K011561 · Wilsa, Inc. · LPN · Nov 20, 2001 · Ophthalmic
Device Facts
Record ID
K011561
Device Name
SEREINE EXTRA STRENGTH DAILY CLEANER
Applicant
Wilsa, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Nov 20, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
Sereine Extra Strength Daily Cleaner is a daily cleaner for use in the removal of film, deposits and other accumulated debris which collects on contact lenses during wear. Sereine Extra Strength Daily Cleaner is intended for use with all soft (hydrophilic) lenses and hard (PMMA) lenses. The cleaner is for use on lenses immediately following removal from the eye and prior to disinfection. Sereine Extra Strength Daily Cleaner is intended for use each time the lenses are removed from the eye. Sereine Extra Strength Daily Cleaner is not intended for use in the eye. The cleaner must be rinsed from the lens after cleaning with either the recommended rinsing solution (soft lenses) or water (hard lenses).
Device Story
Sereine Extra Strength Daily Cleaner is a sterile, viscous, aqueous solution containing an amphoteric surfactant; formulated to remove proteins, lipids, and debris from contact lenses. Used by patients at home; applied to lenses immediately after removal from the eye and prior to disinfection. The cleaner is not for use in the eye; must be rinsed off with appropriate solution or water. Packaged in 1 FL OZ polyethylene bottles. Benefits include maintenance of lens cleanliness and hygiene. Clinical decision-making is not applicable; device is a maintenance product for contact lens wearers.
Clinical Evidence
Bench testing only. Studies included pH measurement, viscosity testing via Gilmont Falling Ball Viscometer, total solids analysis, sterility testing per USP 24, UV-visible spectra comparison, two-year shelf-life stability study, and preservative effectiveness testing with 14-day re-challenge per FDA guidance. Solution compatibility was assessed over 30 cycles on soft and hard lenses with no significant changes in optical parameters or physical appearance.
Technological Characteristics
Sterile, viscous, aqueous solution; pH 8.3; viscosity 74 cp; contains amphoteric surfactant; no added preservatives. Packaged in 1 FL OZ polyethylene bottle with polypropylene over-cap. Standalone chemical cleaner.
Indications for Use
Indicated for daily cleaning of soft (hydrophilic) and hard (PMMA) contact lenses to remove film, deposits, and debris. Compatible with thermal (heat) or chemical (cold) disinfection systems. Not for use in the eye.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
MiraFlow Extra Strength Daily Cleaner (K00155)
Submission Summary (Full Text)
{0}------------------------------------------------
K011561 NOV 20 2001
Optikem International, Incorporated 2172 South Jason Street Denver, Colorado 80223 USA
# 510(K) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS
1. Submitter's Name
Optikem International, Inc. 2172 South Jason Street Denver, CO 80223
Telephone: 303-936-1137 FAX: 303-936-7320
2. Contact Person
Sally Cook President
3. Date Summary was Prepared:
May 8, 2001
Sereine
Contact Lens
Solutions
**Optisoap**
Hand Soap
{1}------------------------------------------------
#### 4. Name of Device
Trade Name: Sereine Extra Strength Daily Cleaner Common Name: Daily Cleaner for Contact Lenses -Classification Name: Contact Lens Cleaner
- 5. Legally Marketed Device to Which Equivalence is Claimed:
MiraFlow Extra Strength Daily Cleaner, CIBA Vision
Cleaner, Carength Ca., 00155 Corporation, Duluth, GA 30155.
Description of Device: 6.
> Sereine Extra Strength Daily Cleaner is a sterile, viscous, aqueous Sereine Extra Strength Dany Cloaner contains no added solution with an alkaline pri. The citains an amphoteric surfactant to aid in
> preservatives. The cleaner contains an amphoteric surfactant to aid in preservatives. The cleaner contains and other debris which accumulate on the removal of proteins, lipids and other dealer is not meant for use in the eye
> contact lenses during wear. The cleaner is not meant for use in the eye contact lenses during weak ed from the lenses.
Sereine Extra Strength Daily Cleaner is packaged in a 1 FL OZ plastic Sereine Extra Strength Daily Citcano is passessorse on white bottle (polyethylene) with red uropper tip (poryon) (and carton. It has an
over-cap (polypropylene). It has an outside cardboard carton. It has an
the exp (polypropylene) over-cap (polypropylene). It som the date of bottling.
