DEN250028 · Spectralmd, Inc. · SHY · May 21, 2026 · General, Plastic Surgery
Device Facts
Record ID
DEN250028
Device Name
DeepView AI® System
Applicant
Spectralmd, Inc.
Product Code
SHY · General, Plastic Surgery
Decision Date
May 21, 2026
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 878.1840
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Deep burn tissue detection
—
—
—
—
—
—
—
Wound area measurement
—
—
—
—
—
—
—
Percent total body surface area (TBSA) burn
—
—
—
—
—
—
—
Indications for Use
The DeepView AI®-Burn software analyzes images obtained using the DeepView AI® System, using artificial intelligence to detect deep burn tissue (deep partial thickness and full thickness) and provides segmentation of specific areas in a burn wound by highlighting areas that are unlikely to heal within 21 days with conservative treatment alone. The absence of a highlighted region on an image indicates the absence of deep burn tissue. The system will calculate wound area measurement and percent total body surface area (TBSA) burn. DeepView AI® System with AI®-Burn software is an aid to physicians in the management of thermal burn wounds and considering various treatment plans. The DeepView AI® System should be used in conjunction with clinical assessment and not as a stand-alone diagnostic device. The device's segmentation contour does not provide definitive delineation of the deep burn margin for precise excision sizing or delineating the precise borders of non-healing tissue.
Device Story
DeepView AI® System captures optical images of thermal burn wounds; AI®-Burn software processes images to identify deep partial-thickness and full-thickness tissue; system highlights areas unlikely to heal within 21 days with conservative treatment; calculates wound area and percent TBSA; used by physicians in clinical settings; provides adjunctive information to support management and treatment planning; not for standalone diagnosis; does not provide definitive excision margins; benefits include improved diagnostic characterization of burn depth to guide clinical decision-making.
Clinical Evidence
Clinical performance validation required to demonstrate improved diagnostic characterization by users compared to unaided assessment. Testing must include standalone performance (sensitivity, specificity) relative to ground truth, segmentation accuracy, and subgroup analysis across relevant risk factors. No specific clinical trial results are provided in the summary document.
Technological Characteristics
Optical imaging system; software-based analysis; non-contact; requires compatibility with specific image acquisition hardware; photobiological safety testing for light sources; electromagnetic compatibility; electrical and thermal safety testing; software verification and validation.
Indications for Use
Indicated for use by physicians as an adjunctive diagnostic aid in the management of thermal burn wounds to detect deep partial-thickness and full-thickness tissue and calculate wound area and percent TBSA.
Regulatory Classification
Identification
The DeepView AI® System is a prescription device that uses artificial intelligence to analyze images of thermal burn wounds. It detects deep burn tissue (deep partial thickness and full thickness), provides segmentation of areas unlikely to heal within 21 days with conservative treatment alone, and calculates wound area measurement and percent total body surface area (TBSA) burn. It is intended as an aid to physicians in the management of thermal burn wounds in conjunction with clinical assessment and is not a stand-alone diagnostic device.
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 21, 2026
Spectral MD, Inc.
Trudy Estridge
Senior Director Regulatory Affairs
2515 Mckinney Ave. Suite 1000
Dallas, Texas 75201
Re: DEN250028
Trade/Device Name: DeepView AI® System
Regulation Number: 21 CFR 878.1840
Regulation Name: Software-aided adjunctive diagnostic device for use by healthcare providers in skin wound assessment.
Regulatory Class: Class II
Product Code: SHY
Dated: June 26, 2025
Received: March 6, 2026
Dear Trudy Estridge:
The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA) has completed its review of your De Novo request for classification of the DeepView AI® System, a prescription device under 21 CFR Part 801.109 with the following indications for use:
The DeepView AI®-Burn software analyzes images obtained using the DeepView AI® System, using artificial intelligence to detect deep burn tissue (deep partial thickness and full thickness) and provides segmentation of specific areas in a burn wound by highlighting areas that are unlikely to heal within 21 days with conservative treatment alone. The absence of a highlighted region on an image indicates the absence of deep burn tissue. The system will calculate wound area measurement and percent total body surface area (TBSA) burn.
DeepView AI® System with AI®-Burn software is an aid to physicians in the management of thermal burn wounds and considering various treatment plans. The DeepView AI® System should be used in conjunction with clinical assessment and not as a stand-alone diagnostic device. The device's segmentation contour does not provide definitive delineation of the deep burn margin for precise excision sizing or delineating the precise borders of non-healing tissue.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
DEN250028 - Trudy Estridge
Page 2
FDA concludes that this device should be classified into Class II. This order, therefore, classifies the DeepView AI® System, and substantially equivalent devices of this generic type, into Class II under the generic name software-aided adjunctive diagnostic device for use by healthcare providers in skin wound assessment.
FDA identifies this generic type of device as:
**Software-aided adjunctive diagnostic device for use by healthcare providers in skin wound assessment.** A software-aided adjunctive diagnostic device for use by healthcare providers in skin wound assessment uses a software algorithm to analyze optical or other physical properties of a skin wound without direct wound contact and returns characterizing information that may include classification, severity, or healing potential of the skin wound. The device is intended for prescription use by healthcare providers to provide adjunctive information to be used in conjunction with clinical assessment. It is not for use as a standalone diagnostic device.
Section 513(f)(2) of the Food, Drug and Cosmetic Act (the FD&C Act) was amended by section 607 of the Food and Drug Administration Safety and Innovation Act (FDASIA) on July 9, 2012. This law provides two options for De Novo classification. First, any person who receives a "not substantially equivalent" (NSE) determination in response to a 510(k) for a device that has not been previously classified under the Act may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act. On December 13, 2016, the 21st Century Cures Act removed a requirement that a De Novo request be submitted within 30 days of receiving an NSE determination. Alternatively, any person who determines that there is no legally marketed device upon which to base a determination of substantial equivalence may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act without first submitting a 510(k). FDA shall, within 120 days of receiving such a request, classify the device. This classification shall be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing the classification.
