ACUMATCH L-SERIES CEMENTED FEMORAL STEM, MODEL L-SERIES CEMENTED
K001335 · Exactech, Inc. · JDI · May 18, 2000 · Orthopedic
Device Facts
Record ID
K001335
Device Name
ACUMATCH L-SERIES CEMENTED FEMORAL STEM, MODEL L-SERIES CEMENTED
Applicant
Exactech, Inc.
Product Code
JDI · Orthopedic
Decision Date
May 18, 2000
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AcuMatch L-Series Cemented Femoral Stem Components are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion. AcuMatch L-Series Cemented Femoral Stem Components are intended to be used with bone cement.
Device Story
Cemented femoral stem component for total hip arthroplasty; manufactured from cast Cobalt Chromium Molybdenum alloy. Designed for surgical implantation by orthopedic surgeons in hospital settings. Device provides structural replacement for femoral head/neck; requires bone cement for fixation. Features proximal-to-distal taper and satin surface finish. Clinical benefit includes restoration of hip joint function and mobility. Substantial equivalence established via mechanical fatigue testing compared to predicate femoral stems.
Clinical Evidence
Bench testing only. Three-point fatigue testing performed to compare stem strength against legally marketed predicate devices; results confirmed strength is within the range of existing devices.
Technological Characteristics
Material: Cast Cobalt Chromium Molybdenum alloy (ASTM F 75-98). Design: Proximal to distal taper with satin surface finish. Energy source: None (mechanical implant). Connectivity: None. Sterilization: Gamma radiation (Cobalt 60), 25–37 kGy, SAL 10^-6.
Indications for Use
Indicated for skeletally mature individuals undergoing primary total hip replacement for osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative hip conditions, or proximal femoral fractures. Also indicated for revision of failed reconstructions with sufficient bone stock and to restore mobility after previous fusion. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing hip joint control, or patients whose weight, age, or activity level suggests early system failure.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Koo1335
# 510(k) Summary of Safety and Effectiveness
| Trade Name: | Exactech® AcuMatch L-Series<br>Cemented Femoral Stem Component |
|----------------------|------------------------------------------------------------------------------------|
| Common Name: | Femoral Stem |
| Classification Name: | Prosthesis, Hip, Semi-Constrained, Metal/Polymer,<br>Cemented, (Femoral Component) |
### Legally Marketed Devices for Substantial Equivalence Comparison:
| Model | Manufacturer |
|-------------|---------------------|
| AuRA | Exactech (#K961304) |
| Conquest FX | Smith & Nephew |
| Spectron | Smith & Nephew |
| PFC | Depuy |
#### Substantial Equivalence Information:
The Exactech AcuMatch L-Series Cemented Femoral component has similar indications and contraindications as other femoral components legally marketed in the United States. In addition the L-Series has similar technological features to other devices, most notably Exactech's AuRA femoral component. The L-Series originated from design modifications made to the Exactech's AuRA design (ref. #K961304). In addition, the proposed L-Series component is similar to femoral components currently marketed by other manufacturers. These include the "Conquest FX" and "Spectron" by Smith & Nephew and the "PFC" by Depuy. The Conquest FX model like the L-Series is manufactured from cast cobalt chrome. Other similarities between the predicate stems and the proposed L-Series design include a satin surface treatment and a proximal to distal taper. All of the components are supplied sterile. Three-Point fatigue testing of the Exactech L-Series device places the strength of the stem in the range of other legally marketed devices.
{1}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
#### Indications for Use:
AcuMatch L-Series Cemented Femoral Stem Components are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
AcuMatch L-Series Cemented Femoral Stem Components are intended to be used with bone cement.
#### Contraindications:
AcuMatch L-Series Cemented Femoral Stem Components are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
#### Device Description:
L-Series Femoral Stem Components are made from cast Cobalt Chromium Molybdenum alloy conforming to ASTM F 75-98. Mechanical properties for ultimate strength, ductility, and grain structure are controlled by this specification. The L-Series components have a satin finish and are intended for cemented applications only.
#### Packaging. Labeling and Sterilization:
L-Series components are first packaged at Exactech in a certified Class 100,000 Cleanroom. The products are then shipped to an ISO/EN certified Irradiation Facility and returned to Exactech where they are quarantined pending a final product inspection. Qualifying implants are then released for distribution. Packaging materials are outlined in the following table.
{2}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| Material | Composition |
|---------------------|-----------------------------------|
| Inner / Outer Trays | PETG – 0.040" thickness |
| Tray Lids | Spun-Bonded Olefin - Tyvek® |
| Inserts | Medium grade LD45 Foam |
| Box | Heavy weight cardboard |
| Outer Shrink-Wrap | Clear, Light-Weight Plastic |
| Shipping Cartons | Heavy-weight Corrugated Cardboard |
Utilization and implantation instructions are included in the package insert provided with each product. The name, size, dimension, material, lot, serial number and sterility status are indicated on the labeling.
### Sterilization Specifications:
Method: Gamma radiation (Cobalt 60 source) Dose: 25 – 37 kGy Sterility Assurance Level (SAL): 10-6
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### MAY 1 8 2000
Ms. Lisa Simpson Regulatory Representative Exactech, Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K001335
Trade Name: Exactech® AcuMatch™ Integrated Hip System L-series Cemented Femoral Stem Component Regulatory Class: II Product Code: JDI Dated: April 26, 2000 Received: April 27,2000
Dear Ms. Simpson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
Page 2 - Ms. Lisa Simpson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Neil RPEyden
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number:
Kool335
Device Name:
Exactech® AcuMatch™ Integrated Hip System L-Series Cemented Femoral Stem Component
#### Indications for Use:
AcuMatch L-Series Cemented Femoral Stem Components are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to SKeloary mature marriad arthritis, osteonecrosis, ankylosing spondylitis, congenital hip Osteourinitie, Includio degenerative problems of the hip, and for treatment of proximal as the fentoral fractures whice pponents of Exactech Hip Systems are also potentially indicated presenta a softailed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
AcuMatch L-Series Cemented Femoral Stem Components are intended to be used with bone cement.
#### Contraindications:
AcuMatch L-Series Cemented Femoral Stem Components are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | NRO for cmw |
|-----------------------------------------|-------------|
| Division of General Restorative Devices | |
| 510(k) Number | K001335 |
| Prescription Use | Yes | or | Over the Counter Use | No |
|------------------|-----|----|----------------------|----|
|------------------|-----|----|----------------------|----|
Section 3 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.