{2}------------------------------------------------
#### Intended Use of the Device: 7.
Sereine Extra Strength Daily Cleaner is a daily cleaner for use Sereine Extra Daving in ids and other accumulated debris which collects on contact lenses during wear.
Sereine Extra Strength Daily Cleaner is intended for use with all soft (hydrophilic) lenses and hard (PMMA) lenses on a an soft (1) arope cleaner is for use on lenses immediately following removal from the eye and prior to disinfection.
Sereine Extra Strength Daily Cleaner is intended for use each time the lenses are removed from the eye.
Sereine Extra Strength Daily Cleaner is not intended for use in Screme Entra Buche insed from the lens after cleaning with either the Cyc and mast cophilic lenses) or water (hydrophobic lenses).
{3}------------------------------------------------
- 8. Technological Characteristics of the Device as Compared to the Predicate Device:
| | MiraFlow | Sereine |
|---------------------|----------|----------------|
| pH | 8.3 | 8.3 |
| Viscosity | 74 cp | 74 cp |
| Total solids | 20.6% | 20.8% |
| Sterility | Sterile | Sterile |
| UV-visible spectrum | | Match Miraflow |
{4}------------------------------------------------
## DESCRIPTION OF TESTS CONTAINED IN TABLE ON PREVIOUS PAGE:
#### 1. pH
The pH of three different lots of MiraFlow Extra Strength Daily File pri of there different lots of Sereine Extra Strength Daily Cleaner was obtained by using a calibrated Piccolo pH meter. The results of the three lots were averaged.
#### 2. Viscosity
The viscosity of the MiraFlow Extra Strength Daily Cleaner and the Sereine Extra Strength Daily Cleaner was determined using a Gilmont Falling Ball Viscometer. Three lots of each cleaner were tested and the results averaged. The Gilmont Falling Ball Viscometer was Size 2 (2-100 cps) with a stainless steel I alling Dur . The viscometer was placed in a constant temperature water bath of 25C. Approximately 5 ml of solution was placed in the viscometer and the stainless steel ball was dropped. The time for the stainless steel ball to descend through the liquid was timed with a calibrated Cole-Parmer Model #94410-20 Stop Watch. The viscosity was calculated by averaging the times for the lots of MiraFlow Extra Strength Daily Cleaner. The procedure was repeated for the lots of Sereine Extra Strength Daily Cleaner.
{5}------------------------------------------------
# DESCRIPTION OF TESTS CONTAINED IN TABLE (CONT.)
#### 3. Total Solids
The total non-volatile solids were determined by weighing an The total non-voluene vished product of MiraFlow Extra Strength Daily Cleaner in a porcelain evaporating dish. Streign Dany Cloater in been previously weighed (empty). The evaporating dish had boing oven at 105C. The dish was The dish was placed in a diffing obtained. The procedure was Three uned using Sereine Extra Strength Daily Cleaner. Three samples of each product were tested and the results averaged. Sumples or courts were comparable as to total non-volatile solids.
#### 4. Sterility
MiraFlow Extra Strength Daily Cleaner is labeled as a sterile product. Sereine Extra Strength Daily Cleaner was tested for sterility by Nelson Laboratories, Inc., Salt Lake City, UT storing of Norther contained in USP 24. Sereine Extra Strength Daily Cleaner was tested for sterility by randomly sampling Daily Cleaner wers from the lot. Sereine Extra Strength Daily Cleaner was found to be sterile.
{6}------------------------------------------------
# DESCRIPTION OF TESTS CONTAINED IN TABLE (CONT).
### 5. UV-Visible Spectra Comparison
MiraFlow Extra Strength Daily Cleaner and Sereine Extra Strength Milal 10W Examer were compared using a Shimadzu UV-visible Dany Cloudiol were Sompl UV-160. The cleaner was placed neat in a 10 mm rectangular silica absorption cell. The reference was ill a 10 min receanganas orbance range was 0-2.5 absorbance units and the spectral range was 200-1000 nm. The spectra of the two cleaners were comparable indicating like components.
