol's design is somewhat geometric, with a sense of movement or flow.
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
## 510(K) Number (if known):
Device Name:
BioSorbFX™ O/M 2.0/2.4 Bioalisorbable Fixation System
Indications for Use:
Indications for Ose.
The BioSorbFX O/M 2:0/2.4 System is intended for use in trauma and recorpstituctive The Blosotor'A One 2.0/2.4 Bystein a mandible. Specifically, the device is 'ndicated for use in surgical repair procedures in the treatment of trauma to the midface, maxilla and mandible, and in orthognathic and reconstructive procedures of the midface, maxilla or mandible. The BioSorbFX O/M 2.0/2.4 System stabilizes borge duling healing, in conjunction with appropriate postoperative immobilization. Especially in mandible BioSorbFX O/M System must be used in conjunction of appropriate maxillomandibular fixation (MMF).
The BioSorbFX O/M 2.0/2.4 System implants are not intended for use in and are contraindicated for:
- 1. Mandibular tumor resection.
- 2. Situations where internal fixation is otherwise contraindicated e.g., active or potential infection; patient conditions, including blood supply limitations, insufficient quantity or quality of bone; and where patient co-operation cannot be guaranteed (e.g., alcoholism).
- 3. Significant comminuted fractures, including significant bone loss; of the mandible.
- Intermax illary fixation without an appropriate external fixation by other means.
. (Please do not write below this line - continue on another page is needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-off Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _
ﺳﺎ Prescription Use_ : (Per 21 CFR 801.109)
Over-The-Counter Use OR
Siven Rung
(Timision Sign-Off) intelan Sign-Um)
Inform of Dental, Infection Control, - oneral Hospital Devices ﺗﻘﺪﻳﻢ ﺍﻟﻤﺘﺎﺑﻌﺔ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ ﻳﺘﻢ ﺍﻟﺘﻲ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.