K003301 · CellaVision AB · JOY · Mar 15, 2001 · Hematology
Device Facts
Record ID
K003301
Device Name
DIFFMASTER OCTAVIA AUTOMATIC HEMATOLOGY ANALYZER
Applicant
CellaVision AB
Product Code
JOY · Hematology
Decision Date
Mar 15, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5260
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
White blood cell differential count
Deterministic artificial neural networks
—
Results equal to the reference method
—
—
Two clinical trials performed according to NCCLS document H-20A
—
Indications for Use
The DiffMaster Octavia™ is an automated cell locating device. DiffMaster Octavia automatically locates and presents images of blood cells on peripheral blood specimens. The operator identifies and verifies the suggested classification of each cell according to type. DiffMaster Octavia is intended to be used by skilled operators, trained in the use of the instrument and in recognition of leukocyte classes.
Device Story
DiffMaster Octavia is an automated hematology analyzer for peripheral blood specimens. System components include a slide positioning system, microscope, camera, and Cytologica software. Device automatically scans stained blood film slides to locate white blood cells (WBCs), captures digital images, and presents them to the operator. Software uses deterministic artificial neural networks (ANNs) to suggest cell classifications. Used in clinical laboratory settings by trained technologists. The operator reviews the suggested classifications, confirming or modifying them as necessary. This process aids the technologist in performing the WBC differential procedure, improving efficiency in cell identification compared to manual microscopic methods.
Clinical Evidence
Two clinical trials were conducted comparing DiffMaster Octavia results to the Romanowski (MGG)-Stain manual reference method, following NCCLS document H-20A (1992). Endpoints included accuracy of suggested cell classification, precision of cell location/display, instrument reproducibility, and sensitivity/specificity (false positives/negatives). Results demonstrated performance equivalent to the manual reference method.
Technological Characteristics
System comprises a positioning system, microscope, camera, and Cytologica software. Analyzes stained peripheral blood film slides (Wright, Wright Giemsa, or May Grünwald Giemsa). Employs deterministic artificial neural networks for cell classification. Operates as an automated cell locating and imaging system requiring human verification.
Indications for Use
Indicated for automated location and presentation of peripheral blood cell images for classification. Intended for use by skilled operators trained in leukocyte recognition to assist in performing white blood cell differentials.
Regulatory Classification
Identification
An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)
Romanowski (MGG)-Stain manual light microscopic process
Submission Summary (Full Text)
{0}------------------------------------------------
K003301
# SECTION 2. SUMMARY AND CERTIFICATION
# A. 510(K) SUMMARY
## Summary of Safety and Effectiveness
In accordance with 21 CFR 807.92, the following information constitutes the 510(k) summary for the DiffMaster Octavia™.
| SUBMITTER'S NAME: | CellaVision AB |
|---------------------|------------------------------------------|
| ADDRESS: | Ideon Research Park, 223 70 Lund, Sweden |
| CONTACT PERSON: | Constance G. Bundy |
| TELEPHONE NUMBER: | 612-574-1976 |
| FAX NUMBER: | 612-571-2437 |
| DATE OF SUBMISSION: | October 19, 2000 |
### Identification of device 1.
Proprietary Name: DiffMaster Octavia™ Automatic Hematology Analyzer Common Name: DiffMaster Octavia™ Classification Status: Class II per regulations 864-5260 Product Codes: JOY
### Equivalent devices 2.
CellaVision AB believes the DiffMaster Octavia™is substantially equivalent to the Romanowski (MGG)-Stain manual light microscopic process for cell classification and the IMI MICRO21 with WBC Diff (White blood cell Differential, 510(k) number K925670/A).
### 3. Description of the Device
The DiffMaster-Octavia™ is an automated cell locating device for differential count of white blood cells and characterization of red morphology. It is based on a CellaVision AB developed software system Cytologica. DiffMaster Octavia™consists of a commercially available positioning system for the slides, a commercially available microscope, a commercially available camera and the software system.
{1}------------------------------------------------
### Intended use 4.
The DiffMaster Octavia™ is an automated cell locating device.
DiffMaster Octavia automatically locates and presents images of blood cells on peripheral blood specimens. The operator identifies and verifies the suggested classification of each cell according to type.
DiffMaster Octavia is intended to be used by skilled operators, trained in the use of the instrument and in recognition of leukocyte classes.
### Technological characteristics, comparison to predicate device 5.
Like the predicate device, the DiffMaster Octavia™ instrument locates WBCs, stores digital images of the cells and displays the images in an organized manner and suggests a cell class for each cell to aid technologists in performing the WBC differential procedure. A competent technologist is required to confirm or modify the suggested classification of each cell. It is intended to be used by skilled operators, trained in the use of the instrument and in recognition of leukocyte classes.
### Brief Discussion of Non clinical factors supporting a determination of substantial equivalence:
The method involves a competent human observer to examine the microscopic images of the cells as does the predicate method and device. See attached substantial equivalence comparisons.
Brief discussion of clinical tests supporting a determination of substantial equivalence: Two clinical trials have been performed to confirm equivalence with the predicate method Romanowski (MGG)-Stain manual light microscopic process for cell classification. The studies have been performed according to the approved standard, NCCLS, vol. 23, no 1, document H-20A, March 1992: Reference Leukocyte Differential Count (proportional) and Evaluation of Instrumental Method. Complementary nonclinical tests have been performed to confirm cell location and display of found cells.
