DEN250007 · Ultrasound AI · SHE · Feb 11, 2026 · Radiology
Device Facts
Record ID
DEN250007
Device Name
Delivery Date AI
Applicant
Ultrasound AI
Product Code
SHE · Radiology
Decision Date
Feb 11, 2026
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 892.8200
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Predicted Delivery Date
—
—
—
—
—
—
—
Indications for Use
Delivery Date AI is a post-processing software designed to analyze pregnancy ultrasound images using machine learning techniques. It analyzes raw images to provide healthcare practitioners (HCPs) with a Predicted Delivery Date (PDD) in women without a reliable estimated delivery date. The software estimates the predicted delivery date for a singleton pregnancy and should be used as an aid in clinical judgment, alongside standard methods for assessing gestational age. This is a prediction of the actual delivery date called the Predicted Delivery Date (PDD). The software processes entire images, which include fetal and maternal characteristics, and targets a specific patient population consisting of women 18 years of age or older with a singleton pregnancy at or beyond 14 0/7 weeks through 36 6/7 weeks gestation who lack a reliable Estimated Delivery Date (EDD) due to: - Unreliable LMP: Examples include irregular menstrual cycles, spotting mistaken for menses, or recent use of hormonal contraceptives. - No First-Trimester Ultrasound: Absence of an ultrasound before 14 weeks gestation, precluding crown-rump length measurement. - Gestational Age Confirmation: Pregnancy within the indicated range verified by clinical indicators such as fundal height or detection of a fetal heartbeat.
Device Story
Delivery Date AI is post-processing software using machine learning to analyze raw pregnancy ultrasound images; inputs include fetal and maternal characteristics from images; outputs Predicted Delivery Date (PDD). Used by healthcare practitioners in clinical settings as an aid to clinical judgment alongside standard gestational age assessment methods. Software processes images to estimate delivery date for singleton pregnancies; output intended to support, not replace, clinician review. Does not identify at-risk pregnancies or complications. Benefits include providing a delivery date estimate for patients lacking reliable EDD data due to irregular cycles or late-term presentation.
Clinical Evidence
Clinical performance validation testing required under anticipated conditions of use. Must demonstrate performance in the intended patient population using an independent test dataset. Documentation must include study objectives, endpoints, statistical hypotheses, performance goals, sample size calculations, and statistical analyses comparing device output to known delivery information (e.g., Bland-Altman plots, error metrics). Testing must capture inter- and intra-patient variability and characterize performance across demographic subgroups and relevant confounders.
Technological Characteristics
Software-only radiological system; utilizes machine learning techniques for image analysis. Processes raw obstetric ultrasound images. Requires integration with clinical workflow and imaging hardware. Subject to special controls regarding software verification, validation, and hazard analysis. Must feature specific limitations and warnings in the user interface.
Indications for Use
Indicated for women 18+ years old with singleton pregnancies between 14 0/7 and 36 6/7 weeks gestation lacking a reliable Estimated Delivery Date (EDD) due to unreliable LMP or absence of first-trimester ultrasound. Requires confirmation of pregnancy within indicated range via clinical indicators (e.g., fundal height, fetal heartbeat).
Regulatory Classification
Identification
Delivery Date AI is a post-processing software designed to analyze pregnancy ultrasound images using machine learning techniques. It analyzes raw images to provide healthcare practitioners (HCPs) with a Predicted Delivery Date (PDD) in women 18 years of age or older with a singleton pregnancy at or beyond 14 0/7 weeks through 36 6/7 weeks gestation who lack a reliable Estimated Delivery Date (EDD) due to unreliable LMP or no first-trimester ultrasound. The software should be used as an aid in clinical judgment, alongside standard methods for assessing gestational age.
Special Controls
In combination with the general controls of the FD&C Act, the radiological software system for delivery date prediction is subject to the following special controls:
- (1) Data obtained from clinical performance validation testing acquired under anticipated conditions of use must demonstrate that the device performs as intended when used in the intended patient population. Documentation must include the following:
- (i) A clear description of the patient population the device is intended to analyze.
- (ii) A clear description of the clinical environment and context of use in which the device is intended to be used.
