K011311 · Pck Electronic Industry and Trade Co, Ltd. · OWB · Jul 10, 2001 · Radiology
Device Facts
Record ID
K011311
Device Name
UROLOGIC
Applicant
Pck Electronic Industry and Trade Co, Ltd.
Product Code
OWB · Radiology
Decision Date
Jul 10, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
UROlogic is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to urologic and endoscopic applications The system may be used for other imaging applications at physician's discretion.
Device Story
UROlogic is a floor-mounted, tilting fluoroscopic X-ray diagnostic system. It utilizes an undertable image intensifier, X-ray generator, X-ray tube assembly, and TV system to capture and display patient images. The system features a motorized patient support table (longitudinal/lateral movement, -15 to +87 degree tilt) and a cranial-moving X-ray tube/intensifier assembly, allowing scanning without patient repositioning. Operated via tableside control panel, foot/handswitches, and X-ray control panel. Includes a triple-mode image intensifier, CCD camera with one-frame memory, stationary grid, and cassette holder. Used in clinical settings (e.g., OR, urology clinics) by physicians/technicians to provide real-time visualization during procedures. Output is displayed on monitors to assist in clinical decision-making and procedural guidance.
Clinical Evidence
bench testing only
Technological Characteristics
Fluoroscopic X-ray imaging system. Components: floor-mounted tilting patient support table, X-ray generator, X-ray tube, triple-mode image intensifier, CCD camera, TV system, and monitors. Connectivity: tableside control panel, foot/handswitches. Motorized longitudinal/lateral table movement and -15 to +87 degree tilt. Stationary grid and cassette holder included.
Indications for Use
Indicated for patients requiring fluoroscopic and radiographic imaging during diagnostic, surgical, and interventional procedures, including urologic and endoscopic applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
510(k) Application
UROlogic
ÍJÚL 1 0 2001
Image /page/0/Picture/3 description: The image shows a logo with a geometric design on the left and the letters "PCK" on the right. The geometric design resembles a wheel or a stylized flower with six petals or sections. A horizontal line extends from the right side of the geometric design, underlining the letters "PCK". The letters are in a bold, sans-serif font and appear to be a part of the company's name or brand.
×011311
ELECTRONIC INDUSTRY AND TRADE CO, LTD
TURAN GUNES BUL KONRAD AD CAD 59/1 SANCAK, CANKAYA, 06550 ANKARA, TURKEY TEL:+90 312 491 6010 FAX:+90 312 491 601 1
30 APR 2001
## 510(k) Summary
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
1. Applicant
PCK ELECTRONIC INDUSTRY AND TRADE CO, LTD TURAN GUNES BUL KONRAD AD CAD 59/1 SANCAK CANKAYA, 06550 ANKARA TURKEY
TEL:+90 312 491 6010 FAX:+90 312 491 6011
## CONTACT PERSON: CENGİZ KABAKCI ASSISTANT GENERAL MANAGER
2. Device Identification
| Proprieatary Device Name: | UROlogic Urological Table |
|-----------------------------|------------------------------------------------|
| Common/Generic Device Name: | Fluorocsopic Imaging System, Urological Table |
| Classification Name: | SYSTEM, X-RAY, FLUOROSCOPIC, IMAGE-INTENSIFIED |
| Product Code: | 90 JAA |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR 892.1650 |
{1}------------------------------------------------
:
## 3. Substantial Equivalence
The UROlogic Urological Table is substantially equivalent to the following currently marketed devices:
- OEC Uroview 2600 (K940295) ●
- Liebel-Flarsheim Hydradjust IV (K943581) .
## 4. Description of Device
UROlogic is a universal fluoroscopic x-ray diagnostic system intended for use in providing x-ray imaging of patient with an undertable image intensifier. The system consists of a floor mounted tilting patient support table, x-ray generator, xsystem or a sembly, image intensifer and the tv system. The system is operated via tableside control panel, foot/handswitches and x-ray control panel. The system comes with a tripple mode image intensifier, a CCD camera with one frame memory, x-ray tube with housing and an image monitor.
The tabletop can be moved motorized in longitudinal and lateral directions. The table can be tilted -15 to +87 degrees. Cranial movement of connected x-ray tube and image intensifer assembly gives the operator the advantage of scanning without moving patient. System has a stationary grid and casette holder for wallow mo films. Patient positioning and other accessories are also provided.
#### 5. Intended Use
UROlogic is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to urologic and endoscopic applications The system may be used for other imaging applications at physician's discretion.
### 6. Technological Characteristics
URO!ogic Urological Table employs the same technological characteristicsas the predicate devices. This device is intended for the same applications as the currently marketed predicate devices. All systems are image intensified x-ray imaging systems with an overtable x-ray tube assembly. Like the predicate devices, UROlogic Urological Table consists of basic the basic patient suport table, and standard system components: x-ray generator, x-ray tube, Image Intensifier, TV system and monitor(s).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are simple and abstract, with a curved line representing the forehead, nose, and chin. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Mr. Cengiz Kabakci Assistant General Manager PCK Electronic Industry and Trade Co, Ltd. Turan Gunes Bul Konrad AD CAD 59/1 Sancak CANKAYA 06550 ANKARA TURKEY
1 6 2012 N VA
Re: K011311
Trade/Device Name: IURO) Logic (Urology x-ray table) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: April 30, 2001 Received: April 30, 2001
Dear Mr. Kabakci:
This letter corrects our substantially equivalent letter of July 10, 2001.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
Applicant:
PCK ELECTRONIC INDUSTRY AND TRADE CO, LTD
TURAN GUNES BUL KONRAD ADENAUER CAD 59/1 SANCAK, CANKAYA O6550, ANKARA TURKIYE
510(k) NUMBER: K011311
UROlogic Urological Table DEVICE NAME:
## INDICATIONS FOR USE:
UROlogic is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures
*Prescription Use*
David A. Seigman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological I 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.