{7}------------------------------------------------
### 9. Summary of Non-clinical tests
The following non-clinical tests were performed on Sereine The following non emileaner. These tests were not performed Latra Strength Daily Cleaner as Optikem On Mind Tow Encra Ses not have access to materials and data necessary to conduct these tests.
Although the non-clinical tests were not performed on MiraFlow Anthough the non cinner for comparison, it is presumed that Extra Strengan Daily Cleaccessfully in order to obtain pre-market approval.
A. Stability Study
The stability or shelf-life study for Sereine Extra Strength Daily Cleaner was conducted using lots stored at ambient temperature for two years. The ambient temperature was between 60 and 80 F and was recorded on a daily basis as part of the quality assurance program. Three lots of Sereine Extra Strength Daily Cleaner were tested for stability. These lots were bottled in 1 FL OZ (30 ml) plastic bottles. Bottles were capped and the caps sealed with tamper-resistant seals. Bottles were not subjected to any specific conditions as far as humidity or light. The relative humidity is recorded on a daily basis as part of the quality assurance program.
{8}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### Summary of Non-clinical tests (cont). 9.
A. Stability Study (cont).
The following characteristics of Sereine Extra Strength Daily Cleaner were monitored:
1. pH
The pH of each of three lots was taken using a calibrated The pH of each of thee lots was compared to the original
Piccolo pH meter. The result was compared to the original Piccolo pH meter. The finished device acceptance pH obtained as part of the humaneng date. The pH was not
standards from the original bottling and storage standards from the original soming and storage.
2. Viscosity
The viscosity of each of three lots was measured using a The viscosity of each of the City of Ciscometer. The viscosity Gilmont Size 2 Falling Ball Type Tibecausessity as determined at the time of bottling.
3. Active Ingredients
The ingredients were determined using a Shimadzu UV-visible The ingredients were determinou uses acception silica absorption Spectrophotometer, Model 100 will a scan of 200-800 mm.
cells. Three stored lots were measured in a scan of 200-800 mm. cells. Three stored lots were was not performed on the lots as part of the original finished device acceptance, the scans were as part of the original imislied on the use. The ingredients were
compared to a recently formulated batch. The ingredied as indicated compared to a recently formations of the storage period as indicated by the UV-visible spectra.
{9}------------------------------------------------
K 0 11 56 1
## 510(k) SUMMARY
# SUMMARY OF SAFETY AND EFFECTIVENESS
Summary of Non-clinical tests (cont.) g.
A. Stability Study (cont.)
- 4. Physical Appearance
Physical appearance of the bottle, tip and cap was observed. Physical appearance of the occases, inc. are discoloration, cracks, Packaging components were Extra Strength Daily Cleaner itself was observed for any precipitates, turbidity or color change. was observed for any precipitates, turelows of the chemical components were noted.
- 5. Sterility
The sterility of the stored solution was tested using USP The Stermly of the Stored Stated States, Inc., Salt Lake City, UT. All three stored lots tested were sterile.
6. Preservative Effectiveness with 14 day re-challenge
Preservative effectiveness testing of three stored lots was performed by Northview Pacific Laboratories, Inc. of Hercules, CA. The testing was conducted according to the FDA Guidelines CA. The testing was conceeded ation (510(k)) Guidance Document . For Contact Lens Care Products". The stored lots passed the requirements for this test.
#### Conclusion:
Shelf-life studies indicated that the Sereine Extra Strength Daily Cleaner is safe and effective for the period of the proposed shelf-life of two years.
{10}------------------------------------------------
# 9. Summary of Non-clinical tests (cont.)
#### B. Chemistry
### 1. Solution Compatibility
The compatibility of Sereine Extra Strength Daily Cleaner with soft (hydrophilic) lenses and hard (PMMA) lenses with soft (figuropining) tofnen ing the "Solution Compatibility was unlined in the "Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products". Ten lenses of Metrotint (enhancing tints of polymacon, 38% water) and ten lenses of Horizon 55 (visibility tints of 55% water) were used to test solution compatibility of hydrophilic lenses. Metrotint lenses are manufactured by Metro Optics, Austin, TX. Horizon 55 lenses are by Medio Opace, Westcon Contact Lenses, Grand Junction, CO. Ten PMMA lenses were also tested for compatibility with Sereine Extra Strength Daily Cleaner. These were blue lenses manufactured by Lens Dynamics, Lakewood, CO. The lens blanks were manufactured by Optikem Polymers of Englewood, CO.
