← Product Code [GCJ](/productcode/GCJ) · K000663

# KRONNER LOW PROFILE SCOPE HOLDER (K000663)

_Kronner Prototypes, Inc. · GCJ · May 2, 2000 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K000663

## Device Facts

- **Applicant:** Kronner Prototypes, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** May 2, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

KSH-4 Kronner Low Profile Scope Holder For abdominal and thoracic endoscopic surgical procedures HPL2-CRL High pressure* Gas Line Set For abdominal and thoracic endoscopic and arthroscopic surgical procedures KSHSA-3 Standard Arm Assembly For abdominal and thoracic endoscopic surgical procedures C-1 Control For abdominal, thoracic, arthroscopic, endoscopic surgery and holding manual surgical instruments ECB-2 Electronic Control Box For abdominal, thoracic, arthroscopic, nasal endoscopic surgical procedures and holding manual surgical instruments GLW-1 Gas line wrench For abdominal, thoracic, arthroscopic, nasal endoscopic surgical procedures and holding manual surgical instruments SEAH-1 Small Endoscope Accessory with Handle For nasal endoscopic surgical procedures HPL1-BL1-CRL High pressure* Branched Gas Line Set For nasal endoscopic surgical procedures and holding manual surgical instruments

## Device Story

Kronner Low Profile Scope Holder system supports endoscopes and manual surgical instruments during endoscopic procedures; abdominal, thoracic, arthroscopic, and nasal. System comprises arm assembly, gas line sets, and control components. Device utilizes electronic control box with solenoids to regulate gas output for positioning/holding; replaces mechanical control of predicate. Operated by surgical staff in OR. Provides stable positioning of instruments/scopes; reduces manual fatigue; improves procedural efficiency. Output is mechanical support/stabilization of surgical tools.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

System includes arm assembly, gas lines, and electronic control box. Uses solenoid-based electronic control for gas output regulation. Designed for endoscopic surgical support. Operates with gas pressures up to 150 PSI.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Kronner Low Profile Scope Holder ([K973543](/device/K973543.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

JUL 2 7 2015

Kronner Prototypes, Inc.

Richard F. Kronner, M.D.

President

1443 Upper Cleveland Rapids Road Roseburg, OR 97470

K000663 Re:

> Trade/Device Name: Low Profile Scope Holder Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated (Date on orig SE Itr): February 22, 2000 Received (Date on orig SE ltr): February 28, 2000

Dear Dr. Kronner,

This letter corrects our substantially equivalent letter of May 2, 2000.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

**Benjamin R. Fisher -S**

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

KOODEES 510(K) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Kronner Low Profile Scope Holder

Indications for Use:

KSH-4 Kronner Low Profile Scope Holder KYOMAET LOW I TOTAC Soupe Works surgical procedures

### HPL2-CRL High pressure® Gas Line Set

H. High pressure - Gas Eine Oot

### KSHSA-3 Standard Arm Assembly

For abdominal and thoracic endoscopic surgical procedures

#### C-1 Control

For abdominal, thoracic, arthroscopic, endoscopic surgery and holding manual surgical instruments

#### ECB-2 Electronic Control Box

For abdominal, thoracic, arthroscopic, nasal endoscopic surgical r of action. Than and holding manual surgical instruments

#### GLW-1 Gas line wrench

Gas this through, anthroscopic, nasal endoscopic surgical For abounting, mording manual surgical instruments

### SEAH-1 Small Endoscope Accessory with Handle

For nasal endoscopic surgical procedures

### HPL1-BL1-CRL. High pressure* Branched Gas Line Set

ET-ORL Thyn procedures and holding and holding manual surgical Instruments

· Tolerates pressures to 150 PSI, do not use with pressures over 150 psi.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

e R. Vochner.

Division Restorative Devices
510(k) Number K600663

Prescription Use
(Per 21 CFR 801.109)

27

1999

Over-The-Counter Use _

(Optional Format 1-2-96)

8-2

{3}------------------------------------------------

MAY - 2 2000

41

# k 000663

# 510(K) SUMMARY as required by 807.92(c)

{4}------------------------------------------------

K00663

### KRONNER PROTOTYPES, INC.

1443 Upper Cleveland Rapids Road
 Phone: (541) 672-2543 

#### Prepared: February 22, 2000

Contact Person: Richard Kronner, President

દર

{5}------------------------------------------------

K 000663

# Trade Name: Kronner Low Profile Scope Holder

## Common Name: Endoscope Holder

# Classification Name: Endoscope holding device (no industry name for this device)

{6}------------------------------------------------

000663

# Equivalent to legally marketed devices

by

# ( K973543 ) Kronner Prototypes, Inc.

4

{7}------------------------------------------------

K000663

### Description:

The Kronner Low Profile Scope Holder with Electronic Control and Control Box is the same as the Kronner Low Profile Scope Holder (K973543) with mechanical control except that it has an electronic control, momentary button, and electronic control box with solenoids to control gas output.

4

{8}------------------------------------------------

## K 000 663

## Intended Usage:

#### KSH-4 Kronner Low Profile Scope Holder

For abdominal and thoracic endoscopic surgical procedures

### HPL2-CRL High pressure* Gas Line Set

For abdominal and thoracic endoscopic and arthroscopic surgical procedures

### KSHSA-3 Standard Arm Assembly

For abdominal and thoracic endoscopic surgical procedures

#### C-1 Control

For abdominal, thoracic, arthroscopic, endoscopic surgery and holding manual surgical instruments

### ECB-2 Electronic Control Box

For abdominal, thoracic, arthroscopic, nasal endoscopic surgical procedures and holding manual surgical instruments

### GLW-1 Gas line wrench

For abdominal, thoracic, arthroscopic, nasal endoscopic surgical procedures and holding manual surgical instruments

### SEAH-1 Small Endoscope Accessory with Handle

For nasal endoscopic surgical procedures

### HPL1-BL1-CRL High pressure* Branched Gas Line Set

For nasal endoscopic surgical procedures and holding manual surgical instruments

· Tolerates pressures to 150 PSI, do not use with pressures over 150 psi.

{9}------------------------------------------------

000663

## Summary of technological characteristics of device compared to predicate devices.

The Kronner Low Profile Scope Holder with Electronic control and Electronic Control Box is essentially equivalent to the Kronner Low Profile Scope Holder (K973543) except that the mechanical control has been replaced with an electronic control and electronic control box with solenoids to control gas output.

---

**Source:** [https://fda-staging.innolitics.com/device/K000663](https://fda-staging.innolitics.com/device/K000663)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).