On March 6, 2026, FDA received your De Novo requesting classification of the DeepView AI® System. The request was submitted under section 513(f)(2) of the FD&C Act. In order to classify the DeepView AI® System into class I or II, it is necessary that the proposed class have sufficient regulatory controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the De Novo request FDA has determined that, for the previously stated indications for use, the DeepView AI® System can be classified in class II with the establishment of special controls for class II. FDA believes that class II (special) controls provide reasonable assurance of the safety and effectiveness of the device type. The identified risks and mitigation measures associated with the device type are summarized in the following table:
| Risks to Health | Mitigation Measures |
| --- | --- |
| False negative results, leading to delayed treatment or failure to treat appropriately, or false positive results, leading to unnecessary referrals, medical therapy (e.g., medication), and/or medical procedures (e.g., surgery) | Clinical performance testing Non-clinical performance testing Labeling |
{2}
DEN250028 - Trudy Estridge
Page 3
| Risks to Health | Mitigation Measures |
| --- | --- |
| False or inaccurate results, or failure to generate a result due to use error or improper device use | Precision testing Human factors testing Labeling |
| False or inaccurate results, or failure to generate a result due to device failure or malfunction | Non-clinical performance testing Precision testing Software verification, validation, and hazard analysis Labeling |
| Electrical, thermal, or light exposure-related injury (e.g., burn, eye damage) | Electrical and thermal safety testing Software verification, validation, and hazard analysis Labeling |
| Interference with other devices | Electromagnetic compatibility testing |
In combination with the general controls of the FD&C Act, the software-aided adjunctive diagnostic device for use by healthcare providers in skin wound assessment is subject to the following special controls:
(1) Data obtained from clinical performance validation testing must:
(i) Demonstrate improvement of device-aided users' diagnostic characterization of the indicated wound compared to the accuracy of unaided users in the intended patient population and under anticipated conditions of use;
(ii) Include an evaluation of patients across risk factors representative of the intended use population under the anticipated conditions of use; and
(iii) Include standalone device performance testing that demonstrates the accuracy of the device output relative to ground truth in the intended patient population and under anticipated conditions of use, including the following:
(A) Testing must include sensitivity and specificity of the device output with clinical justification of the reported results;
(B) For devices that provide mapping of the wound area (e.g., wound borders, extent of wound, etc.), testing must demonstrate segmentation accuracy;
(C) Wounds must be selected by representative users and include a justified quantity and range of wound types;
(D) Justification must be provided for the determination of ground truth; and
(E) Analysis of standalone performance must include subgroup analysis by relevant risk factors.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following testing must be performed:
(i) Device software must be tested for compatibility with specific signal or image acquisition hardware. Testing must include a description of compatible hardware and processes, prespecified compatibility testing protocols, and dataset(s); and
(ii) Performance testing must demonstrate the photobiological safety of any lamp or lamp systems.
(3) Performance testing must demonstrate device precision, including repeatability and reproducibility of device performance, across operators and challenging use conditions.
{3}
DEN250028 - Trudy Estridge
Page 4
(4) Performance testing must demonstrate electromagnetic compatibility and electrical and thermal safety of any electrical components of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) A human factors assessment must demonstrate that the device can be safely and correctly used by intended users based solely on the directions for use.
(7) Labeling must include:
(i) A summary of standalone and clinical performance testing conducted with the device. The summary must describe performance measures, including sensitivity and specificity, and statistical confidence intervals, as well as performance of the device for all clinically relevant subgroups within the intended use population;
(ii) Information related to the limitations of device performance or subpopulations for which the device may not perform as expected;
(iii) Information needed to facilitate interpretation of all device outputs;
(iv) A statement that the device is not intended for use as a standalone diagnostic and is not for use to confirm a clinical diagnosis; and
(v) Warnings to avoid unsafe exposure to any energy-emitting components of the device (e.g., excluding use of the device on wounds close to the eye).
In addition, this is a prescription device and must comply with 21 CFR 801.109.
Although this letter refers to your product as a device, please be aware that some granted products may instead be combination products. If you have questions on whether your product is a combination product, contact CDRHProductJurisdiction@fda.hhs.gov.
Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the FD&C Act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type. FDA has determined premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device type and, therefore, the device is not exempt from the premarket notification requirements of the FD&C Act. Thus, persons who intend to market this device type must submit a premarket notification containing information on the software-aided adjunctive diagnostic device for use by healthcare providers in skin wound assessment they intend to market prior to marketing the device.
Please be advised that FDA's decision to grant this De Novo request does not mean that FDA has made a determination that your device complies with other requirements of the FD&C Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the FD&C Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR
{4}
DEN250028 - Trudy Estridge
Page 5
4, Subpart A) for combination products; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the FD&C Act; 21 CFR 1000-1050).
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System Rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
A notice announcing this classification order will be published in the Federal Register. A copy of this order and supporting documentation are on file in the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Room 1061, Rockville, MD 20852 and are available for inspection between 9 a.m. and 4 p.m., Monday through Friday.
As a result of this order, you may immediately market your device as described in the De Novo request, subject to the general control provisions of the FD&C Act and the special controls identified in this order.
For comprehensive regulatory information about medical devices and radiation-emitting products, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
If you have any questions concerning the contents of the letter, please contact Scott Kominsky at 301-796-9601.
Sincerely,
JULIE A. MORABITO -S
2026.05.21 09:27:30
-04'00'
Julie Morabito, Ph.D.
Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.