The optical parameters were recorded before the test. Lenses were inspected for any physical defects. For hydrophilic lenses, each lens was flexed and then cleaned with Sereine Extra Strength Daily Cleaner as per product directions. After rinsing with sterile saline, the lenses were placed in a lens case and disinfected with Opti-One Solution (Alcon Laboratories, Ft Worth, TX) as per label instructions. After disinfection the lenses were removed from the lens case and the cycle repeated. The procedure was repeated for 30 cycles for each lens. After the 30 proces, the lens parameters and physical appearance of the lenses were again recorded.
{11}------------------------------------------------
#### Summary of Non-clinical tests (cont.) ல்
B. Chemistry (cont.)
## 1. Solution Compatibility
For PMMA lenses, the lenses were cleaned with Sereine Extra For PMMA lenses, the lenses were cicance in The lenses were
Strength Daily Cleaner as per label directions. The lenses were Strength Daily Cleaner as per iaber as per interes were
rinsed with water and placed in lens cases. The lenses were rinsed with water and placed in left soaking Solution as per disinfected with Sereine wetting or boulenges were removed
label directions. After disinfection, the lenses were removed label directions. And the cycle repeated. After 30 cycles,
from the lens cases and the cycle repeated. After 30 cycles, from the lens cases and the Cycle reposee of the lenses were the parameters and physical appearantee on
No significant changes in optical parameters or physical No significant changes no opine Extra Strength Daily appearance was noted. Serence DAGE Burenge
2. Cleaning Effectiveness
The Determination of the critical micelle concentration of the The Determination of the circles tension of the solution done according
surfactant and the surface tension of the solution (510(k)) surfactant and the surface tension of also to the bottfication (510(k))
to the procedure contained in "Premarket North Corp. Products" to the procedure contained in "re Lens Care Products", Guidance Document for Someon Donator Donator Book Testing Appendix B. Testing Company, Inc. of Passaic, NJ.
{12}------------------------------------------------
## 9. SUMMARY OF NON-CLINICAL TESTS (CONT.)
C. Microbiological Tests
1. Sterility
Sterility of Sereine Extra Strength Daily Cleaner was confirmed using USP methodology. Sterility testing of the product was done by Nelson Laboratories, Salt Lake City, UT. All lots tested were sterile. The manufacturing process used by Optikem International, Inc. to produce Sereine Extra Strength Daily Cleaner was adequate to ensure that the solution was sterile.
2. Preservative Effectiveness with 14 day rechallenge
Preservative effectiveness testing was conducted according to the procedure outlined in "Pre-market Notification (510(k)) Guidance Document for Contact Lens Care products". The testing was conducted by Northview Pacific Laboratories, Hercules, CA.
Sereine Extra Strength Daily Cleaner passed the test for preservative effectiveness and is safe for use in a multi-dose container.
{13}------------------------------------------------
### 10. Conclusions
Based on the data outlined on the preceding pages, Sereine Extra
tion of the outlined offered officative for use as a daily clear Based on the data outlined on the preceding pagos, our as a daily cleaner for
Strength Daily Cleaner is safe and effective for use as a daily cleaner for Strength Daily Cleaner is safe un deposits and debris from the removal of accumulated (PMMA) lenses. The expiration date of
soft (hydrophilic) and bard (PMMA) lensessmiste based on data outlined soft (hydrophilic) and hard (PMMA) tenses. "The expropriate based on data outlined on the preceding pages.
