← Product Code [SHO](/productcode/SHO) · DEN240071

# neuropacs (DEN240071)

_Automated Imaging Diagnostics, LLC · SHO · Apr 3, 2026 · Neurology · DENG_

**Canonical URL:** https://fda-staging.innolitics.com/device/DEN240071

## Device Facts

- **Applicant:** Automated Imaging Diagnostics, LLC
- **Product Code:** [SHO](/productcode/SHO.md)
- **Decision Date:** Apr 3, 2026
- **Decision:** DENG
- **Submission Type:** Direct
- **Regulation:** 21 CFR 882.2000
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Software as a Medical Device

## Indications for Use

The neuropacs™ system is a software application intended to receive and analyze diffusion MRI data from patients aged 40 years and older with either Parkinson's Disease (PD), Multiple System Atrophy Parkinsonian variant (MSAp), or Progressive Supranuclear Pulse (PSP). The neuropacs™ system provides a classification report based on degenerative brain patterns of MSAp and PSP, to help neuroradiologists and/or neurologists in differentiating these diseases from PD. The neuropacs™ system is intended to provide supplemental information in conjunction with a standard neurological assessment and other clinical tests. This device does not rule out other potential causes of Parkinsonism symptoms or signs and is intended to be used after all other potential causes of Parkinsonism symptoms (such as dementia with Lewy bodies, vascular Parkinsonism, drug-induced Parkinsonism, and corticobasal degeneration/syndrome) have been ruled out by trained clinicians on the basis of clinical, neurological, and other laboratory tests. Patient management decisions should not be made solely on the basis of analysis by the neuropacs™ system.

## Device Story

Neuropacs is a software application that analyzes diffusion MRI data to assist neuroradiologists and neurologists in differentiating Parkinson's Disease (PD) from Multiple System Atrophy Parkinsonian variant (MSAp) and Progressive Supranuclear Palsy (PSP). The system processes brain imaging to identify degenerative patterns associated with MSAp and PSP. It functions as an adjunctive tool, providing a classification report to supplement standard neurological assessments and clinical tests. It is not a stand-alone diagnostic; clinicians must rule out other causes of Parkinsonism (e.g., vascular Parkinsonism, drug-induced, dementia with Lewy bodies) before use. The output is intended to support clinical decision-making, though patient management decisions should not rely solely on the device analysis.

## Clinical Evidence

Clinical performance validation testing is required as a special control to demonstrate diagnostic accuracy and reproducibility for evaluating Parkinsonian syndromes compared to a clinically relevant reference standard. Objective performance measures must be reported.

## Technological Characteristics

Software application; analyzes diffusion MRI data; provides classification report based on degenerative brain patterns. Subject to special controls requiring software verification, validation, and hazard analysis, including technical description of models/algorithms and inputs/outputs.

## Regulatory Identification

The neuropacs™ system is a software application intended to receive and analyze diffusion MRI data from patients aged 40 years and older with either Parkinson's Disease (PD), Multiple System Atrophy Parkinsonian variant (MSAp), or Progressive Supranuclear Palsy (PSP). The system provides a classification report based on degenerative brain patterns of MSAp and PSP to help neuroradiologists and/or neurologists in differentiating these diseases from PD. It is intended for adjunctive use in conjunction with a standard neurological assessment and other clinical tests and is not a stand-alone diagnostic.

## Special Controls

(1) Clinical performance validation testing acquired under anticipated conditions of use must demonstrate that the device performs as intended when used to analyze data from the intended patient population. Testing must demonstrate the diagnostic accuracy and reproducibility for evaluating Parkinsonian syndromes compared to a clinically relevant reference standard. Objective performance measures must be reported.

(2) Software verification, validation, and hazard analysis must be provided. Software documentation must include a technical description of the model/algorithm(s), and algorithm inputs and outputs.