- (iii) A description of the dataset(s) used, including detailed annotation methods and important cohorts (e.g., subsets defined by patient demographics, clinically relevant confounders, and image acquisition characteristics). Validation must contain a sufficient number of cases from important cohorts (i.e., subsets defined by clinically relevant demographics, confounders, effect modifiers, concomitant diseases, challenging cases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. The test dataset must be independent of the data used in the training/development of the device.
- (iv) A description of prespecified performance testing protocols (including but not limited to the study objectives, study endpoints, statistical hypotheses, performance goals, sample size calculation, and statistical analyses). Performance goals used to determine success of the clinical validation study must be clinically justified and must be based on appropriate objective performance measures comparing the device output to known delivery information on sampled patients (e.g., Bland-Altman plots, error metrics).
- (v) Performance testing should capture inter- and intra- patient variability and ranges of agreement between the device predictions and reference standard, as well as other relevant sources of variation.
- (2) Device design characteristics must ensure that information and limitations in special control (4)(i) and 4(vii) are featured in the software user interface.
- (3) Software verification, validation, and hazard analysis must be provided. Software documentation must include a detailed technical description of all image analysis algorithms, including the algorithm inputs and outputs, each major component or block, and any algorithm limitations.
- (4) Labeling must include:
- (i) A description of the meaning of the date predicted by the device and the relationship of the date to other calculated dates used in clinical practice (e.g., estimated delivery date);
- (ii) A description of the intended patient population, the intended user, clinical environment, and context of use, including information on interpretation of outputs within the intended clinical workflow;
- (iii) A description of the device inputs and outputs;
- (iv) A description of compatible imaging hardware and imaging protocols;
- (v) A summary of the development data and clinical validation data, including sources of data, study sites, samples sizes, demographics and other relevant characteristics of the study participants;
- (vi) A summary of performance testing including test methods, dataset characteristics, reference standard, testing environment, results (with confidence intervals), and a summary of clinical performance for all demographic subgroups and relevant confounders from testing dataset(s);
- (vii) Limiting statements that indicate:
- (A) The output is not intended to predict or assess the risk of pre-term birth;
- (B) A description of situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable; and
- (C) A warning that users should use the device in conjunction with other clinical and diagnostic findings, including information obtained by alternative methods and clinical evaluation, as appropriate.
- (5) The device manufacturer must develop and implement a post-market performance management plan that ensures regular assessment of the generalizability and device performance in the intended patient population in real-world use. The plan must include:
- (i) Data collection, analysis methods, and procedures for:
- (A) Monitoring relevant performance characteristics and detecting changes in performance;
- (B) Identifying sources of performance changes between validation and real-world environment over time; and
- (C) Assessing the results from the performance testing on safety and effectiveness;
- (ii) Procedures for communicating the device's current performance to the users.
Submission Summary (Full Text)
{0}------------------------------------------------
Ultrasound AI % Raymond Kelly Consultant Arazy Group Consultants Inc. 3422 Leonardo Lane NEW SMYRNA FL. 32168
Re: DEN250007
Trade/Device Name: Delivery Date AI Regulation Number: 21 CFR 892.8200
Regulation Name: Radiological software system for delivery date prediction
Regulatory Class: Class II
Product Code: SHE
Dated: February 28, 2025 Received: March 17, 2025
## Dear Raymond Kelly:
The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA) has completed its review of your De Novo request for classification of the Delivery Date AI, a prescription device under 21 CFR Part 801.109 with the following indications for use:
Delivery Date AI is a post-processing software designed to analyze pregnancy ultrasound images using machine learning techniques. It analyzes raw images to provide healthcare practitioners (HCPs) with a Predicted Delivery Date (PDD) in women without a reliable estimated delivery date.
The software estimates the predicted delivery date for a singleton pregnancy and should be used as an aid in clinical judgment, alongside standard methods for assessing gestational age. This is a prediction of the actual delivery date called the Predicted Delivery Date (PDD).
The software processes entire images, which include fetal and maternal characteristics, and targets a specific patient population consisting of women 18 years of age or older with a singleton pregnancy at or beyond 14 0/7 weeks through 36 6/7 weeks gestation who lack a reliable Estimated Delivery Date (EDD) due to:
- Unreliable LMP: Examples include irregular menstrual cycles, spotting mistaken for menses, or recent use of hormonal contraceptives.