### Conclusions drawn from clinical tests:
The following information was obtained from the clinical tests comparing the DiffMaster Octavia™ cell classification results to the manual reference method results:
- the accuracy of the suggested classification for each cell type (DiffMaster Octavia™ . results compared to the light microscope manual diff count results)
- the precision for location and display of the cells found .
- the precision of the instrument (reproducibility) ●
- the sensitivity and the specificity of the instrument (false positive and false negatives ● found)
Results equal to the reference method was provided by use of the DiffMaster Octavia™ instrument.
{2}------------------------------------------------
Comparative features of the DiffMaster Octavia compared with the predicate device and method
| Characteristic | IMI Micro21 | Manual light microscopic process | DiffMaster Octavia™ |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Automated cell location device for cell location and identification, for in-vitro use. Verification of results by human observer | Manual method for cell location and identification, for in-vitro use. Verification of results by human observer | Automated cell locating device for cell location, and identification, for in-vitro use. Verification of results by human observer |
| Sample type | Bloodfilm on glass slides of peripheral whole blood, stained | Bloodfilm on glass slides of peripheral whole blood, stained | Bloodfilm on glass slides of peripheral whole blood, stained |
| Sample preparation | Wright, Wright Giemsa, May Grünwald stain | May Grünwald Giemsa stain (MGG) | May Grünwald Giemsa stain (MGG) |
| Analysis technique | Cells are located using the NCCLS scanning procedure. Cell images are analyzed using standard mathematical methods, including deterministic artificial neural networks (ANN's) trained to distinguish between classes of white blood cells | | Cells are located using the NCCLS scanning procedure. Cell images are analyzed using standard mathematical methods, including deterministic artificial neural networks (ANN's) trained to distinguish between classes of white blood cells |
・
:
{3}------------------------------------------------
### Discussion of performance testing. ર.
Discussion of performance testing.
An extensive collection of tests has been conducted and successfully completed, including tests to ensure safety, effectiveness, and compliance with product requirements.
The following tests were conducted as prototype tests and successfully completed: a) Evaluation of:
a) Draianon on of ...
- Performance (Accuracy and precision, clinical sensitivity, specificity, cell location)
- Safety
b) Inspections to verify that all design features are as intended and that all authorized design changes have been accomplished and recorded.
c) Validation of the system (including efficacy of user interface)
#### 7. Conclusion
Based on extensive performance testing and a comparison to the predicate device, it is the conclusion of CellaVision AB that the DiffMaster Octavia™ is substantially equivalent to devices already on the marked (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized representation of an eagle, with three curved lines forming the body and wings.
# MAR 1 5 2001
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Public Health Service
CellaVision AB c/o Ms. Constance G. Bundy C.G. Bundy Associates, Inc. 6740 Riverview Terrace Minneapolis, Minnesota 55432
Re: K003301
Trade Name: DiffMaster Octavia™ Automatic Hematology Analyzer Regulatory Class: II Product Code: JOY Dated: February 16, 2001 Received: February 20, 2001
Dear Ms. Bundy:
We have reviewed your Section 510(k) notification of intent to market the device referenced We have reviewed your Section 910(x) notifically equivalent (for the indications for use above and we nave determined the actrices marketed predicate devices marketed in interstate commerce stated in the enclosure) to legally maneces processes in the Medical Device Amendments, or to devices that prior to May 26, 1970, the chaouncen and of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosment Act (Act). Tourmaly, morolors, manov as as f the Act include requirements for annual provisions of the Act. "The general controle practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (Fremarket Approval), it they of sayses in one of Federal Regulations, Title 21, Parts 800 to 895. allecting your device our of tournination assumes compliance with the Current Good A Substantially Cquiralient determinates set forth in the Quality System Regulation (QS) for Manufacturing Fridence required in (21 CFR Part 820) and that, through periodic QS Medical Devices. General regulation (FDA) will verify such assumptions. Failure to Inspections, the Food and Drug Temninesult in regulatory action. In addition, FDA may publish Comply with the GMT Togatator may readeral Register. Please note: this response to your premarket notification submission does not affect any obligation you might Its polise to your premarked noursearch start for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket a This letter will allow you to begin marketing your do nee of your device to a legally marketed notification. The FDA Inding of substantial equivalise of your device and thus, permits your device to procced to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at additionally 809.10 for in vitto diagnostions on the promotion and advertising of your device, (301) 594-4588. Additionally, for questions on the predictions of Areas note the regulation please contact the Office of Comphanoo at (301) 37 - 407.97 - 10.07.9 Directed entitled, "Misoranding of telefone to pennation in the Act may be obtained from the Division of Small
information on your responsibilities under the Act may be obtained from information on your responsibilities and its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## B. INDICATIONS FOR USE
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Device Name: DiffMaster Octavia™
Indications for Use:
The DiffMaster Octavia™ is an automated cell locating device.
DiffMaster Octavia automatically locates and presents images of blood cells on Diffiviaster Octavia adtomations focults identifies and verifies the suggested classification of each cell according to type.
DiffMaster Octavia is intended to be used by skilled operators, trained in the use of the instrument and in recognition of leukocyte classes.
J. Roberts
(Division Sign-Off) (Division of Clinical Laboratory Deyices o 3,30 L 510(k) Number -
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) ાર
Over the Counter Use _________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.