(3) Labeling must include:

(i) A detailed summary of the clinical performance testing methods, including results of the performance testing for tested performance measures/metrics, selection criteria, and the patient demographics;

(ii) A description of the patient population that was used in development or training of the device algorithm/model;

(iii) Device limitations or subpopulations for which the device may not perform as expected or for whom the device has not been validated, including information about how to incorporate device use into the clinical diagnostic workflow;

(iv) A statement that the device is not a stand-alone diagnostic tool and that the device output should only be interpreted in the context of all available clinical information; and

(v) Information for interpretation of the device outputs detailing the risks associated with misinterpretation of the device outputs.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a De Novo](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 3, 2026

Automated Imaging Diagnostics, LLC
% Alexia Haralambous
Principal Consultant
Harmonia Insights, LLC
6800 Wisconsin Ave. #1280
Chevy Chase, MD 20815

Re: DEN240071
Trade/Device Name: neuropacs
Regulation Number: 21 CFR 882.2000
Regulation Name: Parkinsonian syndrome diagnostic aid
Regulatory Class: Class II
Product Code: SHO
Dated: December 4, 2024
Received: December 4, 2024

Dear Alexia Haralambous:

The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA) has completed its review of your De Novo request for classification of the neuropacs, a prescription device under 21 CFR Part 801.109 with the following indications for use:

The neuropacs™ system is a software application intended to receive and analyze diffusion MRI data from patients aged 40 years and older with either Parkinson's Disease (PD), Multiple System Atrophy Parkinsonian variant (MSAp), or Progressive Supranuclear Palsy (PSP). The neuropacs™ system provides a classification report based on degenerative brain patterns of MSAp and PSP, to help neuroradiologists and/or neurologists in differentiating these diseases from PD. The neuropacs™ system is intended to provide supplemental information in conjunction with a standard neurological assessment and other clinical tests. This device does not rule out other potential causes of Parkinsonism symptoms or signs and is intended to be used after all other potential causes of Parkinsonism symptoms (such as dementia with Lewy bodies, vascular Parkinsonism, drug-induced Parkinsonism, and corticobasal degeneration/syndrome) have been ruled out by trained clinicians on the basis of clinical, neurological, and other laboratory tests. Patient management decisions should not be made solely on the basis of analysis by the neuropacs™ system.

FDA concludes that this device should be classified into Class II. This order, therefore, classifies the neuropacs, and substantially equivalent devices of this generic type, into Class II under the generic name Parkinsonian syndrome diagnostic aid.

FDA identifies this generic type of device as:

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

DEN240071 - Alexia Haralambous
Page 2

Parkinsonian syndrome diagnostic aid. A Parkinsonian syndrome diagnostic aid is a prescription device that analyzes patient-specific data to aid in the diagnosis of Parkinsonian syndromes. This device is intended for adjunctive use and not intended as a stand-alone diagnostic.

Section 513(f)(2) of the Food, Drug and Cosmetic Act (the FD&amp;C Act) was amended by section 607 of the Food and Drug Administration Safety and Innovation Act (FDASIA) on July 9, 2012. This law provides two options for De Novo classification. First, any person who receives a "not substantially equivalent" (NSE) determination in response to a 510(k) for a device that has not been previously classified under the Act may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act. On December 13, 2016, the 21st Century Cures Act removed a requirement that a De Novo request be submitted within 30 days of receiving an NSE determination. Alternatively, any person who determines that there is no legally marketed device upon which to base a determination of substantial equivalence may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act without first submitting a 510(k). FDA shall, within 120 days of receiving such a request, classify the device. This classification shall be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing the classification.