- No First-Trimester Ultrasound: Absence of an ultrasound before 14 weeks gestation, precluding crown-rump length measurement.
{1}------------------------------------------------
• Gestational Age Confirmation: Pregnancy within the indicated range verified by clinical indicators such as fundal height or detection of a fetal heartbeat.
FDA concludes that this device should be classified into Class II. This order, therefore, classifies the Delivery Date AI, and substantially equivalent devices of this generic type, into Class II under the generic name radiological software system for delivery date prediction.
FDA identifies this generic type of device as:
Radiological software system for delivery date prediction. A radiological software system for delivery date prediction is a software-only device that analyzes obstetric radiological images to predict the delivery date of a pregnancy. The device output does not replace a clinician's review or judgment on the progression of a pregnancy and should be used in conjunction with other clinical and diagnostic findings. The device does not identify at-risk pregnancies or identify specific complications.
Section 513(f)(2) of the Food, Drug and Cosmetic Act (the FD&C Act) was amended by section 607 of the Food and Drug Administration Safety and Innovation Act (FDASIA) on July 9, 2012. This law provides two options for De Novo classification. First, any person who receives a "not substantially equivalent" (NSE) determination in response to a 510(k) for a device that has not been previously classified under the Act may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act. On December 13, 2016, the 21st Century Cures Act removed a requirement that a De Novo request be submitted within 30 days of receiving an NSE determination. Alternatively, any person who determines that there is no legally marketed device upon which to base a determination of substantial equivalence may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act without first submitting a 510(k). FDA shall, within 120 days of receiving such a request, classify the device. This classification shall be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing the classification.
On March 17, 2025, FDA received your De Novo requesting classification of the Delivery Date AI. The request was submitted under section 513(f)(2) of the FD&C Act. In order to classify the Delivery Date AI into class I or II, it is necessary that the proposed class have sufficient regulatory controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the De Novo request, FDA has determined that, for the previously stated indications for use, the Delivery Date AI can be classified in class II with the establishment of special controls for class II. FDA believes that class II (special) controls provide reasonable assurance of the safety and effectiveness of the device type. The identified risks and mitigation measures associated with the device type are summarized in the following table:
| Risks to Health | Mitigation Measures |
|------------------------------------------------|-----------------------------------------------|
| Inappropriate patient management due to | Clinical performance testing |
| inaccurate prediction of actual delivery date, | Software verification, validation, and hazard |
| whether significantly early or late | analysis |
| | Labeling |
{2}------------------------------------------------
| Risks to Health | Mitigation Measures |
|---------------------------------------------------|-----------------------------------------------|
| | Postmarket monitoring plan |
| Inappropriate patient management resulting | Clinical performance testing |
| from inconsistent device output due to | Software verification, validation, and hazard |
| differences in acquisitions, equipment settings, | analysis |
| and other image input characteristics | Labeling |
| | Postmarket monitoring plan |
| User misinterpretation of the device's output as | Labeling |
| being directly reflective of the gestational age | Device Characteristics |
| of the fetus, contributing to: | |
| - Inappropriate patient management | |
| <ul> <li>Overreliance on device output</li> </ul> | |
In combination with the general controls of the FD&C Act, the radiological software system for delivery date prediction is subject to the following special controls:
- (1) Data obtained from clinical performance validation testing acquired under anticipated conditions of use must demonstrate that the device performs as intended when used in the intended patient population. Documentation must include the following:
- (i) A clear description of the patient population the device is intended to analyze.
- (ii) A clear description of the clinical environment and context of use in which the device is intended to be used.
- (iii) A description of the dataset(s) used, including detailed annotation methods and important cohorts (e.g., subsets defined by patient demographics, clinically relevant confounders, and image acquisition characteristics). Validation must contain a sufficient number of cases from important cohorts (i.e., subsets defined by clinically relevant demographics, confounders, effect modifiers, concomitant diseases, challenging cases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. The test dataset must be independent of the data used in the training/development of the device.
- (iv) A description of prespecified performance testing protocols (including but not limited to the study objectives, study endpoints, statistical hypotheses, performance goals, sample size calculation, and statistical analyses). Performance goals used to determine success of the clinical validation study must be clinically justified and must be based on appropriate objective performance measures comparing the device output to known delivery information on sampled patients (e.g., Bland-Altman plots, error metrics).