{14}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## NOV Z 0 2001
Ms. Sally Cook President, Optikem International, Inc. Wilsa, Incorporated 2172 South Jason Street Denver, CO 80223
Re: K011561
K011501
Trade/Device Name: Sereine Extra Strength Daily Cleaner Regulation Number: 21 CFR 886.5928 Regulation Number: 21 Of St Irophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN; HPX Dated: October 1, 2001 Received: October 2, 2001
Dear Ms. Cook:
We have reviewed your Section 510(k) notification of intent to market the device referenced We have reviewed your Section 310(ts) notheration of the indications for the indications for above and we have determined the cence is substainly of commence prior to May 28, 1976, 1976, 1976, 1976, 1976, 1976, 1976, use stated in the enclosure) to device interestments, or to devices that have been the enactment date of the Medical Device runes of Food, Drug, and Cosmetic Act
reclassified in accordance with the provisions of the Federal controls provisions of reclassified in accordance with the provisions on the general controls provisions of the (Act). You may, therefore, market the actives seauirements for annual registration,
Act. The general controls provisions of the Act include requirements mishranding Act. The general controls provisious of the Act include requiremisms against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III
n major If your device is classified (see above) into essues additional controls. Existing major (Premarket Approval), it may be subject to such and in the Code of Federal Regulations, Title 21,
regulations affecting your device can be found in the Code of Federal parth regulations affecting your device can or in the management on assumes compliance with the Parts 800 to 895. A substantially cquivalents, as set forth in the Quality System
Current Good Manufacturing Practice requirements, as set forth in the Rest, 800 and that, Current Good Manufacturing Fractic requiremently is over CFR Part 820) and that, through
Regulation (QS) for Medical Devices: General regulation (21 CFR Parily such Regulation (QS) for Medical Devices. Ocher a voginistration (FDA) will verify such periodic QS inspections, the Food and Drag regulation may result in regulatory action. In the Federal assumptions. Tandre to comply with the Other toge oncerning your device in the Federal addition, FDA may publish lurther announcements on notification submission does not affect
Register. Please note: this response to your premarket not for devices under Register. Flease note: this response to your f 542 of the Act for devices under any obligation you might have under sociologio sportions, or other Federal laws or regulations. I
{15}------------------------------------------------
### Page 2 - Ms. Sally Cook
This letter will allow you to begin marketing your device as described in your Section 510(k)
. The subscription for a Tim FDA freding of sybstantial equivalence of your dev This letter will allow you to begin marketing your device of your device to a would and the research I its retect will and of the FDA finding of substantial equivalence of your device to trigure of the first of the first of the for proceed to the market.
procedure specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your device on our labeling regulation on earther Office of
additionally 21 CFR Part 809.10 for in vited diagnosic devices, please contact t If you attended in the most of on the diagnostic devices); pressoons on the promotion and advertising of
Compliance at (301) 594-4613. Additionally, for questions on the pro Compliance at (301) 594-4613. Additionally, for questions on and the promotes and Ocuped on the Office of Compliation of Compliance an (101). "C/C/R Part 807.97).
your device, please contact the Office of Compliation "(21CFR Part 807.97).
regulation entitl regulation entitled, "Misbranding by responsibilities may be obtained from the Division
Other general information on your responsibilities under the Act may be obtained from rogeneral information on your responsibilities under the Act may 10 of the many of the manager (800)
of Small Manufacturers, International and Consumer Assistance at its tol of Small Manufacturers, International its Internet address .
638-2041 or (301) 443-6597 or at its Internain html 638-2041 of (501) 175-604) 175-607 14-15-2041 11-1
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{16}------------------------------------------------
K011561
NOV 2 0 2001
# INDICATIONS FOR USE STATEMENT
Sereine Extra Strength Daily Cleaner is indicated for use as a daily
our and other that the first be acits and other debris which accums Sereine Extra Strength Daily Cleaner is Indicator for which accumulate
cleaner for removal of film, deposits and other debris which accumulate
s and of the county in and DJ A cleaner for removal of film, deposits and other were are and
Sereine Extra Strength Dailly Cleaner may be used with either thermal
s and the success to belief of ation systems . Follow the recommends Sereine Extra Strength Dailly Cleaner may be used with exammendation
(heat) or chemical (cold) disinfection systems. Follow the recommendation
Services of chemical (cold) di (heat) or chemical (cold) disinicollor systems
of your eye care practitioner when disinfecting your lenses.
Eis O Ph.D.
(Division Sign-Off) Division of Ophthalmic Devices Koll561 510(k) Number_
**Over-the-Counter Use**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.