On March 2, 2026, FDA received your De Novo requesting classification of the neuropacs. The request was submitted under section 513(f)(2) of the FD&amp;C Act. In order to classify the neuropacs into class I or II, it is necessary that the proposed class have sufficient regulatory controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the De Novo request, FDA has determined that, for the previously stated indications for use, the neuropacs can be classified in class II with the establishment of special controls for class II. FDA believes that class II (special) controls provide reasonable assurance of the safety and effectiveness of the device type. The identified risks and mitigation measures associated with the device type are summarized in the following table:

|  Identified Risks to Health | Mitigation Measures  |
| --- | --- |
|  Device failure or incorrect analysis leading to:
• False positives resulting in inappropriate patient treatment and delayed diagnosis of non-Parkinson’s disease conditions
• False negatives resulting in delayed diagnosis and/or patient treatment | Clinical performance testing
Software verification, validation, and hazard analysis
Labeling  |
|  Use error or misinterpretation of results leading to:
• A false positive resulting in inappropriate patient treatment and delayed diagnosis of non-Parkinson’s disease conditions
• A false negative resulting in delayed diagnosis and/or patient treatment
• Overreliance on device output for follow-up or need for additional diagnostic assessment | Labeling  |

In combination with the general controls of the FD&amp;C Act, the Parkinsonian syndrome diagnostic aid is subject to the following special controls:

{2}

DEN240071 - Alexia Haralambous
Page 3

(1) Clinical performance validation testing acquired under anticipated conditions of use must demonstrate that the device performs as intended when used to analyze data from the intended patient population. Testing must demonstrate the diagnostic accuracy and reproducibility for evaluating Parkinsonian syndromes compared to a clinically relevant reference standard. Objective performance measures must be reported.

(2) Software verification, validation, and hazard analysis must be provided. Software documentation must include a technical description of the model/algorithm(s), and algorithm inputs and outputs.

(3) Labeling must include:

(i) A detailed summary of the clinical performance testing methods, including results of the performance testing for tested performance measures/metrics, selection criteria, and the patient demographics;

(ii) A description of the patient population that was used in development or training of the device algorithm/model;

(iii) Device limitations or subpopulations for which the device may not perform as expected or for whom the device has not been validated, including information about how to incorporate device use into the clinical diagnostic workflow;

(iv) A statement that the device is not a stand-alone diagnostic tool and that the device output should only be interpreted in the context of all available clinical information; and

(v) Information for interpretation of the device outputs detailing the risks associated with misinterpretation of the device outputs.

In addition, this is a prescription device and must comply with 21 CFR 801.109.

Although this letter refers to your product as a device, please be aware that some granted products may instead be combination products. If you have questions on whether your product is a combination product, contact CDRHProductJurisdiction@fda.hhs.gov.

Section 510(m) of the FD&amp;C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the FD&amp;C Act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type. FDA has determined premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device type and, therefore, the device is not exempt from the premarket notification requirements of the FD&amp;C Act. Thus, persons who intend to market this device type must submit a premarket notification containing information on the Parkinsonian syndrome diagnostic aid they intend to market prior to marketing the device.

Please be advised that FDA's decision to grant this De Novo request does not mean that FDA has made a determination that your device complies with other requirements of the FD&amp;C Act or any Federal statutes

{3}

DEN240071 - Alexia Haralambous
Page 4

and regulations administered by other Federal agencies. You must comply with all the FD&amp;C Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the FD&amp;C Act; 21 CFR 1000-1050).

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System Rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

A notice announcing this classification order will be published in the Federal Register. A copy of this order and supporting documentation are on file in the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Room 1061, Rockville, MD 20852 and are available for inspection between 9 a.m. and 4 p.m., Monday through Friday.

As a result of this order, you may immediately market your device as described in the De Novo request, subject to the general control provisions of the FD&amp;C Act and the special controls identified in this order.

For comprehensive regulatory information about medical devices and radiation-emitting products, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

If you have any questions concerning the contents of the letter, please contact Liyue Tong at 301-796-6705.

{4}

DEN240071 - Alexia Haralambous
Page 5

Sincerely,

JAY R. GUPTA -S

For Jaime Raben, Ph.D.
Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

---

**Source:** [https://fda-staging.innolitics.com/device/DEN240071](https://fda-staging.innolitics.com/device/DEN240071)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a De Novo](https://innolitics.com/services/regulatory/), [a 510(k)](https://innolitics.com/services/510ks/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