- (v) Performance testing should capture inter- and intra- patient variability and ranges of agreement between the device predictions and reference standard, as well as other relevant sources of variation.
- (2) Device design characteristics must ensure that information and limitations in special control (4)(i) and 4(vii) are featured in the software user interface.
{3}------------------------------------------------
- (3) Software verification, validation, and hazard analysis must be provided. Software documentation must include a detailed technical description of all image analysis algorithms, including the algorithm inputs and outputs, each major component or block, and any algorithm limitations.
- (4) Labeling must include:
- (i) A description of the meaning of the date predicted by the device and the relationship of the date to other calculated dates used in clinical practice (e.g., estimated delivery date);
- (ii) A description of the intended patient population, the intended user, clinical environment, and context of use, including information on interpretation of outputs within the intended clinical workflow;
- (iii) A description of the device inputs and outputs;
- (iv) A description of compatible imaging hardware and imaging protocols;
- (v) A summary of the development data and clinical validation data, including sources of data, study sites, samples sizes, demographics and other relevant characteristics of the study participants;
- (vi) A summary of performance testing including test methods, dataset characteristics, reference standard, testing environment, results (with confidence intervals), and a summary of clinical performance for all demographic subgroups and relevant confounders from testing dataset(s);
- (vii) Limiting statements that indicate:
- (A) The output is not intended to predict or assess the risk of pre-term birth;
- (B) A description of situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable; and
- (C) A warning that users should use the device in conjunction with other clinical and diagnostic findings, including information obtained by alternative methods and clinical evaluation, as appropriate.
- (5) The device manufacturer must develop and implement a post-market performance management plan that ensures regular assessment of the generalizability and device performance in the intended patient population in real-world use. The plan must include:
- (i) Data collection, analysis methods, and procedures for:
- (A) Monitoring relevant performance characteristics and detecting changes in performance;
- (B) Identifying sources of performance changes between validation and real-world environment over time; and
- (C) Assessing the results from the performance testing on safety and effectiveness;
- (ii) Procedures for communicating the device's current performance to the users.
In addition, this is a prescription device and must comply with 21 CFR 801.109.
Although this letter refers to your product as a device, please be aware that some granted products may instead be combination products. If you have questions on whether your product is a combination product, contact CDRHProductJurisdiction@fda.hhs.gov.
Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the FD&C Act, if FDA determines that premarket
{4}------------------------------------------------
notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type. FDA has determined premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device type and, therefore, the device is not exempt from the premarket notification requirements of the FD&C Act. Thus, persons who intend to market this device type must submit a premarket notification containing information on the radiological software system for delivery date prediction they intend to market prior to marketing the device.
Please be advised that FDA's decision to grant this De Novo request does not mean that FDA has made a determination that your device complies with other requirements of the FD&C Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the FD&C Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see <a href="https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products">https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products</a>); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the FD&C Act; 21 CFR 1000-1050).
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System Rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at <a href="https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.">https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.</a>
A notice announcing this classification order will be published in the Federal Register. A copy of this order and supporting documentation are on file in the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Room 1061, Rockville, MD 20852 and are available for inspection between 9 a.m. and 4 p.m., Monday through Friday.
As a result of this order, you may immediately market your device as described in the De Novo request, subject to the general control provisions of the FD&C Act and the special controls identified in this order.
For comprehensive regulatory information about medical devices and radiation-emitting products, please see Device Advice (<a href="https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance">https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance</a>) and CDRH Learn (<a href="https://www.fda.gov/training-and-continuing-education/cdrh-learn">https://www.fda.gov/training-and-continuing-education/cdrh-learn</a>). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (<a href="https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice">https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice">https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice</a>) for more information or contact DICE by email (<a href="mailto:DICE@fda.hhs.gov">DICE@fda.hhs.gov</a>) or phone (1-800-638-2041 or 301-796-7100).
{5}------------------------------------------------
If you have any questions concerning the contents of the letter, please contact Samuel Fielden at 301-796-6964.
Sincerely,
Laurel Burk, Ph.D.
Director
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.