← Product Code [OCW](/productcode/OCW) · K002398

# ENDOPATH ETS45, ETS FLEX45, ETS COMPACT-FLEX45 LINEAR CUTTERS, ETS FLEX45 AND ETS COMPACT-FLEX45 NO-KNIFE ARTICULATING (K002398)

_Ethicon Endo-Surgery, Inc. · OCW · Nov 3, 2000 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K002398

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, Inc.
- **Product Code:** [OCW](/productcode/OCW.md)
- **Decision Date:** Nov 3, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The ENDOPATH® ETS45 Endoscopic Linear Cutters, the ETS-Flex45 Endoscopic Articulating Linear Cutters, and the ETS Compact- Flex45 Articulating Linear Cutter are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing material such as bovine pericardium. The ENDOPATH® ETS Flex45 No-Knife Articulating Linear Staplers and ETS Compact-Flex45 No-Knife Articulating Linear Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic, and pediatric procedures. They can be used with staple line or tissue buttressing material such as bovine pericardium.

## Device Story

The Endopath ETS45 series are sterile, single-patient-use surgical instruments designed for tissue transection, resection, and anastomosis creation. These devices deliver two or three double-staggered rows of staples to tissue. The product line includes articulating and non-articulating linear cutters and no-knife linear staplers. Operated by surgeons in open or minimally invasive surgical settings, the devices function by clamping tissue between an anvil and a cartridge channel. The modified versions feature enhanced clamping force to ensure more consistent staple formation. The devices are used to facilitate surgical procedures across multiple specialties, including general, gynecologic, urologic, thoracic, and pediatric surgery. By providing reliable staple line formation, the devices assist in tissue management and closure, potentially improving surgical outcomes.

## Clinical Evidence

Bench testing only. Preclinical testing was performed to verify device performance in transection, resection, and anastomosis creation, demonstrating satisfactory results and improved staple formation consistency due to increased clamping force.

## Technological Characteristics

Linear cutters/staplers delivering two or three double-staggered rows of staples. Features include articulation mechanisms and no-knife configurations. Materials are compatible with tissue buttressing (e.g., bovine pericardium). Devices are sterile and intended for single-patient use. The design incorporates an anvil and channel mechanism for tissue clamping.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ENDOPATH® ETS45 Linear Cutters, Linear Staplers and Reloads ([K002398](/device/K002398.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three curved lines representing its wings and tail feathers. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is written around the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ethicon Endo-Surgery, Inc. Mr. Edwin O. Billips Senior Associate, Regulatory Affairs 4545 Creek Road Cincinnati, OH 45242-2839

·JUL 2 7 2015

Re: K002398

> Trade/Device Name: Endopath ETS45 Endoscopic Linear cutters: Flex 45 Endoscopic Articulating Linear cutters; ETS Compact-Flex 45 Articulating Linear cutter; ETS Flex 45 No-Knife Articulating Linear staplers, and ETS Compact-Flex 45 No-Knife Articulating Linear staplers

Regulation Number: 21 CFR 876.1500

Regulation Name: Endoscope and accessories

Regulatory Class: II

Product Code: OCW

Dated (Date on orig SE ltr): August 3, 2000

Received (Date on orig SE Itr): August 7, 2000

Dear Mr. Billips,

This letter corrects our substantially equivalent letter of November 3, 2000.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K002398

510(k) Number (if known):

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

The ENDOPATH® ETS45 Endoscopic Linear Cutters, the ETS-Flex45 Endoscopic Articulating Linear Cutters, and the ETS Compact- Flex45 Articulating Linear Cutter are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including appiated on managed and pediatic surgical procedures. They can be used with staple line or tissue buttressing material such as bovine pericardium.

The ENDOPATH® ETS Flex45 No-Knife Articulating Linear Staplers and ETS Compact-Flex45 No-Knife Articulating Linear Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), theracic, and pediatic procedures. They can be used with staple line or tissue buttressing material such as bovine pericardium.

(PLEASE DO NOT WRITE BELOW THIS UNE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|               | <div align="center">Mark N. Millman</div> |
|---------------|-------------------------------------------|
|               | (Division Sign-Off)                       |
|               | Division of General Restorative Devices   |
| 510(k) Number | K002398                                   |

| Prescription Use | X | OR | Over-The-Counter Use |
|------------------|---|----|----------------------|
|------------------|---|----|----------------------|

Presc (Per 21 CFR 801.109)

OR

(Optional Format 1-2-96)

{3}------------------------------------------------

K002398

Ethicon Endo-Surgery, Inc ENDOPATH® ETS45 Linear Cutters, Staplers, and Reloads

### Section L

# 510(k) Summary of Safety and Effectiveness (2 copies)

Company:

Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242

Contact:

Edwin O. Billips Senior Regulatory Associate Telephone: 513- 337-7162 Fax: 513- 337-7134

Date Prepared: August 3, 2000

Modified Devices: ENDOPATH® ETS45 Linear Cutters, Linear Staplers and Reloads

Marketed Devices: ENDOPATH® ETS45 Linear Cutters, Linear Staplers and Reloads

### Device Description:

The ENDOPATH® ETS45 Linear Cutters, Linear Staplers and Reloads are sterile, single patient use instruments that deliver two or three double-staggered rows of staples.

#### Indication For Use:

The ENDOPATH® ETS45 Endoscopic Linear Cutters, the ETS-Flex45 Endoscopic Articulating Linear Cutters, and the ETS Compact- Flex45 Articulating Linear Cutter are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectorny), thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing material such as bovine pericardium.

The ENDOPATH® ETS Flex45 No-Knife Articulating Linear Staplers and ETS Compact-Flex45 No-Knife Articulating Linear Staplers are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic, and

380:\510(k)s\K00_ lets510.doc 8/3/00

{4}------------------------------------------------

# KOO 2398

Ethicon Endo-Surgery, Inc ENDOPATH® ETS45 Linear Cutters, Staplers, and Reloads

pediatric procedures. They can be used with staple line or tissue buttressing material such as bovine pericardium.

## Technological characteristics:

The technological characteristics of the modified devices are the similar to the marketed products. The modified devices are linear cutters/staplers that are used for transection, resection, and/or creation of anastomoses.

# Performance Data:

Preclinical testing was performed to ensure that the devices performed as intended. Testing demonstrated satisfactory performance in transection, resection, and/or creation of anastomoses. The modified device will provide increased clamping force when the anvil and channel (which holds the cartridge) is clamped down across tissue, resulting in more consistent staple formation.

## Conclusion:

Based on 21CFR § 807, we conclude that the modified products are as safe and effective as the marketed devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K002398](https://fda-staging.innolitics.com/device/K002398)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
